| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | Prospera Spinal Cord Stimulation (SCS) System, Resilience Percutaneous Lead, HomeStream Remote Management |
| Generic Name | Stimulator, spinal-cord, totally implanted for pain relief |
| Applicant | Biotronik Nro, Inc. 6024 Jean Rd. Lake Oswego, OR 97035 |
| PMA Number | P210037 |
| Supplement Number | S011 |
| Date Received | 04/04/2025 |
| Decision Date | 06/25/2025 |
| Product Code |
LGW |
| Advisory Committee |
Neurology |
| Supplement Type | Normal 180 Day Track |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement approval for changes made to the Prospera SCS System embedded software (implant software [ISW]) and Neuro Service Center (NSC). Specifically, the Prospera SCS System embedded software (ISW) was updated to version 1.0.3, which included changes to the lead impedance measurements algorithm to provide more information during intraoperative testing and improve the accuracy of calculation of resistive measurements. The NSC was updated to version 1.6.3, which included changes to allow magnetic resonance imaging (MRI) mode entry if the automated lead impedance measurements are within range for the three most recent days and the consideration of touching leads condition as safe for MRI. |
|
|