New Search Back to Search Results
Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceTriClip G4 System
Generic NameTricuspid valve repair device, percutaneously delivered
ApplicantABBOTT MEDICAL
177 County Rd. B E.
St. Paul, MN 55117
PMA NumberP230007
Supplement NumberS018
Date Received09/19/2025
Decision Date10/17/2025
Product Code NPS 
Advisory Committee Cardiovascular
Supplement Type30-Day Notice
Supplement Reason Process Change - Manufacturer/Sterilizer/Packager/Supplier
Expedited Review Granted? No
Combination ProductNo
Predetermined Change Control Plan AuthorizedNo
Approval Order Statement  
the substitution of commercially available DC shaft components used for process monitoring with vendor inspected DC shaft components with specific reject reasons that do not impact bond integrity, and a change in the incoming sampling plan of 8 components from 100% inspection to RQL
-
-