| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | Star S4 IR Excimer Laser System |
| Generic Name | Excimer laser system |
| Applicant | Amo Manufacturing USA, LLC 510 Cottonwood Dr. Milpitas, CA 95035 |
| PMA Number | P930016 |
| Supplement Number | S051 |
| Date Received | 05/16/2017 |
| Decision Date | 07/14/2017 |
| Product Code |
LZS |
| Advisory Committee |
Ophthalmic |
| Supplement Type | Real-Time Process |
| Supplement Reason | Change Design/Components/Specifications/Material |
| Expedited Review Granted? | No |
| Combination Product | No |
| Predetermined Change Control Plan Authorized | No |
Approval Order Statement Approval for the replacement of the existing photo interrupter (3703-0124, manufactured by Hamamatsu) in the BSM Flex assembly (0030-1275) used in the Beam Shaping Module (BSM) assembly (0030-2162) of the STAR S4 IR, with a replacement photo interrupter (3703-0157-L, manufactured by Sharp Microelectronics). |
|
|