| Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information. |
| |
| Device | Medtronic (R) CapSure (R) EPI Pacing Lead |
| Generic Name | Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes |
| Applicant | Medtronic, Inc. 8200 Coral Sea St. N.E Ms Mvs11 Mounds View, MN 55112 |
| PMA Number | P950024 |
| Supplement Number | S114 |
| Date Received | 02/13/2025 |
| Decision Date | 03/20/2025 |
| Product Code |
NVN |
| Advisory Committee |
Cardiovascular |
| Supplement Type | 135 Review Track For 30-Day Notice |
| Supplement Reason | Process Change - Manufacturer/Sterilizer/Packager/Supplier |
| Expedited Review Granted? | No |
| Combination Product | Yes |
| Predetermined Change Control Plan Authorized | Yes |
Approval Order Statement approval for a predetermined change control plan (PCCP) that encompasses the following biocompatibility-related manufacturing changes, provided that they do not impact the properties of the device materials, for the devices under the above PMAs:1. Removal of a processing aid (solvent, lubricant, antifoaming agent, etc.), or cleaning agent2. Removal of a cleaning process3. Change in parameters (temperature, time, sonication, etc.) of a current cleaning process |
|
|