Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceActiva Deep Brain Stimulation Therapy System
Generic NameStimulator, electrical, implanted, for parkinsonian tremor
ApplicantMedtronic, Inc.
7000 Central Ave., NE
Minneapolis, MN 55432
PMA NumberP960009
Supplement NumberS336
Date Received12/18/2018
Decision Date03/01/2019
Product Code MHY 
Advisory Committee Neurology
Supplement TypeReal-Time Process
Supplement Reason Change Design/Components/Specifications/Material
Expedited Review Granted? No
Combination ProductNo
Predetermined Change Control Plan AuthorizedNo
Approval Order Statement  
Approval to add a Polyethylene terephthalate (PET) slip tube over proximal end of each cable conductor to prevent mechanical locking of the build epoxy to the Ethylene tetrafluoroethylene (ETFE) coated wire cables for DBS Extension models 37085 and 37086.
-
-