• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Premarket Approval (PMA)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
Note: This medical device record is a PMA supplement. A supplement may have changed the device description/function or indication from that approved in the original PMA. Be sure to look at the original PMA record for more information.
 
DeviceNatrelle Saline-Filled Breast Implants
Generic NameProsthesis, breast, inflatable, internal, saline
Regulation Number878.3530
ApplicantAllergan
2525 Dupont Dr.
Irvine, CA 92612
PMA NumberP990074
Supplement NumberS041
Date Received10/02/2017
Decision Date11/03/2017
Product Code FWM 
Advisory Committee General & Plastic Surgery
Supplement Type30-Day Notice
Supplement Reason Process Change - Manufacturer/Sterilizer/Packager/Supplier
Expedited Review Granted? No
Combination ProductNo
Predetermined Change Control Plan AuthorizedNo
Approval Order Statement  
Implement an Electronic Data Acquisition (EDA) System to collect cycle parameter, such as temperature and time, for the shell drying after washing and gel curing after shell filling occur in the Blue-M oven. The above change is located 900 Parkway Global Park, La Aurora, Heredia, Costa Rica.
-
-