| | Class 2 Device Recall LEO Scooter |  |
| Date Initiated by Firm | December 07, 2011 |
| Date Posted | January 17, 2012 |
| Recall Status1 |
Terminated 3 on July 02, 2013 |
| Recall Number | Z-0774-2012 |
| Recall Event ID |
60670 |
| 510(K)Number | K090735 |
| Product Classification |
Vehicle, motorized 3-wheeled - Product Code INI
|
| Product | Leo 3-Wheel Scooter and Leo Scooter, Distributed By:
Invacare Corporation, 1200 Taylor Street, Elyria, OH 44035
Intended to provide transportation for a disabled or elderly person. |
| Code Information |
Model # LEO-4B; Serial # 10MCT000110, 10MCT000111, 10MCT000112, 10MCT000113, 10MCT000114, 10MCT000115, 10MCT000116, 10MCT000117, 10MCT000118, 10MCT000119, 10MCT000120, 10MCT000121, 10MCT000122, 10MCT000123, 10MCT000124, 10MCT000125, 10MCT000126, 10MCT000127, 10MCT000128, 10MCT000129, 10MCT000130, 10MCT000131, 10MCT000132, 10MCT000133, 10MCT000134, 10MCT000135, 10MCT000136, 10MCT000137, 10MCT000138, 10MCT000139, 10MCT000140, 10MCT000141, 10MCT000142, 10MCT000143, 10MCT000144, 10MCT000145, 10MCT000146, 10MCT000147, 10MCT000148, 10MCT000149, 10MCT000150, 10MCT000151, 10MCT000152, 10MCT000153, 10MCT000154, 10MCT000155 & 10MCT000156. Model # LEO-4S; Serial # 10MCT000001, 10MCT000002, 10MCT000003, 10MCT000004, 10MCT000005, 10MCT000006, 10MCT000007, 10MCT000008, 10MCT000009, 10MCT000010, 10MCT000011, 10MCT000012, 10MCT000013, 10MCT000014, 10MCT000015, 10MCT000016, 10MCT000017, 10MCT000018, 10MCT000019, 10MCT000020, 10MCT000021, 10MCT000022, 10MCT000023, 10MCT000024, 10MCT000025, 10MCT000026, 10MCT000027, 10MCT000028, 10MCT000029, 10MCT000030, 10MCT000031, 10MCT000032, 10MCT000033, 10MCT000034, 10MCT000035, 10MCT000036, 10MCT000037, 10MCT000038, 10MCT000039, 10MCT000040, 10MCT000041, 10MCT000042, 10MCT000043, 10MCT000044, 10MCT000045, 10MCT000046, 10MCT000047, 10MCT000048, 10MCT000049, 10MCT000050, 10MCT000051, 10MCT000052, 10MCT000053, 10MCT000054, 10MCT000055, 10MCT000056, 10MCT000057, 10MCT000058, 10MCT000059, 10MCT000060, 10MCT000061, 10MCT000062, 10MCT000063, 10MCT000064, 10MCT000065, 10MCT000066, 10MCT000067, 10MCT000068, 10MCT000069, 10MCT000070, 10MCT000071, 10MCT000072, 10MCT000073, 10MCT000074, 10MCT000075, 10MCT000076, 10MCT000077, 10MCT000078, 10MCT000079, 10MCT000080, 10MCT000081, 10MCT000082, 10MCT000083, 10MCT000084, 10MCT000085, 10MCT000086, 10MCT000087, 10MCT000088, 10MCT000089, 10MCT000090, 10MCT000091, 10MCT000092, 10MCT000093, 10MCT000094, 10MCT000095, 10MCT000096, 10MCT000097, 10MCT000098, 10MCT000099, 10MCT000100, 10MCT000101, 10MCT000102, 10MCT000103, 10MCT000104, 10MCT000105, 10MCT000106, 10MCT000107, 10MCT000108, 10MCT000109, 10MDG0001, 10MDG0002, 10MDG0003, 10MDG0004, 10MDG0005, 10MDG0006, 10MDG0007, 10MDG0008, 10MDG0009, 10MDG0010, 10MDG0011, 10MDG0012, 10MDG0013, 10MDG0014, 10MDG0015, 10MDG0016, 10MDG0017, 10MDG0018, 10MDG0019, 10MDG0020, 10MDG0021, 10MDG0022, 10MDG0023, 10MDG0024, 10MDG0025, 10MDG0026, 10MDG0027, 10MDG0028, 10MDG0029, 10MDG0030, 10MDG0047, 10MDG0048, 10MDG0049, 10MDG0050, 10MDG0051, 10MDG0052, 10MDG0053, 10MDG0054, 10MDG0055, 10MDG0056, 10MDG0057, 10MDG0058, 10MDG0059, 10MDG0060, 10MDG0061, 10MDG0062, 10MDG0063, 10MDG0064, 10MDG0065, 10MDG0066, 10MDG0067, 10MDG0068, 10MDG0069, 10MDG0070, 10MDG0071, 10MDG0072, 10MDG0073, 10MDG0074, 10MDG0075, 10MDG0076, 10MDG0077, 10MDG0078, 10MDG0079, 10MDG0080, 10MDG0081, 10MDG0082, 10MDG0083, 10MDG0084, 10MDG0085, 10MDG0086, 10MDG0087, 10MDG0088, 10MDG0089, 10MDG0090, 10MDG0091, 10MDG0092, 10MDG0093, 10MDG0094, 10MDG0095, 10MDG0096, 10MDG0097, 10MDG0098, 10MDG0099, 10MDG0100, 10MDG0101, 10MDG0102, 10MDG0103, 10MDG0104, 10MDG0105, 10MDG0106, 10MDG0107, 10MDG0108, 10MDG0109, 10MDG0110, 10MDG0111, 10MDG0112, 10MDG0113, 10MDG0114, 10MDG0115, 10MDG0116, 10MDG0117, 10MDG0118, 10MDG0119, 10MDG0120, 10MDG0121, 10MDG0122, 10MDG0123, 10MDG0124, 10MDG0125, 10MDG0126, 10MDG0127, 10MDG0128, 10MDG0129, 10MDG0130, 10MDG0131, 10MDG0132, 10MDG0133, 10MDG0134, 10MDG0135, 10MDG0151, 10MDG0152, 10MDG0153, 10MDG0154, 10MDG0155, 10MDG0156, 10MDG0157, 10MDG0158, 10MDG0159, 10MDG0160, 10MDG0161, 10MDG0162, 10MDG0163, 10MDG0164, 10MDG0165, 10MDG0166, 10MDG0167, 10MDG0168, 10MDG0169, 10MDG0170, 10MDG0171, 11ADG0001, 11ADG0002, 11ADG0003, 11ADG0004, 11ADG0005, 11ADG0006, 11ADG0007, 11ADG0008, 11ADG0009, 11ADG0010, 11ADG0011, 11ADG0012, 11ADG0013, 11ADG0014, 11ADG0015, 11ADG0016, 11ADG0017, 11ADG0018, 11ADG0019, 11ADG0020, 11ADG0021, 11ADG0022, 11ADG0023, 11ADG0024, 11ADG0025, 11ADG0026, 11ADG0027, 11ADG0028, 11ADG0029, 11ADG0030, 11ADG0031, 11ADG0032, 11ADG0033, 11ADG0034, 11ADG0035, 11ADG0036, 11ADG0037, 11ADG0038, 11ADG0039, 11ADG0040, 11ADG0041, 11ADG0042, 11ADG0073, 11ADG0074, 11ADG0075, 11ADG0076, 11ADG0077, 11ADG0078, 11ADG0079, 11ADG0080, 11ADG0081, 11ADG0082, 11ADG0083, 11ADG0084, 11ADG0085, 11ADG0086, 11ADG0087, 11ADG0088, 11ADG0089, 11ADG0090, 11ADG0091, 11ADG0092, 11ADG0093, 11ADG0094, 11ADG0095, 11ADG0096, 11ADG0097, 11ADG0098, 11ADG0099, 11ADG0100, 11ADG0101, 11ADG0102, 11ADG0103, 11ADG0104, 11ADG0105, 11ADG0106, 11ADG0107, 11ADG0108, 11ADG0109, 11ADG0110, 11ADG0111, 11ADG0112, 11ADG0113, 11ADG0114, 11ADG0115, 11ADG0116, 11ADG0117, 11ADG0118, 11ADG0119, 11ADG0120, 11ADG0121, 11ADG0122, 11ADG0123, 11ADG0124, 11ADG0125, 11ADG0126, 11ADG0127, 11ADG0128, 11ADG0129, 11ADG0130, 11ADG0131, 11ADG0132, 11ADG0133, 11ADG0134, 11ADG0135, 11ADG0136, 11ADG0137, 11ADG0138, 11ADG0139, 11ADG0140, 11ADG0141 & 11ADG0142. |
| FEI Number |
3002889431
|
Recalling Firm/ Manufacturer |
Invacare Corporation
1 Invacare Way
Elyria
OH
44035-4190
|
| For Additional Information Contact | 440-329-6595 |
Manufacturer Reason for Recall | Invacare Corporation decided to recall the products because the scooter backrest fails unexpectedly due to seat flexing, and this has the potential to result in a consumer falling causing death or serious injury. |
FDA Determined Cause 2 | Device Design |
| Action | On 12/7/2011 the firm sent Urgent: Recall Information Letters to their customers. The letter informed customers of the affected product and the related problem. The letter also states that Invacare has developed replacement seat kits intended to replace the original seats. The replacement kit will be provided free of charge and can be ordered by completing and returning the attached response card. After receiving replacement kits, customers are asked to contact their customers that have potentially affected devices and replace the seats immediately. Questions should be directed to Customer Service at 1-800-333-6900. |
| Quantity in Commerce | 391 scooters |
| Distribution | Worldwide Distribution -- USA, including the states of AK, CA, FL, GA, HI, IA, ID, IL, KS, LA, MD, MI, MN, MO, MT, NC, NY, OH, OK, OR, PA, SC, TX, UT, VA, WA, & WI, and the countries of AUSTRIA, BELGIUM, CZECH REPUBLIC, ESTONIA, FRANCE, GERMANY, IRELAND, ITALY, JERSEY (CHANNEL ISLANDS), LITHUANIA, PORTUGAL, SWEDEN, SWITZERLAND, & UNITED KINGDOM. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = INI
|
|
|
|