| | Class 2 Device Recall Depuy Mitek Lupine (Panalok) Loop Arthro 2 Ethibond |  |
| Date Initiated by Firm | January 27, 2012 |
| Date Posted | March 06, 2012 |
| Recall Status1 |
Terminated 3 on August 05, 2015 |
| Recall Number | Z-1162-2012 |
| Recall Event ID |
61088 |
| 510(K)Number | K040827 |
| Product Classification |
Fastener, fixation, biodegradable, soft tissue - Product Code MAI
|
| Product | Depuy Mitek Lupine (Panalok) Loop Arthro #2 Ethibond
Catalog Number: 210704
The Lupine Anchor System is indicated for use in soft tissue bone fixation. |
| Code Information |
Lot Numbers: 3254455 3254464 3254772 3262313 3270104 3276492 3278110 3278111 3286624 3286916 3286917 3286918 3287309 3287311 3287390 3287392 3295568 3295569 3297682 3297683 3299220 3304047 3304048 3304049 3305321 3305430 3305681 3305682 3305683 3306004 3306391 3306677 3307006 3334454 3334455 3334634 3334635 3345378 3345379 3370177 3370178 3370180 3370783 3371236 3371239 3371241 3371615 3377559 3377561 3377562 3383125 3383127 3383265 3383266 3383267 3383268 3383623 3384221 3384222 3396398 3396503 3396504 3397755 3412266 3412270 3412856 3414484 3414533 3415555 3415714 3416128 3417104 3418602 3420846 3423623 3424661 3425288 3426996 3427237 3427238 3428991 3432156 3433007 3437113 3442445 3443870 3447918 3450122 3451457 3456002 3456269 3456435 3459762 3464533 3466040 3468840 3471144 3473778 3476081 3477621 3478976 3484026 3486327 3489054 3489725 3491267 3494653 3494656 3494917 3501542 3502156 3502967 3503801 3504239 3507680 3508146 3508905 3509441 3509538 3509541 3509956 3510410 3511003 3518413 3523828 3527730 3528739 3531511 3533828 3540939 3541612 3542349 3549274 3551177 3556763 3556770 3560268 3564054 3570366 3570374 3571653 |
| FEI Number |
1221934
|
Recalling Firm/ Manufacturer |
DePuy Mitek, Inc., a Johnson & Johnson Co.
325 Paramount Dr
Raynham
MA
02767
|
| For Additional Information Contact | Same 508-880-8100 |
Manufacturer Reason for Recall | Lupine BR Anchor w/ Orthocord may be assembled with an incorrect suture configuration and result in a loose repair |
FDA Determined Cause 2 | Process control |
| Action | DePuy Mitek sent an Urgent Voluntary Product Recall letter dated January 27, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to immediately check all inventories to determine if they have any affected product. Customers were asked to refer to the instructions to report their inventory status, return and obtain replacement or credit for the product.
For questions regarding this recall call 1-877-379-4871. |
| Quantity in Commerce | 41581 |
| Distribution | Worldwide Distribution - USA (nationwide) and the countries of Canada, AE, Argentina, Australia, Austria,Belgium,Brazil, Bahrain, Chile, China, Columbia,Cyprus, Check Republic, Germany, Denmark, Ecuador, Egypt, Spain, Finland, France, Finland, Great Britain, Guadeloupe, Hong Kong, Hungary,Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Luxembourg,Latvia, Martinique, Montenegro, Mexico, Maylasia, Netherlands,Norway, New Zealand, Oman, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Sweden, Singapore, Slovakia, Thailand, Tunisia, Turkey,Uruguay, Venezuela, and South Africa |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|