 Class 3 Device Recall STERRAD 100S Cassettesee related information
Date Initiated by FirmJune 19, 2012
Date PostedSeptember 07, 2012
Recall Status1 Terminated 3 on November 08, 2013
Recall NumberZ-2337-2012
Recall Event ID 62925
510(K)NumberK991999 
Product Classification Sterilizer, chemical - Product Code MLR
ProductSTERRAD 100S Cassette, Batch # 12A077. The STERRAD 100S and STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture sensitive reusable medical devices. The sterilant used is hydrogen peroxide supplied in cassette form as a separate accessory. The STERRAD Cassettes are used in conjunction with the STERRAD Sterilizers, and each cassette model is for use only in the matching sterilizer model.
Code Information Product Code: 10113
FEI Number 3003662624
Recalling Firm/
Manufacturer
Advanced Sterilization Products
33 Technology Dr
Irvine CA 92618-2346
For Additional Information Contact
949-453-6400
Manufacturer Reason
for Recall
Advanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The STERRAD 100S and STERRAD NX System cases were shipped without verification of proper printing.
FDA Determined
Cause 2
Error in labeling
ActionAdvanced Sterilization Products sent an "URGENT MEDICAL DEVICE RECALL (REMOVAL) letter dated July 16, 2012 to all their customers who purchased the STERRAD 100S and STERRAD NX System cassettes. The letter informed the customers of the problem identified and the actions to be taken. For the STERRAD NX System cassettes, there should be no impact to the functionality of these affected cassettes. Although printing may be smeared, illegible, or missing, the packaging has multiple references to the information, and the cassettes should be usable despite any printing error. For the STERRAD 100S System cassettes, there is a chance that the barcode will be unreadable and the STERRAD 100S Sterilizer will not accept the cassette. Customers were instructed to complete and return the business reply card and packing slip that are included in the letter. Customers were instructed to contact Stericycle at (888) 406-9303 for additional labels. Customers with questions about the recall letter were instructed to contact Stericycle at (888) 406-9303.
Quantity in Commerce7,495 units
DistributionNationwide Distribution.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.