| Date Initiated by Firm | December 05, 2012 |
| Date Posted | May 25, 2013 |
| Recall Status1 |
Terminated 3 on September 20, 2013 |
| Recall Number | Z-1404-2013 |
| Recall Event ID |
65082 |
| Product Classification |
Refractometer, ophthalmic - Product Code HKO
|
| Product | Canon Full Auto Ref-Keratometer RK-F2, Catalog Number 6937B002AA.
Product Usage: This device is intended to be used to measure the refractive power and the radius of corneal curvature of the human eye. |
| Code Information |
Catalog Number 6937B002AA. All serial numbers from the first unit onward. |
| FEI Number |
1000181430
|
Recalling Firm/ Manufacturer |
Canon Inc.
30-2, Shimomaruko 3-chome,
Ohta-ku
Tokyo
Japan
|
Manufacturer Reason for Recall | Canon Inc. had received 17 claims from our foreign
sales companies between October to November
2012 that measurement value of RK-F2s refractive
power (SPH value) sometimes indicated abnormal
value which occurs approximately 5% in probability. |
FDA Determined Cause 2 | Software design |
| Action | Canon initiated recall by sending letters on November 22, 2012 to all affected customers. The letter identified the affected product, problem, actions to be taken and contact information. |
| Quantity in Commerce | 384 units |
| Distribution | Worldwide Distribution - USA Nationwide and countries of: Australia, Japan, Korea, Singapore and United Kingdom. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|