Class 2 Device Recall PB840 Ventilator Printed Circuit Board

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall PB840 Ventilator Printed Circuit Boardsee related information
Date Initiated by FirmJune 04, 2014
Date PostedJune 17, 2014
Recall Status1 Terminated 3 on July 11, 2016
Recall NumberZ-1809-2014
Recall Event ID 68454
510(K)NumberK970460 
Product Classification Ventilator, continuous, facility use - Product Code CBK
ProductModel Number(s): PB840 Ventilator (4-840120DIUU-XX). The product is a component of the Puritan Bennett 840 Ventilator: The Backlight Inverter Printed Circuit Board (BLI PCBA) for the 9.4 Display part number 4-079056-00. The PB840 Ventilator System is intended to provide invasive or non-invasive ventilator support and monitoring for infant, pediatric, and adult patients with respiratory failure or respiratory insufficiency. It is used in hospitals and healthcare facilities
Code Information Display part number 4-079056-00; serial number 4200000001 to 3510012377 that have a 9.4 display. Lots 1032H,1041H,1042H,1043H,1008J,1009J,1010J,1018K,1019K,1020K,1021K,1022K,1023K,1030L,1031L,1032L,1033L,1034L,1035L,
FEI Number 2936999
Recalling Firm/
Manufacturer
Nellcor Puritan Bennett Inc. (dba Covidien LP)
6135 Gunbarrel Ave
Boulder CO 80301-3214
For Additional Information ContactMr. Michael A. Ronningen
303-876-8870
Manufacturer Reason
for Recall
In the case of a loss of GUI display due to a Backlight Inverter PCBA failure, the ventilator continues to provide uninterrupted ventilatory support at the programmed settings for the patient. However, there is a loss of display and thus there is a necessity to move the patient to another ventilator.
FDA Determined
Cause 2
Component design/selection
ActionCovidien will be issuing the Product Notification letter in 2014. Customers are requested to forward the letter to anyone they may have transferred or assigned for use one of the 9.4display configuration ventilators. Service engineers will be replacing the BLI PCBA with a new design containing the 1.0 ampere fuse and then destroying the old PCBA. Ventilators replaced with newer models under the commercial program will be returned to Covidien for disposal. Should you have any questions regarding this letter or to report any issues with the Puritan Bennett" 840 ventilator contact our Technical Support Department at 1-800-255-6774, option 4, then option 1, to ensure proper device reporting procedures are followed. Thank you for your prompt attention to this matter.
Quantity in Commerce7371
DistributionWorldwide Distribution - USA (nationwide) including the states: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and District of Columbia., and the countries of Canada and Japan.
Total Product Life CycleTPLC Device Report