U.S. Department of Health and Human Services

Class 2 Device Recall Esprit V1000 Ventilator

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Esprit V1000 Ventilatorsee related information
Date Initiated by FirmOctober 29, 2014
Date PostedDecember 01, 2014
Recall Status1 Terminated 3 on June 07, 2016
Recall NumberZ-0485-2015
Recall Event ID 69728
510(K)NumberK001208 K034040 K041412 K023350 K981072 
Product Classification Ventilator, continuous, facility use - Product Code CBK
ProductEsprit V1000 Ventilator and Esprit/V200 Conversion Ventilator The Esprit V1000 and Esprit/V200 Conversion Ventilators are microprocessor-controlled, electrically powered mechanical ventilators. Intended for use by medical personnel to provide continuous or intermittent ventilator support for adult, pediatric and neonatal patients.
Code Information All Esprit V1000 ventilators manufactured prior to December 31, 2004 and those units manufactured prior to 12/2004 that have been converted to V200.  Serial Numbers VS3000584 VS3000614 VS3001023 VS3001026 VS3001046 VS3001122 VS3001123 VS3001643 VS3001814 VS3001815 VS3002027 VS3002079 VS3004925 VS3001953 VS3001979 VS3002122 VS3004232 VS3004312 VS3004336 VS3004342 VS3004348 VS3004946 VS3005661 VS3005779 VS3000536 VS3000538 VS3000542 VS3000927 VS3000935 VS3001253 VS3001255 VS3001314 VS3001355 VS3001388 VS3001986 VS3002078 VS3002625 VS3002737 VS3002768 VS3002804 VS3002832 VS3002835 VS3003127 VS3003194 VS3003199 VS3003218 VS3004609 VS3004841 VS3000570 VS3000638 VS3001007 VS3001017 VS3001027 VS3001110 VS3001222 VS3001256 VS3001266 VS3001313 VS3001501 VS3001502 VS3001503 VS3001504 VS3001505 VS3001507 VS3001508 VS3001509 VS3001514 VS3001515 VS3001653 VS3001694 VS3001748 VS3001766 VS3001781 VS3001786 VS3001795 VS3001797 VS3001799 VS3001804 VS3001808 VS3001809 VS3002016 VS3002662 VS3002749 VS3005907 VS3006181 VS3000604 VS3000675 VS3001276 VS3001366 VS3001768 VS3002199 VS3002205 VS3002823 VS3004607 VS3004642 VS3004684 VS3004685 VS3004723 VS3004803 VS3004832 VS3004916 VS3004945 VS3005626 VS3005797 VS3005852 VS3005865 VS3005902 VS3005984 VS3006200 VS3006218 VS3003252 VS3003301 VS3003343 VS3003427 VS3003672 VS3003673 VS3003674 VS3003742 VS3003746 VS3003787 VS3003817 VS3003829 VS3003833 VS3003913 VS3003933 VS3004565 VS3004648 VS3004689 VS3004729 VS3004830 VS3004942 VS3004956 VS3004974 VS3004983 VS3005041 VS3005627 VS3005656 VS3005665 VS3005681 VS3005689 VS3005690 VS3005702 VS3005719 VS3005731 VS3005752 VS3005778 VS3005789 VS3005800 VS3005801 VS3005802 VS3005806 VS3005811 VS3005813 VS3005814 VS3005815 VS3005819 VS3005823 VS3005824 VS3005828 VS3005834 VS3001865 VS3001981 VS3001982 VS3001990 VS3003696 VS3003219 VS3003876 VS3004563 VS3006204 VS3006212 VS3000576 VS3002176 VS3002223 VS3002678 VS3004555 VS3004785 VS3004798 VS3005695 VS3005784 VS3005909 VS3001691 VS3003214 VS3003272 VS3003327 VS3003653 VS3000955 VS3000998 VS3001015 VS3001139 VS3001154 VS3001166 VS3001283 VS3001661 VS3001674 VS3001676 VS3001687 VS3001736 VS3001757 VS3001819 VS3001833 VS3001842 VS3001844 VS3001858 VS3001911 VS3001912 VS3001923 VS3001970 VS3001972 VS3002006 VS3002026 VS3002033 VS3002034 VS3002082 VS3002096 VS3002116 VS3002185 VS3002731 VS3002748 VS3002757 VS3002769 VS3002771 VS3002785 VS3002791 VS3002793 VS3002797 VS3002817 VS3002837 VS3002846 VS3002850 VS3002858 VS3002872 VS3002909 VS3002911 VS3002934 VS3002939 VS3002947 VS3002952 VS3002955 VS3002961 VS3002963 VS3002965 VS3002972 VS3002973 VS3002974 VS3002980 VS3003100 VS3003102 VS3003105 VS3003108 VS3003109 VS3003117 VS3003123 VS3003126 VS3003131 VS3003133 VS3003143 VS3003150 VS3003156 VS3003159 VS3003160 VS3003164 VS3003176 VS3003184 VS3003191 VS3003192 VS3003196 VS3003200 VS3003202 VS3003209 VS3003226 VS3003235 VS3003254 VS3003255 VS3003258 VS3003260 VS3003265 VS3003267 VS3003270 VS3003273 VS3003276 VS3003282 VS3003287 VS3003291 VS3003292 VS3003293 VS3003294 VS3003300 VS3003310 VS3003317 VS3003321 VS3003323 VS3003336 VS3003337 VS3003339 VS3003351 VS3003359 VS3003361 VS3003364 VS3003376 VS3003378 VS3003391 VS3003404 VS3003412 VS3003417 VS3003421 VS3003426 VS3003439 VS3003623 VS3003625 VS3003626 VS3003627 VS3003628 VS3003629 VS3003630 VS3003631 VS3003634 VS3003635 VS3003637 VS3003638 VS3003640 VS3003643 VS3003644 VS3003649 VS3003650 VS3003651 VS3003652 VS3003654 VS3003655 VS3003656 VS3003657 VS3003658 VS3003659 VS3003663 VS3003664 VS3003666 VS3003668 VS3003670 VS3003686 VS3003702 VS3003704 VS3003707 VS3003713 VS3003728 VS3003729 VS3003730 VS3003731 VS3003737 VS3003740 VS3003741 VS3003743 VS3003750 VS3003752 VS3003754 VS3003757 VS3003763 VS3003766 VS3003770 VS3003772 VS3003774 VS3003784 VS3003785 VS3003793 VS3003803 VS3003807 VS3003813 VS3003814 VS3003831 VS3003836 VS3003840 VS3003841 VS3003844 VS3003846 VS3003848 VS3003849 VS3003851 VS3003852 VS3003854 VS3003855 VS3003858 VS3003887 VS3003890 VS3003928 VS3004976 VS3005820 VS3005856 VS3005893 VS3005894 VS3005899 VS3005905 VS3005925 VS3005942 VS3006199 VS3006203 VS3001149 VS3001670 VS3001806 VS3001811 VS3002249 VS3002956 VS3002986 VS3003004 VS3003197 VS3003206 VS3004268 VS3004419 VS3004924 VS3005668 VS3005691 VS3005869 VS3001013 VS3001720 VS3001762 VS3001810 VS3001898 VS3002156 VS3002195 VS3002206 VS3002212 VS3002221 VS3002229 VS3002243 VS3002960 VS3002994 VS3003751 VS3003884 VS3003996 VS3004148 VS3004272 VS3005024 VS3005036 VS3000574 VS3000631 VS3000637 VS3000641 VS3000644 VS3000918 VS3000921 VS3000922 VS3000923 VS3000924 VS3000934 VS3000986 VS3000995 VS3001010 VS3001012 VS3001014 VS3001016 VS3001018 VS3001041 VS3001052 VS3001053 VS3001054 VS3001055 VS3001232 VS3001240 VS3001259 VS3001268 VS3001272 VS3001301 VS3001331 VS3001360 VS3001741 VS3001794 VS3002134 VS3002632 VS3002667 VS3002668 VS3002708 VS3002721 VS3002760 VS3002926 VS3003172 VS3003726 VS3003747 VS3003811 VS3004178 VS3004464 VS3004538 VS3004561 VS3004577 VS3004601 VS3004629 VS3004674 VS3004805 VS3004811 VS3004837 VS3004851 VS3004980 VS3004987 VS3006193 VS3006202 VS3005886 VS3000563 VS3001633 VS3001880 VS3001900 VS3001993 VS3001998 VS3002008 VS3002021 VS3002023 VS3002045 VS3002063 VS3002077 VS3002083 VS3002130 VS3002136 VS3002792 VS3003177 VS3001662 VS3001867 VS3001868 VS3001980 VS3001988 VS3002003 VS3002005 VS3002028 VS3002032 VS3002042 VS3002066 VS3002069 VS3002080 VS3002087 VS3002088 VS3002093 VS3002097 VS3002107 VS3002108 VS3002126 VS3002139 VS3002210 VS3002232 VS3002246 VS3002681 VS3002696 VS3002725 VS3002733 VS3002759 VS3002763 VS3002776 VS3002824 VS3002853 VS3003054 VS3003289 VS3003987 VS3004038 VS3004170 VS3004189 VS3004230 VS3004246 VS3004254 VS3004261 VS3004270 VS3004279 VS3004314 VS3004390 VS3004489 VS3004548 VS3004783 VS3004941 VS3005653 VS3005739 VS3005868 VS3005914 VS3006064 VS3006104 VS3006222 VS3001738 VS3001751 VS3001025 VS3005703 VS3001712 VS3002810 VS3002830 VS3003013 VS3004568 VS3004638 VS3005013 VS3000553 VS3000930 VS3000931 VS3000944 VS3000945 VS3000948 VS3000949 VS3000951 VS3000961 VS3000962 VS3000963 VS3000978 VS3001119 VS3001138 VS3001155 VS3001156 VS3001160 VS3001422 VS3001774 VS3001939 VS3001976 VS3002145 VS3002164 VS3002169 VS3002179 VS3002847 VS3002861 VS3003155 VS3003163 VS3003169 VS3003193 VS3003229 VS3003278 VS3003688 VS3003698 VS3003709 VS3003712 VS3003723 VS3003725 VS3003734 VS3003843 VS3006063 VS3006098 VS3005676 VS3001335 VS3004725 VS3003048 VS3003221 VS3003222 VS3003238 VS3003240 VS3003720 VS3004635 VS3004813 VS3005642 VS3005696 VS3005787 VS3005810 VS3005897 VS3005945 VS3006191 VS3006213 VS3006220 VSTEST VS3004712 VS3001895 VS3001930 VS3002806 VS3003633 VS3005851 VS3005915 VS3005982 VS3005985 VS3006071 VS3006185 VS3000560 VS3000567 VS3000575 VS3000602 VS3000607 VS3000617 VS3000635 VS3000906 VS3000907 VS3000909 VS3000911 VS3000912 VS3000939 VS3000943 VS3000946 VS3000952 VS3000957 VS3000960 VS3000964 VS3000969 VS3000970 VS3000973 VS3001019 VS3001032 VS3001035 VS3001036 VS3001048 VS3001059 VS3001060 VS3001061 VS3001065 VS3001066 VS3001067 VS3001068 VS3001070 VS3001074 VS3001075 VS3001076 VS3001078 VS3001084 VS3001087 VS3001094 VS3001103 VS3001109 VS3001111 VS3001113 VS3001117 VS3001630 VS3001631 VS3001632 VS3001637 VS3001642 VS3001644 VS3001650 VS3001651 VS3001652 VS3001659 VS3001660 VS3001663 VS3001666 VS3001667 VS3001672 VS3001673 VS3001675 VS3001677 VS3001681 VS3001683 VS3001685 VS3001686 VS3001689 VS3001692 VS3001693 VS3001695 VS3001696 VS3001701 VS3001702 VS3001705 VS3001707 VS3001710 VS3001711 VS3001714 VS3001725 VS3001743 VS3002752 VS3003168 VS3003239 VS3003277 VS3003671 VS3003677 VS3003678 VS3003694 VS3003735 VS3003808 VS3003839 VS3004022 VS3004031 VS3004037 VS3004055 VS3004068 VS3004070 VS3004071 VS3004072 VS3004075 VS3004084 VS3004085 VS3004086 VS3004110 VS3004192 VS3004214 VS3004243 VS3004247 VS3004281 VS3004282 VS3004318 VS3004356 VS3004371 VS3004420 VS3004460 VS3004467 VS3004468 VS3004476 VS3004490 VS3004493 VS3004515 VS3004526 VS3004531 VS3004581 VS3004583 VS3004585 VS3004590 VS3004593 VS3004594 VS3004597 VS3004599 VS3004608 VS3004618 VS3004620 VS3004627 VS3004645 VS3004658 VS3004668 VS3004670 VS3004675 VS3004677 VS3004690 VS3004756 VS3004943 VS3004965 VS3004979 VS3004986 VS3005762 VS3005804 VS3005816 VS3005817 VS3005822 VS3005908 VS3006190 VS3006215 VS3000577 VS3000585 VS3000910 VS3001164 VS3001218 VS3001227 VS3001228 VS3001229 VS3001234 VS3001274 VS3001277 VS3001280 VS3001297 VS3001304 VS3001311 VS3001321 VS3001324 VS3001327 VS3001344 VS3001348 VS3001368 VS3001375 VS3001379 VS3001385 VS3001399 VS3001400 VS3001406 VS3001411 VS3001690 VS3001744 VS3001854 VS3001861 VS3005888 VS3004971 VS3004051 VS3004151 VS3004295 VS3004474 VS3004492 VS3004940 VS3000526 VS3000528 VS3000530 VS3000531 VS3000534 VS3000535 VS3000539 VS3000543 VS3000544 VS3000546 VS3000547 VS3000549 VS3000550 VS3000552 VS3000556 VS3000557 VS3000558 VS3000559 VS3000561 VS3000564 VS3000566 VS3000571 VS3000572 VS3000578 VS3000579 VS3000581 VS3000582 VS3000595 VS3000605 VS3000606 VS3000611 VS3000613 VS3000616 VS3000621 VS3000622 VS3000625 VS3000626 VS3000627 VS3000629 VS3000630 VS3000639 VS3000642 VS3000643 VS3000647 VS3000908 VS3000913 VS3000914 VS3000915 VS3000916 VS3000917 VS3000919 VS3000920 VS3000925 VS3000926 VS3000928 VS3000932 VS3000933 VS3000936 VS3000937 VS3000940 VS3000941 VS3000942 VS3000947 VS3000950 VS3000954 VS3000956 VS3000958 VS3000959 VS3000966 VS3000967 VS3000968 VS3000971 VS3000972 VS3000974 VS3000975 VS3000976 VS3000977 VS3000979 VS3000980 VS3000981 VS3000982 VS3000983 VS3000984 VS3000985 VS3000989 VS3000990 VS3000991 VS3000992 VS3000993 VS3000994 VS3000996 VS3000997 VS3000999 VS3001000 VS3001001 VS3001002 VS3001003 VS3001004 VS3001005 VS3001006 VS3001008 VS3001009 VS3001011 VS3001020 VS3001022 VS3001024 VS3001028 VS3001030 VS3001031 VS3001033 VS3001034 VS3001037 VS3001038 VS3001040 VS3001042 VS3001043 VS3001044 VS3001045 VS3001047 VS3001049 VS3001050 VS3001051 VS3001056 VS3001058 VS3001062 VS3001063 VS3001064 VS3001069 VS3001071 VS3001072 VS3001073 VS3001077 VS3001079 VS3001080 VS3001081 VS3001082 VS3001083 VS3001085 VS3001086 VS3001088 VS3001089 VS3001090 VS3001091 VS3001092 VS3001093 VS3001095 VS3001096 VS3001097 VS3001098 VS3001100 VS3001101 VS3001104 VS3001106 VS3001107 VS3001108 VS3001114 VS3001115 VS3001116 VS3001118 VS3001120 VS3001121 VS3001125 VS3001126 VS3001128 VS3001129 VS3001130 VS3001131 VS3001132 VS3001133 VS3001134 VS3001135 VS3001136 VS3001141 VS3001142 VS3001144 VS3001145 VS3001146 VS3001147 VS3001148 VS3001151 VS3001152 VS3001159 VS3001162 VS3001165 VS3001167 VS3001168 VS3001169 VS3001170 VS3001171 VS3001172 VS3001173 VS3001174 VS3001175 VS3001176 VS3001177 VS3001178 VS3001179 VS3001182 VS3001183 VS3001185 VS3001186 VS3001187 VS3001190 VS3001191 VS3001192 VS3001193 VS3001195 VS3001196 VS3001198 VS3001199 VS3001200 VS3001202 VS3001203 VS3001204 VS3001205 VS3001206 VS3001207 VS3001208 VS3001209 VS3001211 VS3001212 VS3001213 VS3001216 VS3001217 VS3001219 VS3001220 VS3001223 VS3001226 VS3001237 VS3001241 VS3001242 VS3001246 VS3001247 VS3001248 VS3001249 VS3001250 VS3001258 VS3001260 VS3001261 VS3001262 VS3001263 VS3001264 VS3001267 VS3001270 VS3001275 VS3001278 VS3001279 VS3001284 VS3001285 VS3001286 VS3001287 VS3001288 VS3001289 VS3001291 VS3001292 VS3001294 VS3001295 VS3001296 VS3001300 VS3001305 VS3001312 VS3001316 VS3001319 VS3001320 VS3001322 VS3001323 VS3001326 VS3001328 VS3001329 VS3001330 VS3001332 VS3001333 VS3001334 VS3001338 VS3001339 VS3001340 VS3001341 VS3001342 VS3001343 VS3001345 VS3001346 VS3001349 VS3001350 VS3001351 VS3001352 VS3001356 VS3001357 VS3001358 VS3001359 VS3001361 VS3001362 VS3001364 VS3001367 VS3001369 VS3001370 VS3001371 VS3001372 VS3001373 VS3001374 VS3001376 VS3001377 VS3001380 VS3001381 VS3001382 VS3001386 VS3001387 VS3001389 VS3001391 VS3001392 VS3001393 VS3001394 VS3001396 VS3001397 VS3001398 VS3001401 VS3001403 VS3001405 VS3001407 VS3001408 VS3001409 VS3001419 VS3001421 VS3001423 VS3001424 VS3001634 VS3001635 VS3001636 VS3001638 VS3001639 VS3001640 VS3001641 VS3001645 VS3001646 VS3001647 VS3001648 VS3001654 VS3001655 VS3001656 VS3001657 VS3001658 VS3001664 VS3001668 VS3001669 VS3001671 VS3001678 VS3001679 VS3001680 VS3001682 VS3001684 VS3001688 VS3001697 VS3001699 VS3001700 VS3001704 VS3001706 VS3001708 VS3001709 VS3001713 VS3001721 VS3001722 VS3001723 VS3001724 VS3001726 VS3001727 VS3001728 VS3001729 VS3001730 VS3001731 VS3001732 VS3001734 VS3001735 VS3001737 VS3001739 VS3001742 VS3001746 VS3001747 VS3001749 VS3001750 VS3001752 VS3001753 VS3001754 VS3001755 VS3001756 VS3001758 VS3001759 VS3001760 VS3001763 VS3001764 VS3001765 VS3001767 VS3001769 VS3001771 VS3001772 VS3001773 VS3001775 VS3001776 VS3001777 VS3001778 VS3001779 VS3001780 VS3001782 VS3001783 VS3001784 VS3001785 VS3001787 VS3001789 VS3001790 VS3001792 VS3001793 VS3001796 VS3001798 VS3001801 VS3001802 VS3001807 VS3001812 VS3001813 VS3001816 VS3001817 VS3001818 VS3001820 VS3001821 VS3001822 VS3001823 VS3001825 VS3001826 VS3001827 VS3001828 VS3001829 VS3001830 VS3001831 VS3001832 VS3001834 VS3001835 VS3001836 VS3001837 VS3001838 VS3001839 VS3001840 VS3001841 VS3001843 VS3001845 VS3001847 VS3001848 VS3001849 VS3001850 VS3001851 VS3001852 VS3001853 VS3001855 VS3001856 VS3001857 VS3001859 VS3001860 VS3001862 VS3001863 VS3001864 VS3001866 VS3001869 VS3001870 VS3001871 VS3001872 VS3001873 VS3001874 VS3001875 VS3001876 VS3001877 VS3001878 VS3001879 VS3001881 VS3001882 VS3001884 VS3001885 VS3001886 VS3001887 VS3001888 VS3001889 VS3001891 VS3001892 VS3001893 VS3001894 VS3001896 VS3001897 VS3001899 VS3001901 VS3001902 VS3001903 VS3001904 VS3001905 VS3001906 VS3001907 VS3001908 VS3001909 VS3001910 VS3001913 VS3001915 VS3001916 VS3001917 VS3001918 VS3001919 VS3001920 VS3001921 VS3001922 VS3001924 VS3001927 VS3001928 VS3001929 VS3001931 VS3001932 VS3001933 VS3001934 VS3001935 VS3001936 VS3001937 VS3001938 VS3001940 VS3001941 VS3001942 VS3001943 VS3001944 VS3001945 VS3001946 VS3001947 VS3001948 VS3001949 VS3001950 VS3001951 VS3001952 VS3001954 VS3001960 VS3001961 VS3001962 VS3001963 VS3001964 VS3001965 VS3001966 VS3001967 VS3001968 VS3001969 VS3001971 VS3001973 VS3001974 VS3001975 VS3001977 VS3001978 VS3001984 VS3001985 VS3001987 VS3001989 VS3001991 VS3001992 VS3001994 VS3001996 VS3001997 VS3001999 VS3002000 VS3002001 VS3002002 VS3002004 VS3002007 VS3002009 VS3002010 VS3002011 VS3002012 VS3002013 VS3002014 VS3002015 VS3002017 VS3002018 VS3002020 VS3002022 VS3002025 VS3002029 VS3002030 VS3002031 VS3002035 VS3002036 VS3002037 VS3002038 VS3002039 VS3002040 VS3002041 VS3002043 VS3002044 VS3002046 VS3002047 VS3002048 VS3002049 VS3002050 VS3002051 VS3002052 VS3002053 VS3002054 VS3002055 VS3002057 VS3002058 VS3002059 VS3002060 VS3002061 VS3002062 VS3002064 VS3002065 VS3002067 VS3002068 VS3002075 VS3002076 VS3002081 VS3002084 VS3002085 VS3002089 VS3002090 VS3002091 VS3002092 VS3002094 VS3002095 VS3002098 VS3002099 VS3002100 VS3002101 VS3002102 VS3002103 VS3002104 VS3002105 VS3002106 VS3002109 VS3002110 VS3002111 VS3002112 VS3002113 VS3002114 VS3002115 VS3002117 VS3002118 VS3002119 VS3002120 VS3002121 VS3002123 VS3002124 VS3002125 VS3002127 VS3002128 VS3002129 VS3002131 VS3002132 VS3002133 VS3002135 VS3002137 VS3002138 VS3002140 VS3002141 VS3002142 VS3002143 VS3002144 VS3002146 VS3002147 VS3002149 VS3002151 VS3002152 VS3002153 VS3002154 VS3002157 VS3002158 VS3002159 VS3002160 VS3002161 VS3002162 VS3002165 VS3002166 VS3002167 VS3002168 VS3002170 VS3002171 VS3002172 VS3002173 VS3002174 VS3002175 VS3002177 VS3002178 VS3002180 VS3002182 VS3002184 VS3002186 VS3002187 VS3002188 VS3002189 VS3002190 VS3002191 VS3002192 VS3002193 VS3002194 VS3002196 VS3002197 VS3002198 VS3002200 VS3002201 VS3002202 VS3002203 VS3002204 VS3002208 VS3002209 VS3002211 VS3002213 VS3002214 VS3002216 VS3002218 VS3002220 VS3002222 VS3002224 VS3002225 VS3002226 VS3002227 VS3002228 VS3002230 VS3002231 VS3002233 VS3002234 VS3002235 VS3002236 VS3002237 VS3002238 VS3002239 VS3002240 VS3002242 VS3002244 VS3002245 VS3002247 VS3002248 VS3002250 VS3002621 VS3002622 VS3002623 VS3002624 VS3002626 VS3002627 VS3002628 VS3002629 VS3002630 VS3002631 VS3002633 VS3002634 VS3002635 VS3002636 VS3002637 VS3002638 VS3002639 VS3002640 VS3002641 VS3002642 VS3002643 VS3002644 VS3002645 VS3002646 VS3002647 VS3002648 VS3002649 VS3002651 VS3002653 VS3002654 VS3002655 VS3002656 VS3002657 VS3002658 VS3002659 VS3002660 VS3002661 VS3002663 VS3002664 VS3002665 VS3002666 VS3002669 VS3002670 VS3002671 VS3002672 VS3002673 VS3002674 VS3002675 VS3002676 VS3002677 VS3002679 VS3002680 VS3002682 VS3002683 VS3002684 VS3002685 VS3002687 VS3002688 VS3002689 VS3002692 VS3002693 VS3002694 VS3002695 VS3002698 VS3002699 VS3002700 VS3002701 VS3002702 VS3002703 VS3002704 VS3002705 VS3002706 VS3002707 VS3002709 VS3002710 VS3002711 VS3002712 VS3002713 VS3002714 VS3002715 VS3002716 VS3002717 VS3002719 VS3002720 VS3002722 VS3002723 VS3002724 VS3002726 VS3002727 VS3002728 VS3002729 VS3002730 VS3002732 VS3002734 VS3002735 VS3002736 VS3002738 VS3002739 VS3002740 VS3002742 VS3002743 VS3002744 VS3002745 VS3002746 VS3002747 VS3002750 VS3002751 VS3002753 VS3002754 VS3002756 VS3002758 VS3002761 VS3002762 VS3002764 VS3002765 VS3002766 VS3002767 VS3002770 VS3002772 VS3002773 VS3002774 VS3002775 VS3002777 VS3002778 VS3002779 VS3002780 VS3002781 VS3002782 VS3002784 VS3002786 VS3002787 VS3002788 VS3002789 VS3002790 VS3002794 VS3002795 VS3002796 VS3002798 VS3002799 VS3002800 VS3002801 VS3002802 VS3002803 VS3002805 VS3002807 VS3002808 VS3002809 VS3002811 VS3002812 VS3002814 VS3002815 VS3002816 VS3002818 VS3002819 VS3002820 VS3002821 VS3002822 VS3002825 VS3002826 VS3002827 VS3002828 VS3002829 VS3002831 VS3002833 VS3002836 VS3002838 VS3002840 VS3002841 VS3002842 VS3002843 VS3002844 VS3002845 VS3002848 VS3002849 VS3002851 VS3002852 VS3002854 VS3002855 VS3002856 VS3002857 VS3002859 VS3002860 VS3002862 VS3002863 VS3002864 VS3002865 VS3002866 VS3002867 VS3002868 VS3002869 VS3002870 VS3002871 VS3002873 VS3002874 VS3002875 VS3002876 VS3002877 VS3002878 VS3002879 VS3002880 VS3002882 VS3002883 VS3002885 VS3002886 VS3002887 VS3002888 VS3002889 VS3002890 VS3002891 VS3002892 VS3002893 VS3002894 VS3002895 VS3002896 VS3002897 VS3002898 VS3002899 VS3002900 VS3002901 VS3002902 VS3002903 VS3002904 VS3002905 VS3002906 VS3002907 VS3002910 VS3002912 VS3002913 VS3002914 VS3002915 VS3002916 VS3002917 VS3002918 VS3002919 VS3002920 VS3002922 VS3002923 VS3002924 VS3002925 VS3002927 VS3002928 VS3002929 VS3002930 VS3002931 VS3002932 VS3002933 VS3002935 VS3002936 VS3002937 VS3002938 VS3002940 VS3002941 VS3002942 VS3002943 VS3002944 VS3002945 VS3002946 VS3002948 VS3002949 VS3002950 VS3002951 VS3002953 VS3002954 VS3002957 VS3002958 VS3002959 VS3002962 VS3002964 VS3002966 VS3002967 VS3002968 VS3002969 VS3002970 VS3002971 VS3002975 VS3002976 VS3002977 VS3002978 VS3002979 VS3002981 VS3002982 VS3002983 VS3002984 VS3002985 VS3002987 VS3002988 VS3002989 VS3002990 VS3002991 VS3002992 VS3002993 VS3002995 VS3002996 VS3002997 VS3002998 VS3002999 VS3003000 VS3003001 VS3003002 VS3003003 VS3003005 VS3003006 VS3003007 VS3003008 VS3003009 VS3003010 VS3003011 VS3003012 VS3003014 VS3003015 VS3003016 VS3003017 VS3003018 VS3003019 VS3003020 VS3003021 VS3003022 VS3003024 VS3003025 VS3003026 VS3003027 VS3003028 VS3003029 VS3003030 VS3003031 VS3003032 VS3003033 VS3003034 VS3003035 VS3003036 VS3003037 VS3003038 VS3003039 VS3003040 VS3003041 VS3003042 VS3003043 VS3003044 VS3003045 VS3003046 VS3003047 VS3003049 VS3003050 VS3003051 VS3003052 VS3003053 VS3003055 VS3003056 VS3003057 VS3003058 VS3003059 VS3003060 VS3003061 VS3003062 VS3003063 VS3003064 VS3003066 VS3003067 VS3003068 VS3003069 VS3003070 VS3003071 VS3003072 VS3003073 VS3003074 VS3003075 VS3003076 VS3003077 VS3003078 VS3003079 VS3003080 VS3003081 VS3003082 VS3003083 VS3003084 VS3003085 VS3003086 VS3003087 VS3003088 VS3003089 VS3003090 VS3003091 VS3003092 VS3003093 VS3003094 VS3003095 VS3003096 VS3003097 VS3003098 VS3003099 VS3003101 VS3003103 VS3003104 VS3003106 VS3003107 VS3003110 VS3003112 VS3003113 VS3003114 VS3003115 VS3003116 VS3003118 VS3003119 VS3003120 VS3003121 VS3003124 VS3003125 VS3003128 VS3003129 VS3003130 VS3003132 VS3003134 VS3003135 VS3003136 VS3003137 VS3003138 VS3003139 VS3003140 VS3003141 VS3003142 VS3003144 VS3003145 VS3003146 VS3003147 VS3003148 VS3003149 VS3003153 VS3003154 VS3003157 VS3003158 VS3003161 VS3003162 VS3003165 VS3003166 VS3003170 VS3003171 VS3003173 VS3003174 VS3003178 VS3003179 VS3003180 VS3003181 VS3003182 VS3003183 VS3003185 VS3003187 VS3003188 VS3003189 VS3003190 VS3003195 VS3003198 VS3003201 VS3003203 VS3003204 VS3003205 VS3003207 VS3003208 VS3003211 VS3003212 VS3003213 VS3003215 VS3003217 VS3003223 VS3003224 VS3003227 VS3003228 VS3003230 VS3003231 VS3003232 VS3003233 VS3003234 VS3003237 VS3003241 VS3003242 VS3003243 VS3003244 VS3003245 VS3003247 VS3003248 VS3003249 VS3003250 VS3003251 VS3003253 VS3003256 VS3003257 VS3003259 VS3003261 VS3003262 VS3003263 VS3003264 VS3003266 VS3003268 VS3003269 VS3003271 VS3003274 VS3003275 VS3003279 VS3003280 VS3003281 VS3003283 VS3003284 VS3003285 VS3003286 VS3003290 VS3003296 VS3003299 VS3003302 VS3003303 VS3003304 VS3003305 VS3003306 VS3003307 VS3003308 VS3003309 VS3003311 VS3003312 VS3003313 VS3003314 VS3003315 VS3003316 VS3003318 VS3003319 VS3003320 VS3003322 VS3003324 VS3003325 VS3003326 VS3003328 VS3003330 VS3003332 VS3003333 VS3003334 VS3003335 VS3003338 VS3003340 VS3003341 VS3003342 VS3003345 VS3003346 VS3003347 VS3003348 VS3003349 VS3003350 VS3003353 VS3003354 VS3003355 VS3003356 VS3003357 VS3003358 VS3003362 VS3003363 VS3003365 VS3003366 VS3003367 VS3003368 VS3003369 VS3003370 VS3003371 VS3003372 VS3003373 VS3003374 VS3003375 VS3003377 VS3003379 VS3003380 VS3003381 VS3003382 VS3003383 VS3003385 VS3003386 VS3003388 VS3003389 VS3003390 VS3003392 VS3003393 VS3003394 VS3003395 VS3003396 VS3003397 VS3003398 VS3003399 VS3003400 VS3003401 VS3003402 VS3003403 VS3003405 VS3003406 VS3003407 VS3003408 VS3003409 VS3003410 VS3003411 VS3003413 VS3003414 VS3003415 VS3003416 VS3003418 VS3003419 VS3003420 VS3003422 VS3003423 VS3003424 VS3003425 VS3003428 VS3003429 VS3003430 VS3003431 VS3003432 VS3003433 VS3003434 VS3003435 VS3003436 VS3003437 VS3003438 VS3003622 VS3003624 VS3003632 VS3003636 VS3003639 VS3003641 VS3003642 VS3003645 VS3003646 VS3003647 VS3003648 VS3003660 VS3003662 VS3003665 VS3003667 VS3003669 VS3003675 VS3003676 VS3003679 VS3003680 VS3003681 VS3003682 VS3003683 VS3003684 VS3003685 VS3003687 VS3003689 VS3003690 VS3003691 VS3003692 VS3003693 VS3003695 VS3003697 VS3003699 VS3003700 VS3003701 VS3003703 VS3003705 VS3003706 VS3003708 VS3003710 VS3003711 VS3003714 VS3003715 VS3003716 VS3003717 VS3003718 VS3003719 VS3003721 VS3003722 VS3003724 VS3003727 VS3003732 VS3003733 VS3003736 VS3003738 VS3003739 VS3003744 VS3003745 VS3003748 VS3003749 VS3003753 VS3003755 VS3003756 VS3003758 VS3003760 VS3003761 VS3003762 VS3003764 VS3003765 VS3003767 VS3003768 VS3003769 VS3003773 VS3003775 VS3003776 VS3003777 VS3003778 VS3003779 VS3003780 VS3003781 VS3003782 VS3003783 VS3003786 VS3003789 VS3003790 VS3003791 VS3003792 VS3003794 VS3003795 VS3003796 VS3003797 VS3003798 VS3003800 VS3003801 VS3003804 VS3003805 VS3003809 VS3003810 VS3003812 VS3003815 VS3003816 VS3003818 VS3003821 VS3003822 VS3003823 VS3003824 VS3003825 VS3003826 VS3003827 VS3003828 VS3003832 VS3003834 VS3003835 VS3003837 VS3003838 VS3003842 VS3003845 VS3003847 VS3003850 VS3003856 VS3003857 VS3003859 VS3003860 VS3003861 VS3003862 VS3003863 VS3003864 VS3003865 VS3003867 VS3003868 VS3003869 VS3003870 VS3003872 VS3003873 VS3003874 VS3003875 VS3003877 VS3003878 VS3003879 VS3003880 VS3003881 VS3003882 VS3003883 VS3003885 VS3003886 VS3003888 VS3003889 VS3003891 VS3003892 VS3003893 VS3003894 VS3003895 VS3003896 VS3003897 VS3003898 VS3003899 VS3003900 VS3003901 VS3003902 VS3003903 VS3003904 VS3003905 VS3003906 VS3003907 VS3003908 VS3003909 VS3003910 VS3003911 VS3003912 VS3003914 VS3003915 VS3003916 VS3003917 VS3003918 VS3003919 VS3003920 VS3003921 VS3003922 VS3003923 VS3003924 VS3003925 VS3003926 VS3003927 VS3003929 VS3003930 VS3003931 VS3003932 VS3003934 VS3003935 VS3003936 VS3003937 VS3003938 VS3003939 VS3003940 VS3003941 VS3003942 VS3003943 VS3003944 VS3003945 VS3003946 VS3003947 VS3003948 VS3003949 VS3003950 VS3003951 VS3003952 VS3003953 VS3003954 VS3003955 VS3003956 VS3003957 VS3003958 VS3003959 VS3003960 VS3003961 VS3003962 VS3003963 VS3003964 VS3003965 VS3003966 VS3003967 VS3003968 VS3003969 VS3003970 VS3003971 VS3003972 VS3003973 VS3003974 VS3003975 VS3003976 VS3003977 VS3003978 VS3003979 VS3003980 VS3003981 VS3003982 VS3003983 VS3003984 VS3003985 VS3003986 VS3003988 VS3003989 VS3003990 VS3003991 VS3003992 VS3003993 VS3003994 VS3003995 VS3003997 VS3003998 VS3003999 VS3004000 VS3004001 VS3004002 VS3004003 VS3004004 VS3004005 VS3004006 VS3004007 VS3004008 VS3004009 VS3004010 VS3004011 VS3004012 VS3004013 VS3004014 VS3004015 VS3004016 VS3004017 VS3004018 VS3004019 VS3004020 VS3004021 VS3004023 VS3004024 VS3004025 VS3004026 VS3004027 VS3004028 VS3004029 VS3004030 VS3004032 VS3004033 VS3004034 VS3004035 VS3004036 VS3004039 VS3004040 VS3004041 VS3004042 VS3004043 VS3004044 VS3004045 VS3004046 VS3004047 VS3004048 VS3004049 VS3004050 VS3004052 VS3004053 VS3004054 VS3004056 VS3004058 VS3004059 VS3004060 VS3004061 VS3004062 VS3004063 VS3004064 VS3004065 VS3004066 VS3004067 VS3004069 VS3004073 VS3004074 VS3004076 VS3004077 VS3004078 VS3004079 VS3004080 VS3004081 VS3004082 VS3004083 VS3004087 VS3004088 VS3004089 VS3004090 VS3004091 VS3004092 VS3004093 VS3004094 VS3004095 VS3004096 VS3004097 VS3004098 VS3004099 VS3004100 VS3004101 VS3004107 VS3004108 VS3004109 VS3004111 VS3004112 VS3004113 VS3004114 VS3004115 VS3004116 VS3004117 VS3004118 VS3004119 VS3004120 VS3004121 VS3004122 VS3004123 VS3004124 VS3004125 VS3004126 VS3004127 VS3004128 VS3004129 VS3004130 VS3004131 VS3004132 VS3004133 VS3004134 VS3004135 VS3004136 VS3004137 VS3004138 VS3004139 VS3004140 VS3004141 VS3004142 VS3004143 VS3004144 VS3004145 VS3004146 VS3004147 VS3004149 VS3004150 VS3004152 VS3004153 VS3004154 VS3004155 VS3004156 VS3004157 VS3004158 VS3004160 VS3004161 VS3004162 VS3004163 VS3004164 VS3004165 VS3004166 VS3004167 VS3004168 VS3004169 VS3004171 VS3004172 VS3004173 VS3004174 VS3004175 VS3004176 VS3004177 VS3004179 VS3004180 VS3004181 VS3004182 VS3004183 VS3004185 VS3004186 VS3004187 VS3004188 VS3004190 VS3004191 VS3004193 VS3004194 VS3004195 VS3004196 VS3004197 VS3004198 VS3004199 VS3004200 VS3004201 VS3004202 VS3004203 VS3004204 VS3004205 VS3004206 VS3004207 VS3004208 VS3004209 VS3004210 VS3004211 VS3004212 VS3004213 VS3004216 VS3004217 VS3004218 VS3004219 VS3004220 VS3004221 VS3004222 VS3004223 VS3004224 VS3004225 VS3004226 VS3004227 VS3004228 VS3004229 VS3004231 VS3004233 VS3004234 VS3004235 VS3004236 VS3004237 VS3004238 VS3004239 VS3004240 VS3004241 VS3004242 VS3004244 VS3004245 VS3004248 VS3004249 VS3004250 VS3004251 VS3004252 VS3004253 VS3004255 VS3004256 VS3004257 VS3004258 VS3004259 VS3004260 VS3004262 VS3004263 VS3004264 VS3004265 VS3004266 VS3004267 VS3004269 VS3004271 VS3004273 VS3004274 VS3004275 VS3004276 VS3004277 VS3004278 VS3004280 VS3004283 VS3004284 VS3004285 VS3004286 VS3004287 VS3004288 VS3004289 VS3004290 VS3004291 VS3004292 VS3004293 VS3004296 VS3004297 VS3004298 VS3004299 VS3004300 VS3004301 VS3004302 VS3004303 VS3004304 VS3004305 VS3004306 VS3004307 VS3004308 VS3004309 VS3004310 VS3004311 VS3004313 VS3004315 VS3004316 VS3004317 VS3004319 VS3004320 VS3004321 VS3004322 VS3004323 VS3004324 VS3004325 VS3004326 VS3004327 VS3004328 VS3004329 VS3004330 VS3004332 VS3004334 VS3004335 VS3004337 VS3004338 VS3004339 VS3004340 VS3004341 VS3004343 VS3004344 VS3004345 VS3004346 VS3004347 VS3004349 VS3004350 VS3004351 VS3004352 VS3004353 VS3004354 VS3004355 VS3004357 VS3004358 VS3004359 VS3004360 VS3004361 VS3004362 VS3004363 VS3004364 VS3004365 VS3004366 VS3004367 VS3004368 VS3004369 VS3004370 VS3004372 VS3004373 VS3004374 VS3004375 VS3004376 VS3004377 VS3004378 VS3004379 VS3004380 VS3004381 VS3004382 VS3004383 VS3004384 VS3004385 VS3004386 VS3004387 VS3004388 VS3004389 VS3004391 VS3004392 VS3004393 VS3004394 VS3004395 VS3004396 VS3004397 VS3004398 VS3004399 VS3004400 VS3004401 VS3004402 VS3004403 VS3004404 VS3004405 VS3004406 VS3004407 VS3004408 VS3004409 VS3004410 VS3004411 VS3004412 VS3004413 VS3004414 VS3004415 VS3004416 VS3004417 VS3004418 VS3004421 VS3004422 VS3004423 VS3004424 VS3004425 VS3004426 VS3004432 VS3004433 VS3004434 VS3004435 VS3004436 VS3004437 VS3004438 VS3004439 VS3004440 VS3004441 VS3004442 VS3004443 VS3004444 VS3004445 VS3004446 VS3004447 VS3004448 VS3004449 VS3004450 VS3004451 VS3004452 VS3004453 VS3004454 VS3004455 VS3004456 VS3004457 VS3004458 VS3004459 VS3004461 VS3004462 VS3004463 VS3004465 VS3004466 VS3004469 VS3004470 VS3004471 VS3004472 VS3004473 VS3004475 VS3004477 VS3004478 VS3004479 VS3004480 VS3004481 VS3004482 VS3004483 VS3004484 VS3004485 VS3004486 VS3004487 VS3004488 VS3004491 VS3004494 VS3004495 VS3004496 VS3004497 VS3004498 VS3004499 VS3004500 VS3004501 VS3004502 VS3004503 VS3004504 VS3004505 VS3004506 VS3004507 VS3004508 VS3004509 VS3004510 VS3004511 VS3004512 VS3004513 VS3004514 VS3004516 VS3004517 VS3004518 VS3004519 VS3004520 VS3004521 VS3004522 VS3004523 VS3004524 VS3004527 VS3004528 VS3004529 VS3004532 VS3004533 VS3004534 VS3004536 VS3004537 VS3004539 VS3004541 VS3004543 VS3004544 VS3004546 VS3004547 VS3004549 VS3004551 VS3004552 VS3004553 VS3004554 VS3004556 VS3004557 VS3004558 VS3004559 VS3004560 VS3004562 VS3004564 VS3004566 VS3004567 VS3004569 VS3004570 VS3004571 VS3004572 VS3004573 VS3004574 VS3004575 VS3004576 VS3004578 VS3004579 VS3004580 VS3004582 VS3004584 VS3004586 VS3004587 VS3004588 VS3004589 VS3004591 VS3004592 VS3004595 VS3004596 VS3004598 VS3004600 VS3004602 VS3004603 VS3004604 VS3004605 VS3004606 VS3004610 VS3004611 VS3004612 VS3004613 VS3004614 VS3004615 VS3004617 VS3004619 VS3004621 VS3004622 VS3004623 VS3004625 VS3004626 VS3004628 VS3004630 VS3004631 VS3004632 VS3004633 VS3004634 VS3004636 VS3004637 VS3004639 VS3004640 VS3004641 VS3004643 VS3004644 VS3004649 VS3004650 VS3004651 VS3004652 VS3004653 VS3004654 VS3004655 VS3004656 VS3004657 VS3004659 VS3004660 VS3004661 VS3004662 VS3004663 VS3004664 VS3004665 VS3004666 VS3004667 VS3004669 VS3004671 VS3004672 VS3004676 VS3004678 VS3004679 VS3004680 VS3004681 VS3004682 VS3004683 VS3004686 VS3004687 VS3004688 VS3004692 VS3004693 VS3004695 VS3004696 VS3004698 VS3004699 VS3004700 VS3004701 VS3004702 VS3004703 VS3004704 VS3004705 VS3004706 VS3004707 VS3004708 VS3004709 VS3004710 VS3004711 VS3004713 VS3004714 VS3004715 VS3004716 VS3004717 VS3004718 VS3004719 VS3004720 VS3004721 VS3004722 VS3004724 VS3004726 VS3004727 VS3004730 VS3004731 VS3004732 VS3004733 VS3004734 VS3004735 VS3004736 VS3004737 VS3004738 VS3004739 VS3004740 VS3004741 VS3004742 VS3004743 VS3004745 VS3004746 VS3004747 VS3004748 VS3004749 VS3004750 VS3004751 VS3004752 VS3004754 VS3004755 VS3004757 VS3004758 VS3004759 VS3004760 VS3004761 VS3004762 VS3004763 VS3004764 VS3004765 VS3004766 VS3004767 VS3004768 VS3004769 VS3004770 VS3004772 VS3004773 VS3004774 VS3004775 VS3004776 VS3004777 VS3004778 VS3004779 VS3004780 VS3004781 VS3004782 VS3004784 VS3004786 VS3004787 VS3004788 VS3004789 VS3004790 VS3004791 VS3004792 VS3004793 VS3004794 VS3004795 VS3004796 VS3004797 VS3004799 VS3004800 VS3004801 VS3004802 VS3004804 VS3004806 VS3004807 VS3004808 VS3004809 VS3004810 VS3004812 VS3004814 VS3004815 VS3004816 VS3004817 VS3004818 VS3004820 VS3004821 VS3004822 VS3004823 VS3004824 VS3004826 VS3004827 VS3004828 VS3004834 VS3004835 VS3004836 VS3004838 VS3004839 VS3004840 VS3004842 VS3004843 VS3004844 VS3004845 VS3004846 VS3004847 VS3004848 VS3004849 VS3004850 VS3004852 VS3004853 VS3004854 VS3004855 VS3004856 VS3004857 VS3004858 VS3004859 VS3004860 VS3004861 VS3004862 VS3004863 VS3004864 VS3004865 VS3004866 VS3004867 VS3004868 VS3004869 VS3004871 VS3004872 VS3004873 VS3004875 VS3004876 VS3004877 VS3004878 VS3004879 VS3004880 VS3004881 VS3004882 VS3004883 VS3004884 VS3004885 VS3004886 VS3004887 VS3004888 VS3004889 VS3004890 VS3004891 VS3004893 VS3004894 VS3004895 VS3004897 VS3004899 VS3004900 VS3004901 VS3004902 VS3004903 VS3004904 VS3004906 VS3004907 VS3004908 VS3004909 VS3004910 VS3004911 VS3004912 VS3004913 VS3004914 VS3004915 VS3004917 VS3004918 VS3004919 VS3004920 VS3004921 VS3004922 VS3004923 VS3004926 VS3004927 VS3004928 VS3004929 VS3004930 VS3004931 VS3004932 VS3004933 VS3004934 VS3004935 VS3004936 VS3004937 VS3004938 VS3004939 VS3004944 VS3004947 VS3004948 VS3004949 VS3004950 VS3004951 VS3004952 VS3004953 VS3004954 VS3004955 VS3004957 VS3004958 VS3004959 VS3004960 VS3004961 VS3004962 VS3004963 VS3004964 VS3004966 VS3004967 VS3004968 VS3004969 VS3004970 VS3004972 VS3004973 VS3004975 VS3004977 VS3004978 VS3004981 VS3004982 VS3004984 VS3004985 VS3004988 VS3004989 VS3004990 VS3004991 VS3004992 VS3004993 VS3004994 VS3004995 VS3004996 VS3004997 VS3004998 VS3004999 VS3005000 VS3005001 VS3005002 VS3005004 VS3005006 VS3005007 VS3005008 VS3005009 VS3005010 VS3005011 VS3005012 VS3005014 VS3005015 VS3005016 VS3005017 VS3005018 VS3005020 VS3005021 VS3005022 VS3005023 VS3005026 VS3005027 VS3005028 VS3005029 VS3005030 VS3005031 VS3005032 VS3005033 VS3005034 VS3005035 VS3005038 VS3005039 VS3005040 VS3005042 VS3005043 VS3005622 VS3005623 VS3005624 VS3005625 VS3005628 VS3005629 VS3005630 VS3005632 VS3005633 VS3005634 VS3005635 VS3005636 VS3005637 VS3005638 VS3005639 VS3005640 VS3005641 VS3005643 VS3005644 VS3005645 VS3005646 VS3005647 VS3005648 VS3005649 VS3005650 VS3005651 VS3005652 VS3005654 VS3005655 VS3005657 VS3005658 VS3005660 VS3005662 VS3005663 VS3005664 VS3005666 VS3005667 VS3005670 VS3005672 VS3005673 VS3005674 VS3005677 VS3005678 VS3005679 VS3005680 VS3005682 VS3005683 VS3005684 VS3005685 VS3005686 VS3005687 VS3005688 VS3005692 VS3005693 VS3005694 VS3005697 VS3005698 VS3005699 VS3005700 VS3005701 VS3005704 VS3005705 VS3005706 VS3005707 VS3005708 VS3005709 VS3005710 VS3005711 VS3005712 VS3005713 VS3005714 VS3005715 VS3005716 VS3005717 VS3005718 VS3005720 VS3005721 VS3005722 VS3005723 VS3005724 VS3005725 VS3005726 VS3005727 VS3005728 VS3005729 VS3005730 VS3005732 VS3005733 VS3005734 VS3005735 VS3005736 VS3005737 VS3005738 VS3005740 VS3005741 VS3005742 VS3005743 VS3005744 VS3005745 VS3005746 VS3005747 VS3005748 VS3005749 VS3005750 VS3005751 VS3005753 VS3005754 VS3005755 VS3005756 VS3005757 VS3005758 VS3005759 VS3005760 VS3005761 VS3005763 VS3005764 VS3005765 VS3005766 VS3005767 VS3005768 VS3005769 VS3005770 VS3005771 VS3005772 VS3005773 VS3005774 VS3005775 VS3005776 VS3005777 VS3005780 VS3005781 VS3005782 VS3005783 VS3005785 VS3005786 VS3005788 VS3005790 VS3005791 VS3005792 VS3005793 VS3005794 VS3005795 VS3005796 VS3005798 VS3005799 VS3005803 VS3005805 VS3005807 VS3005808 VS3005809 VS3005812 VS3005818 VS3005821 VS3005825 VS3005826 VS3005827 VS3005829 VS3005830 VS3005831 VS3005832 VS3005833 VS3005835 VS3005836 VS3005837 VS3005838 VS3005839 VS3005840 VS3005841 VS3005842 VS3005843 VS3005844 VS3005845 VS3005846 VS3005847 VS3005848 VS3005849 VS3005850 VS3005853 VS3005854 VS3005855 VS3005857 VS3005859 VS3005860 VS3005861 VS3005862 VS3005863 VS3005864 VS3005866 VS3005867 VS3005870 VS3005871 VS3005872 VS3005873 VS3005874 VS3005875 VS3005876 VS3005877 VS3005878 VS3005879 VS3005880 VS3005881 VS3005882 VS3005883 VS3005884 VS3005885 VS3005887 VS3005889 VS3005890 VS3005891 VS3005892 VS3005895 VS3005896 VS3005898 VS3005900 VS3005901 VS3005903 VS3005904 VS3005906 VS3005910 VS3005911 VS3005912 VS3005913 VS3005916 VS3005917 VS3005918 VS3005919 VS3005920 VS3005921 VS3005922 VS3005923 VS3005924 VS3005926 VS3005927 VS3005928 VS3005929 VS3005930 VS3005931 VS3005932 VS3005933 VS3005934 VS3005935 VS3005936 VS3005937 VS3005938 VS3005939 VS3005940 VS3005941 VS3005943 VS3005944 VS3005946 VS3005947 VS3005948 VS3005949 VS3005950 VS3005951 VS3005952 VS3005953 VS3005954 VS3005955 VS3005956 VS3005957 VS3005958 VS3005959 VS3005960 VS3005961 VS3005962 VS3005963 VS3005964 VS3005965 VS3005966 VS3005967 VS3005968 VS3005969 VS3005970 VS3005971 VS3005972 VS3005973 VS3005974 VS3005975 VS3005976 VS3005977 VS3005978 VS3005979 VS3005980 VS3005981 VS3005983 VS3005986 VS3005987 VS3005988 VS3005989 VS3005990 VS3005991 VS3005992 VS3005993 VS3005994 VS3005995 VS3005996 VS3005997 VS3005998 VS3005999 VS3006000 VS3006001 VS3006002 VS3006003 VS3006004 VS3006005 VS3006006 VS3006007 VS3006008 VS3006009 VS3006010 VS3006011 VS3006012 VS3006013 VS3006015 VS3006016 VS3006017 VS3006018 VS3006019 VS3006020 VS3006021 VS3006022 VS3006023 VS3006024 VS3006025 VS3006026 VS3006027 VS3006028 VS3006029 VS3006030 VS3006031 VS3006032 VS3006033 VS3006034 VS3006035 VS3006036 VS3006037 VS3006038 VS3006039 VS3006040 VS3006041 VS3006042 VS3006043 VS3006044 VS3006045 VS3006046 VS3006047 VS3006048 VS3006049 VS3006050 VS3006051 VS3006052 VS3006053 VS3006054 VS3006055 VS3006056 VS3006057 VS3006058 VS3006059 VS3006060 VS3006061 VS3006062 VS3006065 VS3006066 VS3006067 VS3006068 VS3006069 VS3006070 VS3006072 VS3006073 VS3006074 VS3006075 VS3006076 VS3006077 VS3006078 VS3006079 VS3006080 VS3006081 VS3006082 VS3006083 VS3006084 VS3006085 VS3006086 VS3006087 VS3006088 VS3006089 VS3006090 VS3006092 VS3006093 VS3006094 VS3006095 VS3006096 VS3006097 VS3006099 VS3006100 VS3006101 VS3006102 VS3006103 VS3006105 VS3006106 VS3006107 VS3006108 VS3006109 VS3006110 VS3006111 VS3006112 VS3006113 VS3006114 VS3006115 VS3006116 VS3006117 VS3006118 VS3006119 VS3006120 VS3006121 VS3006123 VS3006124 VS3006125 VS3006126 VS3006127 VS3006128 VS3006129 VS3006130 VS3006131 VS3006132 VS3006133 VS3006134 VS3006135 VS3006136 VS3006137 VS3006138 VS3006139 VS3006140 VS3006141 VS3006142 VS3006143 VS3006144 VS3006145 VS3006146 VS3006147 VS3006148 VS3006149 VS3006150 VS3006151 VS3006152 VS3006153 VS3006154 VS3006155 VS3006156 VS3006157 VS3006158 VS3006159 VS3006160 VS3006161 VS3006162 VS3006163 VS3006164 VS3006165 VS3006166 VS3006167 VS3006168 VS3006169 VS3006170 VS3006171 VS3006172 VS3006173 VS3006174 VS3006175 VS3006176 VS3006177 VS3006178 VS3006179 VS3006180 VS3006182 VS3006183 VS3006184 VS3006186 VS3006187 VS3006188 VS3006189 VS3006196 VS3006197 VS3006198 VS3006201 VS3006205 VS3006207 VS3006208 VS3006210 VS3006214 VS3006216 VS3006223 VS3006224 VS3006225 VS3006226 VS3006227 VS3006228 VS3006229 VS3006230 VS3006231 VS3006232 VS3006233 VS3006234 VS3006235 VS3006236 VS3006237 VS3006238 VS3006239 VS3006240 VS3006241 VS3006242 VS3006243 VS3006244 VS3006245 VS3006246 VS3006247 VS3006248 VS3006249 VS3006250 VS3006251 VS3006252 VS3006253 VS3006254 VS3006255 VS3006257 VS3006258 VS3006259 VS3006260 VS3006261 VS3006262 VS3006263 VS3006264 VS3006265 VS3006267 VS3006268 VS3006269 VS3006270 VS3006271 VS3006272 VS3006273 VS3006274 VS3006275 VS3006276 VS3006278 VS3006279 VS3006280 VS3006281 VS3006282 VS3006283 VS3006285 VS3006286 VS3006288 VS3006289 VS3006290 VS3006291 VS3006292 VS3006293 VS3006294 VS3006295 VS3006296 VS3006297 VS3006298 VS3006299 VS3006300 VS3006301 VS3006302 VS3006303 VS3006304 VS3006305 VS3006306 VS3006307 VS3006308 VS3006309 VS3006310 VS3006312 VS3006314 VS3006315 VS3006316 VS3006318 VS3006319 VS3006321 VS3006322 VS3006323 VS3006324 VS3006325 VS3006326 VS3006327 VS3006329 VS3006330 VS3006331 VS3006332 VS3006334 VS3006335 VS3006336 VS3006338 VS3006339 VS3006343 VS3006344 VS3006345 VS3006352 VS3006354 VS3006355 VS3006356 VS3006357 VS3006360 VS3006361 VS3006362 VS3006363 VS3006366 VS3006368 VS3006370 VS3006371 VS3006372 VS3006374 VS3006375 VS3006376 VS3006377 VS3006378 VS3006379 VS3006380 VS3006383 VS3006388 VS3006409 VS3010003 VS3010004 VS3010005 VS3001995 VS3002217 VS3002219 VS3002755 VS3000608 VS3000953 VS3001039 VS3001733 VS3003111 VS3003329 VS3003331 VS3003871 VS3004057 VS3006192 VS3006194 VS3006195 VS3006206 VS3006211 VS3006219 VS3006333 VS3006349 
FEI Number 2518422
Recalling Firm/
Manufacturer
Respironics California Inc
2271 Cosmos Ct
Carlsbad CA 92011-1517
For Additional Information Contact
760-918-7300
Manufacturer Reason
for Recall
If the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA) and short to ground. This may prevent the ventilator to switch to battery power if AC power is disrupted.
FDA Determined
Cause 2
Device Design
ActionPhilips Healthcare sent an Urgent - Field Safety Notice to all affected consignees to inform them that Respironics is recalling the Esprit V1000 and Esprit/V200 Conversion ventilators because if the power supply fan mounting screws are installed incorrectly, the ends of the screws may touch the power supply printed circuit board assembly (PCBA) and short to ground. This may prevent the ventilator to switch to battery power if AC power is disrupted. A high urgency audible alarm will sound alerting medical personnel of the loss of power. In the absence of AC power, failure of battery power will cause loss of ventilator support, which may result in hypercarbia or hypoxemia. The potential issue affects ventilators manufactured prior to 12/31/04 due to design of the fan/shroud assembly. Esprit V1000 and Esprit/V200 Conversion Ventilators manufactured after that affected date are not affected. Customers with concerns were instructed to contact a Philips Respironics representative at (800) 722-9377. For questions regarding this recall call 760-918-7300.
Quantity in Commerce3,943 units total (3,071 units in US)
DistributionWorldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Ecuador, Egypt, Greece, Hong Kong, India, Italy, Korea, Lebanon, Malaysia, Mexico, Pakistan, Panama, Peru, Philippines, Saudi Arabia, Singapore, South Africa, Spain, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Vietnam, and Yemen.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.