| Date Initiated by Firm | November 10, 2014 |
| Date Posted | December 12, 2014 |
| Recall Status1 |
Terminated 3 on March 18, 2015 |
| Recall Number | Z-0588-2015 |
| Recall Event ID |
69778 |
| 510(K)Number | K120676 |
| Product Classification |
System, radiation therapy, charged-particle, medical - Product Code LHN
|
| Product | MEVION S250, used for proton radiation therapy. |
| Code Information |
Serial Number: S250-0001 |
| FEI Number |
3007087027
|
Recalling Firm/ Manufacturer |
Mevion Medical Systems, Inc.
300 Foster St
Littleton
MA
01460-2017
|
| For Additional Information Contact | Thomas H. Faris 978-540-1500 |
Manufacturer Reason for Recall | Software defect that causes an incorrect dose compensation function to be applied to the internal dose ionization chamber when either pressure or temperature sensor malfunction. This could result in an incorrect dose delivery of no more than 5%. |
FDA Determined Cause 2 | Software design |
| Action | Mevion Medical Systems sent an Important Safety Notice letter to the affected customer. The letter identified the affected product, problem and actions to be taken. A software update addressing the issue will be provided to the customer on or about December 5, 2014.
For questions or concerns, email lbouchet@mevion.com
or call 978-540-1555. |
| Quantity in Commerce | 1 |
| Distribution | US Distribution in the state of MO |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = LHN
|