 Class 2 Device Recall lnvacare Precise Rx Pediatric Flowmetersee related information
Date Initiated by FirmJune 21, 2016
Date PostedAugust 08, 2016
Recall Status1 Terminated 3 on February 07, 2017
Recall NumberZ-2364-2016
Recall Event ID 74581
510(K)NumberK914488 
Product Classification Conserver, oxygen - Product Code NFB
Productlnvacare Precise Rx Pediatric Flowmeter Model Number: IRCPF16. Packed one per carton. The lnvacare Precise Rx Pediatric Flowmeter adapts to lnvacare Concentrators for low-flow usage.The gauge has a scale printed on the front side and the back side. The front scale is the primary source for monitoring oxygen flow. The rear scale is an aide for monitoring oxygen flow.
Code Information Serial numbers for Model IRCPF16:   16BF023724, 16BF023729, 16BF023734, 16BF023739, 16BF023744,16BF023720, 16BF023725, 16BF023730, 16BF023735, 16BF023740, 16BF023745,16BF023721, 16BF023726, 16BF023731, 16BF023736, 16BF023741, 16BF023746,16BF023722, 16BF023727, 16BF023732, 16BF023737, 16BF023742, 16BF023747,16BF023723, 16BF023728, 16BF023733, 16BF023738, 16BF023743, 16BF023748,16BF023749, 16BF030593, 16CF015205, 16CF019539, 16CF026064, 16CF030100,16BF023750, 16BF030594, 16CF015206, 16CF019540, 16CF026065, 16CF030101,16BF023751, 16CF016169, 16CF015207, 16CF019541, 16CF026066, 16CF030102,16BF023752, 16CF016170, 16CF015208, 16CF019542, 16CF026067, 16CF030103,16BF023753, 16CF016171, 16CF015209, 16CF019543, 16CF026068,16CF030104,16BF023754, 16CF016172, 16CF015210, 16CF019544, 16CF026069, 16CF030105,16BF023755, 16CF016173, 16CFOI5211, 16CF019545, 16CF026070, 16CF030106,16BF025606, 16CF016174, 16CF015212, 16CF019546, 16CF026071, 16CF030107,16BF025607, 16CF016175, 16CFOI5213, 16CF019547, 16CF026072, 16CF030108,16BF025608, 16CF016176, 16CF015214, 16CF019548, I6CF026073, 16CF030109,16BF025609, 16CF016177, 16CF015215, 16CF019549, l6CF026074, 16CF030110,16BF025610, 16CF016178, 16CF015216, l6CF019550, l6CF026075, 16CF0301ll,16BF025611, 16CF016179, 16CF018612, 16CF019551, 16CF026076, 16CF030112,16BF025612, 16CF016180, 16CF018613, 16CF019552, l6CF026077, 16CF030113,16BF025613, 16CF016181, 16CF018614, 16CF019553, 16CF026078, 16CF030114,16BF025614, 16CF016182, 16CF018615, 16CF019554, 16CF026079, 16CF030115,16BF025615, 16CF016183, 16CF018616, 16CF019555, 16CF026080, l6CF030116,16BF025616, 16CF016184, 16CF018617, 16CF019556, 16CF026081, 16CF030117,16BF025617, 16CF016185, 16CF018618, 16CF019557, 16CF026082, 16CF030118,16BF025618, 16CF016186, 16CF018619, 16CF019558, 16CF026083, 16CF030119,16BF025619, 16CF016187, 16CFOI8620, 16CF019559, 16CF026084, 16CF030120,16BF025620, 16CF016188, 16CF018621, 16CF019560, 16CF026085, 16CF030121,16BF025621, 16CF016189, 16CF018622, 16CF019561, 16CF026086, 16CF030122,16BF025622, 16CF016190, 16CF018623, 16CF019562, 16CF026087, 16CF030123,16BF025623, 16CF016191, 16CF018624, 16CF019563, 16CF026088, 16CF030124,16BF025624, 16CF016192, 16CF018625, 16CF019564, 16CF026089, 16CF032012,16BF025625, 16CF016193, 16CF018626, 16CF019565, 16CF026090, 16CF032013,16BF025626, 16CF016194, 16CF018627, 16CF019566, 16CF026091, 16CF032014,16BF025627, 16CF016195, 16CF018628, 16CF023329, 16CF026092, 16CF032015,16BF025628, 16CF016196, 16CF018629, 16CF023330, 16CF027673, 16CF032016,16BF025629, 16CF016197, 16CF018630, 16CF023331, 16CF027674, 16CF032017,16BF025630, 16CF016198, 16CF018631, 16CF023332, 16CF027675, 16CF032018,16BF025631, 16CF015181, 16CF018632, 16CF023333, 16CF027676, 16CF032019,16BF025632, 16CF015182, 16CF018633, 16CF023334, 16CF027677, 16CF032020,16BF025633, 16CF015183, 16CF018634, 16CF023335, 16CF027678, 16CF032021,16BF025634, 16CF015184, 16CF018635, 16CF023336, 16CF027679, 16CF032022,16BF025635, 16CF015185, 16CF018636, 16CF023337, 16CF027680, 16CF032023,16BF025636, 16CF015186, 16CF018637, 16CF023338, 16CF027681, 16CF032024,16BF025637, 16CF015187, 16CF018638, 16CF023339, 16CF027682, 16CF032025,16BF025638, 16CF015188, 16CF018639, 16CF023340, 16CF027683, 16CF032026,16BF025639, 16CF015189, 16CFOI8640, 16CF023341, 16CF027684, 16CF032027,16BF025640, 16CF015190, 16CF018641, 16CF023342, 16CF027685, 16CF032028,16BF025641, 16CF015191, 16CF018642, 16CF023343, 16CF027686, 16CF032029,16BF027950, 16CF015192, 16CF018643, 16CF023344, 16CF027687, 16CF032030,16BF027951, 16CF015193, 16CF018644, 16CF023345, 16CF027688, 16CF040048,16BF027952, 16CF015194, 16CF018645, 16CF023346, 16CF030089, 16CF040049,16BF030583, 16CF015195, 16CF018646, 16CF023347, 16CF030090, 16CF040050,16BF030584, 16CF015196, 16CF018647, 16CF023348, 16CF030091, 16CF040051,16BF030585, 16CF015197, 16CF019531, 16CF023349, 16CF030092, 16CF040052,16BF030586, 16CF015198, 16CF019532, 16CF026057, 16CF030093, 16CF040053,16BF030587, 16CF015199, 16CF019533, 16CF026058, 16CF030094, 16CF040054,16BF030588, 16CF015200, 16CF019534, 16CF026059, 16CF030095, 16CF040055,16BF030589, 16CF015201, 16CF019535, 16CF026060, 16CF030096, 16CF040056,16BF030590, 16CF015202, 16CF019536, 16CF026061, 16CF030097, 16CF040057,16BF030591, 16CF015203, 16CF019537, 16CF026062, 16CF030098, 16CF040058,16BF030592, 16CF015204, 16CF019538, 16CF026063, 16CF030099, 16CF040059,16CF040060,16CF040061,16DF017166, 16DF017167, 16DF017910,16DF017911, 16DF040084 16DF040085, 16DF040632,16DF040633, 16DF045327,16DF045328,16CF040062, 16DF017168, 16DF017912, 16DF040086, 16DF040634, 16DF045329,16CF040063, 16DF017169, 16DF017913, 16DF040087, 16DF040635, 16DF045330,16CF040064, 16DF017170, 16DF017914, 16DF040088, 16DF040636, 16DF045331,16CF040065, 16DF017171, 16DF017915, 16DF040089, 16DF040637, 16DF045332,16CF040066, 16DF017172, 16DF017916, 16DF040090, 16DF040638, 16DF045333,16CF040067, 16DF017173, 16DF017917, 16DF040091, 16DF040639, 16DF045334,16CF040068, 16DF017174, 16DF017918, 16DF040092, 16DF040640, 16DF045335,16CF040069, 16DF017175, 16DF017919, 16DF040093, 16DF040641, 16DF045336,16CF040070, 16DF017176, 16DF017920, 16DF040094, 16DF040642, 16DF045337,16CF040071, 16DF017177, 16DF017921, 16DF040095, 16DF040643, 16DF045338,16CF040072, 16DF017178, 16DF017922, 16DF040096, 16DF040644, 16DF045339,16CF040073, 16DF017179, 16DF017923, 16DF040097, 16DF040645, 16DF045340,16CF040074, 16DF017180, 16DF017924, 16DF040098, 16DF040646, 16DF045341,16CF040075, 16DF017181, 16DF017925, 16DF040099, 16DF040647, 16DF045342,16CF040076, 16DF017182, 16DF017926, 16DF040100, 16DF040648, 16DF045343,16CF040077, 16DF017183, 16DF017927, 16DF040101, 16DF040649, 16EF000999,16CF040078, 16DF017184, 16DF017928, 16DF040102, 16DF040650, 16EF001000,16CF040079, 16DF017185, 16DF017929, 16DF040103, 16DF041832, 16EF001001,16CF040080, 16DF017186, 16DF019605, 16DF040104, 16DF041833, 16EF001002,16CF040081, 16DF017187, 16DF019606, 16DF040615, 16DF041834, 16EF001003,16CF040082, 16DF017894, 16DF019607, 16DF040616, 16DF045311, 16EF001004,16CF040083, 16DF017895, 16DF040069, 16DF040617, 16DF045312, 16EF001005,16DF017152, 16DF017896, 16DF040070, 16DF040618, 16DF045313, 16EF001006,16DF017153, 16DF017897, 16DF040071, 16DF040619, 16DF045314, 16EF001007,16DF017154, 16DF017898, 16DF040072, 16DF040620, 16DF045315, 16EF001008,16DF017155, 16DF017899, 16DF040073, 16DF040621, 16DF045316, 16EF001009,16DF017156, 16DF017900, 16DF040074, 16DF040622, 16DF045317, 16EF001010,16DF017157, 16DF017901, 16DF040075, 16DF040623, 16DF045318, 16EF001011,16DF017158, 16DF017902, 16DF040076, 16DF040624, 16DF045319, 16EF001012,16DF017159, 16DF017903, 16DF040077, 16DF040625, 16DF045320, 16EF001013,16DF017160, 16DF017904, 16DF040078, 16DF040626, 16DF045321, 16EF001014,16DF017161, 16DF017905, 16DF040079, 16DF040627, 16DF045322, 16EF001015,16DF017162, 16DF017906, 16DF040080, 16DF040628, 16DF045323, 16EF001016,16DF017163, 16DF017907, 16DF040081, 16DF040629, 16DF045324, 16EF001017,16DF017164, 16DF017908, 16DF040082, 16DF040630, 16DF045325, 16EF001020,16DF017165, 16DF017909, 16DF040083, 16DF040631, 16DF045326, 16EF001021. 
FEI Number 3002889431
Recalling Firm/
Manufacturer
Invacare Corporation
1 Invacare Way
Elyria OH 44035-4190
For Additional Information Contact
877-413-6008
Manufacturer Reason
for Recall
The rear gradation scale was screen printed upside down on the flowmeter gauge.
FDA Determined
Cause 2
Process control
Actionlnvacare Corporation sent an Urgent Medical Device Recall letter dated June 27, 2016, to all affected consignees with return receipt to notify them of the recall. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees with questions were instructed to call 877-413-6008.
Quantity in Commerce395
DistributionWorldwide Distribution - US to CA, CO, CT, FL, GA, IA, ID, IL,IN,KS,KY,LA,MA,MD,MI,MS,MT,NC,NE,NJ,NM,NY,OH,PA,SC, TN,TX,UT,VA,WA and Internationally to Canada, Germany and New Zealand.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = NFB
-
-