| Date Initiated by Firm | December 01, 2016 |
| Date Posted | December 19, 2016 |
| Recall Status1 |
Terminated 3 on August 31, 2017 |
| Recall Number | Z-0825-2017 |
| Recall Event ID |
75844 |
| 510(K)Number | K110151 |
| Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
| Product | Ingenia MR-OR; 781437, 781438
The MR-OR option is intended for allowing transfer of a neurosurgical patient from OR suite to MR suite and back with tools and instructions for proper positioning and MR imaging of those patients. MR-OR option also facilitates execution of simple neurosurgical procedures in the near proximity of the MR scanner without a need to move the patient back to the OR suite |
| Code Information |
781437: 15001, 15002, 15003 781438: 30001, 30002, 30003, 30004, 30005, 30006 |
| FEI Number |
1218950
|
Recalling Firm/ Manufacturer |
Philips Electronics North America Corporation
3000 Minuteman Rd
Andover
MA
01810-1032
|
| For Additional Information Contact | 800-722-9377 |
Manufacturer Reason for Recall | Possibility that a patient or user may be exposed to a very small level of touch current. |
FDA Determined Cause 2 | Device Design |
| Action | Philips sent an Urgent Field Safety Notice Medical Device Correction letter dated November 2016 to all affected customers. The letter explains the issue and actions planned by Philips. There is no need for action to be taken by customer or user. Customers with questions should contact their local Philips representative Technical Support Line 1-800-722-9377. |
| Quantity in Commerce | 9 |
| Distribution | Worldwide Distribution -US including CA, NY and Internationally to United Kingdom, Australia, Norway, and Thailand |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = LNH
|