 Class 2 Device Recall DJO EMPOWR KNEE LOCKING TIBIAL PUNCH GUIDEsee related information
Date Initiated by FirmDecember 12, 2017
Create DateJanuary 10, 2018
Recall Status1 Terminated 3 on June 13, 2022
Recall NumberZ-0326-2018
Recall Event ID 78727
510(K)NumberK100900 
Product Classification Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code JWH
ProductDJO EMPOWR KNEE LOCKING TIBIAL PUNCH GUIDE, REF 801-05-042, NON STERILE, Qty 1 For use in orthopedic surgery.
Code Information Lot 252567
FEI Number 1000116912
Recalling Firm/
Manufacturer
Encore Medical, Lp
9800 Metric Blvd
Austin TX 78758-5445
For Additional Information Contact
512-832-9500
Manufacturer Reason
for Recall
The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.
FDA Determined
Cause 2
Device Design
ActionThe field safety notice was disseminated by email on 12/12/2017. The notice stated the following: "DJO Sales Agents Action: Please review the bulletin that is provided with this notice. This bulletin contains new instructions on the use of the Locking Punch Guide and Punch. Please ensure your surgeons are familiar with this information. DJO Surgical then requires that you complete the attached acknowledgment and email it to teffany.hutto@djoglobal.com. Please have this form returned by December 15, 2017 Once redesigned guides are available, they will be exchanged through a separate field action."
Quantity in Commerce2 units
DistributionOne medical device distributor in California.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = JWH
-
-