| | Class 2 Device Recall A.T.S. 4000TS Tourniquet Systems, Personalized Pressure |  |
| Date Initiated by Firm | July 27, 2017 |
| Create Date | February 14, 2018 |
| Recall Status1 |
Terminated 3 on August 01, 2019 |
| Recall Number | Z-0612-2018 |
| Recall Event ID |
78759 |
| 510(K)Number | K123553 |
| Product Classification |
Tourniquet, pneumatic - Product Code KCY
|
| Product | A.T.S. 4000TS Tourniquet Systems, Personalized Pressure.
Intended to be used by qualified professionals to temporarily occlude blood flow in a patient's extremities during surgical procedures on those extremities.
. |
| Code Information |
Item No. 60400040400, Lot No. 62510549 (Serial Numbers: 4013JAAA, 4013JAAB, 4013JAAC, 4013JAAD, 4013JAAE, 4013JAAF, 4013JAAG, 4013JAAH, 4013JAAJ, 4013JAAK, 4013JAAL, 4013JAAM, 4013JAAN, 4013JAAP, 4013JAAQ, 4013JAAR, 4013JAAS, 4013JAAT, 4013JAAU, 4013JAAW, 4013JABA, 4013JABB, 4013JABC, 4013JABD, 4013JABE, 4013JABF, 4013JABG); Lot No. 62510845 (Serial Numbers: 4013JABH, 4013JABJ, 4013JABK, 4013JABL, 4013JABM, 4013JABN, 4013JABP, 4013JABQ, 4013JABR, 4013JABS, 4013JABT, 4013JABU, 4013JABW, 4013JACA, 4013JACB, 4013JACC, 4013JACD, 4013JACE, 4013JACF, 4013JACG); Lot No. 62510927 (Serial Numbers: 4013JACH, 4013JACJ, 4013JACK, 4013JACL, 4013JACM, 4013JACN, 4013JACP, 4013JACQ, 4013JACR, 4013JACS, 4013JACT, 4013JACU, 4013JACW, 4013JADA, 4013JADB, 4013JADC, 4013JADD, 4013JADE, 4013JADF, 4013JADG); Lot No. 62510928 (Serial Numbers: 4013JADH, 4013JADJ, 4013JADK, 4013JADL, 4013JADM, 4013JADN, 4013JADP, 4013JADQ, 4013JADR, 4013JADS, 4013JADT, 4013JADU, 4013JADW, 4013JAEA, 4013JAEB, 4013JAEC, 4013JAED, 4013JAEE, 4013JAEF, 4013JAEG); Lot No. 62525817 (Serial Numbers: 4013KAAA, 4013KAAB, 4013KAAC, 4013KAAD, 4013KAAE, 4013KAAF, 4013KAAG, 4013KAAH, 4013KAAJ, 4013KAAK, 4013KAAL, 4013KAAM, 4013KAAN, 4013KAAP, 4013KAAQ, 4013KAAR, 4013KAAS, 4013KAAT, 4013KAAU, 4013KAAW); Lot No. 62525841 (Serial Numbers: 4013KABA, 4013KABB, 4013KABC, 4013KABD, 4013KABE, 4013KABF, 4013KABG, 4013KABH, 4013KABJ, 4013KABK, 4013KABL, 4013KABM, 4013KABN, 4013KABP, 4013KABQ, 4013KABR, 4013KABS, 4013KABT, 4013KABU, 4013KABW); Lot No. 62616141 (Serial Numbers: 4014BAAA, 4014BAAB, 4014BAAC, 4014BAAD, 4014BAAE, 4014BAAF, 4014BAAG, 4014BAAH, 4014BAAJ, 4014BAAK, 4014BAAL, 4014BAAM, 4014BAAN, 4014BAAP, 4014BAAQ, 4014BAAR, 4014BAAS, 4014BAAT, 4014BAAU, 4014BAAW); Lot No. 62616142 (Serial Numbers: 4014BABA, 4014BABB, 4014BABC, 4014BABD, 4014BABE, 4014BABF, 4014BABG, 4014BABH, 4014BABJ, 4014BABK, 4014BABL, 4014BABM, 4014BABN, 4014BABP, 4014BABQ, 4014BABR, 4014BABS, 4014BABT, 4014BABU, 4014BABW); Lot No. 62616143 (Serial Numbers: 4014BACA, 4014BACB, 4014BACC, 4014BACD, 4014BACE, 4014BACF, 4014BACG, 4014BACH, 4014BACJ, 4014BACK, 4014BACL, 4014BACM, 4014BACN, 4014BACP, 4014BACQ, 4014BACR, 4014BACS, 4014BACT, 4014BACU, 4014BACW); Lot No. 62616145 (Serial Number: 4014BADA, 4014BADB, 4014BADC, 4014BADD, 4014BADE, 4014BADF, 4014BADG, 4014BADH, 4014BADJ, 4014BADK, 4014BADL, 4014BADM, 4014BADN, 4014BADP, 4014BADQ, 4014BADR, 4014BADS, 4014BADT, 4014BADU, 4014BADW): Lot No. 62635795 (Serial Numbers: 4014CAAA, 4014CAAB, 4014CAAC, 4014CAAD, 4014CAAE, 4014CAAF, 4014CAAG, 4014CAAH, 4014CAAJ, 4014CAAK, 4014CAAL, 4014CAAM, 4014CAAN, 4014CAAP, 4014CAAQ, 4014CAAR, 4014CAAS, 4014CAAT, 4014CAAU, 4014CAAW); Lot No. 62635796 (Serial Numbers: 4014CABA, 4014CABB, 4014CABC, 4014CABD, 4014CABE, 4014CABF, 4014CABG, 4014CABH, 4014CABJ, 4014CABK, 4014CABL, 4014CABM, 4014CABN, 4014CABP, 4014CABQ, 4014CABR, 4014CABS, 4014CABT, 4014CABU, 4014CABW); Lot No. 62635797 (Serial Numbers: 4014CACA, 4014CACB, 4014CACC, 4014CACD, 4014CACE, 4014CACF, 4014CACG, 4014CACH, 4014CACJ, 4014CACK, 4014CACL, 4014CACM, 4014CACN, 4014CACP, 4014CACQ, 4014CACR, 4014CACS, 4014CACT, 4014CACU, 4014CACW); Lot No. 62642282 (Serial Numbers: 4014CADA, 4014CADB, 4014CADC, 4014CADD, 4014CADE, 4014CADF, 4014CADG, 4014CADH, 4014CADJ, 4014CADK, 4014CADL, 4014CADM, 4014CADN, 4014CADP, 4014CADQ, 4014CADR, 4014CADS, 4014CADT,4014CADU, 4014CADW); Lot No. 62635798 (Serial Numbers: 4014CAEA, 4014CAEB, 4014CAEC, 4014CAED, 4014CAEE, 4014CAEF, 4014CAEG, 4014CAEH, 4014CAEJ, 4014CAEK, 4014CAEL, 4014CAEM, 4014CAEN, 4014CAEP, 4014CAEQ, 4014CAER, 4014CAES, 4014CAET, 4014CAEU, 4014CAEW); Lot No. 62676173 (Serial Numbers: 4014DAAA, 4014DAAB, 4014DAAC, 4014DAAD, 4014DAAE, 4014DAAF, 4014DAAG, 4014DAAH, 4014DAAJ, 4014DAAK, 4014DAAL, 4014DAAM, 4014DAAN, 4014DAAP, 4014DAAQ, 4014DAAR, 4014DAAS, 4014DAAT, 4014DAAU, 4014DAAW); Lot No. 62676174 (Serial Numbers: 4014DABA, 4014DABB, 4014DABC, 4014DABD, 4014DABE, 4014DABF, 4014DABG, 4014DABH, 4014DABJ, 4014DABK, 4014DABL, 4014DABM, 4014DABN, 4014DABP, 4014DABQ, 4014DABR, 4014DABS, 4014DABT, 4014DABU, 4014DABW); Lot No. 62686165 (Serial Numbers: 4014DACA, 4014DACB, 4014DACC, 4014DACD, 4014DACE, 4014DACF, 4014DACG, 4014DACH, 4014DACJ, 4014DACK, 4014DACL, 4014DACM, 4014DACN, 4014DACP, 4014DACQ, 4014DACR, 4014DACS, 4014DACT, 4014DACU, 4014DACW); Lot No. 62686166 (Serial No. 4014DADA, 4014DADB, 4014DADC, 4014DADD, 4014DADE, 4014DADF, 4014DADG, 4014DADH, 4014DADJ, 4014DADK, 4014DADL, 4014DADM, 4014DADN, 4014DADP, 4014DADQ, 4014DADR, 4014DADS, 4014DADT, 4014DADU, 4014DADW); Lot No. 62686167 (Serial Numbers: 4014DAEA, 4014DAEB, 4014DAEC, 4014DAED, 4014DAEE, 4014DAEF, 4014DAEG, 4014DAEH, 4014DAEJ, 4014DAEK, 4014DAEL, 4014DAEM, 4014DAEN, 4014DAEP, 4014DAEQ, 4014DAER, 4014DAES, 4014DAET, 4014DAEU, 4014DAEW); Lot No. 62689574 (Serial Numbers: 4014DAFA, 4014DAFB, 4014DAFC, 4014DAFD, 4014DAFE, 4014DAFF, 4014DAFG, 4014DAFH, 4014DAFJ, 4014DAFK, 4014DAFL, 4014DAFM, 4014DAFN, 4014DAFP, 4014DAFQ, 4014DAFR, 4014DAFS, 4014DAFT, 4014DAFU, 4014DAFW); Lot No. 62689575 (Serial Numbers: 4014DAGA, 4014DAGB, 4014DAGC, 4014DAGD, 4014DAGE, 4014DAGF, 4014DAGG, 4014DAGH, 4014DAGJ, 4014DAGK, 4014DAGL, 4014DAGM, 4014DAGN, 4014DAGP, 4014DAGQ, 4014DAGR, 4014DAGS, 4014DAGT, 4014DAGU, 4014DAGW); Lot No. 62689611 (Serial Numbers: 4014EAAA, 4014EAAB, 4014EAAC, 4014EAAD, 4014EAAE, 4014EAAF, 4014EAAG, 4014EAAH, 4014EAAJ, 4014EAAK, 4014EAAL, 4014EAAM, 4014EAAN, 4014EAAP, 4014EAAQ, 4014EAAR, 4014EAAS, 4014EAAT, 4014EAAU, 4014EAAW); Lot No. 62689602 (Serial Numbers: 4014EABA, 4014EABB, 4014EABC, 4014EABD, 4014EABE, 4014EABF, 4014EABG, 4014EABH, 4014EABJ, 4014EABK, 4014EABL, 4014EABM, 4014EABN, 4014EABP, 4014EABQ, 4014EABR, 4014EABS, 4014EABT, 4014EABU, 4014EABW); Lot No. 62689620 (Serial Numbers: 4014EACA, 4014EACB, 4014EACC, 4014EACD, 4014EACE, 4014EACF, 4014EACG, 4014EACH, 4014EACJ, 4014EACK, 4014EACL, 4014EACM, 4014EACN, 4014EACP, 4014EACQ, 4014EACR, 4014EACS, 4014EACT, 4014EACU, 4014EACW); Lot No. 62689623 (Serial Numbers: 4014EADA, 4014EADB, 4014EADC, 4014EADD, 4014EADE, 4014EADF, 4014EADG, 4014EADH, 4014EADJ, 4014EADK, 4014EADL, 4014EADM, 4014EADN, 4014EADP, 4014EADQ, 4014EADR, 4014EADS, 4014EADT, 4014EADU, 4014EADW); Lot No. 62675189 (Serial Number: 4014DAHA); Lot No. 62688979 (Serial Numbers: 4014DAHB, 4014DAHC, 4014DAHD, 4014DAHE, 4014DAHF, 4014DAHG); Lot No. 62673235 (Serial Number: 4014DAHH); Lot No. 62683264 (Serial Number: 4014DAHJ); Lot No. 62726132 (Serial Numbers: 4014FAAA, 4014FAAB, 4014FAAC, 4014FAAD, 4014FAAE, 4014FAAF, 4014FAAG, 4014FAAH, 4014FAAJ, 4014FAAK, 4014FAAL, 4014FAAM, 4014FAAN, 4014FAAP, 4014FAAQ); Lot No. 62726133 (Serial Numbers: 4014FAAR); Lot No. 62732712 (Serial Numbers: 4014FAAS, 4014FAAT, 4014FAAU, 4014FAAW); Lot No. 62732713 (Serial Numbers: 4014FABA, 4014FABB); Lot No. 62727350 (Serial Numbers: 4014FABC, 4014FABD, 4014FABE, 4014FABF); Lot No. 62728452 (Serial Numbers: 4014FABG, 4014FABH); Lot No. 62723751 (Serial Numbers: 4014FABJ, 4014FABK, 4014FABL, 4014FABM, 4014FABN, 4014FABP, 4014FABQ, 4014FABR, 4014FABS, 4014FABT, 4014FABU, 4014FABW, 4014FACA, 4014FACB, 4014FACC, 4014FACD, 4014FACE, 4014FACF, 4014FACG, 4014FACH, 4014FACJ, 4014FACK, 4014FACL, 4014FACM, 4014FACN, 4014FACP, 4014FACQ, 4014FACR, 4014FACS, 4014FACT, 4014FACU, 4014FACW, 4014FADA, 4014FADB, 4014FADC, 4014FADD, 4014FADE, 4014FADF, 4014FADG, 4014FADH); Lot No. 62723752 (Serial Numbers: 4014FADJ, 4014FADK, 4014FADL, 4014FADM, 4014FADN, 4014FADP, 4014FADQ, 4014FADR, 4014FADS, 4014FADT, 4014FADU, 4014FADW, 4014FAEA, 4014FAEB, 4014FAEC, 4014FAED, 4014FAEE, 4014FAEF, 4014FAEG, 4014FAEH); Lot No. 62745917 (Serial Numbers: 4014FAEJ, 4014FAEK); Lot No. 62745898 (Serial Numbers: 4014FAEL, 4014FAEM, 4014FAEN); Lot No. 62749571 (Serial Numbers: 4014FAEP, 4014FAEQ, 4014FAER, 4014FAES, 4014FAET); Lot No. 62741841 (Serial Numbers: 4014FAEU, 4014FAEW, 4014FAFA, 4014FAFB, 4014FAFC, 4014FAFD, 4014FAFE, 4014FAFF, 4014FAFG, 4014FAFH, 4014FAFJ, 4014FAFK, 4014FAFL, 4014FAFM, 4014FAFN, 4014FAFP, 4014FAFQ, 4014FAFR, 4014FAFS, 4014FAFT); Lot No. 62723753 (Serial Numbers: 4014FAFU, 4014FAFW, 4014FAGA, 4014FAGB, 4014FAGC, 4014FAGD, 4014FAGE, 4014FAGF, 4014FAGG, 4014FAGH, 4014FAGJ, 4014FAGK, 4014FAGL, 4014FAGM, 4014FAGN, 4014FAGP, 4014FAGQ, 4014FAGR, 4014FAGS, 4014FAGT); Lot No. 62723754 (Serial Numbers: 4014FAJW, 4014FAKA, 4014FAKB, 4014FAKC, 4014FAKD, 4014FAKE, 4014FAKF, 4014FAKG, 4014FAKH, 4014FAKJ, 4014FAKK, 4014FAKL, 4014FAKM, 4014FAKN, 4014FAKP, 4014FAKQ, 4014FAKR, 4014FAKS, 4014FAKT, 4014FAKU); Lot No. 62745156 (Serial Numbers: 4014FAGU, 4014FAGW, 4014FAHA, 4014FAHB, 4014FAHC, 4014FAHD, 4014FAHE, 4014FAHF, 4014FAHG, 4014FAHH, 4014FAHJ, 4014FAHK, 4014FAHL, 4014FAHM, 4014FAHN, 4014FAHP, 4014FAHQ, 4014FAHR, 4014FAHS, 4014FAHT); Lot No. 62745158 (Serial Numbers: 4014FAKW, 4014FALA, 4014FALB, 4014FALC, 4014FALD, 4014FALE, 4014FALF, 4014FALG, 4014FALH, 4014FALJ, 4014FALK, 4014FALL, 4014FALM, 4014FALN, 4014FALP, 4014FALQ, 4014FALR, 4014FALS, 4014FALT, 4014FALU); Lot No. 62760910 (Serial Numbers: 4014GAAA, 4014GAAB, 4014GAAC, 4014GAAD, 4014GAAE, 4014GAAF); Lot No. 62769952 (Serial Numbers: 4014GAAG, 4014GAAH); Lot No. 62769958 (Serial Numbers: 4014GAAJ, 4014GAAK); Lot No. 62775967 (Serial Numbers: 4014GAAL, 4014GAAM, 4014GAAN, 4014GAAP); Lot No. 62748382 (Serial Numbers: 4014GAAQ, 4014GAAR, 4014GAAS, 4014GAAT, 4014GAAU, 4014GAAW, 4014GABA, 4014GABB, 4014GABC, 4014GABD, 4014GABE, 4014GABF, 4014GABG, 4014GABH, 4014GABJ, 4014GABK, 4014GABL, 4014GABM, 4014GABN, 4014GABP); Lot No. 62748393 (Serial Numbers: 4014GABQ, 4014GABR, 4014GABS, 4014GABT, 4014GABU, 4014GABW, 4014GACA, 4014GACB, 4014GACC, 4014GACD, 4014GACE, 4014GACF, 4014GACG,4014GACH, 4014GACJ, 4014GACK, 4014GACL, 4014GACM, 4014GACN, 4014GACP); Lot No. 62778431 (Serial Numbers: 4014GACQ, 4014GACR, 4014GACS, 4014GACT, 4014GACU, 4014GACW); Lot No. 62748405 (Serial Numbers: 4014GADA, 4014GADB, 4014GADC, 4014GADD, 4014GADE, 4014GADF, 4014GADG, 4014GADH, 4014GADJ, 4014GADK, 4014GADL, 4014GADM, 4014GADN, 4014GADP, 4014GADQ, 4014GADR, 4014GADS, 4014GADT, 4014GADU, 4014GADW); Lot No. 62779666 (Serial Numbers: 4014GAEA, 4014GAEB, 4014GAEC, 4014GAED); Lot No. 62779665 (Serial Numbers: 4014GAEE, 4014GAEF, 4014GAEG, 4014GAEH, 4014GAEJ); Lot No. 62748415 (Serial Numbers: 4014GAEK, 4014GAEL, 4014GAEM, 4014GAEN, 4014GAEP, 4014GAEQ, 4014GAER, 4014GAES, 4014GAET, 4014GAEU, 4014GAEW, 4014GAFA, 4014GAFB, 4014GAFC, 4014GAFD, 4014GAFE, 4014GAFF, 4014GAFG, 4014GAFH, 4014GAFJ); Lot No. 62797834 (Serial Numbers: 4014HAAA, 4014HAAB, 4014HAAC, 4014HAAD, 4014HAAE); Lot No. 62786181 (Serial Numbers: 4014HAAF, 4014HAAG, 4014HAAH, 4014HAAJ, 4014HAAK, 4014HAAL, 4014HAAM, 4014HAAN, 4014HAAP, 4014HAAQ, 4014HAAR, 4014HAAS, 4014HAAT, 4010HAAU, 4014HAAW, 4014HABA, 4014HABB, 4014HABC, 4014HABD, 4014HABE); Lot No. 62786182 (Serial Numbers: 4014HABF, 4014HABG, 4014HABH, 4014HABJ, 4014HABK, 4014HABL, 4014HABM, 4014HABN, 4014HABP, 4014HABQ, 4014HABR, 4014HABS, 4014HABT, 4014HABU, 4014HABW, 4014HACA, 4014HACB, 4014HACC, 4014HACD, 4014HACE); Lot No. 62791623 (Serial Numbers: 4014HACF, 4014HACG, 4014HACH, 4014HACJ, 4014HACK, 4014HACL, 4014HACM, 4014HACN, 4014HACP, 4014HACQ, 4014HACR, 4014HACS, 4014HACT, 4014HACU, 4014HACW, 4014HADA, 4014HADB, 4014HADC, 4014HADD, 4014HADE); Lot No. 62809993 (Serial Numbers: 4014HADF, 4014HADG, 4014HADH, 4014HADJ, 4014HADK, 4014HADL, 4014HADM, 4014HADN, 4014HADP, 4014HADQ, 4014HADR, 4014HADS, 4014HADT, 4014HADU, 4014HADW, 4014HAEA, 4014HAEB, 4014HAEC, 4014HAED, 4014HAEE); Lot No. 62812519 (Serial Numbers: 4014IAAA, 4014IAAB, 4014IAAC, 4014IAAD, 4014IAAE, 4014IAAF, 4014IAAG, 4014IAAH, 4014IAAJ, 4014IAAK, 4014IAAL, 4014IAAM, 4014IAAN, 4014IAAP, 4014IAAQ, 4014IAAR, 4014IAAS, 4014IAAT, 4014IAAU, 4014IAAW); Lot No. 62812815 (Serial Numbers: 4014IABA, 4014IABB, 4014IABC, 4014IABD, 4014IABE, 4014IABF, 4014IABG, 4014IABH, 4014IABJ, 4014IABK, 4014IABL, 4014IABM, 4014IABN, 4014IABP, 4014IABQ, 4014IABR, 4014IABS, 4014IABT, 4014IABU, 4014IABW); Lot No. 62824837 (Serial Numbers: 4014IACA, 4014IACB, 4014IACC, 4014IACD); Lot No. 62826497 (Serial Numbers: 4014IACE); Lot No. 62812816 (Serial Numbers: 4014IACF, 4014IACG, 4014IACH, 4014IACJ, 4014IACK, 4014IACL, 4014IACM, 4014IACN, 4014IACP, 4014IACQ, 4014IACR, 4014IACS, 4014IACT, 4014IACU, 4014IACW, 4014IADA, 4014IADB, 4014IADC, 4014IADD, 4014IADE); Lot No. 62831377 (Serial Numbers: 4014IADF); Lot No. 62812843 (Serial Numbers: 4014IADG, 4014IADH, 4014IADJ, 4014IADK, 4014IADL, 4014IADM, 4014IADN, 4014IADP, 4014IADQ, 4014IADR, 4014IADS, 4014IADT, 4014IADU, 4014IADW, 4014IAEA, 4014IAEB, 4014IAEC, 4014IAED, 4014IAEE, 4014IAEF); Lot No. 62838045 (Serial Numbers: 4014IAEG, 4014IAEH, 4014IAEJ, 4014IAEK, 4014IAEL, 4014IAEM, 4014IAEN); Lot No. 62812844 (Serial Numbers: 4014IAEP, 4014IAEQ, 4014IAER, 4014IAES, 4014IAET, 4014IAEU, 4014IAEW, 4014IAFA, 4014IAFB, 4014IAFC, 4014IAFD, 4014IAFE, 4014IAFF, 4014IAFG, 4014IAFH, 4014IAFJ, 4014IAFK, 4014IAFL, 4014IAFM, 4014IAFN); Lot No. 62812857 (Serial Numbers: 4014IAFP, 4014IAFQ, 4014IAFR, 4014IAFS, 4014IAFT, 4014IAFU, 4014IAFW, 4014IAGA, 4014IAGB, 4014IAGC, 4014IAGD, 4014IAGE, 4014IAGF, 4014IAGG, 4014IAGH, 4014IAGJ, 4014IAGK, 4014IAGL, 4014IAGM, 4014IAGN); Lot No. 62832912 (Serial Numbers: 4014JAAA, 4014JAAB, 4014JAAC, 4014JAAD, 4014JAAE, 4014JAAF, 4014JAAG, 4014JAAH, 4014JAAJ, 4014JAAK, 4014JAAL, 4014JAAM, 4014JAAN, 4014JAAP, 4014JAAQ, 4014JAAR, 4014JAAS, 4014JAAT 4014JAAU, 4014JAAW); Lot No. 62845553 (Serial Numbers: 4014JABA, 4014JABB, 4014JABC, 4014JABD, 4014JABE, 4014JABF, 4014JABG, 4014JABH, 4014JABJ, 4014JABK, 4014JABL, 4014JABM, 4014JABN, 4014JABP, 4014JABQ, 4014JABR, 4014JABS, 4014JABT, 4014JABU, 4014JABW); Lot No. 62832932 (Serial Numbers: 4014JACA, 4014JACB, 4014JACC, 4014JACD, 4014JACE, 4014JACF, 4014JACG, 4014JACH, 4014JACJ, 4014JACK, 4014JACL, 4014JACM, 4014JACN, 4014JACP, 4014JACQ, 4014JACR, 4014JACS, 4014JACT, 4014JACU, 4014JACW); Lot No. 62852357 (Serial Numbers: 4014JADA, 4014JADB, 4014JADC, 4014JADD, 4014JADE); Lot No. 62845554 (Serial Numbers: 4014JADF, 4014JADG, 4014JADH, 4014JADJ, 4014JADK, 4014JADL, 4014JADM, 4014JADN, 4014JADP, 4014JADQ, 4014JADR, 4014JADS, 4014JADT, 4014JADU, 4014JADW, 4014JAEA, 4014JAEB, 4014JAEC, 4014JAED, 4014JAEE); Lot No. 62832930 (Serial Numbers: 4014JAEF, 4014JAEG, 4014JAEH, 4014JAEJ, 4014JAEK, 4014JAEL, 4014JAEM, 4014JAEN, 4014JAEP, 4014JAEQ, 4014JAER, 4014JAES, 4014JAET, 4014JAEU, 4014JAEW, 4014JAFA, 4014JAFB, 4014JAFC, 4014JAFD, 4014JAFE); Lot No. 62835917 (Serial Numbers: 4014JAFF, 4014JAFG, 4014JAFH, 4014JAFJ, 4014JAFK, 4014JAFL, 4014JAFM, 4014JAFN, 4014JAFP, 4014JAFQ, 4014JAFR, 4014JAFS, 4014JAFT, 4014JAFU, 4014JAFW, 4014JAGA, 4014JAGB, 4014JAGC, 4014JAGD, 4014JAGE); Lot No. 62836063 (Serial Numbers: 4014JAGF, 4014JAGG, 4014JAGH, 4014JAGJ, 4014JAGK, 4014JAGL, 4014JAGM, 4014JAGN, 4014JAGP, 4014JAGQ, 4014JAGR, 4014JAGS, 4014JAGT, 4014JAGU, 4014JAGW, 4014JAHA, 4014JAHB, 4014JAHC, 4014JAHD, 4014JAHE); Lot No. 62836101 (Serial Numbers: 4014JAHF, 4014JAHG, 4014JAHH, 4014JAHJ, 4014JAHK, 4014JAHL, 4014JAHM, 4014JAHN, 4014JAHP, 4014JAHQ, 4014JAHR, 4014JAHS, 4014JAHT, 4014JAHU, 4014JAHW, 4014JAJA, 4014JAJB, 4014JAJC, 4014JAJD, 4014JAJE); Lot No. 62860464 (Serial Numbers: 4014JAJF, 4014JAJG, 4014JAJH, 4014JAJJ, 4014JAJK, 4014JAJL, 4014JAJM, 4014JAJN, 4014JAJP, 4014JAJQ, 4014JAJR, 4014JAJS, 4014JAJT, 4014JAJU, 4014JAJW, 4014JAKA, 4014JAKB, 4014JAKC, 4014JAKD, 4014JAKE); Lot No. 62860465 (Serial Numbers: 4014JAKF, 4014JAKG, 4014JAKH, 4014JAKJ, 4014JAKK, 4014JAKL, 4014JAKM, 4014JAKN, 4014JAKP, 4014JAKQ, 4014JAKR, 4014JAKS, 4014JAKT, 4014JAKU, 4014JAKW, 4014JALA, 4014JALB, 4014JALC, 4014JALD, 4014JALE); Lot No. 62870930 (Serial Numbers: 4014KAAA, 4014KAAB, 4014KAAC, 4014KAAD, 4014KAAE, 4014KAAF, 4014KAAG, 4014KAAH, 4014KAAJ, 4014KAAK, 4014KAAL, 4014KAAM, 4014KAAN, 4014KAAP, 4014KAAQ, 4014KAAR, 4014KAAS, 4014KAAT, 4014KAAU, 4014KAAW); Lot No. 62870931 (Serial Numbers: 4014KABA, 4014KABB, 4014KABC, 4014KABD, 4014KABE, 4014KABF, 4014KABG, 4014KABH, 4014KABJ, 4014KABK, 4014KABL, 4014KABM, 4014KABN, 4014KABP, 4014KABQ, 4014KABR, 4014KABS, 4014KABT, 4014KABU, 4014KABW); Lot No. 62869995 (Serial Numbers: 4014KACA, 4014KACB, 4014KACC, 4014KACD, 4014KACE); Lot No. 62869994 (Serial Numbers: 4014KACF, 4014KACG, 4014KACH, 4014KACJ, 4014KACK) Lot No. 62876470 (Serial Numbers: 4014KACL, 4014KACM, 4014KACN, 4014KACP, 4014KACQ, 4014KACR, 4014KACS, 4014KACT, 4014KACU, 4014KACW, 4014KADA, 4014KADB,4014KADC, 4014KADD, 4014KADE ,4014KADF,4014KADG ,4014KADH,4014KADJ, 4014KADK); Lot No. 62876471 (Serial Numbers: 4014KADL, 4014KADM, 4014KADN, 4014KADP, 4014KADQ, 4014KADR, 4014KADS, 4014KADT, 4014KADU, 4014KADW, 4014KAEA, 4014KAEB, 4014KAEC, 4014KAED, 4014KAEE, 4014KAEF, 4014KAEG, 4014KAEH, 4014KAEJ, 4014KAEK); Lot No. 62877398 (Serial Numbers: 401411110001, 401411110002, 401411110003, 401411110004, 401411110005, 401411110006, 401411110007, 401411110008, 401411110009, 401411110010, 401411110011, 401411110012, 401411110013, 401411110014, 401411110015, 401411110016, 401411110017, 401411110018, 401411110019, 401411110020); Lot No. 62877405 (Serial Numbers: 401411110021, 401411110022, 401411110023, 401411110024, 401411110025, 401411110026, 401411110027, 401411110028, 401411110029, 401411110030, 401411110031, 401411110032, 401411110033, 401411110034, 401411110035, 401411110036, 401411110037, 401411110038, 401411110039, 401411110040); Lot No. 62885591 (Serial Numbers: 401411140001, 401411140002, 401411140003, 401411140004, 401411140005); Lot No. 62877414 (Serial Numbers: 401411140006, 401411140007, 401411140008, 401411140009, 401411140010, 401411140011, 401411140012, 401411140013, 401411140014, 401411140015, 401411140016, 401411140017, 401411140018, 401411140019, 401411140020, 401411140021, 401411140022, 401411140023, 401411140024, 401411140025); Lot No. 62885589 (Serial Numbers: 401411150001, 401411150002, 401411150003, 401411150004, 401411150005); Lot No. 62885588 (Serial Numbers: 401411170001, 401411170002, 401411170003, 401411170004, 401411170005); Lot No. 62877415 (Serial Numbers: 401411180001, 401411180002, 401411180003, 401411180004, 401411180005, 401411180006, 401411180007, 401411180008, 401411180009, 401411180010, 401411180011, 401411180012, 401411180013, 401411180014, 401411180015, 401411180016, 401411180017, 401411180018, 401411180019, 401411180020); Lot No. 62904095 (Serial Numbers: 401411200001, 401411200002, 401411200003, 401411200004, 401411200005, 401411200006, 401411200007, 401411200008, 401411200009, 401411200010, 401411200011, 401411200012, 401411200013, 401411200014, 401411200015, 401411200016, 401411200017, 401411200018, 401411200019, 401411200020); Lot No. 62904096 (Serial Numbers: 401411250001, 401411250002, 401411250003, 401411250004, 401411250005, 401411250006, 401411250007, 401411250008, 401411250009, 401411250010, 401411250011, 401411250012, 401411250013, 401411250014, 401411250015, 401411250016, 401411250017, 401411250018, 401411250019, 401411250020); Lot No. 62904127 (Serial Numbers: 401412020001, 401412020002, 401412020003, 401412020004, 401412020005, 401412020006, 401412020007, 401412020008, 401412020009, 401412020010, 401412020011, 401412020012, 401412020013, 401412020014, 401412020015, 401412020016, 401412020017, 401412020018, 401412020019, 401412020020); Lot No. 62878965 (Serial Numbers: 401412040001, 401412040002, 401412040003, 401412040004, 401412040005); Lot No. 62885590 (Serial Numbers: 401412050001, 401412050002, 401412050003, 401412050004, 401412050005, 401412050006, 401412050007, 401412050008, 401412050009, 401412050010); Lot No. 62894035 (Serial Numbers: 401412080001, 401412080002, 401412080003, 401412080004, 401412080005); Lot No. 62904131 (Serial Numbers: 401412090001, 401412090002, 401412090003, 401412090004, 401412090005, 401412090006, 401412090007, 401412090008, 401412090009, 401412090010, 401412090011, 401412090012, 401412090013, 401412090014, 401412090015, 401412090016, 401412090017, 401412090018, 401412090019, 401412090020); Lot No. 62905841 (Serial Numbers: 401412110001, 401412110002, 401412110003, 401412110004, 401412110005); Lot No. 62904271 (Serial Numbers: 401412120001, 401412120002, 401412120003, 401412120004, 401412120005, 401412120006, 401412120007, 401412120008, 401412120009, 401412120010, 401412120011, 401412120012, 401412120013, 401412120014, 401412120015, 401412120016, 401412120017, 401412120018, 401412120019, 401412120020); Lot No. 62905845 (Serial Numbers: 401412120021, 401412120022, 401412120023, 401412120024, 401412120025); Lot No. 62932765 (Serial Numbers: 401412160001, 401412160002, 401412160003, 401412160004, 401412160005); Lot No. 62904272 (Serial Numbers: 401412170001, 401412170002, 401412170003, 401412170004, 401412170005, 401412170006, 401412170007, 401412170008, 401412170009, 401412170010, 401412170011, 401412170012, 401412170013, 401412170014, 401412170015, 401412170016, 401412170017, 401412170018, 401412170019, 401412170020); Lot No. 62932247 (Serial Numbers: 401501020001, 401501020002, 401501020003, 401501020004, 401501020005, 401501020006, 401501020007, 401501020008, 401501020009, 401501020010, 401501020011, 401501020012, 401501020013, 401501020014, 401501020015, 401501020016, 401501020017, 401501020018, 401501020019, 401501020020); Lot No. 62932251 (Serial Numbers: 401501080001, 401501080002, 401501080003, 401501080004, 401501080005, 401501080006, 401501080007, 401501080008, 401501080009, 401501080010, 401501080011, 401501080012, 401501080013, 401501080014, 401501080015, 401501080016, 401501080017, 401501080018, 401501080019, 401501080020); Lot No. 62932249 (Serial Numbers: 401501080021, 401501080022, 401501080023, 401501080024, 401501080025, 401501080026, 401501080027, 401501080028, 401501080029, 401501080030, 401501080031, 401501080032, 401501080033, 401501080034, 401501080035, 401501080036, 401501080037, 401501080038, 401501080039, 401501080040); Lot No. 62932628 (Serial Numbers: 401501090001, 401501090002, 401501090003, 401501090004, 401501090005, 401501090006, 401501090007, 401501090008, 401501090009, 401501090010, 401501090011, 401501090012, 401501090013, 401501090014, 401501090015, 401501090016, 401501090017, 401501090018, 401501090019, 401501090020); Lot No. 62945084 (Serial Numbers: 401501090021, 401501090022, 401501090023, 401501090024, 401501090025); Lot No. 62932630 (Serial Numbers: 401501120001, 401501120002, 401501120003, 401501120004, 401501120005, 401501120006, 401501120007, 401501120008, 401501120009, 401501120010, 401501120011, 401501120012, 401501120013, 401501120014, 401501120015, 401501120016, 401501120017, 401501120018, 401501120019, 401501120020); Lot No. 62932632 (Serial Numbers: 401501130001, 401501130002, 401501130003, 401501130004, 401501130005, 401501130006, 401501130007, 401501130008, 401501130009, 401501130010, 401501130011, 401501130012, 401501130013, 401501130014, 401501130015, 401501130016, 401501130017, 401501130018, 401501130019, 401501130020); Lot No. 62932651 (Serial Numbers: 401501140001, 401501140002, 401501140003, 401501140004, 401501140005, 401501140006, 401501140007, 401501140008, 401501140009, 401501140010, 401501140011, 401501140012, 401501140013, 401501140014, 401501140015, 401501140016, 401501140017, 401501140018, 401501140019, 401501140020); Lot No. 62932652 (Serial Numbers: 401501150001, 401501150002, 401501150003, 401501150004, 401501150005, 401501150006, 401501150007, 401501150008, 401501150009, 401501150010, 401501150011, 401501150012, 401501150013, 401501150014, 401501150015, 401501150016, 401501150017, 401501150018, 401501150019, 401501150020); Lot No. 62932663 (Serial Numbers: 401501200001, 401501200002, 401501200003, 401501200004, 401501200005, 401501200006, 401501200007, 401501200008, 401501200009, 401501200010, 401501200011, 401501200012, 401501200013, 401501200014, 401501200015, 401501200016, 401501200017, 401501200018, 401501200019, 401501200020); Lot No. 62932666 (Serial Numbers: 401501210001, 401501210002, 401501210003, 401501210004, 401501210005, 401501210006, 401501210007, 401501210008, 401501210009, 401501210010, 401501210011, 401501210012, 401501210013, 401501210014, 401501210015, 401501210016, 401501210017, 401501210018, 401501210019, 401501210020); Lot No. 62962026 (Serial Numbers: 401501230001, 401501230002, 401501230003); Lot No. 62961808 (Serial Numbers: 401501260001, 401501260002, 401501260003, 401501260004, 401501260005, 401501260006, 401501260007, 401501260008, 401501260009, 401501260010, 401501260011, 401501260012, 401501260013, 401501260014, 401501260015, 401501260016, 401501260017
401501260018, 401501260019, 401501260020); Lot No. 62961810 (Serial Numbers: 401501290001, 401501290002, 401501290003, 401501290004, 401501290005, 401501290006, 401501290007, 401501290008, 401501290009, 401501290010, 401501290011, 401501290012, 401501290013, 401501290014, 401501290015, 401501290016, 401501290017, 401501290018, 401501290019, 401501290020); Lot No. 62945079 (Serial Numbers: 401502050001, 401502050002, 401502050003, 401502050004, 401502050005, 401502050006, 401502050007, 401502050008, 401502050009, 401502050010); Lot No. 62961812 (Serial Numbers: 401502060001, 401502060002, 401502060003, 401502060004, 401502060005, 401502060006, 401502060007, 401502060008, 401502060009, 401502060010, 401502060011, 401502060012, 401502060013, 401502060014, 401502060015, 401502060016, 401502060017, 401502060018, 401502060019, 401502060020); Lot No. 62969001 (Serial Numbers: 401502070001, 401502070002, 401502070003, 401502070004, 401502070005); Lot No. 62961815 (Serial Numbers: 401502120001, 401502120002, 401502120003, 401502120004, 401502120005, 401502120006, 401502120007, 401502120008, 401502120009, 401502120010, 401502120011, 401502120012, 401502120013, 401502120014, 401502120015, 401502120016, 401502120017, 401502120018, 401502120019, 401502120020); Lot No. 62961816 (Serial Numbers: 401502170001, 401502170002, 401502170003, 401502170004, 401502170005, 401502170006, 401502170007, 401502170008, 401502170009, 401502170010, 401502170011, 401502170012, 401502170013, 401502170014, 401502170015, 401502170016, 401502170017, 401502170018, 401502170019, 401502170020); Lot No. 62985038 (Serial Numbers: 401502180001, 401502180002, 401502180003, 401502180004, 401502180005, 401502180006, 401502180007, 401502180008, 401502180009, 401502180010); Lot No. 62961817 (Serial Numbers: 401502190001, 401502190002, 401502190003, 401502190004, 401502190005, 401502190006, 401502190007, 401502190008, 401502190009, 401502190010, 401502190011, 401502190012, 401502190013, 401502190014, 401502190015, 401502190016, 401502190017, 401502190018, 401502190019, 401502190020); Lot No. 62983026 (Serial Numbers: 401502190021, 401502190022, 401502190023, 401502190024, 401502190025); Lot No. 62961818 (Serial Numbers: 401502240001, 401502240002, 401502240003, 401502240004, 401502240005, 401502240006, 401502240007, 401502240008, 401502240009, 401502240010, 401502240011, 401502240012, 401502240013, 401502240014, 401502240015, 401502240016, 401502240017, 401502240018, 401502240019, 401502240020); Lot No. 62976482 (Serial Numbers: 401502260001, 401502260002, 401502260003, 401502260004, 401502260005, 401502260006, 401502260007, 401502260008, 401502260009, 401502260010, 401502260011, 401502260012, 401502260013, 401502260014, 401502260015, 401502260016, 401502260017, 401502260018, 401502260019, 401502260020); Lot No. 63000878 (Serial Numbers: 401503130001, 401503130002, 401503130003); Lot No. 63003147 (Serial Numbers: 401503190001, 401503190002, 401503190003, 401503190004, 401503190005); Lot No. 63003149 (Serial Numbers: 401503230001, 401503230002, 401503230003, 401503230004, 401503230005, 401503230006, 401503230007, 401503230008); Lot No. 62976483 (Serial Numbers: 401503310001, 401503310002, 401503310003, 401503310004, 401503310005, 401503310006, 401503310007, 401503310008, 401503310009, 401503310010, 401503310011, 401503310012, 401503310013, 401503310014, 401503310015, 401503310016, 401503310017, 401503310018, 401503310019, 401503310020); Lot No. 63007860 (Serial Numbers: 401504020001, 401504020002, 401504020003, 401504020004, 401504020005, 401504020006, 401504020007, 401504020008, 401504020009, 401504020010, 401504020011, 401504020012, 401504020013, 401504020014, 401504020015, 401504020016, 401504020017, 401504020018, 401504020019, 401504020020); Lot No. 63035646 (Serial Numbers: 401504060001, 401504060002, 401504060003, 401504060004, 401504060005, 401504060006, 401504060007, 401504060008); Lot No. 63035645 (Serial Numbers: 401504060009, 401504060010, 401504060011, 401504060012, 401504060013, 401504060014, 401504060015, 401504060016, 401504060017, 401504060018, 401504060019, 401504060020, 401504060021, 401504060022, 401504060023); Lot No. 63035647 (Serial Numbers: 401504090001, 401504090002, 401504090003, 401504090004, 401504090005, 401504090006, 401504090007, 401504090008, 401504090009, 401504090010); Lot No. 62976488 (Serial Numbers: 401504090011, 401504090012, 401504090013, 401504090014, 401504090015, 401504090016, 401504090017, 401504090018, 401504090019, 401504090020, 401504090021, 401504090022, 401504090023, 401504090024, 401504090025, 401504090026, 401504090027, 401504090028, 401504090029, 401504090030); Lot No. 63007861 (Serial Numbers: 401504150001, 401504150002, 401504150003, 401504150004, 401504150005, 401504150006, 401504150007, 401504150008, 401504150009, 401504150010, 401504150011, 401504150012, 401504150013, 401504150014, 401504150015, 401504150016, 401504150017, 401504150018, 401504150019, 401504150020); Lot No. 63007859 (Serial Numbers: 401504150021, 401504150022, 401504150023, 401504150024, 401504150025, 401504150026, 401504150027, 401504150028, 401504150029, 401504150030, 401504150031, 401504150032, 401504150033, 401504150034, 401504150035, 401504150036, 401504150037, 401504150038, 401504150039, 401504150040); Lot No. 63042185 (Serial Numbers: 401504200001, 401504200002, 401504200003, 401504200004, 401504200005); Lot No. 63007862 (Serial Numbers: 401504210001, 401504210002, 401504210003, 401504210004, 401504210005, 401504210006, 401504210007, 401504210008, 401504210009, 401504210010, 401504210011, 401504210012, 401504210013, 401504210014, 401504210015, 401504210016, 401504210017, 401504210018, 401504210019, 401504210020); Lot No. 63042186 (Serial Numbers: 401504210021, 401504210022, 401504210023, 401504210024, 401504210025, 401504210026, 401504210027, 401504210028, 401504210029, 401504210030, 401504210031, 401504210032, 401504210033, 401504210034, 401504210035, 401504210036, 401504210037, 401504210038, 401504210039, 401504210040); Lot No. 63007864 (Serial Numbers: 401504240001, 401504240002, 401504240003, 401504240004, 401504240005, 401504240006, 401504240007, 401504240008, 401504240009, 401504240010, 401504240011, 401504240012, 401504240013, 401504240014, 401504240015, 401504240016, 401504240017, 401504240018, 401504240019, 401504240020); Lot No. 63037265 (Serial Numbers: 401504270001, 401504270002, 401504270003, 401504270004, 401504270005, 401504270006, 401504270007, 401504270008, 401504270009, 401504270010, 401504270011, 401504270012, 401504270013, 401504270014, 401504270015, 401504270016, 401504270017, 401504270018, 401504270019, 401504270020); Lot No. 63037270 (Serial Numbers: 401504290001, 401504290002, 401504290003, 401504290004, 401504290005, 401504290006, 401504290007, 401504290008, 401504290009, 401504290010, 401504290011, 401504290012, 401504290013, 401504290014, 401504290015, 401504290016, 401504290017, 401504290018, 401504290019, 401504290020); Lot No. 63037272 (Serial Numbers: 401505010001, 401505010002, 401505010003, 401505010004, 401505010005, 401505010006, 401505010007, 401505010008, 401505010009, 401505010010, 401505010011, 401505010012, 401505010013, 401505010014, 401505010015, 401505010016, 401505010017, 401505010018, 401505010019, 401505010020); Lot No. 63037274 (Serial Numbers: 401505070001, 401505070002, 401505070003, 401505070004, 401505070005, 401505070006, 401505070007, 401505070008, 401505070009, 401505070010, 401505070011, 401505070012, 401505070013, 401505070014, 401505070015, 401505070016, 401505070017, 401505070018, 401505070019, 401505070020); Lot No. 63037276 (Serial Numbers: 401505110001, 401505110002, 401505110003, 401505110004, 401505110005, 401505110006, 401505110007, 401505110008, 401505110009, 401505110010, 401505110011, 401505110012, 401505110013, 401505110014, 401505110015, 401505110016, 401505110017, 401505110018, 401505110019, 401505110020); Lot No. 63072308 (Serial Numbers: 401505300001, 401505300002, 401505300003, 401505300004, 401505300005, 401505300006, 401505300007, 401505300008, 401505300009, 401505300010, 401505300011, 401505300012, 401505300013, 401505300014, 401505300015, 401505300016, 401505300017, 401505300018, 401505300019, 401505300020); Lot No. 63072309 (Serial Numbers: 401506020001, 401506020002, 401506020003, 401506020004, 401506020005, 401506020006, 401506020007, 401506020008, 401506020009, 401506020010, 401506020011, 401506020012, 401506020013, 401506020014, 401506020015, 401506020016, 401506020017, 401506020018, 401506020019, 401506020020); Lot No. 63072326 (Serial Numbers: 401506050001, 401506050002, 401506050003, 401506050004, 401506050005, 401506050006, 401506050007, 401506050008, 401506050009, 401506050010, 401506050011, 401506050012, 401506050013, 401506050014, 401506050015, 401506050016, 401506050017, 401506050018, 401506050019, 401506050020); Lot No. 63072327 (Serial Numbers: 401506060001, 401506060002, 401506060003, 401506060004, 401506060005, 401506060006, 401506060007, 401506060008, 401506060009, 401506060010, 401506060011, 401506060012, 401506060013, 401506060014, 401506060015, 401506060016, 401506060017, 401506060018, 401506060019, 401506060020); Lot No. 63092211 (Serial Numbers: 401506080001, 401506080002, 401506080003, 401506080004, 401506080005); Lot No. 63092212 (Serial Numbers: 401506100001, 401506100002, 401506100003); Lot No. 63072338 (Serial Numbers: 401506100004, 401506100005, 401506100006, 401506100007, 401506100008, 401506100009, 401506100010, 401506100011, 401506100012, 401506100013, 401506100014, 401506100015, 401506100016, 401506100017, 401506100018, 401506100019, 401506100020, 401506100021, 401506100022, 401506100023); Lot No. 63103592 (Serial Numbers: 401506130001, 401506130002, 401506130003, 401506130004, 401506130005); Lot No. 63072339 (Serial Numbers: 401506150001, 401506150002, 401506150003, 401506150004, 401506150005, 401506150006, 401506150007, 401506150008, 401506150009, 401506150010, 401506150011, 401506150012, 401506150013, 401506150014, 401506150015, 401506150016, 401506150017, 401506150018, 401506150019, 401506150020); Lot No. 63072354 (Serial Numbers: 401506200001, 401506200002, 401506200003, 401506200004, 401506200005, 401506200006, 401506200007, 401506200008, 401506200009, 401506200010, 401506200011, 401506200012, 401506200013, 401506200014, 401506200015, 401506200016, 401506200017, 401506200018, 401506200019, 401506200020); Lot No. 63096364 (Serial Numbers: 401506230001, 401506230002, 401506230003, 401506230004, 401506230005, 401506230006, 401506230007, 401506230008, 401506230009, 401506230010, 401506230011, 401506230012, 401506230013, 401506230014, 401506230015, 401506230016, 401506230017, 401506230018, 401506230019, 401506230020); Lot No. 63096365 (Serial Numbers: 401506260001, 401506260002, 401506260003, 401506260004, 401506260005, 401506260006, 401506260007, 401506260008, 401506260009, 401506260010, 401506260011, 401506260012, 401506260013, 401506260014, 401506260015, 401506260016, 401506260017, 401506260018, 401506260019, 401506260020); Lot No. 63095208 (Serial Numbers: 401507070001, 401507070002, 401507070003, 401507070004, 401507070005); Lot No. 63096437 (Serial Numbers: 401507080001, 401507080002, 401507080003, 401507080004, 401507080005, 401507080006, 401507080007, 401507080008, 401507080009, 401507080010, 401507080011, 401507080012, 401507080013, 401507080014, 401507080015, 401507080016, 401507080017, 401507080018, 401507080019, 401507080020); Lot No. 63096469 (Serial Numbers: 401507100001, 401507100002, 401507100003, 401507100004, 401507100005, 401507100006, 401507100007, 401507100008, 401507100009, 401507100010, 401507100011, 401507100012, 401507100013, 401507100014, 401507100015, 401507100016, 401507100017, 401507100018, 401507100019, 401507100020); Lot No. 63096484 (Serial Numbers: 401507140001, 401507140002, 401507140003, 401507140004, 401507140005, 401507140006, 401507140007, 401507140008, 401507140009, 401507140010, 401507140011, 401507140012, 401507140013, 401507140014, 401507140015, 401507140016, 401507140017, 401507140018, 401507140019, 401507140020); Lot No. 63116363 (Serial Numbers: 401507170001, 401507170002, 401507170003); Lot No. 63096485 (Serial Numbers:401507180001, 401507180002, 401507180003, 401507180004, 401507180005, 401507180006, 401507180007, 401507180008, 401507180009, 401507180010, 401507180011, 401507180012, 401507180013, 401507180014, 401507180015, 401507180016, 401507180017, 401507180018, 401507180019, 401507180020); Lot No. 63116361 (Serial Numbers: 401507210001, 401507210002, 401507210003, 401507210004, 401507210005); Lot No. 63123310 (Serial Numbers: 401507220001, 401507220002, 401507220003, 401507220004, 401507220005, 401507220006, 401507220007, 401507220008, 401507220009, 401507220010, 401507220011, 401507220012, 401507220013, 401507220014, 401507220015, 401507220016, 401507220017, 401507220018, 401507220019, 401507220020); Lot No. 63123678 (Serial Numbers: 401507230001, 401507230002, 401507230003, 401507230004, 401507230005); Lot No. 63123311 (Serial Numbers: 401507290001, 401507290002, 401507290003, 401507290004, 401507290005, 401507290006, 401507290007, 401507290008, 401507290009, 401507290010, 401507290011, 401507290012, 401507290013, 401507290014, 401507290015, 401507290016, 401507290017, 401507290018, 401507290019, 401507290020); Lot No. 63123347 (Serial Numbers: 401507310001, 401507310002, 401507310003, 401507310004, 401507310005, 401507310006, 401507310007, 401507310008, 401507310009, 401507310010, 401507310011, 401507310012, 401507310013, 401507310014, 401507310015, 401507310016, 401507310017, 401507310018, 401507310019, 401507310020); Lot No. 63123346 (Serial Numbers: 401508030001, 401508030002, 401508030003, 401508030004, 401508030005, 401508030006, 401508030007, 401508030008, 401508030009, 401508030010, 401508030011, 401508030012, 401508030013, 401508030014, 401508030015, 401508030016, 401508030017, 401508030018, 401508030019, 401508030020); Lot No. 63123373 (Serial Numbers: 401508050001, 401508050002, 401508050003, 401508050004, 401508050005, 401508050006, 401508050007, 401508050008, 401508050009, 401508050010, 401508050011, 401508050012, 401508050013, 401508050014, 401508050015, 401508050016, 401508050017, 401508050018, 401508050019, 401508050020); Lot No. 63151053 (Serial Numbers: 401508060001, 401508060002, 401508060003, 401508060004, 401508060005); Lot No. 63123378 (Serial Numbers: 401508080001, 401508080002, 401508080003, 401508080004, 401508080005, 401508080006, 401508080007, 401508080008, 401508080009, 401508080010, 401508080011, 401508080012, 401508080013, 401508080014, 401508080015, 401508080016, 401508080017, 401508080018, 401508080019, 401508080020); Lot No. 63123411 (Serial Numbers: 401508120001, 401508120002, 401508120003, 401508120004, 401508120005, 401508120006, 401508120007, 401508120008, 401508120009, 401508120010, 401508120011, 401508120012, 401508120013, 401508120014, 401508120015, 401508120016, 401508120017, 401508120018, 401508120019, 401508120020); Lot No. 63138529 (Serial Numbers: 401508190001, 401508190002, 401508190003, 401508190004, 401508190005, 401508190006, 401508190007, 401508190008, 401508190009, 401508190010, 401508190011, 401508190012, 401508190013, 401508190014, 401508190015, 401508190016, 401508190017, 401508190018, 401508190019, 401508190020); Lot No. 63138530 (Serial Numbers: 401508220001, 401508220002, 401508220003, 401508220004, 401508220005, 401508220006, 401508220007, 401508220008, 401508220009, 401508220010, 401508220011, 401508220012, 401508220013, 401508220014, 401508220015, 401508220016, 401508220017, 401508220018, 401508220019, 401508220020); Lot No. 63138555 (Serial Numbers: 401508280001, 401508280002, 401508280003, 401508280004, 401508280005, 401508280006, 401508280007, 401508280008, 401508280009, 401508280010, 401508280011, 401508280012, 401508280013, 401508280014, 401508280015, 401508280016, 401508280017, 401508280018, 401508280019, 401508280020); Lot No. 63138591 (Serial Numbers: 401508310001, 401508310002, 401508310003, 401508310004, 401508310005, 401508310006, 401508310007, 401508310008, 401508310009, 401508310010, 401508310011, 401508310012, 401508310013, 401508310014, 401508310015, 401508310016, 401508310017, 401508310018, 401508310019, 401508310020); Lot No. 63138592 (Serial Numbers: 401509030001, 401509030002, 401509030003, 401509030004, 401509030005, 401509030006, 401509030007, 401509030008, 401509030009, 401509030010, 401509030011, 401509030012, 401509030013, 401509030014, 401509030015, 401509030016, 401509030017, 401509030018, 401509030019, 401509030020); Lot No. 63138613 (Serial Numbers: 401509100001, 401509100002, 401509100003, 401509100004, 401509100005, 401509100006, 401509100007, 401509100008, 401509100009, 401509100010, 401509100011, 401509100012, 401509100013, 401509100014, 401509100015, 401509100016, 401509100017, 401509100018, 401509100019, 401509100020); Lot No. 63138635 (Serial Numbers: 401509150001, 401509150002, 401509150003, 401509150004, 401509150005, 401509150006, 401509150007, 401509150008, 401509150009, 401509150010, 401509150011, 401509150012, 401509150013, 401509150014, 401509150015, 401509150016, 401509150017, 401509150018, 401509150019, 401509150020); Lot No. 63183292 (Serial Numbers: 401509170001, 401509170002, 401509170003, 401509170004, 401509170005, 401509170006, 401509170007, 401509170008); Lot No. 63171062 (Serial Numbers:401509220001, 401509220002, 401509220003, 401509220004, 401509220005, 401509220006, 401509220007, 401509220008, 401509220009, 401509220010, 401509220011, 401509220012, 401509220013, 401509220014, 401509220015, 401509220016, 401509220017, 401509220018, 401509220019, 401509220020); Lot No. 63171063 (Serial Numbers: 401509250001, 401509250002, 401509250003, 401509250004, 401509250005, 401509250006, 401509250007, 401509250008, 401509250009, 401509250010, 401509250011, 401509250012, 401509250013, 401509250014, 401509250015, 401509250016, 401509250017, 401509250018, 401509250019, 401509250020); Lot No. 63171091 (Serial Numbers: 401509300001, 401509300002, 401509300003, 401509300004, 401509300005, 401509300006, 401509300007, 401509300008, 401509300009, 401509300010, 401509300011, 401509300012, 401509300013, 401509300014, 401509300015, 401509300016, 401509300017, 401509300018, 401509300019, 401509300020); Lot No. 63171130 (Serial Numbers: 401510060001, 401510060002, 401510060003, 401510060004, 401510060005, 401510060006, 401510060007, 401510060008, 401510060009, 401510060010, 401510060011, 401510060012, 401510060013, 401510060014, 401510060015, 401510060016, 401510060017, 401510060018, 401510060019, 401510060020); Lot No. 63171131 (Serial Numbers: 401510150001, 401510150002, 401510150003, 401510150004, 401510150005, 401510150006, 401510150007, 401510150008, 401510150009, 401510150010, 401510150011, 401510150012, 401510150013, 401510150014, 401510150015, 401510150016, 401510150017, 401510150018, 401510150019, 401510150020); Lot No. 63205822 (Serial Numbers: 401510150021, 401510150022, 401510150023, 401510150024, 401510150025); Lot No. 63207374 (Serial Numbers: 401510200001, 401510200002, 401510200003); Lot No. 63171148 (Serial Numbers: 401510210001, 401510210002, 401510210003, 401510210004, 401510210005, 401510210006, 401510210007, 401510210008, 401510210009, 401510210010, 401510210011, 401510210012, 401510210013, 401510210014, 401510210015, 401510210016, 401510210017, 401510210018, 401510210019, 401510210020); Lot No. 63216486 (Serial Numbers: 401510280001, 401510280002, 401510280003, 401510280004, 401510280005, 401510280006, 401510280007, 401510280008, 401510280009, 401510280010, 401510280011, 401510280012, 401510280013, 401510280014, 401510280015); Lot No. 63218419 (Serial Numbers: 401510290001, 401510290002, 401510290003, 401510290004, 401510290005); Lot No. 63216489 (Serial Numbers: 401510290006, 401510290007, 401510290008, 401510290009, 401510290010); Lot No. 63216488 (Serial Numbers: 401510290011, 401510290012, 401510290013, 401510290014, 401510290015, 401510290016, 401510290017, 401510290018, 401510290019, 401510290020, 401510290021, 401510290022, 401510290023, 401510290024, 401510290025); Lot No. 63196945 (Serial Numbers: 401510290026, 401510290027, 401510290028, 401510290029, 401510290030, 401510290031, 401510290032, 401510290033, 401510290034, 401510290035, 401510290036, 401510290037, 401510290038, 401510290039, 401510290040, 401510290041, 401510290042, 401510290043, 401510290044, 401510290045); Lot No. 63196943 (Serial Numbers: 401511050001, 401511050002, 401511050003, 401511050004, 401511050005, 401511050006, 401511050007, 401511050008, 401511050009, 401511050010, 401511050011, 401511050012, 401511050013, 401511050014, 401511050015, 401511050016, 401511050017, 401511050018, 401511050019, 401511050020); Lot No. 63196979 (Serial Numbers: 401511060001, 401511060002, 401511060003, 401511060004, 401511060005, 401511060006, 401511060007, 401511060008, 401511060009, 401511060010, 401511060011, 401511060012, 401511060013, 401511060014, 401511060015, 401511060016, 401511060017, 401511060018, 401511060019, 401511060020); Lot No. 63202479 (Serial Numbers: 401511110001, 401511110002, 401511110003, 401511110004, 401511110005, 401511110006, 401511110007, 401511110008, 401511110009, 401511110010, 401511110011, 401511110012, 401511110013, 401511110014, 401511110015, 401511110016, 401511110017, 401511110018, 401511110019, 401511110020); Lot No. 63223990 (Serial Numbers: 401511120001, 401511120002, 401511120003, 401511120004, 401511120005, 401511120006, 401511120007, 401511120008, 401511120009, 401511120010, 401511120011, 401511120012, 401511120013, 401511120014, 401511120015, 401511120016, 401511120017, 401511120018, 401511120019, 401511120020); Lot No. 63197009 (Serial Numbers: 401511130001, 401511130002, 401511130003, 401511130004, 401511130005, 401511130006, 401511130007, 401511130008, 401511130009, 401511130010, 401511130011, 401511130012, 401511130013, 401511130014, 401511130015, 401511130016, 401511130017, 401511130018, 401511130019, 401511130020); Lot No. 63197010 (Serial Numbers: 401511140001, 401511140002, 401511140003, 401511140004, 401511140005, 401511140006, 401511140007, 401511140008, 401511140009, 401511140010, 401511140011, 401511140012, 401511140013, 401511140014, 401511140015, 401511140016, 401511140017, 401511140018, 401511140019, 401511140020); Lot No. 63223991 (Serial Numbers: 401511160001, 401511160002, 401511160003, 401511160004, 401511160005, 401511160006, 401511160007, 401511160008, 401511160009, 401511160010, 401511160011, 401511160012, 401511160013, 401511160014, 401511160015, 401511160016, 401511160017, 401511160018, 401511160019, 401511160020); Lot No. 63223992 (Serial Numbers: 401511170001, 401511170002, 401511170003, 401511170004, 401511170005, 401511170006, 401511170007, 401511170008, 401511170009, 401511170010, 401511170011, 401511170012, 401511170013, 401511170014, 401511170015, 401511170016, 401511170017, 401511170018, 401511170019, 401511170020); Lot No. 63197099 (Serial Numbers: 401511180001, 401511180002, 401511180003, 401511180004, 401511180005, 401511180006, 401511180007, 401511180008, 401511180009, 401511180010, 401511180011, 401511180012, 401511180013, 401511180014, 401511180015, 401511180016, 401511180017, 401511180018, 401511180019, 401511180020); Lot No. 63212346 (Serial Numbers: 401511200001, 401511200002, 401511200003, 401511200004, 401511200005, 401511200006, 401511200007, 401511200008, 401511200009, 401511200010, 401511200011, 401511200012, 401511200013, 401511200014, 401511200015, 401511200016, 401511200017, 401511200018, 401511200019, 401511200020); Lot No. 63212347 (Serial Numbers: 401511210001, 401511210002, 401511210003, 401511210004, 401511210005, 401511210006, 401511210007, 401511210008, 401511210009, 401511210010, 401511210011, 401511210012, 401511210013, 401511210014, 401511210015, 401511210016, 401511210017, 401511210018, 401511210019, 401511210020); Lot No. 63226932 (Serial Numbers: 401511210021, 401511210022, 401511210023, 401511210024, 401511210025, 401511210026, 401511210027, 401511210028, 401511210029, 401511210030, 401511210031, 401511210032, 401511210033, 401511210034, 401511210035, 401511210036, 401511210037, 401511210038, 401511210039, 401511210040); Lot No. 63212372 (Serial Numbers: 401511230001, 401511230002, 401511230003, 401511230004, 401511230005, 401511230006, 401511230007, 401511230008, 401511230009, 401511230010, 401511230011, 401511230012, 401511230013, 401511230014, 401511230015, 401511230016, 401511230017, 401511230018, 401511230019, 401511230020); Lot No. 63212373 (Serial Numbers: 401511240001, 401511240002, 401511240003, 401511240004, 401511240005, 401511240006, 401511240007, 401511240008, 401511240009, 401511240010, 401511240011, 401511240012, 401511240013, 401511240014, 401511240015, 401511240016, 401511240017, 401511240018, 401511240019, 401511240020); Lot No. 63226943 (Serial Numbers: 401511250001 401511250002 401511250003 401511250004 401511250005 401511250006 401511250007 401511250008 401511250009 401511250010 401511250011 401511250012 401511250013 401511250014 401511250015 401511250016 401511250017 401511250018 401511250019 401511250020 Lot No. 63212409 (Serial Numbers: 401511300001 401511300002 401511300003 401511300004 401511300005 401511300006 401511300007 401511300008 401511300009 401511300010 401511300011 401511300012 401511300013 401511300014 401511300015 401511300016 401511300017 401511300018 401511300019 401511300020 Lot No. 63212410 (Serial Numbers: 401512010001 401512010002 401512010003 401512010004 401512010005 401512010006 401512010007 401512010008 401512010009 401512010010 401512010011 401512010012 401512010013 401512010014 401512010015 401512010016 401512010017 401512010018 401512010019 401512010020 Lot No. 63235444 (Serial Numbers: 401512020001 401512020002 401512020003 401512020004 401512020005 401512020006 401512020007 401512020008 401512020009 401512020010 401512020011 401512020012 401512020013 401512020014 401512020015 401512020016 401512020017 401512020018 401512020019 401512020020 Lot No. 63235445 (Serial Numbers: 401512030001 401512030002 401512030003 401512030004 401512030005 401512030006 401512030007 401512030008 401512030009 401512030010 401512030011 401512030012 401512030013 401512030014 401512030015 401512030016 401512030017 401512030018 401512030019 401512030020 Lot No. 63212420 (Serial Numbers: 401512040001 401512040002 401512040003 401512040004 401512040005 401512040006 401512040007 401512040008 401512040009 401512040010 401512040011 401512040012 401512040013 401512040014 401512040015 401512040016 401512040017 401512040018 401512040019 401512040020 Lot No. 63234566 (Serial Numbers: 401512040021 401512040022 401512040023 401512040024 401512040025 Lot No. 63244826 (Serial Numbers: 401512040026 401512040027 401512040028 401512040029 Lot No. 63223993 (Serial Numbers: 401512080001 401512080002 401512080003 401512080004 401512080005 401512080006 401512080007 401512080008 401512080009 401512080010 401512080011 401512080012 401512080013 401512080014 401512080015 401512080016 401512080017 401512080018 401512080019 401512080020 Lot No. 63235446 (Serial Numbers: 401512090001 401512090002 401512090003 401512090004 401512090005 401512090006 401512090007 401512090008 401512090009 401512090010 401512090011 401512090012 401512090013 401512090014 401512090015 401512090016 401512090017 401512090018 401512090019 401512090020 Lot No. 63247372 (Serial Numbers: 401512110001 401512110002 401512110003 401512110004 401512110005 401512110006 401512110007 401512110008 401512110009 401512110010 401512110011 401512110012 401512110013 401512110014 401512110015 401512110016 401512110017 401512110018 401512110019 401512110020 Lot No. 63250538 (Serial Numbers: 401512110021 401512110022 401512110023 401512110024 401512110025 Lot No. 63245753 (Serial Numbers: 401512110026 401512110027 401512110028 401512110029 401512110030 401512110031 401512110032 401512110033 401512110034 401512110035 Lot No. 63247393 (Serial Numbers: 401512170001 401512170002 401512170003 401512170004 401512170005 401512170006 401512170007 401512170008 401512170009 401512170010 401512170011 401512170012 401512170013 401512170014 401512170015 401512170016 401512170017 401512170018 401512170019 401512170020 Lot No. 63247415 (Serial Numbers: 401512210001 401512210002 401512210003 401512210004 401512210005 401512210006 401512210007 401512210008 401512210009 401512210010 401512210011 401512210012 401512210013 401512210014 401512210015 401512210016 401512210017 401512210018 401512210019 401512210020 Lot No. 63261578 (Serial Numbers: 401601050001 401601050002 401601050003 401601050004 401601050005 401601050006 401601050007 401601050008 401601050009 401601050010 401601050011 401601050012 401601050013 401601050014 401601050015 401601050016 401601050017 401601050018 401601050019 401601050020 Lot No. 63261585 (Serial Numbers: 401601080001 401601080002 401601080003 401601080004 401601080005 Lot No. 63264446 (Serial Numbers: 401601120001 401601120002 401601120003 401601120004 401601120005 401601120006 401601120007 401601120008 401601120009 401601120010 401601120011 401601120012 401601120013 401601120014 401601120015 401601120016 401601120017 401601120018 401601120019 401601120020 Lot No. 63247427 (Serial Numbers: 401601180001 401601180002 401601180003 401601180004 401601180005 401601180006 401601180007 401601180008 401601180009 401601180010 401601180011 401601180012 401601180013 401601180014 401601180015 401601180016 401601180017 401601180018 401601180019 401601180020 Lot No. 63271204 (Serial Numbers: 401601200001 401601200002 401601200003 Lot No. 63264369 (Serial Numbers: 401601210001 401601210002 401601210003 401601210004 401601210005 401601210006 401601210007 401601210008 401601210009 401601210010 401601210011 401601210012 401601210013 401601210014 401601210015 401601210016 401601210017 401601210018 401601210019 401601210020 Lot No. 63257989 (Serial Numbers: 401601250001 401601250002 401601250003 401601250004 401601250005 401601250006 401601250007 401601250008 401601250009 401601250010 401601250011 401601250012 401601250013 401601250014 401601250015 401601250016 401601250017 401601250018 401601250019 401601250020 Lot No. 63271203 (Serial Numbers: 401601260001 401601260002 401601260003 401601260004 401601260005 Lot No. 63257990 (Serial Numbers: 401601270001 401601270002 401601270003 401601270004 401601270005 401601270006 401601270007 401601270008 401601270009 401601270010 401601270011 401601270012 401601270013 401601270014 401601270015 401601270016 401601270017 401601270018 401601270019 401601270020 Lot No. 63277184 (Serial Numbers: 401601280001 401601280002 401601280003 Lot No. 63258008 (Serial Numbers: 401601290001 401601290002 401601290003 401601290004 401601290005 401601290006 401601290007 401601290008 401601290009 401601290010 401601290011 401601290012 401601290013 401601290014 401601290015 401601290016 401601290017 401601290018 401601290019 401601290020 Lot No. 63282048 (Serial Numbers: 401602010001 401602010002 401602010003 401602010004 401602010005 401602010006 401602010007 401602010008 401602010009 401602010010 401602010011 401602010012 401602010013 401602010014 401602010015 401602010016 401602010017 401602010018 401602010019 401602010020 Lot No. 63258009 (Serial Numbers: 401602020001 401602020002 401602020003 401602020004 401602020005 401602020006 401602020007 401602020008 401602020009 401602020010 401602020011 401602020012 401602020013 401602020014 401602020015 401602020016 401602020017 401602020018 401602020019 401602020020 Lot No. 63282049 (Serial Numbers: 401602030001 401602030002 401602030003 401602030004 401602030005 401602030006 401602030007 401602030008 401602030009 401602030010 401602030011 401602030012 401602030013 401602030014 401602030015 401602030016 401602030017 401602030018 401602030019 401602030020 Lot No. 63258024 (Serial Numbers: 401602040001 401602040002 401602040003 401602040004 401602040005 401602040006 401602040007 401602040008 401602040009 401602040010 401602040011 401602040012 401602040013 401602040014 401602040015 401602040016 401602040017 401602040018 401602040019 401602040020 Lot No. 63264372 (Serial Numbers: 401602040021 401602040022 401602040023 401602040024 401602040025 401602040026 401602040027 401602040028 401602040029 401602040030 401602040031 401602040032 401602040033 401602040034 401602040035 401602040036 401602040037 401602040038 401602040039 401602040040 Lot No. 63282051 (Serial Numbers: 401602050001 401602050002 401602050003 401602050004 401602050005 401602050006 401602050007 401602050008 401602050009 401602050010 401602050011 401602050012 401602050013 401602050014 401602050015 401602050016 401602050017 401602050018 401602050019 401602050020 Lot No. 63258041 (Serial Numbers: 401602090001 401602090002 401602090003 401602090004 401602090005 401602090006 401602090007 401602090008 401602090009 401602090010 401602090011 401602090012 401602090013 401602090014 401602090015 401602090016 401602090017 401602090018 401602090019 401602090020 Lot No. 63294145 (Serial Numbers: 401602100001 401602100002 401602100003 401602100004 401602100005 401602100006 401602100007 401602100008 401602100009 401602100010 401602100011 401602100012 401602100013 401602100014 401602100015 401602100016 401602100017 401602100018 401602100019 401602100020 401602100021 401602100022 Lot No. 63258042 (Serial Numbers: 401602120001 401602120002 401602120003 401602120004 401602120005 401602120006 401602120007 401602120008 401602120009 401602120010 401602120011 401602120012 401602120013 401602120014 401602120015 401602120016 401602120017 401602120018 401602120019 401602120020 Lot No. 63282053 (Serial Numbers: 401602170001 401602170002 401602170003 401602170004 401602170005 401602170006 401602170007 401602170008 401602170009 401602170010 401602170011 401602170012 401602170013 401602170014 401602170015 401602170016 401602170017 401602170018 401602170019 401602170020 Lot No. 63293873 (Serial Numbers: 401602170021 401602170022 401602170023 Lot No. 63278368 (Serial Numbers: 401602190001 401602190002 401602190003 401602190004 401602190005 401602190006 401602190007 401602190008 401602190009 401602190010 401602190011 401602190012 401602190013 401602190014 401602190015 401602190016 401602190017 401602190018 401602190019 401602190020 Lot No. 63282054 (Serial Numbers: 401602200001 401602200002 401602200003 401602200004 401602200005 401602200006 401602200007 401602200008 401602200009 401602200010 401602200011 401602200012 401602200013 401602200014 401602200015 401602200016 401602200017 401602200018 401602200019 401602200020 Lot No. 63278370 (Serial Numbers: 401602220001 401602220002 401602220003 401602220004 401602220005 401602220006 401602220007 401602220008 401602220009 401602220010 401602220011 401602220012 401602220013 401602220014 401602220015 401602220016 401602220017 401602220018 401602220019 401602220020 Lot No. 63278391 (Serial Numbers: 401602260001 401602260002 401602260003 401602260004 401602260005 401602260006 401602260007 401602260008 401602260009 401602260010 401602260011 401602260012 401602260013 401602260014 401602260015 401602260016 401602260017 401602260018 401602260019 401602260020 Lot No. 63278392 (Serial Numbers: 401602270001 401602270002 401602270003 401602270004 401602270005 401602270006 401602270007 401602270008 401602270009 401602270010 401602270011 401602270012 401602270013 401602270014 401602270015 401602270016 401602270017 401602270018 401602270019 401602270020 Lot No. 63282055 (Serial Numbers: 401603010001 401603010002 401603010003 401603010004 401603010005 401603010006 401603010007 401603010008 401603010009 401603010010 401603010011 401603010012 401603010013 401603010014 401603010015 401603010016 401603010017 401603010018 401603010019 401603010020 Lot No. 63278407 (Serial Numbers: 401603010021 401603010022 401603010023 401603010024 401603010025 401603010026 401603010027 401603010028 401603010029 401603010030 401603010031 401603010032 401603010033 401603010034 401603010035 401603010036 401603010037 401603010038 401603010039 401603010040 Lot No. 63278408 (Serial Numbers: 401603030001 401603030002 401603030003 401603030004 401603030005 401603030006 401603030007 401603030008 401603030009 401603030010 401603030011 401603030012 401603030013 401603030014 401603030015 401603030016 401603030017 401603030018 401603030019 401603030020 Lot No. 63282056 (Serial Numbers: 401603040001 401603040002 401603040003 401603040004 401603040005 401603040006 401603040007 401603040008 401603040009 401603040010 401603040011 401603040012 401603040013 401603040014 401603040015 401603040016 401603040017 401603040018 401603040019 401603040020 Lot No. 63314123 (Serial Numbers: 401603080001 401603080002 401603080003 401603080004 401603080005 Lot No. 63278422 (Serial Numbers: 401603100001 401603100002 401603100003 401603100004 401603100005 401603100006 401603100007 401603100008 401603100009 401603100010 401603100011 401603100012 401603100013 401603100014 401603100015 401603100016 401603100017 401603100018 401603100019 401603100020 Lot No. 63278423 (Serial Numbers: 401603110001 401603110002 401603110003 401603110004 401603110005 401603110006 401603110007 401603110008 401603110009 401603110010 401603110011 401603110012 401603110013 401603110014 401603110015 401603110016 401603110017 401603110018 401603110019 401603110020 Lot No. 63282057 (Serial Numbers: 401603140001 401603140002 401603140003 401603140004 401603140005 401603140006 401603140007 401603140008 401603140009 401603140010 401603140011 401603140012 401603140013 401603140014 401603140015 401603140016 401603140017 401603140018 401603140019 401603140020 Lot No. 63314119 (Serial Numbers: 401603140021 401603140022 401603140023 401603140024 401603140025 Lot No. 63314120 (Serial Numbers: 401603160001 401603160002 401603160003 401603160004 401603160005 Lot No. 63278438 (Serial Numbers: 401603170001 401603170002 401603170003 401603170004 401603170005 401603170006 401603170007 401603170008 401603170009 401603170010 401603170011 401603170012 401603170013 401603170014 401603170015 401603170016 401603170017 401603170018 401603170019 401603170020 Lot No. 63278439 (Serial Numbers: 401603180001 401603180002 401603180003 401603180004 401603180005 401603180006 401603180007 401603180008 401603180009 401603180010 401603180011 401603180012 401603180013 401603180014 401603180015 401603180016 401603180017 401603180018 401603180019 401603180020 Lot No. 63324624 (Serial Numbers: 401603180021 401603180022 401603180023 401603180024 401603180025 Lot No. 63282058 (Serial Numbers: 401603220001 401603220002 401603220003 401603220004 401603220005 401603220006 401603220007 401603220008 401603220009 401603220010 401603220011 401603220012 401603220013 401603220014 401603220015 401603220016 401603220017 401603220018 401603220019 401603220020 Lot No. 63324623 (Serial Numbers: 401603230001 401603230002 401603230003 Lot No. 63314013 (Serial Numbers: 401603240001 401603240002 401603240003 401603240004 401603240005 401603240006 401603240007 401603240008 401603240009 401603240010 401603240011 401603240012 401603240013 401603240014 401603240015 401603240016 401603240017 401603240018 401603240019 401603240020 Lot No. 63314015 (Serial Numbers: 401604010001 401604010002 401604010003 401604010004 401604010005 401604010006 401604010007 401604010008 401604010009 401604010010 401604010011 401604010012 401604010013 401604010014 401604010015 401604010016 401604010017 401604010018 401604010019 401604010020 Lot No. 63314034 (Serial Numbers: 401604040001 401604040002 401604040003 401604040004 401604040005 401604040006 401604040007 401604040008 401604040009 401604040010 401604040011 401604040012 401604040013 401604040014 401604040015 401604040016 401604040017 401604040018 401604040019 401604040020 Lot No. 63314035 (Serial Numbers: 401604060001 401604060002 401604060003 401604060004 401604060005 401604060006 401604060007 401604060008 401604060009 401604060010 401604060011 401604060012 401604060013 401604060014 401604060015 401604060016 401604060017 401604060018 401604060019 401604060020 Lot No. 63345667 (Serial Numbers: 401604260001 401604260002 401604260003 401604260004 401604260005 401604260006 401604260007 401604260008 401604260009 401604260010 401604260011 401604260012 401604260013 401604260014 401604260015 Lot No. 63345668 (Serial Numbers: 401604270001 401604270002 401604270003 401604270004 401604270005 401604270006 401604270007 401604270008 401604270009 401604270010 401604270011 401604270012 401604270013 401604270014 401604270015 Lot No. 63352350 (Serial Numbers: 401604270016 401604270017 401604270018 Lot No. 63345670 (Serial Numbers: 401604280001 401604280002 401604280003 401604280004 401604280005 401604280006 401604280007 401604280008 401604280009 401604280010 401604280011 401604280012 401604280013 401604280014 401604280015 Lot No. 63345669 (Serial Numbers: 401604300001 401604300002 401604300003 Lot No. 63352351 (Serial Numbers: 401605040001 401605040002 401605040003 Lot No. 63345673 (Serial Numbers: 401605050001 401605050002 401605050003 401605050004 401605050005 401605050006 401605050007 401605050008 401605050009 401605050010 401605050011 401605050012 401605050013 401605050014 401605050015 Lot No. 63369288 (Serial Numbers: 401605060001 401605060002 401605060003 401605060004 401605060005 Lot No. 63314089 (Serial Numbers: 401605090001 401605090002 401605090003 401605090004 401605090005 401605090006 401605090007 401605090008 401605090009 401605090010 401605090011 401605090012 401605090013 401605090014 401605090015 401605090016 401605090017 401605090018 401605090019 401605090020 Lot No. 63314090 (Serial Numbers: 401605100001 401605100002 401605100003 401605100004 401605100005 401605100006 401605100007 401605100008 401605100009 401605100010 401605100011 401605100012 401605100013 401605100014 401605100015 401605100016 401605100017 401605100018 401605100019 401605100020 Lot No. 63314110 (Serial Numbers: 401605120001 401605120002 401605120003 401605120004 401605120005 401605120006 401605120007 401605120008 401605120009 401605120010 401605120011 401605120012 401605120013 401605120014 401605120015 401605120016 401605120017 401605120018 401605120019 401605120020 Lot No. 63314111 (Serial Numbers: 401605130001 401605130002 401605130003 401605130004 401605130005 401605130006 401605130007 401605130008 401605130009 401605130010 401605130011 401605130012 401605130013 401605130014 401605130015 401605130016 401605130017 401605130018 401605130019 401605130020 Lot No. 63339446 (Serial Numbers: 401605160001 401605160002 401605160003 401605160004 401605160005 401605160006 401605160007 401605160008 401605160009 401605160010 401605160011 401605160012 401605160013 401605160014 401605160015 401605160016 401605160017 401605160018 401605160019 401605160020 Lot No. 63339449 (Serial Numbers: 401605200001 401605200002 401605200003 401605200004 401605200005 401605200006 401605200007 401605200008 401605200009 401605200010 401605200011 401605200012 401605200013 401605200014 401605200015 401605200016 401605200017 401605200018 401605200019 401605200020 Lot No. 63339477 (Serial Numbers: 401605250001 401605250002 401605250003 401605250004 401605250005 401605250006 401605250007 401605250008 401605250009 401605250010 401605250011 401605250012 401605250013 401605250014 401605250015 401605250016 401605250017 401605250018 401605250019 401605250020 Lot No. 63339478 (Serial Numbers: 401605260001 401605260002 401605260003 401605260004 401605260005 401605260006 401605260007 401605260008 401605260009 401605260010 401605260011 401605260012 401605260013 401605260014 401605260015 401605260016 401605260017 401605260018 401605260019 401605260020 Lot No. 63339502 (Serial Numbers: 401606030001 401606030002 401606030003 401606030004 401606030005 401606030006 401606030007 401606030008 401606030009 401606030010 401606030011 401606030012 401606030013 401606030014 401606030015 401606030016 401606030017 401606030018 401606030019 401606030020 Lot No. 63339503 (Serial Numbers: 401606060001 401606060002 401606060003 401606060004 401606060005 401606060006 401606060007 401606060008 401606060009 401606060010 401606060011 401606060012 401606060013 401606060014 401606060015 401606060016 401606060017 401606060018 401606060019 401606060020 Lot No. 63339527 (Serial Numbers: 401606090001 401606090002 401606090003 401606090004 401606090005 401606090006 401606090007 401606090008 401606090009 401606090010 401606090011 401606090012 401606090013 401606090014 401606090015 401606090016 401606090017 401606090018 401606090019 401606090020 Lot No. 63339529 (Serial Numbers: 401606130001 401606130002 401606130003 401606130004 401606130005 401606130006 401606130007 401606130008 401606130009 401606130010 401606130011 401606130012 401606130013 401606130014 401606130015 401606130016 401606130017 401606130018 401606130019 401606130020 Lot No. 63403016 (Serial Numbers: 401606140001 401606140002 401606140003 401606140004 401606140005 401606140006 401606140007 401606140008 401606140009 401606140010 401606140011 401606140012 401606140013 401606140014 401606140015 Lot No. 63369556 (Serial Numbers: 401606150001 401606150002 401606150003 401606150004 401606150005 401606150006 401606150007 401606150008 401606150009 401606150010 401606150011 401606150012 401606150013 401606150014 401606150015 401606150016 401606150017 401606150018 401606150019 401606150020 Lot No. 63408712 (Serial Numbers: 401606170001 401606170002 401606170003 401606170004 401606170005 401606170006 401606170007 401606170008 401606170009 401606170010 401606170011 401606170012 401606170013 401606170014 401606170015 401606170016 401606170017 401606170018 401606170019 401606170020 Lot No. 63403014 (Serial Numbers: 401606170021 401606170022 401606170023 401606170024 401606170025 Lot No. 63369618 (Serial Numbers: 401606220001 401606220002 401606220003 401606220004 401606220005 401606220006 401606220007 401606220008 401606220009 401606220010 401606220011 401606220012 401606220013 401606220014 401606220015 401606220016 401606220017 401606220018 401606220019 401606220020 Lot No. 63394137 (Serial Numbers: 401606270001 401606270002 401606270003 401606270004 401606270005 401606270006 401606270007 401606270008 401606270009 401606270010 401606270011 401606270012 401606270013 401606270014 401606270015 401606270016 401606270017 401606270018 401606270019 401606270020 Lot No. 63394138 (Serial Numbers: 401606300001 401606300002 401606300003 401606300004 401606300005 401606300006 401606300007 401606300008 401606300009 401606300010 401606300011 401606300012 401606300013 401606300014 401606300015 401606300016 401606300017 401606300018 401606300019 401606300020 Lot No. 63410586 (Serial Numbers: 401607010001 401607010002 401607010003 401607010004 401607010005 Lot No. 63410587 (Serial Numbers: 401607070001 401607070002 401607070003 401607070004 401607070005 Lot No. 63394151 (Serial Numbers: 401607120001 401607120002 401607120003 401607120004 401607120005 401607120006 401607120007 401607120008 401607120009 401607120010 401607120011 401607120012 401607120013 401607120014 401607120015 401607120016 401607120017 401607120018 401607120019 401607120020 Lot No. 63403021 (Serial Numbers: 401607120021 401607120022 401607120023 Lot No. 63422765 (Serial Numbers: 401607130001 401607130002 401607130003 401607130004 401607130005 401607130006 401607130007 401607130008 401607130009 401607130010 401607130011 401607130012 401607130013 401607130014 401607130015 401607130016 401607130017 401607130018 401607130019 401607130020 Lot No. 63394150 (Serial Numbers: 401607190001 401607190002 401607190003 401607190004 401607190005 401607190006 401607190007 401607190008 401607190009 401607190010 401607190011 401607190012 401607190013 401607190014 401607190015 401607190016 401607190017 401607190018 401607190019 401607190020 Lot No. 63410590 (Serial Numbers: 401607200001 401607200002 401607200003 401607200004 401607200005 Lot No. 63410591 (Serial Numbers: 401607220001 401607220002 401607220003 401607220004 401607220005 Lot No. 63394167 (Serial Numbers: 401607230001 401607230002 401607230003 401607230004 401607230005 401607230006 401607230007 401607230008 401607230009 401607230010 401607230011 401607230012 401607230013 401607230014 401607230015 401607230016 401607230017 401607230018 401607230019 401607230020 Lot No. 63410592 (Serial Numbers: 401607250001 401607250002 401607250003 401607250004 401607250005 Lot No. 63436339 (Serial Numbers: 401607270001 401607270002 401607270003 401607270004 401607270005 401607270006 401607270007 401607270008 401607270009 401607270010 Lot No. 63394168 (Serial Numbers: 401607280001 401607280002 401607280003 401607280004 401607280005 401607280006 401607280007 401607280008 401607280009 401607280010 401607280011 401607280012 401607280013 401607280014 401607280015 401607280016 401607280017 401607280018 401607280019 401607280020 Lot No. 63394178 (Serial Numbers: 401608010001 401608010002 401608010003 401608010004 401608010005 401608010006 401608010007 401608010008 401608010009 401608010010 401608010011 401608010012 401608010013 401608010014 401608010015 401608010016 401608010017 401608010018 401608010019 401608010020 Lot No. 63394179 (Serial Numbers: 401608020001 401608020002 401608020003 401608020004 401608020005 401608020006 401608020007 401608020008 401608020009 401608020010 401608020011 401608020012 401608020013 401608020014 401608020015 401608020016 401608020017 401608020018 401608020019 401608020020 Lot No. 63430479 (Serial Numbers: 401608030001 401608030002 401608030003 401608030004 401608030005 401608030006 401608030007 401608030008 401608030009 401608030010 401608030011 401608030012 401608030013 401608030014 401608030015 401608030016 401608030017 401608030018 401608030019 401608030020 Lot No. 63410594 (Serial Numbers: 401608030021 401608030022 401608030023 401608030024 401608030025 Lot No. 63445184 (Serial Numbers: 401608040001 401608040002 401608040003 401608040004 401608040005 401608040006 401608040007 401608040008 401608040009 401608040010 401608040011 401608040012 401608040013 401608040014 401608040015 401608040016 401608040017 401608040018 401608040019 401608040020 Lot No. 63430504 (Serial Numbers: 401608050001 401608050002 401608050003 401608050004 401608050005 401608050006 401608050007 401608050008 401608050009 401608050010 401608050011 401608050012 401608050013 401608050014 401608050015 401608050016 401608050017 401608050018 401608050019 401608050020 Lot No. 63445185 (Serial Numbers: 401608080001 401608080002 401608080003 401608080004 401608080005 401608080006 401608080007 401608080008 401608080009 401608080010 401608080011 401608080012 401608080013 401608080014 401608080015 401608080016 401608080017 401608080018 401608080019 401608080020 Lot No. 63437614 (Serial Numbers: 401608060001 401608060002 401608060003 401608060004 401608060005 Lot No. 63430521 (Serial Numbers: 401608090001 401608090002 401608090003 401608090004 401608090005 401608090006 401608090007 401608090008 401608090009 401608090010 401608090011 401608090012 401608090013 401608090014 401608090015 401608090016 401608090017 401608090018 401608090019 401608090020 Lot No. 63430522 (Serial Numbers: 401608110001 401608110002 401608110003 401608110004 401608110005 401608110006 401608110007 401608110008 401608110009 401608110010 401608110011 401608110012 401608110013 401608110014 401608110015 401608110016 401608110017 401608110018 401608110019 401608110020 Lot No. 63430543 (Serial Numbers: 401608130001 401608130002 401608130003 401608130004 401608130005 401608130006 401608130007 401608130008 401608130009 401608130010 401608130011 401608130012 401608130013 401608130014 401608130015 401608130016 401608130017 401608130018 401608130019 401608130020 Lot No. 63445186 (Serial Numbers: 401608160001 401608160002 401608160003 401608160004 401608160005 401608160006 401608160007 401608160008 401608160009 401608160010 401608160011 401608160012 401608160013 401608160014 401608160015 401608160016 401608160017 401608160018 401608160019 401608160020 Lot No. 63430556 (Serial Numbers: 401608180001 401608180002 401608180003 401608180004 401608180005 401608180006 401608180007 401608180008 401608180009 401608180010 401608180011 401608180012 401608180013 401608180014 401608180015 401608180016 401608180017 401608180018 401608180019 401608180020 Lot No. 63455045 (Serial Numbers: 401608220001 401608220002 401608220003 401608220004 401608220005 401608220006 401608220007 401608220008 401608220009 401608220010 401608220011 401608220012 401608220013 401608220014 401608220015 401608220016 401608220017 401608220018 401608220019 401608220020 Lot No. 63455047 (Serial Numbers: 401608240001 401608240002 401608240003 401608240004 401608240005 401608240006 401608240007 401608240008 401608240009 401608240010 401608240011 401608240012 401608240013 401608240014 401608240015 401608240016 401608240017 401608240018 401608240019 401608240020 Lot No. 63463530 (Serial Numbers: 401608250001 401608250002 401608250003 401608250004 401608250005 401608250006 401608250007 401608250008 401608250009 401608250010 401608250011 401608250012 401608250013 401608250014 401608250015 401608250016 401608250017 401608250018 401608250019 401608250020 Lot No. 63463531 (Serial Numbers: 401608260001 401608260002 401608260003 401608260004 401608260005 Lot No. 63455048 (Serial Numbers: 401608270001 401608270002 401608270003 401608270004 401608270005 401608270006 401608270007 401608270008 401608270009 401608270010 401608270011 401608270012 401608270013 401608270014 401608270015 401608270016 401608270017 401608270018 401608270019 401608270020 Lot No. 63455067 (Serial Numbers: 401608290001 401608290002 401608290003 401608290004 401608290005 401608290006 401608290007 401608290008 401608290009 401608290010 401608290011 401608290012 401608290013 401608290014 401608290015 401608290016 401608290017 401608290018 401608290019 401608290020 Lot No. 63455068 (Serial Numbers: 401608300001 401608300002 401608300003 401608300004 401608300005 401608300006 401608300007 401608300008 401608300009 401608300010 401608300011 401608300012 401608300013 401608300014 401608300015 401608300016 401608300017 401608300018 401608300019 401608300020 Lot No. 63455069 (Serial Numbers: 401609020001 401609020002 401609020003 401609020004 401609020005 401609020006 401609020007 401609020008 401609020009 401609020010 401609020011 401609020012 401609020013 401609020014 401609020015 401609020016 401609020017 401609020018 401609020019 401609020020 Lot No. 63455070 (Serial Numbers: 401609070001 401609070002 401609070003 401609070004 401609070005 401609070006 401609070007 401609070008 401609070009 401609070010 401609070011 401609070012 401609070013 401609070014 401609070015 401609070016 401609070017 401609070018 401609070019 401609070020 Lot No. 63455089 (Serial Numbers: 401609090001 401609090002 401609090003 401609090004 401609090005 401609090006 401609090007 401609090008 401609090009 401609090010 401609090011 401609090012 401609090013 401609090014 401609090015 401609090016 401609090017 401609090018 401609090019 401609090020 Lot No. 63455090 (Serial Numbers: 401609120001 401609120002 401609120003 401609120004 401609120005 401609120006 401609120007 401609120008 401609120009 401609120010 401609120011 401609120012 401609120013 401609120014 401609120015 401609120016 401609120017 401609120018 401609120019 401609120020 Lot No. 63455091 (Serial Numbers: 401609130001 401609130002 401609130003 401609130004 401609130005 401609130006 401609130007 401609130008 401609130009 401609130010 401609130011 401609130012 401609130013 401609130014 401609130015 401609130016 401609130017 401609130018 401609130019 401609130020 Lot No. 63474286 (Serial Numbers: 401609130021 401609130022 401609130023 Lot No. 63474277 (Serial Numbers: 401609140001 401609140002 401609140003 401609140004 401609140005 401609140006 401609140007 401609140008 401609140009 401609140010 401609140011 401609140012 401609140013 401609140014 401609140015 401609140016 401609140017 401609140018 401609140019 401609140020 Lot No. 63474278 (Serial Numbers: 401609160001 401609160002 401609160003 401609160004 401609160005 401609160006 401609160007 401609160008 401609160009 401609160010 401609160011 401609160012 401609160013 401609160014 401609160015 401609160016 401609160017 401609160018 401609160019 401609160020 Lot No. 63474288 (Serial Numbers: 401609160021 401609160022 401609160023 Lot No. 63455105 (Serial Numbers: 401609200001 401609200002 401609200003 401609200004 401609200005 401609200006 401609200007 401609200008 401609200009 401609200010 401609200011 401609200012 401609200013 401609200014 401609200015 401609200016 401609200017 401609200018 401609200019 401609200020 Lot No. 63455107 (Serial Numbers: 401609210001 401609210002 401609210003 401609210004 401609210005 401609210006 401609210007 401609210008 401609210009 401609210010 401609210011 401609210012 401609210013 401609210014 401609210015 401609210016 401609210017 401609210018 401609210019 401609210020 Lot No. 63455108 (Serial Numbers: 401609220001 401609220002 401609220003 401609220004 401609220005 401609220006 401609220007 401609220008 401609220009 401609220010 401609220011 401609220012 401609220013 401609220014 401609220015 401609220016 401609220017 401609220018 401609220019 401609220020 Lot No. 63474279 (Serial Numbers: 401609220021 401609220022 401609220023 401609220024 401609220025 401609220026 401609220027 401609220028 401609220029 401609220030 401609220031 401609220032 401609220033 401609220034 401609220035 401609220036 401609220037 401609220038 401609220039 401609220040 Lot No. 63474280 (Serial Numbers: 401609260001 401609260002 401609260003 401609260004 401609260005 401609260006 401609260007 401609260008 401609260009 401609260010 401609260011 401609260012 401609260013 401609260014 401609260015 401609260016 401609260017 401609260018 401609260019 401609260020 Lot No. 63482438 (Serial Numbers: 401609270001 401609270002 401609270003 401609270004 401609270005 401609270006 401609270007 401609270008 401609270009 401609270010 401609270011 401609270012 401609270013 401609270014 401609270015 401609270016 401609270017 401609270018 401609270019 401609270020 401609270021 401609270022 401609270023 401609270024 401609270025 401609270026 401609270027 401609270028 401609270029 401609270030 Lot No. 63482440 (Serial Numbers: 401609280001 401609280002 401609280003 401609280004 401609280005 401609280006 401609280007 401609280008 401609280009 401609280010 401609280011 401609280012 401609280013 401609280014 401609280015 401609280016 401609280017 401609280018 401609280019 401609280020 401609280021 401609280022 401609280023 401609280024 401609280025 401609280026 401609280027 401609280028 401609280029 401609280030 Lot No. 63482441 (Serial Numbers: 401609300001 401609300002 401609300003 401609300004 401609300005 401609300006 401609300007 401609300008 401609300009 401609300010 401609300011 401609300012 401609300013 401609300014 401609300015 401609300016 401609300017 401609300018 401609300019 401609300020 401609300021 401609300022 401609300023 401609300024 401609300025 401609300026 401609300027 401609300028 401609300029 401609300030 Lot No. 63482443 (Serial Numbers: 401610040001 401610040002 401610040003 401610040004 401610040005 401610040006 401610040007 401610040008 401610040009 401610040010 401610040011 401610040012 401610040013 401610040014 401610040015 401610040016 401610040017 401610040018 401610040019 401610040020 401610040021 401610040022 401610040023 401610040024 401610040025 401610040026 401610040027 401610040028 401610040029 401610040030 Lot No. 63482445 (Serial Numbers: 401610070001 401610070002 401610070003 401610070004 401610070005 401610070006 401610070007 401610070008 401610070009 401610070010 401610070011 401610070012 401610070013 401610070014 401610070015 401610070016 401610070017 401610070018 401610070019 401610070020 401610070021 401610070022 401610070023 401610070024 401610070025 401610070026 401610070027 401610070028 401610070029 401610070030 Lot No. 63474290 (Serial Numbers: 401610070031 401610070032 401610070033 401610070034 401610070035 401610070036 401610070037 401610070038 401610070039 401610070040 Lot No. 63474289 (Serial Numbers: 401610100001 401610100002 401610100003 401610100004 401610100005 401610100006 401610100007 401610100008 Lot No. 63482450 (Serial Numbers: 401610120001 401610120002 401610120003 401610120004 401610120005 401610120006 401610120007 401610120008 401610120009 401610120010 401610120011 401610120012 401610120013 401610120014 401610120015 401610120016 401610120017 401610120018 401610120019 401610120020 401610120021 401610120022 401610120023 401610120024 401610120025 401610120026 401610120027 401610120028 401610120029 401610120030 Lot No. 63474291 (Serial Numbers: 401610130001 401610130002 401610130003 401610130004 401610130005 401610130006 401610130007 401610130008 401610130009 401610130010 401610130011 401610130012 401610130013 401610130014 401610130015 Lot No. 63474293 (Serial Numbers: 401610170001 401610170002 401610170003 401610170004 401610170005 401610170006 401610170007 401610170008 401610170009 401610170010 401610170011 401610170012 401610170013 401610170014 401610170015 Lot No. 63482453 (Serial Numbers: 401610190001 401610190002 401610190003 401610190004 401610190005 401610190006 401610190007 401610190008 401610190009 401610190010 401610190011 401610190012 401610190013 401610190014 401610190015 401610190016 401610190017 401610190018 401610190019 401610190020 401610190021 401610190022 401610190023 401610190024 401610190025 401610190026 401610190027 401610190028 401610190029 401610190030 Lot No. 63482454 (Serial Numbers: 401610240001 401610240002 401610240003 401610240004 401610240005 401610240006 401610240007 401610240008 401610240009 401610240010 401610240011 401610240012 401610240013 401610240014 401610240015 401610240016 401610240017 401610240018 401610240019 401610240020 401610240021 401610240022 401610240023 401610240024 401610240025 401610240026 401610240027 401610240028 401610240029 401610240030 Lot No. 63482455 (Serial Numbers: 401610250001 401610250002 401610250003 401610250004 401610250005 401610250006 401610250007 401610250008 401610250009 401610250010 401610250011 401610250012 401610250013 401610250014 401610250015 401610250016 401610250017 401610250018 401610250019 401610250020 401610250021 401610250022 401610250023 401610250024 401610250025 401610250026 401610250027 401610250028 401610250029 401610250030 Lot No. 63482457 (Serial Numbers: 401610270001 401610270002 401610270003 401610270004 401610270005 401610270006 401610270007 401610270008 401610270009 401610270010 401610270011 401610270012 401610270013 401610270014 401610270015 401610270016 401610270017 401610270018 401610270019 401610270020 401610270021 401610270022 401610270023 401610270024 401610270025 401610270026 401610270027 401610270028 401610270029 401610270030 Lot No. 63482458 (Serial Numbers: 401610280001 401610280002 401610280003 401610280004 401610280005 401610280006 401610280007 401610280008 401610280009 401610280010 401610280011 401610280012 401610280013 401610280014 401610280015 401610280016 401610280017 401610280018 401610280019 401610280020 401610280021 401610280022 401610280023 401610280024 401610280025 401610280026 401610280027 401610280028 401610280029 401610280030 Lot No. 63482461 (Serial Numbers: 401611020001 401611020002 401611020003 401611020004 401611020005 401611020006 401611020007 401611020008 401611020009 401611020010 401611020011 401611020012 401611020013 401611020014 401611020015 401611020016 401611020017 401611020018 401611020019 401611020020 401611020021 401611020022 401611020023 401611020024 401611020025 401611020026 401611020027 401611020028 401611020029 401611020030 Lot No. 63482481 (Serial Numbers: 401611030001 401611030002 401611030003 401611030004 401611030005 401611030006 401611030007 401611030008 401611030009 401611030010 401611030011 401611030012 401611030013 401611030014 401611030015 401611030016 401611030017 401611030018 401611030019 401611030020 Lot No. 63482462 (Serial Numbers: 401611070001 401611070002 401611070003 401611070004 401611070005 401611070006 401611070007 401611070008 401611070009 401611070010 401611070011 401611070012 401611070013 401611070014 401611070015 401611070016 401611070017 401611070018 401611070019 401611070020 401611070021 401611070022 401611070023 401611070024 401611070025 401611070026 401611070027 401611070028 401611070029 401611070030 Lot No. 63490404 (Serial Numbers: 401611080001 401611080002 401611080003 401611080004 401611080005 Lot No. 63482463 (Serial Numbers: 401611100001 401611100002 401611100003 401611100004 401611100005 401611100006 401611100007 401611100008 401611100009 401611100010 401611100011 401611100012 401611100013 401611100014 401611100015 401611100016 401611100017 401611100018 401611100019 401611100020 401611100021 401611100022 401611100023 401611100024 401611100025 401611100026 401611100027 401611100028 401611100029 401611100030 Lot No. 63482464 (Serial Numbers: 401611110001 401611110002 401611110003 401611110004 401611110005 401611110006 401611110007 401611110008 401611110009 401611110010 401611110011 401611110012 401611110013 401611110014 401611110015 401611110016 401611110017 401611110018 401611110019 401611110020 401611110021 401611110022 401611110023 401611110024 401611110025 401611110026 401611110027 401611110028 401611110029 401611110030 Lot No. 63482465 (Serial Numbers: 401611140001 401611140002 401611140003 401611140004 401611140005 401611140006 401611140007 401611140008 401611140009 401611140010 401611140011 401611140012 401611140013 401611140014 401611140015 401611140016 401611140017 401611140018 401611140019 401611140020 401611140021 401611140022 401611140023 401611140024 401611140025 401611140026 401611140027 401611140028 401611140029 401611140030 Lot No. 63482470 (Serial Numbers: 401611150001 401611150002 401611150003 401611150004 401611150005 401611150006 401611150007 401611150008 401611150009 401611150010 401611150011 401611150012 401611150013 401611150014 401611150015 401611150016 401611150017 401611150018 401611150019 401611150020 401611150021 401611150022 401611150023 401611150024 401611150025 401611150026 401611150027 401611150028 401611150029 401611150030 Lot No. 63484748 (Serial Numbers: 401611170001 401611170002 401611170003 401611170004 401611170005 401611170006 401611170007 401611170008 401611170009 401611170010 401611170011 401611170012 Lot No. 63490403 (Serial Numbers: 401611170013 401611170014 401611170015 Lot No. 63482471 (Serial Numbers: 401611180001 401611180002 401611180003 401611180004 401611180005 401611180006 401611180007 401611180008 401611180009 401611180010 401611180011 401611180012 401611180013 401611180014 401611180015 401611180016 401611180017 401611180018 401611180019 401611180020 401611180021 401611180022 401611180023 401611180024 401611180025 401611180026 401611180027 401611180028 401611180029 401611180030 Lot No. 63482473 (Serial Numbers: 401611220001 401611220002 401611220003 401611220004 401611220005 401611220006 401611220007 401611220008 401611220009 401611220010 401611220011 401611220012 401611220013 401611220014 401611220015 401611220016 401611220017 401611220018 401611220019 401611220020 401611220021 401611220022 401611220023 401611220024 401611220025 401611220026 401611220027 401611220028 401611220029 401611220030 Lot No. 63531692 (Serial Numbers: 401611290001 401611290002 401611290003 401611290004 401611290005 401611290006 401611290007 401611290008 401611290009 401611290010 401611290011 401611290012 401611290013 401611290014 401611290015 401611290016 401611290017 401611290018 401611290019 401611290020 401611290021 401611290022 401611290023 401611290024 401611290025 401611290026 401611290027 401611290028 401611290029 401611290030 Lot No. 63501495 (Serial Numbers: 401611290031 401611290032 401611290033 401611290034 401611290035 Lot No. 63501494 (Serial Numbers: 401611300001 401611300002 401611300003 401611300004 401611300005 Lot No. 63482474 (Serial Numbers: 401612010001 401612010002 401612010003 401612010004 401612010005 401612010006 401612010007 401612010008 401612010009 401612010010 401612010011 401612010012 401612010013 401612010014 401612010015 401612010016 401612010017 401612010018 401612010019 401612010020 401612010021 401612010022 401612010023 401612010024 401612010025 401612010026 401612010027 401612010028 401612010029 401612010030 Lot No. 63482475 (Serial Numbers: 401612090001 401612090002 401612090003 401612090004 401612090005 401612090006 401612090007 401612090008 401612090009 401612090010 401612090011 401612090012 401612090013 401612090014 401612090015 401612090016 401612090017 401612090018 401612090019 401612090020 401612090021 401612090022 401612090023 401612090024 401612090025 401612090026 401612090027 401612090028 401612090029 401612090030 Lot No. 63501462 (Serial Numbers: 401612120001 401612120002 401612120003 401612120004 401612120005 401612120006 401612120007 401612120008 401612120009 401612120010 401612120011 401612120012 401612120013 401612120014 401612120015 401612120016 401612120017 401612120018 401612120019 401612120020 401612120021 401612120022 401612120023 401612120024 401612120025 401612120026 401612120027 401612120028 401612120029 401612120030 Lot No. 63501489 (Serial Numbers: 401612150001 401612150002 401612150003 401612150004 401612150005 401612150006 401612150007 401612150008 401612150009 401612150010 401612150011 401612150012 401612150013 401612150014 401612150015 401612150016 401612150017 401612150018 401612150019 401612150020 401612150021 401612150022 401612150023 401612150024 401612150025 401612150026 401612150027 401612150028 401612150029 401612150030 Lot No. 63538139 (Serial Numbers: 401612160001 401612160002 Lot No. 63501488 (Serial Numbers: 401612190001 401612190002 401612190003 401612190004 401612190005 401612190006 401612190007 401612190008 401612190009 401612190010 401612190011 401612190012 401612190013 401612190014 401612190015 401612190016 401612190017 401612190018 401612190019 401612190020 401612190021 401612190022 401612190023 401612190024 401612190025 401612190026 401612190027 401612190028 401612190029 401612190030 Lot No. 63501514 (Serial Numbers: 401612200001 401612200002 401612200003 401612200004 401612200005 401612200006 401612200007 401612200008 401612200009 401612200010 401612200011 401612200012 401612200013 401612200014 401612200015 401612200016 401612200017 401612200018 401612200019 401612200020 401612200021 401612200022 401612200023 401612200024 401612200025 401612200026 401612200027 401612200028 401612200029 401612200030 Lot No. 63538140 (Serial Numbers: 401612200031 401612200032 401612200033 Lot No. 63501516 (Serial Numbers: 401612220001 401612220002 401612220003 401612220004 401612220005 401612220006 401612220007 401612220008 401612220009 401612220010 401612220011 401612220012 401612220013 401612220014 401612220015 401612220016 401612220017 401612220018 401612220019 401612220020 401612220021 401612220022 401612220023 401612220024 401612220025 401612220026 401612220027 401612220028 401612220029 401612220030 Lot No. 63501517 (Serial Numbers: 401612270001 401612270002 401612270003 401612270004 401612270005 401612270006 401612270007 401612270008 401612270009 401612270010 401612270011 401612270012 401612270013 401612270014 401612270015 401612270016 401612270017 401612270018 401612270019 401612270020 401612270021 401612270022 401612270023 401612270024 401612270025 401612270026 401612270027 401612270028 401612270029 401612270030 Lot No. 63501532 (Serial Numbers: 401612290001 401612290002 401612290003 401612290004 401612290005 401612290006 401612290007 401612290008 401612290009 401612290010 401612290011 401612290012 401612290013 401612290014 401612290015 401612290016 401612290017 401612290018 401612290019 401612290020 401612290021 401612290022 401612290023 401612290024 401612290025 401612290026 401612290027 401612290028 401612290029 401612290030 |
| FEI Number |
1526350
|
Recalling Firm/ Manufacturer |
Zimmer Surgical Inc
200 W Ohio Ave
Dover
OH
44622-9642
|
| For Additional Information Contact | 330-343-8801 |
Manufacturer Reason for Recall | Updated on-device label and 3 pages of an updated Operator/Service Manual. Also, decrease from 3.8 meters (m) to 1.0 separation distance between the device and RF communication equipment. |
FDA Determined Cause 2 | Use error |
| Action | The firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE CORRECTIONS" letter dated 7/27/2017 to its customers via email or mail. The letter describes the product, problem and actions to be taken. Instructions for distributors include ensure all affected personnel are aware of the recall, confirm receipt of new device label and updated Operator/Service Manual, place the label on the device in the location specified in the notice, replace the 3 pages of the Operator/Service Manual with the updated pages, review the hospital list of customers that accompanied the notice and return contact information for any additional hospitals product was distributed to.
Instructions for Risk Managers at hospitals include ensure all affected personnel are aware of the recall, confirm receipt of new device label and updated Operator/Service Manual, place the label on the device in the location specified in the notice, replace the 3 pages of the Operator/Service Manual with the updated pages, and complete and return the Attachment 1 form via email to corporatequality.postmarket@zimmerbiomet.com or ZimmerBiomet8287@stericycle.com or fax to 866-338-2611.
If after reviewing this notification you have further questions or concerns please call +1(330) 364-0989 between 8:00 am and 5:00 pm EST, Monday through Friday.Calls received outside of the call center operating hours will receive a prompt to record a voicemail or be transferred to an on-call representative in the case of an emergency. |
| Quantity in Commerce | 8,231 devices total |
| Distribution | Worldwide Distribution: US (nationwide) to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, including Guam; and countries to Australia, Brazil, Canada, Chile, China, Denmark, Dominican, El Salvador, England, Finland, France, Hong Kong, India, Italy, Japan, Malaysia, Mexico, Netherlands, Panama, Singapore, Sweden, Taiwan, and Thailand.. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = KCY
|
|
|
|