• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Implant Direct, INTERACTIVE HEALING COLLAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Implant Direct, INTERACTIVE HEALING COLLARsee related information
Date Initiated by FirmJanuary 23, 2018
Create DateFebruary 06, 2018
Recall Status1 Terminated 3 on August 09, 2018
Recall NumberZ-0539-2018
Recall Event ID 78917
510(K)NumberK130572 
Product Classification Abutment, implant, dental, endosseous - Product Code NHA
ProductImplant Direct, INTERACTIVE HEALING COLLAR, PART NUMBER 6530-15, 5.0mmL: 3.0mmD Platform
Code Information Lot Number 104203
FEI Number 3001617766
Recalling Firm/
Manufacturer
Implant Direct Sybron Manufacturing, LLC
3050 E Hillcrest Dr
Westlake Village CA 91362-3171
For Additional Information ContactCustomer Care Team
818-444-3300 Ext. 3323
Manufacturer Reason
for Recall
InterActive Healing Collar, Lot Number 104203, labeled as sterile with distributed prior to being sterilized.
FDA Determined
Cause 2
Process control
ActionThe firm initiated their recall on 01/23/2018 by letter. the recall notices requested the following actions by the distributors: "1. Please review your inventory for the affected product. 2. Please complete and return the Acknowledgement and Recall Return Form with the affected product, if available, within 48 hours. 3. If you are an authorized Implant Direct Sybron Manufacturing distributor, we request that you identify those customers that may have been shipped the affected product lot and contact these customers to inform them of this issue within fortyeight (48) hours of receipt of this notification in order to provide the customers with replacement product." The letter disseminated to the physician requested the following actions: "1. Please review your inventory for the affected product. 2. Please complete and return the Acknowledgement and Recall Return Form with the affected product, if available, within 48 hours."
Quantity in Commerce55 units
DistributionCO, SC, IA, NJ, FL, CA, VA, ID, Netherlands
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.