| | Class 2 Device Recall IMRIS IMRI 1.5T AND 3T S System |  |
| Date Initiated by Firm | April 04, 2018 |
| Create Date | May 04, 2018 |
| Recall Status1 |
Terminated 3 on July 22, 2020 |
| Recall Number | Z-1645-2018 |
| Recall Event ID |
79816 |
| 510(K)Number | K133692 |
| Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
| Product | IMRIS IMRI 1.5T AND 3T S System
Product Usage:
The IMRIS intraoperative MRI systems are traditional MRI systems that have been suspended on an overhead rail system. and is designated to operative inside an RF shielded room to facilitate intraoperative and multi room use. |
| Code Information |
Serial Numbers: 10111-00, 10073-01, 10043-01, 10097-01, 10006-11, 10076-21, 10081-01, 10089-01, 10101-01, 10093-01 |
| FEI Number |
3010326005
|
Recalling Firm/ Manufacturer |
Deerfield Imaging, Inc.
5101 Shady Oak Rd S
Minnetonka
MN
55343-4100
|
| For Additional Information Contact | 763-203-6300 |
Manufacturer Reason for Recall | The DC power to the IFOSS collision detector control panel at your facility is being adversely affected by the magnetic field. Our investigation shows the magnetic field is causing the IFOSS control panel to malfunction. |
FDA Determined Cause 2 | Device Design |
| Action | IMRIS sent Service bulletin T200627-B01 Rev dated March 14, 2018. - through email to all affected customers. The letter identified the affected product, the problem, and the actions required to be taken by the customer. For questions contact IMRIS Customer Support 866-475-0525. |
| Quantity in Commerce | 10 units |
| Distribution | Worldwide Distribution - US Nationwide in the states of OH, FL, NH, TX, Sweden, Canada |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|