| Date Initiated by Firm | May 31, 2018 |
| Create Date | August 12, 2018 |
| Recall Status1 |
Terminated 3 on April 11, 2019 |
| Recall Number | Z-2791-2018 |
| Recall Event ID |
80492 |
| 510(K)Number | K011812 |
| Product Classification |
Catheter, ventricular (containing antibiotic or antimicrobial agents) - Product Code NHC
|
| Product | VentriClear Ventricular Drainage Catheter Set, Cat. No. N-VVDC-01-ABRM
Product Usage:
The VentriClear Ventricular Drainage Catheter Set has been designed for obtaining access to a ventricular cavity of the brain for short-term use to externally drain fluid for the purpose of relieving elevated intracranial pressure or fluid volume. |
| Code Information |
lot 8639185 |
| FEI Number |
1820334
|
Recalling Firm/ Manufacturer |
Cook Inc.
750 N Daniels Way
Bloomington
IN
47404-9120
|
| For Additional Information Contact | Cook Medical Customer Relations Department 800-457-4500 |
Manufacturer Reason for Recall | This lot of VentriClear failed endotoxin testing. Potential adverse events include immune responses ranging from non-specific febrile reaction to life-threatening systemic inflammatory response syndrome (SIRS). |
FDA Determined Cause 2 | Release of Material/Component prior to receiving test results |
| Action | On May 31, 2018, the firm contacted the single customer (a distributor) who received the entirety of the affected lot, via email. The email alerted the customer that the lot failed endotoxin testing and instructed the customer to quarantine the lot immediately. The firm then conducted a teleconference with the distributor to further discuss interim controls for future lot release. The recalling firm requested return of the affected lot. The distributor returned the lot to the firm. The recalling firm stated that none of the product was distributed to end users. |
| Quantity in Commerce | 250 |
| Distribution | One distributor in Indiana; product was not further distributed to end users. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = NHC
|