| Date Initiated by Firm | May 16, 2018 |
| Create Date | September 19, 2018 |
| Recall Status1 |
Terminated 3 on June 30, 2020 |
| Recall Number | Z-3201-2018 |
| Recall Event ID |
80685 |
| Product Classification |
Integrated Continuous Glucose Monitoring system, factory calibrated - Product Code QBJ
|
| Product | Contour Plus Link 2.4, Product Catalog Number: MMT-1151SK. wireless blood glucose monitoring system
Contour Plus LINK 2.4 wireless blood glucose monitoring system (glucometer, test strips and control solution) is designed for self-testing by persons with diabetes and the monitoring of glucose concentrations for quantitative glucose measurement in fresh capillary full blood removed from the user's fingertips or palms. It is only for in vitro diagnostic use. |
| Code Information |
Lot# MD04M051P S/N# BG1245623B and S/N# BG1245624B |
| FEI Number |
3003166194
|
Recalling Firm/ Manufacturer |
Medtronic Inc.
18000 Devonshire St
Northridge
CA
91325-1219
|
| For Additional Information Contact | 818-576-4700 |
Manufacturer Reason for Recall | CONTOUR PLUS LINK 2.4 meter with an incorrect unit of measure was included into meter kits and distributed. This meter contains the incorrect unit of measure mg/dL instead of mmol/L. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | The firm, Medtronic, sent an "URGENT MEDICAL DEVICE RECALL- CONTOUR"PLUS LINK 2.4 Meter (MMT-1151SK - Batch# MD04M051P) Incorrect Unit of Measure" letter to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following:
A. Confirm if the Model (REF) of your meter is MMT-1151SK and the batch (LOT) number is MD04M051P.
B. If the REF and LOT match this notice, stop using the CONTOUR"PLUS LINK 2.4 meter immediately.
C. Call your local Helpline and let the customer service team know you received a notice stating you may have a CONTOUR"PLUS LINK 2.4 meter with the incorrect unit of measure.
D. Return your recalled CONTOUR"PLUS LINK 2.4 meter to Medtronic by following the instructions provided by the Helpline representative. Medtronic will provide a free replacement.
If you have any questions, contact the Program Manager, Field Corrective Action at 818-576-4700. |
| Quantity in Commerce | 2 |
| Distribution | International Distribution to Slovakia only. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|