Class 2 Device Recall Contour Plus Link 2.4

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Contour Plus Link 2.4see related information
Date Initiated by FirmMay 16, 2018
Create DateSeptember 19, 2018
Recall Status1 Terminated 3 on June 30, 2020
Recall NumberZ-3201-2018
Recall Event ID 80685
Product Classification Integrated Continuous Glucose Monitoring system, factory calibrated - Product Code QBJ
ProductContour Plus Link 2.4, Product Catalog Number: MMT-1151SK. wireless blood glucose monitoring system Contour Plus LINK 2.4 wireless blood glucose monitoring system (glucometer, test strips and control solution) is designed for self-testing by persons with diabetes and the monitoring of glucose concentrations for quantitative glucose measurement in fresh capillary full blood removed from the user's fingertips or palms. It is only for in vitro diagnostic use.
Code Information Lot# MD04M051P S/N# BG1245623B and S/N# BG1245624B
FEI Number 3003166194
Recalling Firm/
Manufacturer
Medtronic Inc.
18000 Devonshire St
Northridge CA 91325-1219
For Additional Information Contact
818-576-4700
Manufacturer Reason
for Recall
CONTOUR PLUS LINK 2.4 meter with an incorrect unit of measure was included into meter kits and distributed. This meter contains the incorrect unit of measure mg/dL instead of mmol/L.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionThe firm, Medtronic, sent an "URGENT MEDICAL DEVICE RECALL- CONTOUR"PLUS LINK 2.4 Meter (MMT-1151SK - Batch# MD04M051P) Incorrect Unit of Measure" letter to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: A. Confirm if the Model (REF) of your meter is MMT-1151SK and the batch (LOT) number is MD04M051P. B. If the REF and LOT match this notice, stop using the CONTOUR"PLUS LINK 2.4 meter immediately. C. Call your local Helpline and let the customer service team know you received a notice stating you may have a CONTOUR"PLUS LINK 2.4 meter with the incorrect unit of measure. D. Return your recalled CONTOUR"PLUS LINK 2.4 meter to Medtronic by following the instructions provided by the Helpline representative. Medtronic will provide a free replacement. If you have any questions, contact the Program Manager, Field Corrective Action at 818-576-4700.
Quantity in Commerce2
DistributionInternational Distribution to Slovakia only.