• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 1 Device Recall Quantum Pump Console

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 1 Device Recall Quantum Pump Consolesee related information
Date Initiated by FirmMay 29, 2019
Date PostedAugust 23, 2019
Recall Status1 Terminated 3 on August 21, 2020
Recall NumberZ-2238-2019
Recall Event ID 83352
510(K)NumberK173834 
Product Classification Console, heart-lung machine, cardiopulmonary bypass - Product Code DTQ
ProductQuantum Pump Console
Code Information Software version 6.2.1.9; Serial Numbers PH1000005, PH1000020, PH1000027, PH1000030, PH1000077, PH1000081, PH1000082, PH1000086, PH1000087, PH1000093, PH1000100, PH1000101, PH1000102, PH1000103, PH1000109, PH1000110, PH1000112, PH1000113, PH1000114, PH1000115, PH1000120, PH1000127, PH1000129, PH1000131, PH1000132, PH1000134, PH1000135, PH1000137, PH1000139, PH1000140, PH1000141, PH1000144, PH1000145, PH1000150, PH1000156, PH1000158, PH1000159, PH1000160, PH1000161, PH1000162, PH1000165, PH1000168, PH1000170, PH1000171, PH1000172, PH1000175, PH1000176, PH1000177, PH1000178, PH1000179, PH1000180, PH1000182, PH1000201, PH1000208, PH1000209, PH1000210, PH1000211, PH1000212, PH1000213, PH1000215, PH1000233, PH1000256, PH1000257, PH1000259, PH1000438, PH1000439, PH1000440, PH1000441, PH1000442, PH1000445, PH1000455, PH1000495, PH1000498, PH1000522, PH1000615, PH1000644, PH1000645, PH1000646, PH1000647, PH1000648, PH1000649, PH1000668, PH1000669, PH1000670, PH1000682, PH1000683, PH1000684, PH1000686, PH1000687, PH1000688, PH1000689, PH1000692, PH1000722, PH1000726, PH1000727, PH1000740, PH1000742, PH1000743, PH1000744, PH1000879, PH1000880, PH1000882, PH1000885, PH1000886, PH1000895, PH1000897, PH1000939, PH1000940 
FEI Number 3006073153
Recalling Firm/
Manufacturer
Spectrum Medical Ltd.
Harrier 4, Meteor Business Park
Cheltenham Road East
Gloucester United Kingdom
Manufacturer Reason
for Recall
Potential for Quantum Pump Console, part of the Quantum Perfusion Systems, to unexpectedly shut down while in use.
FDA Determined
Cause 2
Software design
ActionSpectrum Medical notified customers on about 05/29/2019 via "URGENT FIELD SAFETY CORRECTION" letter. Customers were informed that Spectrum Medical is in the process of installing the most recent software update v6.2.3.16 to the Quantum Perfusion Systems, (Quantum Pump Console). This software update corrects an error that could potentially lead to power loss. All Quantum Perfusion Systems are in scope of this issue. Customers were instructed to be aware that user intervention may be necessary with continued use before the software update, assure that all users receive notice of this issue, confirm receipt of the Urgent Medical Device Correction by returning the customer response form, and a Spectrum Medical service team will contact each customer and arrange a time and date to update the systems with the most recent software update. Additionally, customers were instructed to notify all relevant personnel and to customers if the devices have been further distributed.
Quantity in Commerce108 units (8 US, 100 OUS)
DistributionUS distribution to NY, OH, VA, WA. International distribution to the Netherlands.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = DTQ
-
-