• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Comprehensive Humeral Tray Taper Extraction Pliers Replacement Tips

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Comprehensive Humeral Tray Taper Extraction Pliers Replacement Tipssee related information
Date Initiated by FirmMarch 13, 2020
Create DateApril 20, 2020
Recall Status1 Terminated 3 on July 21, 2021
Recall NumberZ-1750-2020
Recall Event ID 85272
510(K)NumberK080642 
Product Classification shoulder prosthesis, reverse configuration - Product Code PHX
ProductComprehensive Humeral Tray Taper Extraction Pliers - Replacement Tips, Model Number 110028522 - Product Usage: Intended to be used in surgery to relieve pain and to replace the glenoid and humeral side of the shoulder joint in order to restore mobility in a grossly rotator cuff deficient joint with severe arthropathy or a previous joint replacement with a grossly rotator cuff deficient joint.
Code Information Lot Numbers: 494350 494350R 233080 364910 489000 079630 516230 742760 362370 465800 197900 175110 175100 
FEI Number 1825034
Recalling Firm/
Manufacturer
Biomet, Inc.
56 E Bell Dr
Warsaw IN 46582-6989
For Additional Information Contact
574-371-3071
Manufacturer Reason
for Recall
There were unknown bioburden levels prior to sterilization, which may affect sterility of the product.
FDA Determined
Cause 2
Under Investigation by firm
ActionURGENT MEDICAL DEVICE RECALL notification letters dated 3/13/20 were sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1  Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form will be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1  Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Mon
Quantity in Commerce125
DistributionWorldwide distribution - US Nationwide distribution including the states of AL, CO, KS, MA, MD, MI, MO, NH, NJ, and PA. The countries of Australia, Canada, India, and Netherlands.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
510(K) Database510(K)s with Product Code = PHX
-
-