Class 2 Device Recall Philips HeartStart MRx

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Philips HeartStart MRxsee related information
Date Initiated by FirmJuly 16, 2020
Create DateAugust 13, 2020
Recall Status1 Terminated 3 on August 03, 2021
Recall NumberZ-2794-2020
Recall Event ID 85743
510(K)NumberK031187 
Product Classification Automated external defibrillators (non-wearable) - Product Code MKJ
ProductPhilips HeartStart MRx, Monitor/Defibrillators with model numbers M3535A, M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, M3536M9 - Product Usage: is intended for use in hospital and pre-hospital settings by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support or defibrillation.
Code Information All devices are affected
FEI Number 1218950
Recalling Firm/
Manufacturer
Philips North America, LLC
3000 Minuteman Rd
Andover MA 01810-1032
For Additional Information ContactPhilips Customer Services
800-722-9377
Manufacturer Reason
for Recall
The therapy switch may fail, resulting in the device exhibiting the following behaviors: the device may not perform the selected function, the therapy knob may not change to the energy setting selected, or the device may deliver a shock with an energy level different from the setting selected by the user. These device behaviors could result in a delay in therapy or a failure to deliver the intended therapy.
FDA Determined
Cause 2
Process control
ActionOn July 16, 2020, the firm distributed an Urgent Medical Device Correction letter to all affected customers. Customers were informed of the product issue and were instructed to do the following: - The MRx is safe to use and can remain in service if the device does not exhibit any of the behaviors described in this Notice (the device may not perform the selected function, the therapy knob may not change to the energy setting selected, or the device may deliver a shock with an energy level different from the setting selected by the user.) - If an MRx device experiences a switch failure as described, contact your local Philips representative, who can arrange for a replacement of the switch at no charge. - Continue to perform the Shift and Operational Checks described in the Instructions for Use, Maintenance Chapter - To detect possible switch failures as early as possible, conduct Weekly Shock Tests (part of the Shift Check in the IFU) on a daily basis - It is not necessary to perform a Shock Test on days when an Operational Check is performed, since an Operational Check includes a Shock Test. - Keep a copy of the customer letter with each HeartStart MRx Instructions for Use. - To acknowledge receipt of this notification, please complete and fax the Customer Reply Form to the firm. If the MRx Therapy Switch fails, Philips will arrange for the replacement of the switch at no charge to the customer.
Quantity in Commerce95,212 units
DistributionWorldwide distribution.
Total Product Life CycleTPLC Device Report