| Date Initiated by Firm | December 22, 2020 |
| Create Date | February 08, 2021 |
| Recall Status1 |
Terminated 3 on June 07, 2022 |
| Recall Number | Z-1099-2021 |
| Recall Event ID |
87090 |
| Product Classification |
Reagents, specific, analyte - Product Code MVU
|
| Product | Aquarius LPH 511-A CRLF2 Distal Probe Green, model no. LPH 511-A
The CRLF2 Distal probe, labelled in green, consists of two probes (71kb and 131kb) that are situated distal (telomeric)
to the CRLF2 gene |
| Code Information |
Lots 072240, 071390, and 072693 |
| FEI Number |
3002646096
|
Recalling Firm/ Manufacturer |
Cytocell Ltd.
418 Cambridge Science Park, Milton Road
Oxford Gene Technology
Cambridge
United Kingdom
|
Manufacturer Reason for Recall | Users may observe very faint additional locus specific signals at 11q13. There is a low risk of a normal sample being reported as not normal in a resulting laboratory developed test (LDT) that utilizes the LPH511-A CRLF2 Distal Probe Green. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | The firm began distributing recall notifications to customers on December 22, 2020. Customers were informed of the product issue.
Customers were instructed to do the following:
Immediately examine your inventory and quarantine all product subject to recall. The firm requests that you destroy the remaining inventory. The firm also suggests that laboratories review any results obtained with the affected devices and check that signal patterns were not misinterpreted as a result of any additional locus specific signal at 11q13.
If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or vigilance@ogt.com. |
| Quantity in Commerce | 72 |
| Distribution | Domestic distribution to California, Illinois, Michigan, Minnesota, New Jersey, Texas, and Washington. Foreign distribution to Canada. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|