| Date Initiated by Firm | October 14, 2021 |
| Create Date | November 02, 2021 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-0189-2022 |
| Recall Event ID |
88840 |
| 510(K)Number | K191233 |
| Product Classification |
System, imaging, pulsed doppler, ultrasonic - Product Code IYN
|
| Product | Arietta 750 Ultrasound
Software Version: V1.0.0 through V2.1.3 |
| Code Information |
Serial Numbers:
G3075073 (01)04562122959841(11)201214(21)G3075073;
G3110777 (01)04562122959841(11)210421(21)G3110777;
G3110771 (01)04562122959841(11)210421(21)G3110771;
G3113780 (01)04562122959841(11)210510(21)G3113780;
G3113778 (01)04562122959841(11)210510(21)G3113778;
G3113779 (01)04562122959841(11)210510(21)G3113779;
G3113777 (01)04562122959841(11)210510(21)G3113777;
G3071857 (01)04562122959841(11)210420(21)G3071857;
G3108384 (01)04562122959841(11)210402(21)G3108384;
G3114220 (01)04562122959841(11)210510(21)G3114220;
G3110776 (01)04562122959841(11)210421(21)G3110776;
G3108385 (01)04562122959841(11)210402(21)G3108385;
G3114223 (01)04562122959841(11)210518(21)G3114223;
G3108401 (01)04562122959841(11)210420(21)G3108401;
G3114959 (01)04562122959841(11)210524(21)G3114959
|
| FEI Number |
1528028
|
Recalling Firm/ Manufacturer |
FujiFilm Healthcare Americas Corporation
1959 Summit Commerce Park
Twinsburg
OH
44087-2371
|
| For Additional Information Contact | Aaron Pierce 330-425-1313 Ext. 3720 |
Manufacturer Reason for Recall | Diagnostic ultrasound system with the specified
software version and used in conjunction with a Fujifilm (formally Hitachi) transducer does not display
the measurement results correctly with two cardiology measurement functions:
Mitral Regurgitation Flow (MR Flow) Measurement;
Mitral Valve (MV) Measurement.
In addition there is an error in the instruction manual. |
FDA Determined Cause 2 | Software design |
| Action | Fujifi Film issued Medical Device Correction Notice letter dated 10/07/21 via courier on 10/14/21.
Letter states reason for recall, health risk and action to take:
Customers/Users are advised to accept the corrected software and manual as soon as available.
Advisory Notification applies to all customers and distributors of the ARIETTA 850 systems with software version V1.0.0 through V4.1.3, See following page for complete list of affected systems.
We will contact you to set up time to update the software and provide the revised manual. (7) Other Information:
For questions, comments, or concerns, please contact:
Aaron Pierce
Director, QA/RA
1959 Summit Commerce Park
Twinsburg, OH 44087
Phone (330) 425-1313 x2934, Fax (330) 963-0749
Email aaron.pierce@fujifilm.com |
| Quantity in Commerce | 15 units |
| Distribution | Nationwide
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = IYN
|