• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Cytocell

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Cytocellsee related information
Date Initiated by FirmOctober 27, 2022
Date PostedNovember 30, 2022
Recall Status1 Terminated 3 on December 27, 2024
Recall NumberZ-0341-2023
Recall Event ID 91119
Product Classification Reagents, specific, analyte - Product Code MVU
ProductCytoCell BCL11B Distal in Texas Red Spectrum- Analyte Specific Reagent Ref: MPD39782
Code Information Lot Number: RD22/230/11
FEI Number 3002646096
Recalling Firm/
Manufacturer
Cytocell Ltd.
418 Cambridge Science Park, Milton Road
Oxford Gene Technology
Cambridge United Kingdom
Manufacturer Reason
for Recall
An error with the chromomap on version 1 of the package insert which is incorrect. The nucleotide locations specified in the insert are correct but the map has been created using the wrong coordinates.
FDA Determined
Cause 2
Under Investigation by firm
ActionCytoCell issued Urgent Medical Device Recall letter on 10/27/22 to end user via email. The letter states reason for recall, health risk and action to take: Immediately examine your inventory and quarantine all product subject to recall. Cytocell requests that all users review and implement the enclosed new package insert (v2) for this device. Please dispose of any copies of v1 of the package insert. Please complete and return the enclosed response form as soon as possible. If you have any questions, contact Cytocell on Tel: +44 (0)1223 294048, Fax +44(0)1223 294986 or vigilance@ogt.com.
Quantity in Commerce3 units
DistributionTX
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-