| | Class 2 Device Recall Abbott Alinity |  |
| Date Initiated by Firm | November 11, 2022 |
| Date Posted | April 06, 2023 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1349-2023 |
| Recall Event ID |
91777 |
| PMA Number | P190025 |
| Product Classification |
real time Nucleic acid amplification system - Product Code OOI
|
| Product | Abbott Alinity m System, REF 08N53-002, containing Amplification Detetction Unit (ADU) #4. |
| Code Information |
Alinity m System Serial #01015, UDI: (01)00884999048034(11)220505(21)01015(240)08N53-002; ADU Serial #AMP05139. |
| FEI Number |
3005248192
|
Recalling Firm/ Manufacturer |
Abbott Molecular, Inc.
1300 E Touhy Ave
Des Plaines
IL
60018-3315
|
| For Additional Information Contact | Ray Bastian 224-361-7000 |
Manufacturer Reason for Recall | The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returned for use without calibration. This would result in a potential for delay of results and incorrect results. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | The field service engineer installed the missing plunger assembly in the clamp bar while at the consignee location on 11/11/2022.
The recalling firm issued the recall letter on 3/1/2023 via overnight mail to their sole consignee. The letter explains the issue, the potential impact, and the necessary actions, which includes placing the impacted ADU out of service and contacting Abbott technical support regarding concerns about the performance of the instrument or results. The letter is to be reviewed with the Medical Director or Laboratory Management and the consignee is to follow their laboratory protocol regarding the need for review of previously reported patient results using the affected system. A customer reply form is enclosed with the letter to acknowledge receipt and to indicate it was disseminated to all users, understood, and implemented.
If you have any questions regarding this communication, please contact your local Abbott representative or call DVP Quality Assurance, at 224-361-7000. |
| Quantity in Commerce | 1 |
| Distribution | US Distribution to state of: PA, a government account. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|