| | Class 2 Device Recall Allura Xper FD; Allura Xper OR Table |  |
| Date Initiated by Firm | January 02, 2024 |
| Date Posted | February 16, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1144-2024 |
| Recall Event ID |
93783 |
| 510(K)Number | K033737 K130638 K130842 K133292 K141979 K162859 |
| Product Classification |
Interventional fluoroscopic x-ray system - Product Code OWB
|
| Product | Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of affected systems.
Affected PCs: FlexVision PC
Allura Xper Model Name - Model Number:
Allura Xper FD10
722003
722010
722026;
Allura Xper FD10/10
722005
722011
722027;
Allura Xper FD20
722006
722012
722028;
Allura Xper FD20 Biplane
722008
722013;
Allura Xper FD20 Biplane OR Table
722025;
Allura Xper FD20 OR Table
722023
722035;
Allura Xper FD20/10
722029;
Allura Xper FD20/15
722058;
Allura Xper FD20/15 OR Table
722059;
Allura Xper FD20/20
722038;
Allura Xper FD20/20 OR Table
722039; |
| Code Information |
Allura Xper FD10
722003 - no UDI
Serial Numbers
1181
1187
1311
1349
1435
1594
1700
1788
268
285
368
390
445
467
472
485
913
940
,
722010 - UDI-DI 00884838059030
Serial Numbers
1005
1009
1011
1025
1028
1040
1041
1043
1048
1069
107
1078
1082
1084
1108
1109
1118
1123
1126
1128
1132
1149
1170
1173
1176
1181
1183
1209
1210
1230
1233
1234
1235
1236
1244
1261
1263
1275
1279
1283
1287
1292
1297
1298
1302
1309
1310
1328
133
1336
1344
1347
1365
1369
1370
1373
1395
1396
1397
1398
1424
1425
1432
1441
1442
175
190
215
244
252
280
33
343
344
365
40
414
42
43
468
488
500
527
585
646
667
668
679
68
690
714
728
754
755
763
767
780
785
787
788
791
818
821
826
830
838
839
873
882
896
902
905
917
920
921
931
932
939
965
968
970
976
986
987
992
999
,
722026 - UDI-DI 00884838054189
Serial Numbers
100
1000
101
102
1050
1051
1052
1068
1072
109
110
112
1126
1132
114
115
116
117
1182
1189
1192
1205
121
1215
122
1239
1248
1268
130
142
151
155
156
185
191
192
204
215
220
222
223
224
233
234
244
245
253
258
259
265
323
324
335
368
378
381
391
394
396
400
431
432
436
458
462
484
485
531
544
558
559
560
561
562
600
601
602
604
608
612
617
622
627
633
634
646
648
653
654
663
664
683
694
704
713
716
725
73
731
738
74
75
765
770
774
778
78
781
786
79
796
805
825
826
832
838
841
843
845
87
877
908
909
93
931
935
942
957
959
965
966
967
969
973
98
985
99
990
;
Allura Xper FD10/10
722005 - no UDI
Serial Number
259
261
270
329
343
449
456
480
49
523
63
,
722011 - no UDI
Serial Number
123
140
15
172
173
177
23
234
237
255
256
267
268
269
270
275
308
319
324
325
332
344
345
358
362
375
380
49
76
79
82
83
,
722027 - UDI-DI 00884838054196
Serial Number
1
104
106
107
112
116
120
123
127
134
153
156
164
174
175
179
183
202
203
216
222
230
231
234
240
241
255
263
265
266
280
284
288
29
298
305
306
307
32
332
335
342
351
39
41
42
43
49
50
56
71
;
Allura Xper FD20
722006 - UDI-DI 00884838059115
Serial Numbers
1000
1093
1114
1127
1132
1174
1221
1276
1318
1364
1383
1407
1420
1439
1444
1446
1476
1507
1588
1589
1627
1648
1661
1729
1761
1807
1830
1845
1848
1870
1875
1913
1921
1922
1942
1955
2122
2126
2143
2147
2183
2200
2203
2268
2274
2323
2404
39
518
559
560
587
641
680
683
700
722
767
774
810
886
904
,
722012 - UDI-DI 00884838059054, 00884838059191, or 00884838059207
Serial Numbers
100
101
1017
1020
1032
1034
1036
1038
1054
1058
1077
1083
1084
1097
1098
1099
1104
1113
1120
1140
1141
1166
1189
1192
1207
1210
1211
1212
1219
1224
1238
1239
1240
1249
1250
1254
1265
1269
1270
1277
1282
1285
1293
1294
1305
1306
1316
1317
1321
1324
1328
1329
1334
1335
1344
1345
1356
1357
1396
1399
1401
1404
1405
1406
1407
1408
141
1422
1426
1430
1440
1441
1447
1455
1461
1467
1478
1506
1507
1508
1513
1515
1516
1527
1532
1533
1537
1538
1542
1545
1549
1550
1558
1559
1560
1561
1570
1581
1588
1589
1593
1594
1595
1596
1597
1606
1607
1608
1610
1611
1616
1617
1622
1625
1635
1649
1656
1658
1659
1660
1661
1664
1672
1683
1688
1690
1692
1712
1713
1714
1715
1716
1723
1724
1726
1731
1732
1733
1734
1735
1743
1746
1792
1793
1796
1798
1803
1806
1807
1823
1832
1842
1848
1852
1854
1858
1868
1869
1871
1889
1904
1906
1908
1919
1922
1923
1924
1941
1943
1946
1953
1961
1964
197
1975
1989
1992
2002
2003
2013
2014
2015
2022
2046
2054
2057
2063
2065
2068
208
2085
2086
2092
2133
2135
2136
2138
2145
2146
2148
2149
2150
2151
2156
2166
2171
2202
2203
2207
2215
2219
2221
2222
2227
2234
224
2254
2264
2265
2269
2276
2286
2288
2289
2292
2293
2303
2305
2307
2308
2309
2311
2313
2319
2323
233
234
2353
239
2454
2469
2484
2506
2508
2511
2512
2514
2515
2516
2521
2522
2607
2609
2610
2614
2629
2645
2656
2657
2693
2696
2697
2707
2713
2722
2723
2726
2735
2736
2740
2741
2742
2755
2760
2774
2821
2828
2829
2835
2840
2841
2842
2843
2844
2849
2857
2868
2869
2873
2874
2875
2881
2883
2888
2889
2892
2896
2898
2899
2900
2901
2912
2920
2929
2930
2940
2943
314
323
336
379
380
383
394
412
424
430
435
440
442
446
447
450
452
470
484
511
518
520
527
560
570
580
589
596
605
633
644
649
650
655
666
675
676
685
686
69
694
695
718
760
787
788
790
791
792
828
84
856
859
893
92
920
923
925
926
943
95
950
952
953
954
963
974
975
986
99
,
722028 - UDI-DI 00884838054202, 00884838059191, 00884838059269
Serial Numbers
1
1000
1001
1004
1011
1014
1017
1022
1024
1030
1032
1034
1035
1041
1043
1050
1051
1054
1058
1068
1069
1074
1076
1077
1081
1094
1098
1099
11
110
1103
1111
1116
1122
1129
1130
1132
1136
1142
1149
1154
1155
1156
1168
1171
1188
1189
1191
1200
1201
121
1211
1212
1213
1214
122
1232
1233
1244
125
1256
1261
1264
1266
1267
1269
127
1275
1279
128
1286
1288
129
130
131
1310
1311
132
1326
1327
133
1333
1340
1343
1349
1350
1351
1352
1353
1359
136
1368
137
1370
1373
1374
138
1385
1387
139
1393
1394
1395
1396
1397
140
1405
1407
1408
141
1413
142
1424
1428
143
1439
1441
1446
1447
1448
145
1456
1457
1461
1463
1466
1472
1481
1482
1483
1484
1498
1499
1508
151
1514
1522
1538
1549
155
1556
1566
1567
1568
1569
1570
158
159
1598
1600
1619
1629
1633
1634
1639
1646
1649
1650
1656
1657
166
167
1677
169
1690
1705
1710
1716
172
1723
173
1736
1751
1755
1756
176
1774
1783
180
1811
1820
1822
1824
1826
1828
1834
1835
1836
1837
184
1845
1846
185
1850
1854
1855
1856
1869
1875
1880
1889
1890
190
1909
1920
1931
194
1945
1948
1949
195
1956
1957
197
1978
1980
1983
1984
1995
2007
2008
2020
203
204
205
2070
2071
2083
210
2103
211
2110
213
215
2151
216
217
2173
218
2192
2196
2200
2206
224
2252
2261
227
2275
228
2284
229
2293
2295
2310
2335
2348
236
237
238
242
254
257
260
261
262
265
267
272
273
274
275
2760
279
280
282
284
289
292
2922
2933
303
305
31
313
315
317
318
324
326
329
333
336
340
343
346
347
350
356
363
365
368
372
373
375
376
377
379
382
386
394
402
406
407
408
416
422
424
426
428
431
435
437
445
447
449
451
453
456
459
462
464
473
475
476
479
48
481
488
489
491
493
494
496
498
499
516
518
519
520
525
526
528
530
531
533
534
535
538
539
540
541
543
547
549
550
552
553
556
558
559
560
561
564
565
570
571
579
580
582
586
590
591
592
596
601
618
620
625
627
630
632
634
638
647
654
655
66
666
668
67
671
672
674
677
678
680
684
692
695
696
698
71
710
717
720
721
729
731
741
743
751
757
759
768
779
781
793
80
805
806
808
809
811
818
819
821
825
840
842
861
864
866
867
869
874
877
879
880
882
887
888
889
891
892
895
896
900
901
903
904
906
908
909
910
911
917
922
924
925
928
929
930
932
934
938
940
945
946
949
952
956
961
962
966
967
968
971
973
976
978
981
982
983
987
989
991
994
995
998
;
Allura Xper FD20 Biplane
722008 - no UDI
Serial Numbers
115
124
142
1789
19
231
246
264
288
306
319
349
378
419
420
445
463
465
472
487
516
,
722013 - UDI-DI 00884838059061 or 00884838080805
Serial Numbers
100
115
124
129
140
148
150
151
152
175
203
205
214
217
218
253
264
268
274
278
297
30
303
307
313
327
328
330
342
347
357
360
375
377
379
383
401
402
410
422
424
426
427
428
435
441
442
45
461
462
467
475
479
480
481
491
492
495
496
497
500
505
511
513
518
519
550
556
562
564
573
575
580
599
601
603
626
631
634
637
69
74
84
;
Allura Xper FD20 Biplane OR Table
722025 - no UDI
Serial Numbers
2
3
;
Allura Xper FD20 OR Table
722023 - UDI-DI 00884838059085
Serial Numbers
109
110
113
116
117
118
12
124
127
128
138
142
147
160
175
179
35
46
64
66
71
72
75
82
83
98
,
722035 - UDI-DI 00884838054240
Serial Numbers
100
101
117
119
121
123
124
125
126
127
131
143
145
161
165
169
172
183
187
192
196
21
212
216
222
236
24
242
243
244
245
246
247
252
30
34
38
46
52
57
59
60
72
74
85
88
89
;
Allura Xper FD20/10
722029 - UDI-DI 00884838054219
Serial Numbers
110
126
129
21
22
31
36
55
56
57
60
62
63
7
71
84
88
91
94
95
97
;
Allura Xper FD20/15
722058 - UDI-DI 00884838059115 or 00884838059191
Serial Numbers
100
109
111
123
124
129
134
135
138
146
149
156
167
168
171
172
178
186
190
196
197
199
200
202
203
205
212
215
219
22
23
238
239
24
249
254
257
259
264
265
274
277
278
279
285
288
290
304
305
319
325
337
340
365
366
373
378
38
405
5
56
59
61
65
71
82
85
94
;
Allura Xper FD20/15 OR Table
722059 - UDI-DI 00884838059122
Serial Numbers
33
7
;
Allura Xper FD20/20
722038 - UDI-DI 00884838054226
Serial Numbers
106
108
115
118
119
120
126
130
137
141
157
167
173
35
37
38
39
41
42
46
49
55
60
63
67
7
70
71
;
Allura Xper FD20/20 OR Table
722039 - No UDI
Serial Number 3
; |
| FEI Number |
3002807880
|
Recalling Firm/ Manufacturer |
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Veenpluis 4-6
Best
Netherlands
|
Manufacturer Reason for Recall | Framegrabber card may not perform as intended. Monitor may show no viewports, or show one or more viewports may show no image, a distorted image, or a frozen image. Switching between viewports on the monitor may not be possible. Issue may result in delay of procedure. |
FDA Determined Cause 2 | Process design |
| Action | On January 2, 2024, the firm notified affected consignees via URGENT Medical Device Correction letters. These letters covered issues with three components of the PCs that may not perform as intended and which may therefore lead to delay in treatment: 1) DIMMs, 2) Disk Bay, and 3) Framegrabber Card.
This recall record is for the Framegrabber Card issue.
Customers should circulate the notice to all users and establish an emergency protocol for if you experience the issue during a procedure. Customers were provided with monitoring activities to undertake while waiting for their system to be corrected. Philips will be replacing the three components in all affected systems. Philips will contact customers to schedule a visit to replace the affected components. If you need additional information or support concerning this issue, please contact your local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions Center). |
| Quantity in Commerce | 1,993 US; 2,662 OUS |
| Distribution | Worldwide - US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|