| | Class 2 Device Recall DAILIES TOTAL1 soft contact lens TOTAL30 soft contact lens |  |
| Date Initiated by Firm | August 08, 2024 |
| Date Posted | September 30, 2024 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-3313-2024 |
| Recall Event ID |
95249 |
| 510(K)Number | K232921 K233856 |
| Product Classification |
Lenses, soft contact, daily wear - Product Code LPL
|
| Product | DAILIES TOTAL1,
DAILIES TOTAL1 for Astigmatism,
DAILIES TOTAL1 Multifocal,
TOTAL30 Sphere, |
| Code Information |
Product/UDI-DI(GTIN)/Expiration: Lots:
DAILIES TOTAL1/00630175507808, 00630175507754, 00630175507181, 00630175507662, 00630175508423, 00630175507679, 00630175507778, 00630175507075, 00630175507730, 00630175507839, 00630175508362, 00630175508348, 00630175508393, 00630175508263, 00630175508324, 00630175507242, 00630175508256, 00630175507815, 00630175507266, 00630175508287, 00630175507846/ 3/31/2028: N1275784, N1275785, N1275787, N1275788, N1275790, N1275791, N1275792, N1275793, N1275794, N1275795, N1275796, N1275797, N1275798, N1275799, N1275800, N1275801, N1275802, N1275803, N1275804, N1276252, N1276256, N1276261, N1276262, N1276267, N1276268, N1276269, N1276270, N1276271, N1276272, N1276273, N1276766, N1276770, N1276771, N1276789, N1276790; 4/30/2028: N1277278, N1277291, N1277292, N1277293, N1277294, N1277319, N1277322, N1277795, N1277799, N1277805.
DAILIES TOTAL1 for Astigmatism/00846566901746, 00846566874699, 00846566870714, 00846566903863, 00846566864935, 00846566861699, 00846566899418, 00846566866588, 00846566900930, 00846566868322, 00730822262408, 00846566906239, 00846566866069, 00846566865383, 00846566906222, 00846566865611, 00846566865369, 00846566862030, 00730822269025, 00846566901869, 00846566872954, 00846566901654, 00846566863358, 00730822266772, 00730822269476, 00846566900282, 00846566904297, 00730822262545, 00846566865314, 00846566879083, 00846566915132, 00730822264785, 00846566879212, 00730822264792, 00846566870240, 00846566875160, 00846566869428, 00846566874705, 00730822262095, 00846566878680, 00846566904068, 00846566906246, 00846566878314, 00730822264631, 00846566876952, 00846566868681, 00730822258272, 00846566878628, 00730822264990, 00846566869794, 00846566877812, 00846566867837, 00846566866434, 00846566879076, 00846566910779, 00846566879168, 00730822258241, 00846566898534, 00846566878307, 00846566868735, 00846566873876, 00846566879144, 00846566875115, 00846566877850, 00846566866038, 00730822260497, 00846566905799, 00846566862160, 00846566867530, 00730822266840, 00846566897698,00846566901715, 00846566902064, 00846566869770, 00846566863860, 00846566903931, 00846566906741, 00846566867981, 00846566905775, 00846566901630, 00846566869282, 00730822264914, 00730822258265, 00846566866632, 00730822256018, 00846566871070, 00846566905782, 00846566866281, 00846566902071, 00846566900763, 00846566868384, 00846566866786, 00846566863693, 00846566869589, 00846566866007, 00846566869053, 00846566871834, 00846566873180, 00846566862344, 00846566870035, 00846566870974, 00846566897278, 00846566879052, 00846566872282, 00846566870486, 00846566865567, 00846566865130, 00846566865901, 00846566871872, 00846566873630, 00846566865116, 00846566862436, 00846566864560, 00846566874538, 00846566868582, 00846566864218, 00846566869602, 00846566868155, 00846566864119, 00846566874088, 00846566869954, 00846566878154, 00846566872732, 00846566870950, 00846566879038, 00846566878932, 00846566868605, 00846566870509, 00846566869503, 00846566861439, 00846566878130, 00846566861514, 00846566867257, 00846566871407, 00846566878581, 00846566864621, 00846566865291, 00846566910687, 00730822266604, 00846566901210, 00730822264358, 00846566910656, 00846566865185, 00846566906116, 00846566906314, 00846566901616, 00846566908387, 00846566875139, 00846566901784, 00846566901708, 00846566877836, 00846566910632, 00846566901661, 00846566906185, 00730822266734, 00846566864492, 00846566879359, 00846566902187, 00846566867288, 00846566906260, 00846566901241, 00846566871674, 00846566899838, 00730822264334, 00846566869718, 00846566865154, 00846566898954, 00846566906277, 00846566910663, 00846566867103, 00846566901845, 00846566901647, 00846566906338, 00846566878642, 00846566870929, 00846566870417, 00846566869947, 00846566879021, 00846566906284, 00846566874231, 00846566874255/ 3/31/2029: N1275904, N1275908, N1275909, N1275913, N1275914, N1275918, N1275919, N1275921, N1275928, N1275929, N1275932, N1275935, N1275942, N1275943, N1275946, N1275954, N1275955, N1275956, N1275958, N1275991.N1275992, N1275997, N1275999, N1276003, N1276004, N1276005, N1276009, N1276010, N1276013, N1276018, N1276019, N1276025, N1276028, N1276043, N1276047, N1276048, N1276050, N1276051, N1276055, N1276057, N1276068, N1276069, N1276071, N1276073, N1276074, N1276076, N1276078, N1276079, N1276082, N1276083, N1276084, N1276087, N1276088, N1276091, N1276093, N1276094, N1276096, N1276098, N1276104, N1276107, N1276116, N1276121, N1276122, N1276124, N1276126, N1276130, N1276135, N1276136, N1276142, N1276143, N1276146, N1276150, N1276153, N1276154, N1276156, N1276157, N1276160, N1276164, N1276166, N1276167, N1276331, N1276336, N1276339, N1276342, N1276343, N1276345, N1276347, N1276354, N1276357, N1276360, N1276362, N1276365, N1276367, N1276371, N1276372, N1276376, N1276377, N1276379, N1276381, N1276385, N1276390, N1276393, N1276394, N1276395, N1276397, N1276398, N1276399, N1276402, N1276404, N1276407, N1276409, N1276411, N1276413, N1276428, N1276429, N1276431, N1276432, N1276435, N1276440, N1276441, N1276443, N1276445, N1276446, N1276449, N1276450, N1276451, N1276453, N1276454, N1276455, N1276458, N1276461, N1276463, N1276464, N1276468, N1276469, N1276472, N1276473, N1276475, N1276480, N1276482, N1276483, N1276484, N1276486, N1276488, N1276491, N1276494, N1276496, N1276505, N1276508, N1276509, N1276510, N1276511, N1276516, N1276523, N1276526, N1276537, N1276547, N1276551, N1276572, N1276575, N1276578, N1276579, N1276582, N1276589, N1276612, N1276617, N1276621, N1276628, N1276649, N1276653, N1276863, N1276866, N1276876, N1276878, N1276890, N1276891, N1276897, N1276932, N1276942, N1276944, N1276950, N1277084, N1277123, N1277132, N1277136, N1277137, N1277150, N1277160, N1277164, N1277367; 4/30/2029: N1277411, N1277427, N1277443, N1277449, N1277518, N1277564, N1277598, N1277611, N1277624, N1277625, N1277654, N1277897, N1277911, N1278101, N1278309, N1278317, N1278371, N1278378, N1278621, N1278644, N1278649, N1278658, N1278728, N1278729, N1278742, N1278750, N1278751, N1278763, N1278764, N1278765, N1278770, N1278783, N1278795, N1278814, N1278828, N1278850, N1278986, N1279005, N1279021, N1279094, N1279142, N1279156, N1279346, N1279370, N1279437, N1279441, N1279446, N1279455, N1279465.
DAILIES TOTAL1 Multifocal/00846566847112, 00846566850341, 00846566850235, 00846566850273, 00846566850259, 00846566850280, 00846566849581, 00846566851355, 00846566849383, 00846566845668, 00846566847983, 00846566849932, 00846566850686, 00846566850570, 00846566851386, 00846566851331, 00846566848881, 00846566848324, 00846566848386, 00846566848898, 00846566848348, 00846566847587, 00846566850839, 00846566850846, 00846566850297, 00846566850266, 00846566847488, 00846566847280, 00846566850563, 00846566850372, 00846566849789, 00846566850426, 00846566849888, 00846566851058, 00846566851188, 00846566849048, 00846566850921, 00846566850242, 00846566849598, 00846566845729, 00846566848676, 00846566848591, 00846566846382, 00846566850518, 00846566850440, 00846566850662, 00846566850624, 00846566849178, 00846566847846, 00846566848652, 00846566851096, 00846566849208, 00846566845927, 00846566849956, 00846566851126, 00846566846375, 00846566850600, 00846566846214, 00846566850891, 00846566847228, 00846566850853, 00846566850198, 00846566850310, 00846566850204, 00846566848485, 00846566846528, 00846566849901, 00846566849093, 00846566848607, 00846566849741, 00846566851041, 00846566851089, 00846566851201, 00846566850433, 00846566850938, 00846566850914, 00846566846320, 00846566850464, 00846566847051, 00846566849895, 00846566846436,00846566846153, 00846566846146, 00846566849734, 00846566850549, 00846566850389, 00846566848263, 00846566850228, 00846566850907, 00846566850860, 00846566850211, 00846566850457, 00846566848447, 00846566850396, 00846566851195, 00846566849222, 00846566848416, 00846566851171, 00846566849871, 00846566849109, 00846566850495, 00846566847860, 00846566851140, 00846566850884, 00846566847730, 00846566848393, 00846566849123, 00846566848492, 00846566850501, 00846566851102, 00846566851317/ 3/31/2029: N1275728, N1275730, N1275731, N1275732, N1275733, N1275736, N1275739, N1275740, N1275741, N1275744, N1275751, N1275752, N1275753, N1275754, N1275761, N1275767, N1275769, N1275770, N1275773, N1275805, N1275806, N1275807, N1275808, N1275809, N1275811, N1275812, N1275813, N1275814, N1275815, N1275816, N1275817, N1275818, N1275819, N1275820, N1275821, N1275822, N1275823, N1275824, N1275827, N1275829, N1275830, N1275831, N1275847, N1275849, N1275852, N1275855, N1275857, N1275858, N1275860, N1276178, N1276180, N1276181, N1276183, N1276189, N1276198, N1276199, N1276201, N1276203, N1276209, N1276210, N1276212, N1276213, N1276216, N1276219, N1276220, N1276221, N1276222, N1276223, N1276224, N1276232, N1276233, N1276234, N1276235, N1276237, N1276239, N1276242, N1276243, N1276275, N1276276, N1276281, N1276286, N1276288, N1276291, N1276292, N1276293, N1276294, N1276303, N1276304, N1276305, N1276306, N1276308, N1276311, N1276313, N1276314, N1276316, N1276317, N1276320, N1276321, N1276323, N1276325, N1276327, N1276689, N1276695, N1276703, N1276714, N1276728, N1276793, N1276848, N1277330, N1277331, N3500526, N3500528, N3500530, N3500531, N3500532, N3500534, N3500536; 4/30/2029: N1277754, N1278436, N1278440, N1278444, N1278446, N1278448, N1278454, N1278455, N1278457, N1278460, N1278461, N1278464, N1278472, N1278475, N1278476, N1278478, N1278482, N1278485, N1278489, N1278506, N1278507, N1278508, N1278586, N1278587, N1278588, N1278597, N1278598, N1278600, N1278606, N1279162, N1279163, N1279165, N1279166, N1279190, N1279193, N1279194, N1279217, N1279221, N1279223, N1279311, N1279314, N1279319.
TOTAL30 Sphere/00730822287708, 00730822287739, 00730822287654, 00730822287616, 00730822287630, 00730822287746, 00730822287791, 00730822287685, 00730822287593, 00730822287692, 00730822287678, 00730822287647, 00730822287760, 00730822287906, 00730822288477, 00730822287753, 00730822287487, 00730822287449, 00730822287401, 00730822287807, 00730822287333, 00730822287395, 00730822287777, 00730822287463, 00730822287784, 00730822287562, 00730822287623, 00730822288385, 00730822287722, 00730822287609, 00730822287715, 00730822287814, 00730822287661, 00730822288415, 00730822288293, 00730822288309, 00730822288361, 00730822288323/ 3/31/2031: N4159824, N4159826, N4159828, N4159830, N4159832, N4159834, N4159838, N4159840, N4159842, N4159844, N4159846, N4159848, N4159854, N4159856, N4159858, N4159860, N4159866, N4159870, N4159872, N4160926, N4160927, N4160928, N4160929, N4160930, N4160930, N4160931, N4160931, N4160932, N4160933, N4160934, N4160934, N4160936, N4160938, N4160939, N4160941, N4160942, N4160943, N4160944, N4160945, N4160946, N4160947, N4160948, N4160952, N4160953, N4160954, N4160955, N4160957, N4162814, N4162815, N4162817, N4162818, N4162819, N4162820, N4162821, N4162822, N4162823, N4162825, N4162826, N4162827, N4162828, N4162829, N4162830, N4162831, N4162832, N4162833, N4162834, N4162835, N4162836, N4162837, N4162838, N4162839, N4162840, N4162841, N4162842, N4162843, N4162844, N4162845, N4162846, N4162847, N4162848, N4162849, N4162850, N4162852, N4162853, N4162854, N4162855, N4162857, N4162858, N4162859, N4162860, N4162861, N4162862, N4162863, N4162864, N4162865, N4162866, N4162867, N4162868, N4162869, N4162870, N4162873, N4162874, N4162876, N4162877
|
| FEI Number |
1610287
|
Recalling Firm/ Manufacturer |
Alcon Research LLC
6201 South Fwy
Aspex Facility
Fort Worth
TX
76134-2099
|
| For Additional Information Contact | Heather Attra 817-551-8937 |
Manufacturer Reason for Recall | Material with a quality issue was used in contact lens production, so contact lenses may not meet quality and/or performance standards for the entirety of their labeled shelf life, and wearing these lenses may lead to temporary issues such as irritation, redness, or blurred vision. |
FDA Determined Cause 2 | Process change control |
| Action | On 8/8/24, recall notices were sent to customers who were asked to do the following:
1) Discontinue any further distribution of affected product.
2) Dispose of all affected product. If immediate destruction is not feasible, ensure the identified lots are segregated from unaffected products.
3) Forward this notification to departments within your organization who may be in possession of the affected product.
4) Contact other organizations, eye care providers and/or individuals to whom you have transferred or distributed the identified lots, and patients, and inform them of this Voluntary Medical Device Recall, and request they dispose of contact lenses from the identified lots. For your convenience, a notification template is enclosed for communicating this issue to your customers/patients.
3) Complete and return the response form via email to market.actions@alcon.com Alternately, respond via QR code: myalcon.com/notification-response-form
If you have questions about the notice, or replacement process, or need electronic copies of the notification letter, call 1-888-266-7974 |
| Quantity in Commerce | 92,835 Packs |
| Distribution | US: UT, TX, NC, MT, AL, WA, MS, PA, LA, MI, GA, FL, KY, OH, AZ, IL, CA, NJ, TN, NY, AR, MN, WI, MD, OK, SC, OR, MA, CO, VA, CT, IA, DE, WV, MO, ID, SD, NE, NV, IN, ME, VT, NM, NH, AK, HI, KS, RI, WY, PR, DC.
OUS: GB, NL, HK, ES, PL, IT, FR, CA, CH, PT, AT, DE, NO, GG, LU, CZ, SK, DK, FI, SE, IN, OM, BE, QA, KW, SA, AE, LB, BH, HU, MY, IE, SG, IS, XI, MK, GI, MT, KR, SI, NZ, GR, GF, GP, MQ, TW, EE, GL, GE, IL, UA, FO, LV, AD, RS, LT, RO, SM, AZ, AU, JE, HR, AX, LI, CY, BG
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LPL
|
|
|
|