| | Class 2 Device Recall Quantum Perfusion |  |
| Date Initiated by Firm | March 13, 2023 |
| Date Posted | October 24, 2024 |
| Recall Status1 |
Completed |
| Recall Number | Z-0224-2025 |
| Recall Event ID |
95428 |
| 510(K)Number | K173834 |
| Product Classification |
Console, heart-lung machine, cardiopulmonary bypass - Product Code DTQ
|
| Product | Quantum Perfusion Systems for CPB, Roller Pumps use peristaltic action to propel blood through the venous and arterial circuit tubing. Catalog Numbers: a) 51-000013-00, b) 51-000008-00 |
| Code Information |
UDI/DI 05060434420770: Serial Numbers: a) AU6004040, AU6004041, AU6004042, AU6004043, AU6004044, AU6004046, AU6004050, AU6004051, AU6004052, AU6004053, AU6004054, AU6004056, AU6004057, AU6004058, AU6004060, AU6004061, AU6004062, AU6004063, AU6004064, AU6004065, AU6004066, AU6004069, AU6004070, AU6004071, AU6004072, AU6004073, AU6004074, AU6004075, AU6004136, AU6004137, AU6004138, AU6004139, AU6004140, AU6004141, AU6004142, AU6004143, AU6004144, AU6004145, AU6004146, AU6004147, AU6004148, AU6004159, AU6004160, AU6004161, AU6004162, AU6004167, AU6004168, AU6004169, AU6004170, AU6004171, AU6004172, AU6004173, AU6004174, AU6004175, AU6004176, AU6004177, AU6004178, AU6004179, AU6004180, AU6004181, AU6004182, AU6004183, AU6004184, AU6004185, AU6004186, AU6004187, AU6004188, AU6004195, AU6004197, AU6004198, AU6004200, AU6004201, AU6004202, AU6004203, AU6004204, AU6004205, AU6004207, AU6004332, AU6004333, AU6004348, AU6004349, AU6004350, AU6004352, AU6004353, AU6004354, AU6004532, AU6004577, AU6004578, AU6004579, AU6004580, AU6004587, AU6004588, AU6004590, AU6004591, AU6004603, AU6004604, AU6004605, AU6004606, AU6004607, AU6004609, AU6004610, AU6004613, AU6004731, AU6004771, AU6004773, AU6004776, AU6004777, AU6004778, AU6004779, AU6004781, AU6004782, AU6004919, AU6004920, AU6004921, AU6004922, AU6004923, AU6004924, AU6004925, AU6004926, AU6004927, AU6004929, AU6004930, AU6004955, AU6004957, AU6004958, AU6004960, AU6004963, AU6004966, AU6004968, AU6004970, AU6004971, AU6004972, AU6005010, AU6005021, AU6005022, AU6005023, AU6005024, AU6005025, AU6005026, AU6005027, AU6005028, AU6005029, AU6005030, AU6005037, AU6005038, AU6005039, AU6005040, AU6005041, AU6005042, AU6005057, AU6005065, AU6005109, AU6005110, AU6005118, AU6005124, AU6005125, AU6005355, AU6006351;
Serial Numbers: b) AU8001273, AU8001275, AU8001346, AU8001347, AU8001348, AU8001359, AU8001360, AU8001366, AU8001403, AU8001406, AU8001407, AU8001446, AU8001447, AU8001449 |
| FEI Number |
3006073153
|
Recalling Firm/ Manufacturer |
Spectrum Medical Ltd.
Harrier 4, Meteor Business Park
Cheltenham Road East
Gloucester
United Kingdom
|
Manufacturer Reason for Recall | The ratchet and pawl had an out of tolerance geometry which exhibited as the left bobbin would not lock. |
FDA Determined Cause 2 | Component design/selection |
| Action | All potentially affected product was identified. The firm corrected all products between 03/13/2024 and 08/07/2024 during service visits. |
| Quantity in Commerce | 172 units |
| Distribution | US Nationwide distribution including in the states of Alabama, California, Florida, Kentucky, Maine, Massachusetts, Minnesota, Missouri, New Jersey, New York, North Carolina, Ohio, Oklahoma, Pennsylvania, Tennessee, Texas, Utah, West Virginia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
|