| | Class 2 Device Recall Atellica IM Homocysteine Calibrator |  |
| Date Initiated by Firm | April 15, 2025 |
| Date Posted | May 27, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1821-2025 |
| Recall Event ID |
96693 |
| Product Classification |
Calibrator, primary - Product Code JIS
|
| Product | Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498 |
| Code Information |
Siemens Material Number: 10995498.
UDI-DI: 00630414598253
Kit lot numbers: 86236A74, 44206A75.
***Updated 5/8/26***
Additional Lots 25675A77, 44505A78, 52364A79. |
| FEI Number |
1219913
|
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics, Inc.
333 Coney St
East Walpole
MA
02032-1516
|
| For Additional Information Contact | 508-668-5000 |
Manufacturer Reason for Recall | Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias. |
FDA Determined Cause 2 | Under Investigation by firm |
| Action | An Urgent Medical Device Correction (UMDC) was mailed to all US affected customers by FedEx, and an Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens reporting sites with affected customers on about 04/15/2025. Customers were instructed to review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, discontinue use of and discard the affected devices, and complete and return the Effectiveness Check questionnaire. Customers were also instructed to forward the letter to those who may have received the affected devices.
***Updates:
***On 10/21/25, a second notice was mailed to all affected US customers to inform them that an additional lot was added to the scope of the recall. Kit lots ending in 77 are affected.
***On 5/1/26, a third notice was mailed to all affected US customers to inform them that an additional affected lot was added. Kit lots ending in 78 and 79 are affected. |
| Quantity in Commerce | 2,212 packs (4,424 units) - 448 packs US, 1,764 packs OUS |
| Distribution | Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, VT, WI, WY and the countries of Austria, Belgium, Croatia, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lithuania, Nepal, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Vatikan City, Zaire.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
|
|
|