| Date Initiated by Firm | April 25, 2025 |
| Date Posted | June 10, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1938-2025 |
| Recall Event ID |
96830 |
| 510(K)Number | K992637 |
| Product Classification |
Detector and alarm, arrhythmia - Product Code DSI
|
| Product | MUSE 5 Systems, Model Numbers 2021274-001, 2023032-005, 2027478-001, Large capacity client-server-based computer system that accesses stores and manages cardiovascular information. |
| Code Information |
Serial Numbers: ABF04400266FA
SAD04390481GA
SAD05091885GA
SAD05091907GA
SAD05091908GA
PA2155054
PA2155057
PA2155059
PA2155061
PA2155063
SAD05302639GA
SAD05302646GA
SAD05322687GA
SAD05322723GA
SAD05322735GA
SAD06524784GA
SAD06204075GA
SAD06204076GA
SAD06204077GA
SAD04380416GA
SAD04380427GA
SAD04430607GA
SAD04440716GA
SAD04440717GA
SAD04440718GA
SAD04440719GA
SAD05453127GA
SAD05453128GA
SAD05453129GA
SAD05453130GA
SAD05453172GA
SAD08105319GA
SAD08105320GA
SAD04501236GA
SAD04501237GA
SAD04501252GA
SAD04501257GA
SAD04501258GA
SAD04350305GA
SAD04350317GA
SAD04350318GA
SAD04350319GA
SAD04350320GA
SAD04350326GA
SAD06424592GA
SAD06444630GA
SAD06454662GA
SAD04491128GA
SAD04491129GA
SAD04491130GA
SAD04491131GA
SAD04491132GA
SAD04491151GA
SAD04501216GA
SAD04501217GA
SAD04501218GA
SAD04501220GA
SAD04501221GA
SAD04501222GA
SAD04501223GA
SAD04501224GA
SAD04501225GA
SAD04501226GA
SAD04501227GA
SAD04501228GA
SAD04501229GA
PA2020935
PA2020936
SAD05081763GA
SAD05081836GA
SAD05091867GA
SAD05091868GA
SAD0932
SAD09335621GA
SAD09335622GA
SAD05302635GA
SAD06344526GA
SAD05091896GA
SAD05091905GA
SAD05091906GA
SAD06424610GA
SAD06444638GA
SAD06194039GA
SAD06194040GA
SAD06194041GA
SAD05172239GA
SAD05172240GA
SAD05172241GA
SAD05172242GA
SAD05172243GA
SAD05172268GA
SAD05172269GA
SAD05172271GA
SAD06204073GA
SAD05382932GA
SAD05392967GA
SAD07455199GA
SAD08285434GA
SAD06113777GA
SAD06113778GA
SAD06113779GA
SAD06113780GA
PA2297273
SAD07205030GA
SAD07205031GA
SAD07215068GA
SAD07215069GA
SAD05302649GA
SAD05322725GA
SAD05332753GA
SAD05352769GA
SAD05473247GA
SAD06023533GA
SAD06023574GA
SAD06474699GA
SAD05513429GA
SAD05513433GA
SAD05513435GA
SAD05513436GA
SAD05523459GA
SAD08355476GA
SAD05021340GA
SAD05021369GA
SAD05021370GA
SAD05021371GA
SAD05352768GA
SAD05352770GA
SAD05352776GA
SAD05352777GA
SAD05503390GA
SAD05503394GA
SAD05503395GA
SAD05523447GA
SAD05523448GA
SAD05523449GA
SAD06033597GA
SAD06033598GA
SAD06033599GA
SAD06033600GA
SAD06033601GA
SAD04491155GA
SAD04491156GA
SAD04491157GA
SAD06444629GA
SAD06474693GA
SAD05061599GA
SAD04450783GA
SAD04470920GA
SAD04470921GA
SAD04470922GA
SAD04470923GA
SAD04430609GA
SAD04430610GA
SAD04440662GA
SAD04440663GA
SAD04390495GA
SAD04390496GA
SAD04390528GA
SAD04390529GA
SAD04440723GA
SAD04440724GA
SAD04440725GA
SAD04390482GA
SAD04390483GA
SAD04390484GA
SAD05503399GA
SAD06204071GA
SAD06214093GA
SAD06214094GA
SAD06214095GA
SAD04450735GA
SAD04450736GA
SAD06454656GA
SAD06454659GA
SAD06073742GA
SAD04511286GA
SAD04511314GA
SAD04511315GA
SAD04511317GA
SAD06204062GA
SAD06204063GA
SAD06234131GA
SAD06234132GA
SAD06234133GA
SAD06234134GA
SAD06394581GA
SAD06524786GA
SAD06524787GA
SAD06524788GA
SAD07024812GA
SAD06113788GA
SAD04410533GA
SAD04410534GA
SAD04410535GA
SAD04420575GA
SAD04420577GA
SAD04420578GA
PA2305331
SAD06113828GA
SAD06123870GA
SAD06123877GA
SAD06123878GA
SAD05031412GA
SAD05031413GA
SAD05031414GA
SAD05031415GA
SAD05031416GA
SAD05041424GA
SAD05041425GA
SAD05041426GA
SAD05041427GA
SAD05041428GA
SAD05041429GA
SAD05041430GA
SAD05041431GA
SAD05041432GA
SAD05041433GA
SAD05041434GA
SAD05041435GA
SAD05041436GA
SAD05041437GA
SAD05041438GA
SAD05041439GA
SAD05041440GA
SAD05041446GA
SAD05041447GA
SAD05041448GA
SAD05041449GA
SAD05051450GA
SAD05051451GA
SAD05051452GA
SAD05051453GA
SAD05051454GA
SAD05051455GA
SAD05051456GA
SAD05051457GA
SAD05051458GA
SAD05051459GA
SAD05051460GA
SAD05051461GA
SAD05051462GA
SAD05051463GA
SAD05051464GA
SAD05051465GA
SAD05051466GA
SAD05051467GA
SAD05051468GA
SAD05051469GA
SAD05051470GA
SAD05051471GA
SAD05051472GA
SAD05051473GA
SAD05051474GA
SAD05051475GA
SAD05392976GA
SAD05392978GA
SAD05392979GA
SAD05392980GA
SAD05392981GA
SAD05403010GA
SAD05403011GA
SAD05403012GA
SAD05392977GA
SAD05292610GA
SAD05302640GA
SAD05302641GA
SAD05182284GA
SAD05182297GA
SAD05182299GA
SAD05182300GA
SAD05182302GA
SAD04380402GA
SAD04380403GA
SAD04390443GA
SAD04390444GA
SAD04390447GA
SAD04390449GA
SAD04440671GA
PA2020328
SAD06143930GA
SAD06143931GA
SAD06143932GA
SAD05111959GA
SAD05142093GA
SAD06073699GA
SAD06073700GA
SAD06073704GA
SAD06073705GA
SAD06073706GA
SAD06073707GA
SAD06073702GA
SAD06073738GA
SAD06073739GA
SAD06073740GA
SAD07084847GA
SAD05081857GA
SAD05091880GA
SAD05091881GA
SAD04460880GA
SAD08515553GA
SBR08522301GA
SAD04450738GA
SAD04450739GA
SAD04450740GA
SAD05111954GA
SAD05111961GA
SAD05051548GA
SAD05051549GA
SAD05061554GA
SAD05061555GA
SAD05061556GA
SAD05061557GA
SAD05061558GA
SAD05061559GA
SAD05061560GA
SAD05061561GA
SAD05061562GA
SAD05061563GA
SAD05061564GA
SAD05061565GA
SAD05061566GA
SAD05061567GA
SAD05061568GA
SAD05061569GA
SAD05061570GA
SAD05061571GA
SAD05061572GA
SAD05061573GA
SAD05061574GA
SAD05061575GA
SAD05061580GA
SAD05061581GA
SAD05061582GA
SAD05061583GA
SAD05061584GA
SAD05061585GA
SAD05061586GA
SAD05061587GA
SAD05061588GA
SAD05061589GA
SAD05061590GA
SAD05061591GA
SAD05061592GA
SAD05061593GA
SAD05061594GA
SAD05061595GA
SAD05061596GA
SAD05061597GA
SAD05061598GA
SAD05061600GA
SAD05061601GA
SAD05061602GA
SAD05061603GA
SAD05061604GA
SAD05061605GA
SAD06344539GA
SAD06474694GA
SAD07215060GA
SAD07325138GA
SAD07325135GA
SAD04330300GA
SAD04330301GA
SAD05332754GA
SAD05352771GA
SAD05352772GA
SAD05443097GA
SAD05443113GA
SAD05443114GA
SAD05443115GA
SAD05443116GA
SAD05443117GA
SAD05443118GA
SAD05443119GA
SAD05443120GA
SAD05021324GA
SAD05021366GA
SAD05021367GA
SAD05021368GA
SAD06524761GA
SAD06524762GA
SAD06524763GA
SAD06524772GA
SAD06524776GA
SAD06524777GA
SAD06524778GA
SAD06073736GA
SAD06073744GA
SAD06093751GA
SAD06093752GA
SAD06093763GA
SAD06093764GA
SAD06123845GA
SAD06123846GA
SAD06123847GA
SAD06123848GA
SAD06123849GA
SAD06123850GA
SAD06123851GA
SAD06123852GA
SAD06123854GA
SAD06123855GA
PA2021683
SAD05423038GA
SAD05423039GA
SAD05172237GA
SAD05172238GA
SAD05172262GA
SAD05172263GA
SAD05172264GA
SAD05172265GA
SAD05172266GA
SAD05172274GA
SAD05172277GA
SAD07475203GA
SAD06113789GA
SAD06113790GA
SAD06113791GA
SAD05423043GA
SAD04440687GA
SAD04511300GA
SAD06334353GA
SAD05302632GA
SAD05302633GA
SAD05322721GA
SAD05332751GA
SAD05352803GA
SAD05362816GA
SAD05362837GA
SAD05423029GA
SAD06294292GA
SAD04450745GA
SAD05111947GA
SAD05121967GA
SAD05121968GA
SAD05121980GA
SAD06143915GA
SAD06143916GA
SAD04460793GA
SAD04460794GA
SAD04460795GA
SAD04460796GA
SAD04460797GA
SAD04460798GA
SAD04460799GA
SAD04460800GA
SAD04460801GA
SAD04460805GA
SAD05111953GA
SAD05111955GA
SAD05111956GA
SAD05111957GA
SAD07475213GA
PA2156447
SAD06073703GA
SAD05322722GA
SAD05332752GA
SAD05352779GA
SAD05352780GA
SAD04460843GA
SAD05503400GA
SAD05503401GA
SAD05503402GA
SAD05503403GA
SAD05503404GA
SAD05513430GA
SAD05513437GA
SAD05513438GA
SAD05081843GA
SAD05121963GA
SAD05121964GA
SAD05121965GA
SAD05121966GA
SAD05121978GA
SAD05121979GA
SAD05132030GA
SAD05523456GA
SAD06053651GA
SAD06053666GA
SAD06123829GA
SAD06123830GA
SAD06123838GA
SAD05443098GA
SAD05142107GA
SAD05142108GA
SAD05142109GA
SAD05142110GA
SAD05142116GA
SAD05152183GA
SAD06123839GA
SAD05051481GA
SAD05051482GA
SAD05051487GA
SAD05051491GA
SAD05051505GA
SAD05051507GA
SAD05051508GA
SAD05051523GA
SAD05051524GA
SAD05051525GA
SAD05051526GA
SAD05051527GA
SAD04491020GA
SAD09495631GA
SAD05352762GA
SAD05352763GA
SAD05352764GA
SAD05362836GA
SAD05362850GA
SAD05252554GA
SAD05292630GA
SAD05302675GA
PA2464753
SAD04330202GA
SAD04330203GA
SAD04370333GA
SAD04370337GA
SAD04380353GA
SAD04380354GA
SAD04380355GA
SAD04380356GA
SAD04491091GA
SAD04491135GA
SAD04491141GA
SAD04491142GA
SAD04491143GA
SAD04491144GA
SAD04491145GA
SAD04491146GA
SAD04491147GA
SAD05392960GA
SAD05392961GA
SAD05392962GA
SAD05392964GA
SAD05392995GA
SAD07385168GA
SAD05071625GA
SAD05071626GA
SAD05071627GA
SAD05071628GA
SAD05071629GA
SAD05071630GA
SAD05071631GA
SAD05332740GA
SAD05332741GA
SAD05332742GA
SAD05332758GA
SAD05332759GA
SAD04330275GA
SAD04330276GA
SAD04330277GA
SAD04330278GA
SAD04330279GA
SAD04330296GA
SAD04330297GA
SAD05071693GA
SAD05081800GA
SAD05081801GA
SAD05081834GA
SAD05081837GA
SAD05483288GA
SAD05483289GA
SAD05483290GA
SAD05483291GA
SAD05483292GA
SAD05493328GA
SAD05463188GA
SAD05463189GA
SAD05473242GA
SAD05483269GA
SAD05483302GA
SAD05483308GA
SAD05493311GA
SAD05493342GA
SAD05212410GA
SAD05222466GA
SAD05222467GA
SAD05222469GA
SAD05222470GA
SAD08205342GA
SAD05252559GA
SAD05252560GA
SAD05262580GA
SAD05262581GA
ABF04290227FA
ABF04450276FA
ABF04450277FA
SAD06424601GA
ABF05070319FA
SAD05071610GA
SAD06514758GA
SAD05172244GA
SAD05172245GA
SAD05192355GA
SAD06204074GA
SAD06204081GA
SAD06204082GA
SAD06204083GA
SAD06204084GA
SAD06204085GA
SAD06204086GA
SAD06234117GA
SAD05473252GA
SAD05473253GA
SAD05473254GA
SAD05473255GA
SAD05473256GA
SAD05483279GA
SAD06093765GA
SAD06093766GA
SAD06093767GA
SAD06093768GA
SAD06023519GA
SAD06023520GA
SAD06194054GA
SAD07335152GA
SAD07335153GA
SAD07335154GA
SAD07335155GA
SAD07335156GA
SAD07335157GA
SAD07375163GA
SAD07375164GA
SAD06093760GA
SAD06093761GA
SAD06093762GA
SAD06093769GA
SAD06093770GA
SAD04330272GA
SAD05493318GA
SAD05493354GA
SAD05493355GA
SAD06264265GA
SAD07325133GA
SAD08105316GA
SAD08105318GA
SAD08175328GA
SAD08175329GA
SAD08175330GA
SAD08175331GA
SAD08175332GA
SAD08175333GA
SAD08205335GA
SAD08205337GA
SAD08205338GA
SAD08205340GA
SAD08205343GA
SAD08205345GA
SAD08205346GA
SAD08205348GA
SAD08205351GA
SAD08205353GA
SAD10465642GA
SAD05483298GA
ABF05010305FA
ABF05010306FA
ABF05010307FA
SAD04511304GA
SAD06194049GA
SAD06194055GA
SAD06194057GA
SAD06194058GA
SAD06194059GA
SAD06204087GA
SAD09255604GA
SAD09255605GA
SAD05111960GA
SAD05142074GA
SAD05212432GA
SAD05212433GA
SAD05252555GA
SAD05252556GA
PA2021453
SAD05443101GA
SAD05443102GA
SAD05443103GA
SAD05453144GA
SAD05453145GA
SAD05453146GA
SAD05453147GA
SAD05453148GA
SAD05453149GA
SAD05453150GA
SAD05453151GA
SAD04420606GA
SAD05051532GA
SAD05051533GA
SAD06073731GA
SAD06093753GA
SAD06093754GA
SAD06093755GA
SAD06344528GA
SAD06113818GA
SAD05071633GA
SAD05071640GA
SAD05071657GA
SAD05091871GA
SAD05091872GA
SAD05091875GA
SAD05051513GA
SAD05051500GA
SAD05051501GA
SAD05051506GA
SAD05051512GA
SAD05051518GA
SAD08055298GA
SAD08055299GA
SAD08055309GA
SAD04450756GA
SAD07124950GA
SAD04460822GA
SAD04460825GA
SAD04460829GA
SAD04460871GA
SAD04460872GA
SAD04460881GA
SAD04470943GA
SAD04470944GA
SAD04470945GA
SAD04470946GA
SAD04330264GA
SAD06344515GA
SAD05372908GA
PA2305487
SAD05473232GA
SAD05503382GA
SAD05503383GA
PA2166977
SAD05493353GA
SAD08305445GA
SAD04390458GA
SAD04470939GA
SAD07335148GA
PA2020349
PA2020350
SAD06234121GA
SAD06133897GA
SAD06133899GA
SAD06344514GA
SAD06424605GA
SAD06424606GA
SAD06424607GA
SAD06424612GA
SAD06424613GA
SAD06254215GA
SAD06254216GA
SAD06254217GA
SAD05252570GA
SAD05262584GA
SAD07175018GA
SAD07205027GA
PA2021454
SAD06254196GA
SAD06254197GA
SAD06254198GA
SAD06254199GA
SAD06254200GA
SAD06254201GA
SAD06254202GA
SAD08205339GA
SAD08355467GA
SAD08355469GA
SAD09335614GA
SAD09335616GA
SAD09335618GA
SAD06334340GA
SAD06334341GA
SAD06334348GA
SAD06334349GA
SAD06334350GA
SAD06334351GA
SAD07175016GA
SAD08285441GA
SAD06023572GA
SAD06023573GA
SAD06053630GA
SAD07475208GA
SAD08235374GA
SAD08285440GA
ABF05140338FA
SAD05132015GA
SAD05132016GA
SAD05132017GA
SAD05132018GA
SAD05132019GA
SAD05132020GA
SAD05132021GA
SAD05132022GA
SAD05142087GA
SAD05142132GA
SAD05362845GA
SAD05362846GA
SAD06073718GA
ABF05200350FA
SAD05182287GA
SAD05182291GA
SAD05182293GA
SAD05182322GA
SAD05182323GA
SAD05182324GA
SAD05182325GA
SAD06284283GA
SAD06284284GA
SAD06284285GA
SAD08095312GA
SAD06133884GA
SAD06133885GA
SAD06133886GA
SAD06133887GA
SAD06254218GA
SAD06474686GA
SAD06474687GA
SAD07285071GA
SAD07305105GA
SAD04430640GA
SAD04410543GA
SAD04330214GA
PA2021185
PA2021455
SKW14530253GA
SAD08515556GA
SAD09255599GA
SAD09255600GA
SAD06324317GA
SAD06324319GA
SAD06324328GA
SAD04491063GA
SAD04491064GA
ABF04500295FA
ABF04500296FA
SAD04491096GA
SAD04491097GA
SAD04491098GA
SAD04491099GA
SAD04491102GA
SAD04491103GA
SAD07175017GA
SAD09185594GA
PA0000277
SAD05232510GA
SAD05232511GA
SAD05232512GA
SAD07134964GA
PA1311009
SAD06344529GA
SAD07124909GA
SAD07124951GA
SAD05352784GA
SAD04390464GA
SAD04390468GA
SAD04390469GA
SAD04390470GA
SAD04390471GA
SAD05322710GA
SAD05322727GA
SAD06344520GA
SAD07144971GA
SAD07375159GA
SAD07375160GA
ABF04480282FA
SAD04470949GA
SAD04470952GA
SAD04470953GA
SAD04470956GA
SAD04491028GA
SAD04491029GA
SAD04491030GA
SAD04491031GA
SAD04491032GA
SAD07124922GA
SAD06053646GA
SAD05091869GA
SAD05091870GA
SAD05091873GA
SAD06334354GA
PA2295549
ABF05110329FA
ABF05110330FA
SAD05091895GA
SAD05091897GA
SAD05212426GA
SAD05212427GA
SAD05212428GA
SAD05222440GA
SAD06524785GA
SAD05302662GA
SAD05302663GA
SAD05322706GA
SAD08355472GA
SAD04511280GA
SAD05021329GA
SAD05021330GA
SAD05021348GA
SAD06234165GA
SAD04390472GA
SAD07084863GA
PA2021145
PA2021146
SAD05142086GA
SAD05142117GA
SAD05142146GA
SAD05142147GA
SAD05142153GA
SAD05142154GA
SAD05142155GA
SAD06444640GA
SAD06454652GA
SAD07124910GA
SAD07144970GA
SAD05302661GA
SAD04280166GA
SAD04491190GA
SAD04460879GA
SAD05292620GA
SAD06143928GA
SAD06143929GA
SAD04380386GA
SAD07515276GA
SAD09335623GA
SAD04330304GA
SAD04370340GA
SAD04370342GA
SAD04370343GA
SAD04501256GA
SAD04501264GA
SAD04501265GA
SAD04511296GA
SAD04511297GA
SAD06234114GA
SAD06254192GA
SAD06254219GA
SAD06254220GA
PA2166167
PA2166168
PA2166169
PA2166170
PA2166171
SAD05443104GA
SAD05443112GA
SAD05453160GA
SAD05453171GA
SAD05453173GA
SAD05453174GA
SAD05453175GA
SAD05453176GA
SAD05453177GA
SAD05453178GA
SAD05443111GA
SAD05453170GA
SAD07435184GA
SAD04390459GA
SAD04390460GA
PA2305484
PA2305506
SAD05423033GA
SAD06194060GA
SAD06204072GA
SAD05392988GA
SAD05402997GA
SAD05402998GA
SAD05402999GA
SAD05403000GA
SAD05403001GA
SAD05403002GA
SAD05403003GA
SAD05403004GA
SAD05403005GA
SAD06023563GA
SAD06023564GA
SAD06023565GA
SAD06023566GA
SAD06023567GA
ABF05150344FA
SAD05081846GA
SAD05081858GA
SAD05142072GA
SAD05372873GA
SAD05372874GA
SAD07124932GA
SAD07124933GA
SAD06033613GA
SAD05443107GA
SAD05443108GA
SAD05443109GA
SAD05453132GA
SAD05453133GA
SAD05453134GA
SAD05453136GA
SAD05453138GA
SAD05453139GA
SAD05453140GA
SAD05453141GA
SAD05453142GA
SAD05051480GA
SAD05051484GA
SAD05302659GA
SAD04491043GA
SAD04491071GA
SAD05362866GA
SAD05352788GA
SAD05352789GA
SAD05362841GA
SAD04370327GA
SAD04370328GA
SAD04370329GA
SAD04370330GA
SAD04370331GA
SAD04370332GA
SAD04370335GA
SAD04460884GA
SAD04491191GA
SAD04491193GA
SAD05523500GA
SAD05523502GA
SAD05523503GA
SAD05523506GA
SAD05523507GA
SAD05523508GA
SAD05523509GA
SAD06053647GA
SAD06053648GA
SAD06053649GA
SAD06053650GA
SAD06053652GA
SAD06053653GA
SAD06053654GA
SAD06053655GA
SAD06053656GA
SAD06053657GA
SAD06053658GA
SAD06053659GA
SAD06053660GA
SAD06053661GA
SAD06053662GA
SAD06053663GA
SAD06053664GA
SAD06053665GA
SAD06053667GA
SAD06053668GA
SAD06053669GA
SAD06053670GA
SAD06053671GA
SAD06053672GA
SAD06063690GA
SAD06063691GA
SAD06063692GA
SAD06063693GA
SAD08095310GA
SAD07435183GA
SAD06123871GA
SAD06123872GA
SAD06123873GA
SAD06123874GA
SAD05302676GA
SAD05302677GA
SAD05302680GA
SAD05322684GA
SAD05322685GA
SAD05322686GA
SAD05322709GA
SAD05322730GA
SAD05322731GA
SAD05332743GA
SAD05332744GA
SAD05332745GA
SAD05332746GA
SAD05332750GA
SAD04460866GA
SAD04460876GA
SAD06424595GA
SAD07104892GA
SAD04491037GA
SAD04491038GA
SAD04491039GA
SAD04491040GA
SAD04491041GA
SAD04491042GA
SAD04491047GA
PA2296523
SAD07205025GA
SAD07205050GA
SAD08205347GA
SAD05142076GA
SAD05142077GA
SAD05142078GA
SAD05142079GA
SAD05142081GA
ABF05210352FA
SAD05182281GA
SAD05182286GA
SAD05182312GA
SAD05182313GA
SAD05182329GA
SAD05192351GA
SAD05192352GA
SAD05483300GA
SAD05483301GA
SAD05483307GA
SAD04470987GA
SAD04470988GA
SAD04470989GA
SAD04470990GA
SAD04470991GA
SAD04470992GA
SAD04470993GA
SAD04471005GA
SAD04471006GA
SAD06123863GA
SAD06123875GA
SAD06123876GA
SAD06123879GA
SAD06123880GA
SAD07335151GA
SAD05382936GA
SAD05443089GA
SAD05443090GA
SAD05443096GA
SAD05463212GA
SAD05473234GA
SAD05473243GA
SAD05473244GA
SAD05483304GA
SAD05483305GA
SAD04511298GA
SAD04460885GA
SAD04470903GA
SAD04470904GA
SAD04470905GA
SAD04470906GA
SAD04470907GA
SAD04470908GA
SAD04470909GA
SAD04470910GA
SAD04470911GA
SAD04470912GA
SAD04470915GA
SAD04470916GA
SAD04491138GA
SAD04491168GA
SAD04491169GA
SAD04491170GA
SAD04491171GA
SAD04491172GA
SAD04491173GA
SAD04491174GA
SAD04491175GA
SAD05192368GA
SAD05192369GA
SAD05222468GA
SAD08255392GA
SAD08255395GA
SAD08255397GA
SAD08255398GA
SAD04280133GA
SAD04280171GA
SAD04280184GA
SAD04420597GA
SAD05453169GA
SAD06474696GA
SAD07104891GA
SAD05372884GA
SAD05382942GA
SAD05111958GA
SAD05222471GA
SAD05222472GA
SAD05222473GA
SAD05292623GA
SAD05292613GA
SAD05292614GA
SAD07084874GA
SAD07084875GA
SAD07104888GA
SAD07104889GA
SAD05372867GA
SAD05372868GA
SAD05372869GA
SAD05372870GA
PA2297408
PA2297409
SAD04420598GA
SAD04430616GA
SAD06334343GA
SAD05362842GA
SAD05362843GA
SAD06524765GA
SAD06524766GA
SAD07024803GA
SAD07024804GA
SAD07024811GA
SAD10465644GA
SAD06444633GA
SAD07104893GA
SAD05152173GA
SAD05152178GA
SAD05152179GA
SAD05162224GA
SAD05162225GA
SAD05172228GA
SAD05172229GA
SAD05172230GA
SAD05172231GA
SAD05172232GA
SAD05172233GA
SAD05172235GA
SAD05172236GA
SAD05172252GA
SAD05172267GA
SAD07475209GA
SAD05302648GA
SAD05362819GA
SAD05372875GA
SAD05372876GA
SAD05372877GA
SAD05372878GA
SAD05372879GA
SAD05372880GA
SAD05372881GA
SAD05372882GA
SAD05372883GA
SAD05372894GA
SAD05051509GA
SAD05051510GA
SAD05051511GA
SAD05051534GA
SAD05051535GA
SAD05051536GA
SAD05071616GA
SAD08305442GA
SAD08305443GA
SAD08305444GA
SHF12460206TA
SAD05443080GA
SAD05443081GA
SAD05443082GA
SAD05443083GA
SAD05443105GA
SAD05443106GA
PA2167824
PA2296520
SAD05031381GA
SAD05192375GA
SAD05192376GA
SAD05192377GA
SAD05192378GA
SAD05192379GA
SAD05192380GA
SAD05192381GA
SAD06454664GA
SAD05071650GA
SAD05071651GA
SAD05071652GA
SAD05071653GA
SAD05071661GA
SAD05071660GA
SAD05372885GA
SAD05372886GA
SAD05382938GA
SAD06424614GA
SAD06424615GA
SAD07395177GA
SAD05172234GA
SAD05382920GA
SAD05382921GA
SAD05382922GA
SAD05382923GA
SAD05382924GA
SAD05382925GA
SAD05382926GA
SAD05382927GA
SAD05382928GA
SAD05382929GA
SAD05382930GA
SAD05382931GA
SAD05423045GA
SAD06053625GA
SAD06053626GA
SAD06053627GA
SAD06053628GA
SAD06053629GA
SAD06053632GA
SAD06053633GA
SAD06063678GA
SAD06063687GA
SAD07175023GA
SAD06063674GA
SAD06113824GA
SAD06123831GA
SAD06143907GA
SAD06143908GA
SAD06143909GA
SAD06143910GA
SAD06143911GA
SAD06143918GA
SAD06143922GA
SAD06143923GA
SAD06143924GA
SAD06183946GA
SAD06183984GA
SAD06184035GA
SAD06194053GA
SAD08385498GA
SAD08385499GA
SAD08385500GA
SAD08385501GA
SAD08385503GA
SAD05463195GA
SAD05463196GA
SAD05463197GA
SAD05463198GA
SAD05463199GA
SAD05463200GA
SAD06424598GA
SAD04380368GA
SAD04380369GA
SAD04380370GA
SAD04380371GA
SAD04380372GA
SAD05372871GA
SAD05372872GA
SAD04440657GA
SAD04440666GA
SAD04440667GA
SAD04440668GA
SAD04440669GA
SAD04440670GA
SAD04440674GA
SAD04440701GA
SAD06444631GA
ABF04340232FA
SAD04280168GA
SAD04280187GA
SAD06474698GA
SAD07205044GA
SAD09495627GA
SAD06023544GA
SAD06023550GA
SAD06023551GA
SAD06053634GA
SAD06053635GA
SAD06053636GA
PA2305325
ABF04350237FA
ABF04350238FA
SAD05192373GA
SAD05192374GA
SAD05212408GA
SAD05212409GA
SAD05212416GA
SAD04280169GA
SAD04280170GA
SAD04280173GA
SAD04280181GA
SAD04280196GA
SAD04330198GA
SAD04330199GA
SAD07455200GA
SAD06123841GA
SAD06123842GA
SAD06023518GA
SAD06023559GA
SAD06023560GA
SAD06023561GA
SAD06023562GA
SAD06023575GA
SAD06023576GA
SAD06023577GA
SAD06023585GA
SAD05423031GA
SAD05443055GA
SAD05443058GA
SAD06194047GA
SAD06194048GA
SAD06194050GA
SAD06194051GA
SAD06194056GA
SAD06234138GA
SAD06234139GA
SAD05322701GA
SAD05322704GA
SAD06113822GA
SAD07054828GA
SAD06334336GA
SAD06334337GA
SAD06334338GA
SAD06334346GA
SAD06334352GA
SAD06344435GA
SAD06344461GA
SAD06344521GA
SAD06344522GA
SAD06344523GA
SAD06344524GA
SAD06344525GA
SAD06344530GA
SAD06344531GA
SAD04491140GA
SAD04491150GA
SAD04501219GA
SAD04501231GA
SAD04501232GA
SAD04501233GA
SAD04501234GA
SAD04501235GA
SAD04330243GA
SAD04491035GA
SAD04491036GA
SAD04491044GA
ABF05200349FA
SAD05182292GA
SAD05192347GA
SAD05192348GA
SAD06254203GA
SAD06254204GA
SAD06264246GA
SAD06264247GA
SAD06264258GA
SAD06264259GA
SAD06264260GA
SAD06264261GA
SAD06264262GA
SAD06264263GA
SAD06264264GA
SAD06264276GA
SAD06264277GA
SAD06524767GA
SAD06524768GA
ABF04360243FA
SAD04280127GA
SAD04280129GA
SAD04280131GA
SAD04280176GA
SAD04280180GA
SAD04280185GA
SAD07104886GA
SAD07435181GA
SAD05021325GA
SAD05021341GA
SAD04491008GA
SAD04491012GA
SAD04491013GA
SAD04491048GA
SAD04491049GA
SAD07475201GA
SAD08355474GA
SAD06023548GA
SAD06063679GA
SAD06063680GA
SAD06063685GA
SAD06063686GA
SAD06073717GA
SAD07205036GA
SAD06143912GA
SAD06143913GA
SAD06113823GA
SAD06113825GA
SAD06133892GA
SAD06133893GA
SAD06133894GA
SAD04470886GA
SAD05423023GA
SAD05423032GA
SAD05443084GA
SAD05443085GA
SAD05443086GA
SAD05443087GA
SAD05443088GA
SAD06524769GA
SAD06524770GA
SAD07024809GA
SAD07024810GA
SAD07335143GA
SAD07455198GA
SAD06294290GA
SAD06294294GA
SAD06294295GA
SAD06294296GA
SAD06294297GA
SAD06444641GA
SAD06454653GA
SAD08205344GA
SAD06454644GA
SAD06454645GA
SAD06454646GA
SAD06454647GA
SAD06454654GA
SAD08255406GA
SUW32506S8
SAD06053637GA
SAD06053638GA
SAD06053639GA
SAD06053640GA
SAD06053641GA
SAD06053642GA
SAD06053643GA
SAD06053644GA
SAD06053645GA
SAD06023529GA
SAD06023578GA
SAD06023580GA
SAD06023589GA
SAD06023590GA
SAD06023591GA
SAD06023592GA
SAD06023593GA
SAD06023594GA
SAD06023595GA
SAD05443110GA
SAD05443126GA
SAD06023524GA
PA2305332
PA2305333
PA2305334
PA2305335
PA2305873
SAD06073714GA
SAD06294293GA
SAD08235368GA
SAD08235369GA
SAD08235370GA
SAD08235371GA
SAD08235372GA
SAD07084862GA
SAD07124920GA
SAD07054836GA
SAD07054837GA
SAD07054842GA
SAD07084851GA
SAD07084852GA
SAD07084853GA
SAD07084854GA
SAD06424604GA
SAD05051478GA
SAD05051486GA
SAD05051488GA
SAD05051494GA
SAD05132000GA
SAD06444628GA
SAD05101915GA
SAD05101920GA
SAD05101921GA
SAD05362830GA
SAD05362865GA
SAD05071619GA
SAD05071637GA
SAD05071672GA
SAD09095566GA
SAD09095567GA
SAD05292607GA
SAD05292608GA
SAD05302637GA
SAD05302638GA
SAD05302647GA
SAD06113782GA
SAD06113783GA
SAD06113784GA
SAD06254206GA
SAD06254207GA
SAD06254208GA
SAD06254209GA
SAD06254210GA
SAD06254211GA
SAD06254212GA
SAD06254213GA
SAD07215058GA
SAD07215061GA
SAD07285072GA
SAD07285077GA
SAD07295083GA
SAD04440731GA
SAD06444619GA
SAD07385167GA
SAD05222437GA
SAD05222438GA
SAD05232519GA
SAD05232520GA
SAD05232521GA
SAD05232522GA
SAD05232523GA
SAD05232524GA
SAD05232525GA
SAD05232526GA
SAD05232527GA
SAD06514743GA
SAD08105324GA
SAD08105325GA
SAD08105326GA
SAD09335613GA
SAD08055300GA
SAD08055301GA
SAD08055302GA
SAD08055303GA
ABF04380255FA
ABF04380256FA
ABF04380257FA
ABF04380258FA
SAD04330298GA
SAD07285078GA
SAD04440711GA
SAD04440712GA
SAD04440713GA
SAD04440714GA
SAD04440715GA
SAD06494710GA
SAD07215051GA
SAD07215052GA
SAD07215054GA
SAD07215055GA
SAD07285076GA
SAD07285081GA
SAD07285082GA
SAD05222446GA
SAD05222464GA
SAD05222465GA
SAD05352804GA
SAD05352805GA
SAD06514744GA
SAD10465653GA
SAD04410536GA
SAD04410548GA
SAD04410550GA
SAD04410551GA
SAD04420571GA
SAD04420572GA
SAD04420573GA
SAD04420576GA
SAD04420579GA
SAD04420580GA
SAD06494711GA
SAD06494712GA
SAD06514753GA
SAD07024793GA
SAD07024794GA
SAD07024795GA
SAD07024798GA
SAD07024799GA
SAD07024800GA
SAD07024801GA
SAD07024802GA
SAD07084848GA
SAD07104895GA
SAD07104896GA
SAD07104897GA
SAD07104898GA
SAD07124914GA
SAD07124915GA
SAD07124923GA
SAD07124924GA
SAD07455192GA
SAD07455193GA
SAD08215355GA
SAD08235377GA
SAD08355470GA
SAD08355471GA
SAD07385169GA
SAD07395171GA
SAD05222439GA
SAD05252540GA
SAD05252541GA
SAD05252575GA
SAD05252576GA
SAD05252577GA
PA2165096
PA2165099
SAD04460877GA
SAD07144995GA
SAD07205037GA
SAD07205038GA
SAD07205043GA
SAD07205045GA
SAD07205046GA
SAD07205047GA
SAD07205048GA
SAD07205049GA
SAD06123844GA
SAD06063689GA
SAD05503407GA
SAD05513432GA
SAD05523465GA
SAD05523466GA
SAD05523467GA
SAD05523468GA
SAD05523469GA
SAD05523470GA
SAD05523471GA
SAD05523472GA
SAD05523473GA
SAD05523474GA
SAD05523475GA
SAD05523476GA
SAD05523477GA
SAD05523478GA
SAD05523479GA
SAD05523480GA
SAD05523481GA
SAD05523482GA
SAD05523483GA
SAD05523484GA
SAD05523486GA
SAD05523487GA
SAD05523488GA
SAD05523489GA
SAD05523490GA
SAD05523491GA
SAD05523492GA
SAD05523493GA
SAD05523494GA
SAD05523495GA
SAD05523496GA
SAD05523497GA
SAD05523498GA
SAD05523499GA
SAD05523517GA
SAD04440699GA
SAD04511289GA
SAD05362827GA
SAD05362828GA
SAD05372887GA
SAD05372888GA
SAD05372889GA
SAD05372890GA
SAD05372891GA
SAD05382919GA
SAD05382941GA
SAD05382944GA
SAD04430622GA
SAD04430623GA
SAD04430624GA
SAD04430625GA
SAD04430629GA
SAD04430631GA
SAD05503405GA
SAD05503406GA
SAD06023554GA
SAD06023555GA
SAD06033609GA
SAD06033610GA
SAD06033611GA
SAD06033612GA
SAD07315127GA
SAD05182278GA
SAD05182279GA
SAD05182280GA
SAD05182288GA
SAD05182304GA
SAD05182305GA
SAD05182306GA
SAD05182307GA
SAD05182308GA
SAD05182309GA
SAD05182310GA
SAD05182311GA
SAD05182314GA
SAD05182315GA
SAD05182316GA
SAD05182317GA
SAD05182321GA
SAD05192354GA
SAD05302656GA
SAD05493343GA
SAD05503361GA
SAD05503362GA
SAD05503363GA
SAD05503364GA
SAD05503365GA
SAD05503366GA
SAD05503378GA
SAD05503384GA
SAD05503385GA
SAD06424593GA
SAD04440677GA
SAD04440678GA
SAD05252563GA
SAD05252568GA
SAD05252569GA
SAD05252578GA
SAD05252579GA
PA2020949
SAD06234119GA
SAD06234120GA
SAD06514760GA
SAD05362829GA
SAD05372893GA
SAD05372895GA
SAD05372896GA
SAD05372897GA
SAD05372898GA
SAD07455197GA
SAD07475207GA
SAD07475210GA
SAD07475212GA
SAD07515216GA
SAD07515217GA
SAD07515218GA
SAD07515219GA
SAD07515220GA
SAD07515221GA
SAD07515222GA
SAD07515223GA
SAD07515224GA
SAD07515225GA
SAD07515226GA
SAD07515227GA
SAD07515228GA
SAD07515229GA
SAD07515231GA
SAD07515232GA
SAD07515233GA
SAD07515234GA
SAD07515235GA
SAD07515236GA
SAD07515237GA
SAD07515238GA
SAD07515239GA
SAD07515240GA
SAD07515241GA
SAD07515242GA
SAD07515243GA
SAD07515244GA
SAD07515245GA
SAD07515246GA
SAD07515247GA
SAD07515248GA
SAD07515249GA
SAD07515250GA
SAD07515251GA
SAD07515252GA
SAD07515253GA
SAD07515254GA
SAD07515255GA
SAD07515256GA
SAD07515257GA
SAD07515258GA
SAD07515259GA
SAD07515260GA
SAD07515261GA
SAD07515262GA
SAD07515263GA
SAD07515264GA
SAD07515265GA
SAD07515266GA
SAD07515267GA
SAD07515268GA
SAD07515269GA
SAD07515270GA
SAD07515271GA
SAD07515272GA
SAD07515273GA
SAD07515274GA
SAD07515275GA
SAD04491212GA
SAD04491213GA
SAD04501241GA
SAD05302658GA
SAD05302671GA
SAD05302672GA
SAD05302673GA
SAD05111949GA
SAD05111950GA
SAD05121969GA
SAD05121970GA
SAD05121971GA
SAD05121972GA
SAD05121973GA
SAD05121974GA
SAD05121975GA
SAD05121976GA
SAD05121977GA
SAD05142138GA
SAD05142139GA
SAD05142140GA
SAD05142141GA
SAD05142142GA
SAD05142143GA
SAD05162210GA
SAD05162211GA
SAD05162212GA
SAD05162213GA
SAD05162214GA
SAD05162215GA
SAD05162216GA
SAD05162217GA
SAD05162218GA
SAD08105322GA
SAD05292622GA
SAD09255601GA
SAD05081765GA
SAD05081764GA
SAD06194042GA
SAD06334339GA
SAD05051538GA
SAD05051539GA
SAD05051540GA
SAD05051541GA
SAD05051542GA
SAD05051543GA
SAD05051544GA
SAD05061576GA
SAD05061577GA
SAD05061578GA
SAD05061579GA
SAD05071611GA
SAD05071612GA
SAD05071613GA
SAD05071614GA
SAD05071615GA
SAD05192334GA
SAD05192335GA
SAD05192338GA
SAD05192382GA
SAD05192389GA
SAD05192390GA
SAD05192391GA
SAD05192392GA
SAD05192393GA
SAD05192394GA
SAD05192395GA
SAD05192396GA
SAD05192397GA
SAD05192398GA
SAD05192399GA
SAD05192400GA
SAD05192401GA
SAD05192402GA
SAD05192403GA
SAD05192404GA
SAD05192405GA
SAD05192406GA
SAD04380359GA
SAD04470898GA
SAD04460863GA
SAD04460874GA
SAD04460875GA
SAD04460883GA
SAD04470890GA
SAD04470891GA
SAD04470896GA
SAD04470899GA
SAD04470901GA
SAD04470902GA
SAD05523501GA
SAD05523505GA
SAD05523516GA
SAD06113797GA
SAD06113798GA
SAD06113799GA
SAD07395175GA
SAD05081762GA
SAD05081794GA
SAD05081795GA
SAD05081796GA
SAD05081804GA
SAD05081805GA
SAD05081806GA
SAD05081807GA
SAD05081808GA
SAD05081809GA
SAD05081810GA
SAD05081811GA
SAD05081812GA
SAD05081813GA
SAD05081856GA
SAD05523510GA
SAD05523511GA
SAD05523512GA
SAD05523513GA
SAD05523514GA
SAD06023522GA
SAD06023535GA
SAD06023536GA
SAD06023537GA
SAD06023538GA
SAD06023540GA
SAD06023542GA
SAD06023568GA
ABF04390264FA
SAD04390473GA
SAD04430632GA
SAD05292611GA
SAD05292612GA
SAD06023545GA
SAD06023581GA
SAD06023582GA
SAD06023583GA
SAD06053622GA
SAD06053623GA
SAD06053624GA
SAD04470984GA
SAD04470998GA
SAD04470999GA
SAD04471000GA
SAD04471001GA
SAD04471002GA
SAD04471003GA
SAD04471004GA
SAD06194036GA
SAD06194044GA
SAD06194052GA
SAD06194061GA
SAD04430617GA
SAD04430618GA
SAD04430619GA
SAD04430620GA
SAD04430621GA
SAD04430628GA
SAD04430633GA
SAD04430634GA
SAD04430635GA
SAD04430636GA
SAD04430637GA
SAD04430638GA
SAD04430639GA
SAD04430642GA
SAD04430643GA
SAD04430644GA
SAD04430645GA
SAD04430646GA
SAD04430647GA
SAD04430648GA
SAD04430649GA
SAD04430650GA
SAD04430651GA
SAD04430652GA
SAD04430653GA
SAD04430654GA
SAD04430655GA
SAD06514745GA
SAD06514752GA
SAD06514754GA
SAD06514755GA
SAD06514756GA
SAD06514757GA
SAD04420599GA
SAD04420600GA
SAD06023546GA
SAD06023547GA
SAD06053621GA
SAD06053631GA
SAD06334356GA
SAD07315128GA
SAD07315129GA
SAD07285079GA
SAD07285080GA
SAD05192366GA
SAD05192367GA
SAD05222447GA
SAD05222448GA
SAD05222449GA
SAD05222450GA
SAD05222451GA
SAD05222452GA
SAD05222453GA
SAD05222454GA
SAD05222455GA
SAD05222456GA
SAD05222457GA
SAD05222458GA
SAD05222459GA
SAD05222460GA
SAD05222461GA
SAD05222462GA
SAD05222463GA
SAD05222474GA
SAD05222485GA
SAD05302650GA
SAD05302651GA
SAD05302652GA
SAD05322713GA
SAD05322714GA
SAD05322715GA
SAD05322716GA
SAD05322717GA
SAD05322718GA
SAD05322719GA
SAD04450747GA
SAD06474697GA
SAD05483299GA
SAD08255400GA
SAD08255403GA
SAD05031378GA
SAD05031379GA
SAD05031380GA
SAD05031390GA
SAD06123864GA
SAD06123865GA
SAD06123866GA
SAD06123867GA
SAD06474671GA
SAD05252534GA
SAD05252535GA
SAD05252536GA
SAD05252537GA
SAD05252538GA
SAD05262586GA
SAD06133901GA
SAD06133904GA
SAD05322732GA
SAD05352786GA
SAD05352787GA
SAD05071721GA
SAD05071722GA
SAD05071723GA
SAD05252528GA
SAD05252529GA
PA2021953
SAD05192371GA
SAD05192372GA
SAD05212421GA
SAD05212422GA
SAD05212423GA
SAD05212424GA
SAD05212425GA
SAD04250042GA
SAD04250043GA
SAD04250054GA
SAD04260102GA
SAD04280104GA
SAD04280105GA
SAD04280106GA
SAD04280107GA
SAD04280108GA
SAD04280109GA
SAD04280110GA
SAD04280111GA
SAD04280112GA
SAD04280113GA
SAD04280114GA
SAD04280115GA
SAD04280116GA
SAD04280117GA
SAD04280118GA
SAD04280124GA
SAD04280125GA
SAD04330201GA
SAD06143925GA
SAD06143926GA
SAD05051492GA
SAD05051493GA
SAD07515280GA
SAD07515281GA
SAD07515282GA
SAD07515283GA
SAD07515284GA
SAD06334355GA
SAD04460836GA
SAD06444627GA
SAD06454663GA
SAD04330215GA
SAD07124926GA
SAD07124934GA
SAD08515557GA
SAD04250046GA
SAD04250076GA
SAD04250078GA
SAD04250079GA
SAD04250080GA
SAD04250081GA
SAD04250082GA
SAD04250083GA
SAD04250084GA
SAD05031421GA
ABF04460279FA
SAD07054839GA
SAD07084849GA
SAD07175014GA
SAD04330273GA
SAD04370344GA
SAD04370345GA
SAD04370346GA
SAD04370349GA
SAD04370350GA
SAD07084876GA
SAD07084877GA
SAD07084878GA
SAD09255606GA
ABF04130165FA
SAD05453167GA
SAD05453168GA
SAD05463181GA
SAD05463186GA
SAD05463187GA
SAD05252557GA
SAD05252558GA
SAD05262585GA
SAD05292631GA
SAD06033617GA
SAD07124908GA
SAD07124952GA
SAD07144966GA
SAD07144968GA
SAD07144969GA
SAD07144972GA
SAD07144973GA
SAD07144974GA
ABF04480283FA
SAD04470951GA
SAD04491055GA
SAD04491056GA
SAD10465649GA
SAD05523457GA
SAD06023584GA
SAD06073710GA
SAD06073711GA
SAD06073712GA
SAD06073713GA
SAD06073724GA
SAD05503408GA
SAD05503409GA
SAD05503410GA
SAD05513434GA
SAD09085564GA
SAD09085565GA
ABF05090321FA
SAD06454655GA
SAD06454665GA
SAD06454666GA
SAD06524774GA
SAD07054822GA
SAD07124928GA
SAD04330213GA
SAD07205029GA
ABF05190347FA
ABF05190348FA
SAD05132001GA
SAD05132002GA
SAD05152169GA
SAD05152170GA
SAD05152171GA
SAD05152172GA
SAD05162222GA
SAD05162223GA
SAD05172247GA
SAD05172248GA
SAD05302643GA
SAD05302644GA
SAD05302653GA
SAD05302657GA
ABF05160345FA
ABF05160346FA
SAD05232502GA
SAD05232503GA
SAD06023521GA
SAD06023557GA
SAD06023558GA
SAD06494706GA
SAD04380393GA
SAD04380394GA
SAD04380395GA
SAD04380400GA
SAD06334344GA
SAD06334345GA
SAD06344358GA
SAD06344359GA
SAD06344360GA
SAD06344361GA
SAD06344362GA
SAD06344363GA
SAD06344364GA
SAD06344365GA
SAD06344366GA
SAD06344367GA
SAD06344368GA
SAD06344369GA
SAD06344370GA
SAD06344371GA
SAD06344372GA
SAD06344373GA
SAD06344374GA
SAD06344375GA
SAD06344376GA
SAD06344377GA
SAD06344378GA
SAD06344379GA
SAD06344380GA
SAD06344381GA
SAD06344382GA
SAD06344383GA
SAD06344384GA
SAD06344385GA
SAD06344386GA
SAD06344387GA
SAD06344388GA
SAD06344389GA
SAD06344390GA
SAD06344391GA
SAD06344392GA
SAD06344393GA
SAD06344394GA
SAD06344395GA
SAD06344396GA
SAD06344397GA
SAD06344398GA
SAD06344399GA
SAD06344400GA
SAD06344401GA
SAD06344402GA
SAD06344403GA
SAD06344404GA
SAD06344405GA
SAD06344406GA
SAD06344407GA
SAD06344408GA
SAD06344409GA
SAD06344410GA
SAD06344411GA
SAD06344412GA
SAD06344413GA
SAD06344414GA
SAD06344415GA
SAD06344416GA
SAD06344417GA
SAD06344418GA
SAD06344419GA
SAD06344420GA
SAD06344421GA
SAD06344422GA
SAD06344423GA
SAD06344424GA
SAD06344425GA
SAD06344426GA
SAD06344427GA
SAD06344428GA
SAD06344429GA
SAD06344430GA
SAD06344431GA
SAD06344432GA
SAD06344433GA
SAD06344434GA
SAD06344436GA
SAD06344437GA
SAD06344438GA
SAD06344439GA
SAD06344440GA
SAD06344441GA
SAD06344442GA
SAD06344443GA
SAD06344444GA
SAD06344445GA
SAD06344446GA
SAD06344447GA
SAD06344448GA
SAD06344449GA
SAD06344450GA
SAD06344451GA
SAD06344452GA
SAD06344453GA
SAD06344454GA
SAD06344455GA
SAD06344456GA
SAD06344457GA
SAD06344458GA
SAD06344459GA
SAD06344460GA
SAD06344462GA
SAD06344463GA
SAD06344464GA
SAD06344465GA
SAD06344466GA
SAD06344467GA
SAD06344468GA
SAD06344469GA
SAD06344470GA
SAD06344471GA
SAD06344472GA
SAD06344473GA
SAD06344474GA
SAD06344475GA
SAD06344476GA
SAD06344477GA
SAD06344478GA
SAD06344479GA
SAD06344480GA
SAD06344481GA
SAD06344482GA
SAD06344483GA
SAD06344484GA
SAD06344485GA
SAD06344486GA
SAD06344487GA
SAD06344488GA
SAD06344489GA
SAD06344490GA
SAD06344491GA
SAD06344492GA
SAD06344493GA
SAD06344494GA
SAD06344495GA
SAD06344496GA
SAD06344497GA
SAD06344498GA
SAD06344499GA
SAD06344500GA
SAD06344501GA
SAD06344502GA
SAD06344503GA
SAD06344504GA
SAD06344505GA
SAD06344506GA
SAD06344507GA
SAD06344508GA
SAD06344509GA
SAD06344510GA
SAD06344511GA
SAD06344512GA
SAD06344513GA
SAD04380380GA
SAD04380381GA
SAD04380382GA
SAD04380383GA
SAD04380384GA
SAD04380385GA
SAD04380415GA
SAD04380418GA
SAD04380419GA
SAD04380420GA
SAD04380421GA
SAD04380422GA
SAD04380423GA
SAD04380424GA
SAD04380425GA
SAD04380426GA
SAD04390440GA
SAD04390441GA
SAD04390442GA
SAD04390450GA
SAD04390451GA
SAD04390453GA
SAD06324312GA
SAD05493346GA
SAD05493347GA
SAD05493348GA
SAD05493360GA
SAD08285437GA
SAD05192333GA
SAD05222443GA
SAD06334342GA
ABF04380263FA
SAD05172249GA
SAD05172250GA
SAD05172251GA
SAD05252572GA
SAD05252573GA
SAD05252574GA
SAD05392945GA
SAD04380412GA
SAD05473250GA
SAD05473251GA
SAD05483278GA
SAD07175022GA
SAD05453158GA
SAD05453159GA
SAD05453165GA
SAD05453166GA
SAD05463201GA
SAD05463202GA
SAD05463225GA
SAD05182319GA
SAD05182320GA
PA2295556
PA2296339
SAD06454651GA
SAD06524771GA
SAD07375162GA
SAD08105321GA
SAD08095311GA
SAD04430611GA
SAD04430612GA
SAD04430613GA
SAD06474689GA
SAD06474690GA
SAD06474691GA
SAD06474692GA
SAD06474695GA
SAD06113810GA
SAD06113811GA
SAD06113812GA
SAD06113813GA
SAD06113814GA
SAD06113815GA
SAD06113816GA
SAD06123832GA
SAD06123833GA
SAD06123834GA
SAD06123835GA
SAD06123836GA
SAD06123837GA
SAD06123856GA
SAD06023556GA
SAD04380351GA
SAD06344527GA
SAD06344537GA
SAD05352795GA
SAD05352801GA
SAD07205028GA
SAD05322726GA
SAD08515555GA
SAD09095571GA
SAD05182296GA
SAD08205334GA
SAD08285436GA
SAD08285439GA
SAD05392949GA
SAD05392983GA
SAD05392984GA
SAD05392985GA
SAD06234155GA
SAD07215066GA
SAD05192359GA
SAD05192384GA
SAD05403007GA
SAD05382937GA
SAD05382939GA
SAD05392963GA
SAD05392966GA
SAD05403006GA
SAD08105323GA
SAD05423034GA
SAD05423035GA
SAD05483259GA
SAD05483261GA
SAD05483262GA
SAD05483272GA
SAD05483273GA
SAD05473246GA
SAD05523504GA
SAD07215056GA
PA2021187
PA2021188
SAD05322693GA
SAD05483295GA
SAD05493324GA
SAD05493325GA
SAD05493326GA
SAD04390510GA
SAD07395176GA
SAD07435189GA
SAD07435190GA
SAD04390511GA
SAD05493327GA
SAD04420601GA
SAD04420602GA
SAD04420596GA
SAD04430641GA
SAD05302660GA
SAD05322690GA
SAD05322691GA
SAD05322692GA
SAD04380364GA
SAD04380365GA
SAD04380366GA
SAD04380367GA
SAD10465643GA
SAD06234140GA
SAD06234141GA
SAD06234142GA
SAD06234143GA
SAD06234144GA
SAD06234145GA
SAD06234146GA
SAD06234147GA
SAD06234148GA
SAD06234149GA
SAD06234150GA
SAD06234151GA
SAD06234152GA
SAD06234153GA
SAD06234154GA
SAD06234156GA
SAD06234157GA
SAD06234158GA
SAD06234159GA
SAD06234160GA
SAD06234161GA
SAD06234163GA
SAD06244166GA
SAD06244167GA
SAD06244168GA
SAD06244169GA
SAD06244170GA
SAD06244171GA
SAD06244172GA
SAD06244173GA
SAD06244174GA
SAD06244175GA
SAD06244176GA
SAD06244177GA
SAD06244178GA
SAD06244179GA
SAD06244180GA
SAD06244181GA
SAD06244182GA
SAD06244183GA
SAD06244184GA
SAD06244185GA
SAD06244186GA
SAD06244187GA
SAD06244188GA
SAD06244189GA
SAD06244190GA
SAD06244191GA
SAD06254193GA
SAD06254194GA
SAD06254195GA
SAD06254214GA
SAD05252531GA
SAD05252532GA
SAD05262591GA
SAD05262592GA
SAD05262587GA
SAD05262588GA
SAD05262589GA
SAD05262590GA
SAD07215053GA
SAD06264257GA
SAD06264268GA
SAD06264269GA
SAD06264272GA
SAD06264273GA
SAD06264274GA
SAD06264275GA
SAD06394540GA
SAD06394541GA
SAD06394542GA
SAD06394543GA
SAD06394544GA
SAD06394560GA
SAD06394561GA
R1H04390746FA
SAD04420563GA
SAD04420564GA
SAD04420565GA
SAD04420566GA
SAD04420567GA
SAD04420568GA
SAD04420569GA
SAD04420570GA
SAD04440672GA
SAD05081845GA
SAD05091894GA
SAD05091901GA
SAD05091902GA
SAD05091903GA
SAD05091904GA
SAD05423041GA
SAD05423042GA
SAD05423044GA
SAD05212430GA
SAD05212431GA
SAD06194037GA
SAD06194046GA
SAD06073734GA
SAD06073741GA
SAD06093756GA
SAD06093757GA
SAD06093758GA
SAD06093759GA
SAD06113826GA
SAD06113827GA
SAD05403013GA
SAD05403014GA
SAD05403015GA
SAD05403016GA
ABF04510299FA
SAD04420605GA
SAD04430608GA
SAD07305106GA
SAD05292615GA
SAD08215365GA
SAD06474683GA
SAD06474685GA
SAD07215057GA
PA2020326
SAD06494700GA
SAD05192356GA
SAD05192360GA
SAD05192361GA
SAD05192362GA
SAD05192363GA
SAD05031406GA
SAD05031407GA
SAD05031409GA
SAD05523485GA
SAD06023539GA
SAD06023541GA
SAD09495628GA
SAD09495629GA
ABF05030311FA
SAD05021338GA
SAD05021339GA
SAD07054838GA
SAD07084843GA
SAD07084844GA
SAD07084845GA
SAD07084846GA
PA2021952
SAD10245635GA
SAD10245636GA
SAD05182326GA
SAD05182327GA
SAD05192353GA
SAD08285438GA
SAD06234164GA
SAD08355475GA
SAD08385493GA
PA2426938
PA2426939
SAD08055308GA
SAD08105327GA
ABF04420270FA
ABF04420271FA
SAD07315130GA
SAD07315131GA
SAD07315132GA
SAD10465652GA
SAD08255408GA
SAD08255413GA
SAD08255416GA
SAD08255417GA
SAD08255418GA
SAD08255419GA
SAD08255420GA
SAD08255421GA
SAD08255422GA
SAD08255423GA
SAD08255424GA
SAD08255425GA
SAD08285426GA
SAD08285428GA
SAD08285431GA
SAD08285433GA
SAD08215354GA
SAD08215356GA
SAD08215357GA
SAD08215358GA
SAD08215359GA
SAD08215361GA
SAD08215362GA
SAD08215363GA
SAD08215364GA
SAD08235376GA
SAD07054817GA
SAD07054820GA
SAD07054821GA
SAD07054823GA
SAD07054824GA
SAD07054835GA
SAD07054825GA
SAD05362844GA
SAD06023549GA
ABF05150339FA
ABF05150340FA
ABF05150341FA
ABF05150342FA
SAD05132003GA
SAD05132004GA
SAD05132005GA
SAD05142080GA
SAD05172270GA
SAD05392972GA
SAD05392973GA
ABF04440275FA
SAD04491014GA
SAD04491015GA
SAD04491016GA
SAD05081819GA
SAD05091891GA
SAD05091892GA
ABF04470281FA
ABF05020308FA
ABF05020309FA
SAD04511295GA
ABF04350239FA
ABF04350240FA
SAD04330219GA
SAD07215067GA
SAD06444621GA
SAD06454648GA
SAD05423024GA
SAD05292616GA
SAD05292617GA
SAD05292618GA
SAD05292625GA
SAD05292626GA
SAD05292627GA
SAD05292628GA
SAD05292629GA
SAD07124918GA
SAD05101910GA
SAD05101911GA
SAD05101912GA
SAD05101913GA
SAD05101914GA
SAD05101919GA
SAD05111962GA
SAD05142071GA
SAD07104894GA
SAD04450771GA
SAD04450772GA
SAD04470947GA
SAD06394587GA
SAD06454649GA
SAD06514739GA
SAD06514740GA
SAD07205041GA
SAD07134953GA
SAD07134954GA
SAD07134955GA
SAD07134956GA
SAD07134957GA
SAD07134958GA
SAD07134959GA
SAD07134960GA
SAD07134961GA
SAD07144980GA
SAD07144981GA
SAD08105317GA
SAD08385502GA
SAD05081848GA
SAD05101922GA
SAD05081736GA
SAD05081788GA
SAD05081789GA
SAD05081790GA
SAD05081791GA
SAD05081792GA
SAD05081793GA
SAD05081802GA
SAD05081820GA
SAD05081821GA
SAD05081822GA
SAD05081823GA
SAD05081824GA
SAD05081825GA
SAD05081826GA
SAD05081827GA
SAD05081828GA
SAD05081829GA
SAD05081830GA
SAD05081831GA
SAD06063673GA
SAD06063696GA
SAD06063697GA
SAD06063698GA
SAD06073743GA
SAD06474670GA
SAD06474672GA
SAD06474673GA
SAD06474674GA
SAD06474675GA
SAD06474676GA
SAD06474677GA
SAD06474678GA
SAD06474684GA
SAD05463205GA
SAD05463206GA
SAD05463207GA
SAD05463209GA
SAD05463210GA
SAD05463217GA
SAD05463218GA
SAD05463219GA
SAD05463220GA
SAD05463221GA
SAD05463223GA
SAD05463224GA
SAD05453153GA
SAD05453154GA
SAD05453155GA
SAD05453156GA
SAD05453157GA
SAD05463190GA
SAD05463203GA
SAD05463204GA
SAD05463227GA
SAD05463228GA
SAD05463229GA
SAD05463230GA
SAD06073735GA
SAD06344538GA
SAD07315123GA
SAD07335144GA
SAD07375161GA
SAD06474679GA
SAD06444636GA
SAD07124917GA
SAD07124942GA
SAD07124943GA
SAD07124944GA
SAD07124945GA
SAD07124946GA
SAD07124947GA
SAD07124948GA
SAD07124949GA
SAD07134962GA
SAD07134963GA
SBP06511585GA
SBP06511590GA
SBP06511591GA
SBP06511592GA
SBP06511593GA
SBP06511594GA
SBP06511595GA
SBP06511596GA
SBP06511597GA
SBP06521633GA
SBP06521634GA
SBP06521643GA
SAD07084879GA
SAD05322708GA
SAD05322733GA
SAD05322734GA
SAD05332736GA
SAD05332737GA
SAD05332738GA
SAD05332739GA
SAD05332747GA
SAD05332748GA
SAD05332749GA
SAD08515548GA
SAD06324313GA
SAD05463208GA
SAD06324314GA
SAD06324315GA
SAD06324316GA
SAD06324324GA
SAD06324330GA
SAD04280191GA
SAD04330208GA
SAD05222442GA
SAD05503387GA
SAD05503388GA
SAD05503389GA
SAD05513431GA
SAD05523450GA
SAD05523451GA
SAD10465648GA
PA2305600
PA1343554
SAD05463215GA
SAD05473245GA
SAD05483257GA
SAD05483258GA
ABF05210351FA
SAD05182294GA
SAD05322703GA
PA2021723
SAD06284286GA
SAD06284287GA
SAD05463213GA
SAD05463214GA
SAD05483266GA
SAD05483267GA
SAD05483268GA
SAD05081861GA
SAD05081862GA
SAD05091877GA
SAD06063688GA
SAD06424599GA
SAD05463184GA
SAD05463185GA
SAD05463211GA
SAD05463222GA
PA2297998
SAD05152161GA
SAD05152162GA
SAD05152163GA
SAD05152164GA
SAD05152165GA
SAD05152166GA
SAD05152176GA
SAD05162219GA
ABF04510298FA
SAD04460824GA
SAD04460827GA
SAD05392968GA
SAD05392969GA
SAD05392970GA
SAD05322688GA
SAD05322694GA
SAD05031399GA
SAD05031417GA
SAD05041444GA
SAD05041445GA
SAD05423036GA
SAD05423037GA
SAD07385170GA
SAD10465655GA
SAD10245637GA
SAD05483283GA
SAD05483296GA
SAD05483297GA
SAD05483303GA
SAD05493323GA
SAD05493329GA
SAD05493330GA
SAD05493331GA
SAD05493332GA
SAD05493333GA
SAD05493334GA
SAD06444637GA
SAD06474688GA
SAD07315124GA
SAD07315125GA
SAD07315126GA
SAD05483274GA
SAD06113817GA
SAD05463182GA
SAD05463183GA
SAD05463226GA
SAD04501253GA
SAD04501254GA
SAD04501255GA
SAD05453143GA
SAD05443076GA
SAD05443075GA
SAD05443077GA
SAD05443092GA
SAD05443093GA
SAD05443094GA
SAD05443095GA
SAD05443123GA
SAD05443124GA
SAD05443125GA
SAD06454657GA
SAD07455191GA
SAD07515285GA
SAD04390492GA
SAD04390493GA
SAD04390494GA
SAD04390520GA
SAD04390521GA
SAD04390524GA
SAD04390525GA
SAD04380404GA
SAD04380405GA
SAD04390448GA
SAD04390465GA
SAD04390476GA
SAD06454658GA
SAD05192343GA
SAD05423030GA
SAD06023552GA
SAD06334332GA
SAD06394569GA
ABF05110331FA
SAD04501248GA
SAD05081852GA
SAD05081853GA
SAD05091884GA
SAD05111936GA
SAD06123858GA
SAD06123859GA
SAD06123860GA
SAD06123861GA
SAD06123862GA
SAD06093774GA
SAD05142123GA
SAD04390445GA
SAD04390446GA
SAD04390452GA
SAD04390466GA
SAD05382933GA
SAD05392950GA
SAD05392951GA
SAD05392952GA
SAD05392953GA
SAD05392954GA
SAD05392955GA
SAD05393955GA
SAD07124919GA
SAD05081832GA
SAD05081833GA
SAD05081849GA
SAD05081850GA
SAD05091883GA
SAD05091890GA
SAD06073748GA
SAD05222475GA
SAD05222476GA
SAD05222477GA
SAD05222478GA
SAD05222479GA
SAD05222480GA
SAD05111937GA
SAD05111938GA
SAD05111939GA
SAD05111940GA
SAD05111941GA
SAD05111942GA
SAD05111943GA
SAD05111944GA
SAD05111945GA
SAD05111946GA
SAD05111948GA
SAD05111951GA
SAD05111952GA
SAD05121981GA
SAD05121982GA
SAD05121983GA
SAD05121984GA
SAD05121985GA
SAD05121986GA
SAD05121987GA
SAD05121988GA
SAD05142120GA
SAD05142121GA
SAD05252543GA
SAD05252544GA
SAD05252545GA
SAD05252546GA
SAD05252547GA
SAD05252548GA
PA2305336
PA2305337
PA2305338
PA2305339
PA2305340
SAD06093772GA
PA2021452
PA2021724
PA2020988
PA2020989
PA2020990
PA2021189
PA2021190
SAD06394576GA
ABF05150343FA
SAD05142073GA
SAD05142089GA
SAD05142122GA
SAD06264232GA
SAD09495632GA
SAD05302681GA
SAD05302683GA
SAD05322697GA
SAD05322698GA
SAD05322699GA
SAD05322700GA
SAD06113803GA
SAD06113804GA
SAD06113805GA
SAD06113806GA
SAD06113807GA
SAD06113808GA
SAD06113809GA
SAD06123869GA
SAD08285427GA
SAD06204069GA
SAD06204070GA
SAD06254205GA
SAD06264222GA
SAD06264223GA
SAD06264224GA
SAD06264225GA
SAD06264226GA
SAD06264227GA
SAD06264228GA
SAD06264229GA
SAD06264230GA
SAD06264231GA
SAD06264233GA
SAD06264234GA
SAD06264235GA
SAD06264236GA
SAD06264237GA
SAD06264238GA
SAD06264239GA
SAD06264240GA
SAD06264241GA
SAD06264242GA
SAD06264248GA
SAD06264249GA
SAD06264250GA
SAD06264251GA
SAD06264252GA
SAD06264253GA
SAD06264254GA
SAD06264255GA
SAD06264256GA
SAD07295084GA
SAD07295085GA
SAD07295086GA
SAD06234135GA
SAD06234136GA
SAD05172226GA
SAD05172227GA
SAD05182298GA
SAD05182301GA
SAD05182303GA
SAD06093775GA
SAD06093776GA
ABF05040313FA
PA2425459
PA2425460
SAD05362831GA
SAD05362833GA
SAD05362834GA
SAD05362835GA
SAD05362855GA
SAD05362856GA
SAD05362857GA
SAD05362858GA
SAD05362859GA
SAD05362863GA
SAD05362864GA
SAD06344532GA
SAD06394566GA
SAD06394567GA
SAD06394574GA
SAD06424609GA
SAD06424616GA
SAD07325142GA
SAD09085560GA
SAD09085561GA
SAD09095573GA
SAD09095574GA
SAD09095575GA
SAD06444632GA
SAD06073730GA
SAD06073745GA
SAD06073746GA
SAD06073747GA
SAD06073750GA
SAD06093771GA
SAD08205349GA
SAD08205350GA
SAD08205352GA
SAD06394555GA
SAD06394556GA
SAD06394557GA
SAD06394570GA
SAD06394571GA
SAD06394572GA
SAD06394573GA
SAD06394575GA
SAD06394577GA
SAD06394578GA
SAD06394579GA
SAD06394586GA
SAD06424608GA
ABF05250356FA
SAD05362821GA
SAD05362847GA
SAD05362848GA
SAD05362849GA
SAD06234118GA
PA2021950
SAD05182318GA
SAD05192336GA
SAD05192337GA
SAD06494705GA
SAD04380406GA
SAD04380407GA
SAD04380408GA
SAD04380409GA
SAD04380410GA
SAD06093773GA
SAD07335147GA
SAD06394558GA
SAD06394563GA
SAD06394568GA
SAD06394588GA
SAD06394589GA
SAD06394590GA
SAD06394591GA
SAD06514741GA
SAD06514742GA
SAD05392959GA
PA2305326
PA2305327
SAD05081841GA
SAD05101926GA
SAD05101935GA
SAD06394582GA
SAD06444620GA
SAD07054829GA
SAD07084850GA
SAD07084864GA
SAD05142091GA
SAD05142092GA
SAD05212412GA
SAD05232509GA
SAD05252530GA
SAD04501259GA
SAD04501260GA
SAD09255602GA
SAD09255609GA
SAD09255610GA
ABF05040317FA
SAD05352802GA
SAD07084868GA
SAD04470955GA
SAD04491034GA
SAD04491139GA
SAD05081797GA
SAD05081798GA
SAD05081799GA
SAD05081842GA
SAD06454667GA
SAD06514713GA
SAD06514723GA
SAD08205336GA
SAD08205341GA
SAD05392986GA
SAD05403017GA
SAD05403018GA
SAD05403019GA
SAD05403020GA
SAD05403021GA
SAD05403022GA
SAD05423025GA
SAD05423026GA
SAD05423027GA
SAD05423028GA
SAD05423046GA
SAD05423047GA
SAD05423048GA
SAD05463179GA
SAD06444635GA
SAD06514738GA
SAD07024796GA
SAD09185595GA
SAD09185596GA
SAD09185597GA
SAD09185598GA
SAD09185599GA
SAD05362818GA
SAD05362820GA
SAD05372901GA
SAD05372902GA
SAD05372903GA
SAD05372904GA
SAD05372905GA
SAD05372906GA
SAD05372907GA
SAD05372909GA
SAD05372910GA
SAD05372911GA
SAD05372912GA
SAD05372913GA
SAD05372914GA
SAD05372915GA
SAD05372916GA
SAD05372917GA
SAD04390454GA
SAD04390461GA
SAD04390462GA
SAD04390467GA
SAD04491111GA
SAD06073721GA
SAD09335615GA
SAD06494707GA
SAD06494709GA
ABF05090320FA
SAD05071634GA
SAD09495624GA
SAD06133895GA
SAD06133896GA
SAD06143914GA
SAD06143919GA
SAD06143920GA
SAD05152159GA
SAD05152160GA
SAD05162200GA
SAD05162201GA
SAD05162202GA
SAD05162203GA
SAD05162204GA
SAD05162205GA
SAD05162206GA
SAD05162208GA
SAD05162209GA
SAD05172260GA
SAD05162207GA
PA2021451
SAD05222481GA
SAD05222482GA
SAD05222483GA
SAD05222484GA
SAD06324306GA
SAD06324307GA
SAD06324308GA
SAD06324309GA
SAD06324310GA
SAD06324318GA
SAD06324320GA
SAD06324321GA
SAD06324322GA
SAD06324323GA
SAD06324325GA
SAD06324326GA
SAD06324327GA
SAD06324331GA
SAD08235373GA
SAD04250044GA
SAD04250045GA
SAD04250047GA
SAD04250048GA
SAD04250049GA
SAD04250050GA
SAD04250051GA
SAD04250052GA
SAD04250053GA
SAD04280132GA
SAD06524780GA
SAD06524781GA
SAD07054818GA
SAD07054819GA
SAD07054832GA
SAD07054833GA
SAD07054834GA
SAD07054840GA
SAD07054841GA
SAD05172258GA
SAD05172259GA
SAD05212419GA
SAD05212420GA
SAD08355473GA
SAD06514759GA
SAD06344533GA
SAD06344534GA
SAD06344535GA
SAD06344536GA
SAD06394554GA
SAD06394564GA
SAD06394565GA
SAD07515277GA
SAD07515278GA
SAD07515279GA
SAD04380352GA
SAD04380373GA
SAD04380387GA
SAD04380389GA
SAD04380390GA
SAD04380391GA
SAD04380392GA
SAD04380396GA
SAD04380397GA
SAD04380398GA
SAD05142148GA
SAD05142149GA
SAD05142150GA
SAD05142151GA
SAD05142152GA
SAD05142156GA
SAD05142157GA
SAD05142158GA
SAD05332755GA
SAD05332756GA
SAD05352773GA
SAD06073719GA
SAD05322712GA
SAD05352785GA
SAD04390504GA
SAD04390505GA
SAD04390506GA
SAD04390507GA
SAD04390514GA
SAD04390515GA
SAD04390516GA
SAD04390517GA
SAD04390518GA
SAD04390519GA
SAD06023523GA
SAD06023525GA
SAD06023526GA
SAD06023543GA
SAD06023569GA
SAD06023586GA
SAD06023587GA
SAD06023588GA
SAD06033596GA
SAD06033618GA
SAD06033619GA
SAD06033620GA
SBR08351993GA
SAD04491179GA
SAD04491180GA
SAD07104887GA
SAD07124925GA
SAD05322695GA
SAD07175019GA
SAD07144965GA
SAD07144967GA
SAD07144996GA
SAD07144997GA
SAD07144998GA
SAD07174999GA
SAD07175000GA
SAD07175001GA
SAD07175002GA
SAD07175003GA
SAD07175004GA
SAD07175005GA
SAD07175006GA
SAD07175007GA
SAD07175008GA
SAD07175009GA
SAD07175010GA
SAD07175011GA
SAD07175012GA
SAD07175013GA
SAD07175024GA
SBT08254777GA
SAD07205035GA
SAD08385495GA
SAD08385496GA
SAD08385497GA
SAD07054830GA
SAD07054831GA
SAD07084855GA
SAD07084856GA
SAD07084857GA
SAD07084858GA
SAD07084859GA
SAD07084860GA
SAD07084861GA
SAD07084880GA
SAD07084881GA
SAD07084882GA
SAD07084883GA
SAD07084884GA
SAD04460854GA
SAD04460855GA
SAD04460856GA
SAD04460857GA
SAD04460859GA
SAD04460860GA
ABF05030312FA
SAD06234116GA
SAD06454660GA
ABF04340236FA
SAD04280175GA
SAD06204080GA
ABF04340235FA
SAD06073701GA
SAD06073708GA
SAD06073709GA
SAD06073722GA
SAD06204079GA
SAD08385492GA
SAD08385494GA
SAD05443073GA
SAD05493320GA
SAD05493321GA
SAD05493322GA
SAD05493344GA
SAD05493356GA
SAD05503367GA
SAD05503368GA
SAD05503369GA
SAD05503370GA
SAD05503371GA
SAD05503372GA
SAD05503373GA
SAD05503374GA
SAD05503375GA
SAD05503376GA
SAD05503377GA
SAD05503379GA
SAD05503380GA
SAD05503386GA
SAD05503411GA
SAD05503412GA
SAD05513413GA
SAD05513414GA
SAD05513415GA
SAD05513416GA
SAD05513417GA
SAD05513418GA
SAD05513419GA
SAD05513420GA
SAD05513421GA
SAD05513422GA
SAD05513423GA
SAD05513424GA
SAD05513425GA
SAD05513426GA
SAD05513427GA
SAD05513428GA
SAD06073720GA
SAD04450759GA
SAD04460790GA
SAD04460791GA
SAD04460792GA
SAD06063695GA
SAD07435185GA
SAD07435186GA
SAD07435187GA
SAD07435188GA
SAD07475205GA
SAD09495630GA
SAD04390456GA
SAD04390480GA
SAD04390487GA
SAD04390488GA
SAD04390489GA
SAD04390490GA
SAD04390491GA
SAD04390502GA
SAD04390503GA
SAD06033602GA
PA2165097
PA2165098
SAD06123857GA
SAD06133881GA
SAD06133882GA
SAD06133883GA
SAD06133900GA
SAD06133902GA
SAD06133903GA
SAD06133905GA
SAD06133906GA
SAD06234129GA
SAD05352766GA
SAD05352767GA
SAD05352811GA
SAD05362814GA
SAD05362815GA
SAD05362838GA
SAD05362839GA
SAD05362840GA
SAD06113820GA
SAD06033603GA
SAD06033604GA
SAD06033605GA
SAD06033606GA
SAD06033607GA
SAD06033608GA
SAD06033614GA
SAD06033615GA
SAD06033616GA
SAD05081840GA
SAD05081844GA
SAD05101924GA
SAD05101925GA
SAD05132027GA
SAD05132028GA
SAD05132029GA
SAD06334333GA
SAD09095568GA
SAD08515551GA
SAD08515552GA
SAD08305446GA
SAD04440722GA
SAD04450749GA
SAD04450750GA
SAD04450751GA
SAD04450752GA
SAD04450753GA
SAD04450754GA
SAD06424596GA
SAD06424597GA
SAD06444624GA
SAD06494703GA
SAD06334347GA
SAD06334357GA
SAD07205034GA
PA2021951
SAD04450742GA
ABF04420267FA
ABF04420268FA
ABF04420269FA
SAD04410530GA
SAD06394562GA
SAD05302634GA
SAD06113821GA
SBP06160272GA
SAD09095569GA
SAD09095570GA
SAD04230005GA
SAD04230007GA
SAD04230008GA
SAD04230010GA
SAD04240011GA
SAD04240012GA
SAD04240015GA
SAD04240016GA
SAD04240017GA
SAD04240018GA
SAD04240019GA
SAD04240041GA
SAD04250067GA
SAD04250068GA
SAD04250069GA
SAD04250070GA
SAD04250071GA
SAD04250072GA
SAD04250073GA
SAD04250074GA
SAD04250075GA
SAD04250077GA
SAD05352790GA
SAD05352813GA
SAD04390432GA
SAD04390433GA
SAD04390436GA
SAD04390437GA
SAD04390438GA
SAD04390439GA
SAD04390477GA
SAD04390478GA
SAD04390479GA
SAD04460882GA
SAD04470892GA
SAD04470893GA
SAD04470894GA
SAD04470895GA
SAD04470897GA
SAD10245640GA
SAD04450764GA
SAD04450765GA
SAD04450766GA
SAD04450767GA
SAD04460865GA
SAD04470888GA
SAD04470900GA
SAD04470913GA
SAD04470914GA
SAD06234162GA
ABF05140332FA
ABF05140333FA
ABF05140334FA
ABF05140335FA
ABF05140336FA
ABF05140337FA
SAD05132006GA
SAD05132007GA
SAD05132008GA
SAD05132009GA
SAD05132010GA
SAD05132011GA
SAD05132012GA
SAD05132013GA
SAD05132014GA
SAD05132023GA
SAD05132024GA
SAD05132025GA
SAD05132026GA
SAD05142082GA
SAD05142083GA
SAD05142084GA
SAD05142085GA
SAD05142094GA
SAD06444634GA
ABF05040315FA
SAD07395180GA
SAD07395179GA
SAD06334335GA
PA2305436
PA2305437
SAD05262582GA
SAD05262583GA
SAD05392957GA
SAD05392958GA
SAD06073732GA
SAD06073733GA
SAD08515554GA
SAD06514735GA
SAD06514736GA
SAD07024791GA
ABF05040316FA
SAD05091909GA
SAD06474680GA
SAD07054826GA
SAD07054827GA
SAD07175015GA
SAD06204065GA
SAD06204066GA
SAD06204067GA
SAD06204068GA
SAD06214088GA
SAD06214089GA
SAD06214090GA
SAD06214091GA
SAD06214092GA
SAD06214096GA
SAD06214097GA
SAD06214098GA
SAD06214099GA
SAD06214100GA
SAD06214101GA
SAD06214102GA
SAD06214103GA
SAD06214104GA
SAD06214105GA
SAD06214106GA
SAD06214107GA
SAD06214108GA
SAD06214109GA
SAD06214110GA
SAD06214111GA
SAD06214112GA
SAD06214113GA
SAD06234130GA
SAD05252539GA
SAD05382934GA
SAD05382935GA
SAD06454661GA
SAD06424594GA
SAD05172276GA
SAD05182282GA
SAD05182283GA
SAD05182285GA
SAD05182328GA
SAD05192332GA
SAD05192344GA
SAD05192345GA
SAD05192407GA
ABF05090322FA
PA2297995
SAD06334334GA
PA2305342
PA2305343
PA2305344
SAD07024806GA
SAD07024807GA
SAD07024808GA
SAD07024813GA
SAD07024814GA
SAD07024815GA
SAD07024816GA
SAD05021320GA
SAD05021321GA
SAD05021322GA
SAD05021323GA
SAD05021331GA
SAD05021332GA
SAD05031385GA
SAD05031386GA
SAD05322728GA
SAD05322729GA
SAD05332757GA
SAD05352781GA
SAD05352782GA
SAD05392994GA
SAD04491095GA
SAD09335612GA
SAD05142090GA
SAD05142095GA
SAD05142096GA
SAD05142111GA
SAD05142112GA
SAD05142113GA
SAD05142114GA
SAD05142133GA
SAD05051495GA
SAD05051496GA
SAD05051497GA
SAD05051498GA
SAD05051499GA
SAD05051537GA
SAD05051551GA
SAD05051552GA
SAD05051553GA
SAD05021346GA
SAD05021347GA
SAD04491214GA
SAD04501262GA
SAD04511319GA
SAD07325134GA
SAD06234122GA
SAD06234123GA
SAD08515550GA
SAD04380374GA
SAD04380375GA
SAD04380376GA
SAD04380377GA
SAD04380378GA
SAD04380379GA
SAD04420557GA
SAD04420558GA
SAD04420559GA
SAD04420560GA
SAD04420561GA
SAD04440676GA
SAD07395178GA
SAD05352791GA
SAD05352792GA
SAD05352793GA
SAD05352794GA
SAD05352797GA
SAD05352798GA
SAD05352799GA
SAD05352800GA
SAD05362826GA
SAD05362860GA
SAD05362861GA
SAD05362862GA
SAD06113792GA
SAD06113793GA
SAD06113794GA
SAD06113796GA
SAD06113795GA
SAD06264266GA
SAD06264267GA
SAD08095314GA
SAD05443059GA
SAD05443060GA
SAD05443061GA
SAD05443062GA
SAD05443063GA
SAD05443064GA
SAD05443065GA
SAD05443066GA
SAD05443067GA
SAD05443068GA
SAD05443069GA
SAD05443070GA
SAD05443071GA
SAD05443072GA
SAD06444639GA
SAD04470930GA
SAD04470931GA
SAD04470932GA
SAD04470933GA
SAD04470934GA
SAD04470935GA
SAD06444618GA
SAD04460812GA
SAD04460813GA
SAD04460814GA
SAD04460815GA
SAD05252561GA
SAD05252562GA
SAD05252564GA
SAD05252565GA
SAD05252566GA
SAD05252567GA
SAD05262593GA
SAD05262594GA
SAD05262595GA
SAD05262596GA
SAD05262597GA
SAD05262598GA
SAD05262599GA
SAD05262600GA
SAD05262601GA
SAD05262602GA
SAD05262603GA
SAD05262604GA
SAD05262605GA
SAD05322702GA
SAD04491211GA
SAD04501250GA
SAD04511301GA
SAD04511305GA
SAD04511306GA
SAD04511307GA
SAD04511308GA
SAD04511309GA
SAD04511310GA
SAD04511311GA
SAD06444642GA
SAD07335146GA
SAD04450762GA
SAD04450763GA
SAD05081863GA
SAD05081864GA
SAD05081865GA
SAD05091866GA
SAD06424602GA
SAD06424611GA
SAD05503396GA
SAD05302664GA
SAD05302665GA
SAD05302666GA
SAD05302667GA
SAD05302668GA
SAD05302669GA
SAD05302670GA
SAD05302678GA
SAD05302679GA
SAD07124916GA
SAD04330216GA
SAD04350308GA
SAD04350309GA
SAD04350310GA
SAD04350311GA
SAD04350312GA
SAD04350313GA
SAD04350314GA
SAD04350315GA
SAD04350316GA
SAD04240013GA
SAD04240014GA
SAD04240020GA
SAD04240031GA
SAD04240032GA
SAD04240033GA
SAD04240034GA
SAD04240035GA
SAD04240036GA
SAD04240037GA
SAD04240038GA
SAD04240039GA
SAD04240040GA
SAD04250056GA
SAD04250057GA
SAD05453152GA
SAD05453162GA
SAD05453163GA
SAD05463191GA
SAD05463192GA
SAD05463193GA
SAD05463194GA
SAD05473238GA
SAD05503391GA
SAD05503392GA
SAD05503393GA
SAD05503397GA
SAD05503398GA
SAD05523439GA
SAD05523440GA
SAD05523441GA
SAD05523442GA
SAD05523443GA
SAD05523444GA
SAD05523445GA
SAD05523446GA
SAD05523452GA
SAD05523453GA
SAD05523454GA
SAD05523455GA
SAD05523463GA
SAD05523464GA
SAD06133888GA
SAD06133889GA
SAD06133890GA
SAD06133891GA
SAD06183933GA
SAD06183934GA
SAD06183935GA
SAD06183936GA
SAD06183937GA
SAD06183938GA
SAD06183939GA
SAD06183940GA
SAD06183941GA
SAD06183942GA
SAD06183943GA
SAD06183944GA
SAD06183945GA
SAD06183947GA
SAD06183948GA
SAD06183949GA
SAD06183950GA
SAD06183951GA
SAD06183952GA
SAD06183953GA
SAD06183954GA
SAD06183955GA
SAD06183956GA
SAD06183957GA
SAD06183958GA
SAD06183959GA
SAD06183960GA
SAD06183961GA
SAD06183962GA
SAD06183963GA
SAD06183964GA
SAD06183965GA
SAD06183966GA
SAD06183967GA
SAD06183968GA
SAD06183969GA
SAD06183970GA
SAD06183971GA
SAD06183972GA
SAD06183973GA
SAD06183974GA
SAD06183975GA
SAD06183976GA
SAD06183977GA
SAD06183978GA
SAD06183979GA
SAD06183980GA
SAD06183981GA
SAD06183982GA
SAD06183983GA
SAD06183985GA
SAD06183986GA
SAD06183987GA
SAD06183988GA
SAD06183989GA
SAD06183990GA
SAD06183991GA
SAD06183992GA
SAD06183993GA
SAD06183994GA
SAD06183995GA
SAD06183996GA
SAD06183997GA
SAD06183998GA
SAD06183999GA
SAD06184000GA
SAD06184001GA
SAD06184002GA
SAD06184003GA
SAD06184004GA
SAD06184005GA
SAD06184006GA
SAD06184007GA
SAD06184008GA
SAD06184009GA
SAD06184010GA
SAD06184011GA
SAD06184012GA
SAD06184013GA
SAD06184014GA
SAD06184015GA
SAD06184016GA
SAD06184017GA
SAD06184018GA
SAD06184019GA
SAD06184020GA
SAD06184021GA
SAD06184022GA
SAD06184023GA
SAD06184024GA
SAD06184025GA
SAD06184026GA
SAD06184027GA
SAD06184028GA
SAD06184029GA
SAD06184030GA
SAD06184031GA
SAD06184032GA
SAD06184033GA
SAD06184034GA
SAD06444623GA
SAD06444626GA
SAD10245638GA
SAD10245639GA
SAD06073737GA
SAD06113785GA
SAD06113786GA
SAD06123853GA
SAD05071606GA
SAD05071607GA
SAD05071638GA
SAD05071639GA
SAD05071641GA
SAD05071645GA
SAD05302636GA
SAD05252533GA
ABF05040314FA
SAD07475206GA
ABF04360245FA
ABF04360246FA
ABF04360247FA
ABF04360248FA
ABF04360249FA
SAD08285429GA
SAD08285435GA
SAD08305447GA
SAD08305448GA
SAD08305453GA
SAD05382943GA
SAD05392956GA
SAD05392971GA
SAD05392989GA
SAD05443122GA
SAD08305450GA
SAD08305451GA
SAD08305452GA
SAD05352806GA
SAD05352807GA
SAD05352808GA
SAD05352809GA
SAD05352810GA
SAD05392947GA
SAD06264270GA
SAD05453161GA
SAD05453164GA
SAD05302645GA
SAD05443049GA
SAD05443074GA
SAD07124912GA
SAD07215059GA
SAD07124929GA
PA2156504
PA2156505
PA2156506
PA2156507
PA2156508
PA2156509
PA2156510
PA2156511
PA2305323
PA2305341
SAD05443057GA
SAD07335145GA
SAD05071656GA
SAD05071694GA
SAD05071695GA
SAD05071701GA
SAD05071703GA
SAD05071704GA
SAD05081838GA
SAD05081839GA
SAD06284288GA
SAD08515549GA
SAD05071665GA
SAD05473235GA
SAD05473236GA
SAD05473248GA
SAD05473249GA
SAD05372892GA
SAD07124941GA
SAD05071623GA
SAD05081759GA
SAD05071684GA
SAD06113787GA
SAD06143927GA
SAD05071680GA
SAD05443050GA
SAD05443051GA
SAD05142115GA
PA2298583
PA2298584
PA2298585
PA2298586
PA2298587
PA2298588
PA2298589
PA2368979
PA2368980
PA2368311
PA2368312
PA2368313
PA2368314
PA2368315
PA2368316
PA2368317
PA2368073
PA2368074
PA2368075
PA2368076
PA2368077
PA2368769
PA2368770
PA2368771
PA2368772
PA2368773
PA2367560
PA2367561
PA2367562
PA2367563
PA2367564
PA2368306
PA2368307
PA2368308
PA2368309
PA2368310
PA2299122
PA2299123
PA2299124
PA2299125
PA2299126
PA2298582
SAD06324329GA
SAD04390435GA
SAD05473237GA
SAD06063694GA
ABF04500293FA
PA2305328
PA2305329
PA2427457
PA0000098
PA0000627
PA0000628
PA2427456
PA2427357
PA2427358
SBP14070147TA
SBP14070149TA
PA2020281
SAD07305109GA |
| FEI Number |
3012359877
|
Recalling Firm/ Manufacturer |
GE Medical Systems Information Technologies Inc
9900 W Innovation Dr
Milwaukee
WI
53226-4856
|
| For Additional Information Contact | GE HealthCare Service 800-437-1171 |
Manufacturer Reason for Recall | The MUSE Administrator Account active directory or local credential could potentially be used from your local network to log into a GE HealthCare MUSE 5 system and potentially manipulate recorded patient data. |
FDA Determined Cause 2 | Software design |
| Action | An URGENT MEDICAL DEVICE CORRECTION notification letter dated 4/25/25 was sent to customers.
Actions to be taken by Customer/User:
MUSE 5 systems have reached their End of Sales/Service Support as of November 1, 2014. Please see attached for the End of Sales/Service Support letter that was previously distributed. GE HealthCare recommends that customers replace or upgrade these systems.
If you wish to continue using your MUSE 5 system, you should ensure your IT Department provides ongoing prevention of unauthorized access to the system such as not using a well-known password.
If passwords are well-known, please update the credentials according to your organization's password policies or contact GE HealthCare Service or your local Service Representative if necessary.
Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions.
Please retain this document for your records.
Please complete and return the attached Medical Device Notification Acknowledgement Response Form to DCAR.Recall@gehealthcare.com.
If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. |
| Quantity in Commerce | 4,277 systems |
| Distribution | Worldwide - US Nationwide distribution. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = DSI
|