| | Class 2 Device Recall Achieva |  |
| Date Initiated by Firm | June 30, 2025 |
| Date Posted | July 30, 2025 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-2200-2025 |
| Recall Event ID |
97172 |
| 510(K)Number | K193215 |
| Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
| Product | Achieva 3.0T;
Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345; |
| Code Information |
Model Numbers (REF): (1) 781177, (2) 781277, (3) 781278, (4) 781344, (5) 781345;
UDI-DI: (1) None, (2) None, (3) 00884838004139, (4) None, (5) None;
Serial Numbers: (1) 17008, 17011, 17017, 17030, 17035, 17038, 17039, 17043, 17049, 17051, 17056, 17061, 17072, 17075, 17079, 17093, 17094, 17099, 17114, 17130, 17134, 17139, 17142, 17144, 17146, 17147, 17149, 17161, 17167, 17168, 17170, 17171, 17178, 17179, 17182, 17183, 17185, 17187, 17189, 17216, 17217, 17226, 17233, 17234, 17235, 17243, 17251, 17256, 17270, 17279, 17295, 17296, 17304, 17316, 17319, 17324, 17334, 17347, 17349, 17358, 17362, 17365, 17374, 17379, 17384, 17392, 17393, 17393, 17402, 17404, 17414, 17418, 17426, 17427, 17431, 17432, 17443, 17445, 17446, 17448, 17450, 17452, 17453, 17456, 17469, 17471, 17494, 17497, 17500, 17523, 17535, 17558, 17561, 17596, 17599, 17607, 24007, 24009, 24010, 24019, 24037, 24038, 24049, 24050, 24053, 24056, 24059, 24060, 24066, 24068, 24070, 24078, 24079, 24081, 24089, 24091, 24096, 24097, 24099, 24100, 24107, 24114, 24116, 24120, 24125, 24130, 24131, 24138, 24154, 34148, (2) 24115, 34002, 34011, 34020, 34028, 34029, 34030, 34033, 34034, 34038, 34043, 34065, 34066, 34071, 34073, 34074, 34077, 34082, 34086, 34095, 34106, 34109, 34113, 34115, 34118, 34120, 34123, 34135, 34138, 34139, 34140, 34144, 34159, 34160, 34165, 34200, 34201, 34203, 34207, 34213, 34220, 34225, 34230, 34231, 34232, 34234, 34237, 34239, 34241, 34242, 34247, 34253, 34254, 34255, 34256, 34258, 34263, 34264, 34268, 34270, 34271, 34272, 34273, 34274, 34275, 34276, 34281, 34282, 34285, 34286, 34287, 34288, 34289, 34290, 34291, 34293, 34294, 34295, 34297, 34298, 34299, 34300, 34301, 34305, 34306, 34309, 34311, 73012, (3) 38002, 38003, 38007, 38012, 38017, 38018, 38021, 38022, 38033, 38034, 38043, 38044, 38047, 38054, 38055, 38056, 38057, 38059, 38060, 38061, 38064, 38069, 38072, 38079, 38085, 38086, 38087, 38090, 38091, 38093, 38095, 38098, 38099, 38103, 38104, 38108, 38112, 38115, 38130, 38133, 38137, 38139, 38140, 38141, 38143, 38144, 38145, 38150, 38151, 38152, 38157, 38158, 38160, 38167, 38169, 38171, 38173, 38174, 38181, 38185, 38186, 38187, 38188, 38190, 38191, 38194, 38196, 38200, 38206, 38207, 38209, 38210, 38212, 38215, 38216, 38220, 38223, 38227, 38228, 38229, 38231, 38232, 38233, 38236, 38238, 38240, 38241, 38245, 38246, 38247, 38248, 38249, 38251, 38252, 38253, 38254, 38255, 38256, 38257, 38258, 38260, 38261, 38262, 38264, 38265, 38266, 38267, 38268, 38269, 38270, 38271, 38272, 38273, 38274, 38276, 38278, 38279, 38280, 38282, 38283, 38284, 38287, 38288, 38291, 38292, 38293, 38295, 38296, 38297, 38298, 38299, 38302, 38303, 38304, 38307, 38312, 38314, 38315, 38318, 38319, 38320, 38321, 38322, 38323, 38324, 38326, 38327, 38328, 38329, 38330, 38331, 38332, 38335, 38337, 38338, 38339, 38340, 38341, 38343, 38344, 38345, 38346, 38347, 38349, 38351, 38355, 38358, 38359, 38360, 38361, 38363, 38367, 38368, 38370, 38372, 38401, 38402, 38405, 38407, 38408, 38410, 38412, 38414, 38415, 38420, 38421, 38423, 38424, 38425, 38426, 38427, 38429, 38430, 38431, 38432, 38433, 38436, 38437, 38438, 38439, 38441, 38442, 38443, 38445, 38447, 38448, 38450, 38453, 38454, 38456, 38457, 38458, 382320, (4) 73003, 73005, 73006, 73008, 73009, 73010, (5) 74001, 74005, 74007, 74009, 74010, 74020, 74021, 74026, 74027, 74029, 74030, 74032, 74033, 74034, 74035, 74037, 74039, 74041, 74044, 74046, 74047, 74048, 74052; |
| FEI Number |
3006648320
|
Recalling Firm/ Manufacturer |
Philips North America
222 Jacobs St
Cambridge
MA
02141-2296
|
| For Additional Information Contact | Customer Care Solution Center 1-800-722-9377 |
Manufacturer Reason for Recall | The potential for component failures in the Gradient Coil of the affected MR systems may act as a heat source with a potential to produce smoke and/or fire. |
FDA Determined Cause 2 | Device Design |
| Action | On June 30, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers.
4. Actions that should be taken by the customer / user in order to prevent risks for patients or users
4.1. When using any affected system identified according to the information in section 3, follow
the instructions below.
4.2. Pay particular attention if a scan interruption occurs and a scan abort symbol is
encountered, which may appear in the user interface (UI) with the symbols shown in Figure 4
below:
Figure 4. Scan Abort Symbols
4.3. When scan aborts occur the symbol is shown and applicable messages are presented in the
message box (see Figure 5 and Figure 6 examples)
4.3.1. Check the status of the error messages in the User Interface (UI) screen. Click to
expand the message box in case previous messages are hidden, see orange box in
Figure 5 and Figure 6 below. If possible, correct the issue that is relevant to the error
shown and continue scanning, for example:
RF door must be closed while performing scan. Scan aborted.
Patient support is moved while scanning, scan stopped.
Coil A (or B) failure, coil disconnected? .
4.3.2. If, after a scan abort, one of the following conditions occur five times in a row stop
scanning immediately and contact Philips Service to describe the problem:
If there is no error message in the message box
If the error message meaning is not clear
4.3.3. If the error message Gradient amplifier Rack Fault is received two times in a row,
or if the error message Gradient amplifier Rack Fault is preceded or followed by an
aborted scan with no message, then stop scanning immediately and contact Philips
Service to describe the problem.
4.3.4. If you encounter any of these conditions described in 4.3.2 or 4.3.3 do not attempt
any additional/further scans, including without limitation do not attempt any
phantom scanning until your system has been checked and released by a Philips
Service representative.
4.4. If smoke and/or fire is detected |
| Quantity in Commerce | 460 units |
| Distribution | Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, French Guiana, FrenchPolynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Lesotho, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Trinidad,Tobago, Tunisia, T¿rkiye, Turkmenistan, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen, Zambia. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = LNH
|
|
|
|