| | Class 1 Device Recall TRUE METRIX SelfMonitoring Blood Glucose System |  |
| Date Initiated by Firm | February 06, 2026 |
| Date Posted | March 12, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1545-2026 |
| Recall Event ID |
98317 |
| 510(K)Number | K140100 |
| Product Classification |
System, test, blood glucose, over the counter - Product Code NBW
|
| Product | Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item:
Walgreens TRUE METRIX Kit/RE4001-01,
Leader TRUE METRIX Kit/RE4002-01,
Leader TRUE METRIX Meter Only/RE4002-40,
CVS TRUE METRIX Kit/RE4007-01SB,
GNP TRUE METRIX Meter Only/RE4011-00,
GNP TRUE METRIX Kit /RE4011-01,
MEIJER TRUE METRIX Kit/RE4019-01,
FRED'S TRUE METRIX Kit/RE4023-01,
DISCOUNT DRUG TRUE METRIX Kit/RE4025-01,
KINRAY TRUE METRIX Kit/RE4027-01,
HEB TRUE METRIX Kit/RE4031-01,
McKesson TRUE METRIX Starter Kit/RE4051-43,
SUNMARK TRUE METRIX Meter Only/RE4054-00,
SUNMARK TRUE METRIX Kit/RE4054-01,
Signature Care TRUE METRIX Kit/RE4060-01,
CareOne TRUE METRIX Kit/RE4061-01,
RITE AID TRUE METRIX Kit/RE4066-01,
Liberty TRUE METRIX Meter Only/RE4078-40,
HUMANA TRUE METRIX Starter Kit/RE4081-03,
Wegmans TRUE METRIX Kit/RE4087-01,
HEALTHMART TRUE METRIX Meter Only/RE4089-00,
HEALTHMART TRUE METRIX Kit/RE4089-01,
ARRIVA TRUE METRIX Meter Only/RE4095-40,
HealthyAccents TRUE METRIX Kit/RE4097-01,
Shopko TRUE METRIX Kit/RE4098-01,
Henry Schein TRUE METRIX Kit/RE4099-40,
Henry Schein TRUE METRIX Meter Only/RE4099-45,
Millennium Pharmay Services TRUE METRIX Kit/RE4202-43,
Kroger TRUE METRIX Kit/RE4203-01,
TOPCARE TRUE METRIX Kit/RE4209-01,
Foster and Thrive TRUE METRIX Kit/RE4211-01,
TRUE METRIX Meter Only/RE4H01-00,
TRUE METRIX Kit/RE4H01-01,
TRUE METRIX Starter Kit/RE4H01-04,
TRUE METRIX NFRS Meter Only/RE4H01-40,
TRUE METRIX NFRS Starter Kit/RE4H01-43,
TRUE METRIX Kit/RE4H01-51,
TRUE METRIX Meter Only (mg/dL)/RE4i01-00,
TRUE METRIX Kit (mg/dL)/RE4i01-01,
TRUE METRIX Kit (mg/dL)/RE4i01-02,
TRUE METRIX Starter Kit (mg/dL)/RE4i03-02,
TRUE METRIX Kit (mmol/L)/RE4i03-11,
Farmacia Benavides TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i05-00,
Farmacia del Ahorro TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i07-00,
TRUE METRIX Meter Only (Mexico - mg/dL)/RE4i23-00,
TRUE METRIX Meter Only (Jamaica - mmol/L)/RE4i29-11,
TRUE METRIX Kit (mg/dL)/RE4i61-02,
TRUE METRIX Kit (mg/dL)/RE4i61-02THI,
TRUE METRIX Kit (United Kingdom - mmol/L)/RE4i82-11,
TRUE METRIX Starter Kit (United Kingdom - mmol/L)/RE4i82-12. |
| Code Information |
Device Item/UDI-DI: RE4001-01/00311917166896, RE4002-01/96295126600, RE4002-40/96295126655, RE4007-01SB/50428275283, RE4011-00/87701426520, RE4011-01/87701426254, RE4019-01/713733069816, RE4023-01/21292007256, RE4025-01/93351018992, RE4027-01/21292007065, RE4031-01/41220001286, RE4051-43/20612479197231, RE4054-00/10939722447, RE4054-01/10939719447, RE4060-01/321130788669, RE4061-01/41520147745, RE4066-01/11822445610, RE4078-40/21292007614, RE4081-03/21292014797, RE4087-01/77890394113, RE4089-00/52569138557, RE4089-01/52569138526, RE4095-40/21292007041, RE4097-01/725439988933, RE4098-01/21292008901, RE4099-40/21292014902, RE4099-45/21292008697, RE4202-43/21292007454, RE4203-01/41260016028, RE4209-01/36800234529, RE4211-01/10939958693, RE4H01-00/21292006082, RE4H01-01/21292006051, RE4H01-04/21292008895, RE4H01-40/21292006068, RE4H01-43/21292006075, RE4H01-51/21292014940.
International: RE4i01-00/N/A, RE4i01-01/N/A, RE4i01-02/N/A, RE4i03-02/N/A, RE4i03-11/N/A, RE4i05-00/N/A, RE4i07-00/N/A, RE4i23-00/21292010133, RE4i29-11/N/A, RE4i61-02/N/A, RE4i61-02THI/N/A, RE4i82-11/21292012205, RE4i82-12/21292009977.
Owner's Booklet or IFU Description/Part and Revision:
TRUE METRIX Owner's Booklet/ RE4TVH03 Rev 56 and prior,
TRUE METRIX Owner's Booklet (English)/ RE4TVH35 Rev 51 and prior,
TRUE METRIX Owner's Booklet (Spanish)/ RE4TVH35S Rev 51 and prior,
McKesson Med Surg TRUE METRIX Self Monitoring Owners Booklet/ RE4SUN03 Rev 55 and prior,
McKesson Med Surg TRUE METRIX PRO Owners Booklet/ RE4SUNP03 Rev 56 and prior,
TRUE METRIX Blood Glucose Monitoring System IFU Trividia House Brand/ RE4ITV16 Rev 53 and prior,
TRUE METRIX System Instructions for Use (IFU) for Trividia Health UK Limited/RE4UKT16 Rev 55 and prior,
Trividia Health LATAM - TRUE METRIX System IFU mg/dL/ RE4LAT16 Rev 50 and prior,
TRUE METRIX System Instructions for Use (IFU) for Trividia Health Australia/ RE4AUT16 Rev 56 and prior.
Device serial numbers: Prefix "T" followed by an eight-digit numeric sequence.
|
| FEI Number |
1000113657
|
Recalling Firm/ Manufacturer |
Trividia Health, Inc.
2400 Nw 55th Ct
Fort Lauderdale
FL
33309-2672
|
| For Additional Information Contact | Annmarie Ramos 954-677-9201 |
Manufacturer Reason for Recall | The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose. |
FDA Determined Cause 2 | Software design |
| Action | On 2/6/2026, press was issued and correction notices were emailed to customers.
E-5 Error Code in the Messages section of the Owner's Booklets/System Instructions for Use emphasizes that users must seek medical attention immediately if they receive an E-5 error code and are experiencing symptoms of high glucose.
Customers: notify your customers immediately of the correction and to include the letter in their notification, and to ask that they notify their customers or facilities if they have further distributed the Products.
Retail pharmacies: post the notice in areas where the Products are sold to notify patients.
Multiple Patient Facilities: post the notice in areas where the Products are stored to notify personnel.
People with Diabetes are asked to follow the updated E-5 error code instructions Visit www.trividiahealth.com for updated Owner's Booklets and help resources.
Complete and return the acknowledgement form via email to trividia0126CS@trividiahealth.com
Firm Customer Care Department: 1-888-835-2723, trividia0126CC@trividiahealth.com or visit www.trividiahealth.com/E-5productnotice , which has links to Owner's Booklets, correction notice, and press.
On 5/5/2026, updated correction notices were mailed to customers informing them of the following:
For people with diabetes, if possible, consider transitioning to an alternative method of testing your blood glucose. People with Diabetes who rely on intensive insulin therapy, sulfonylureas, or glucose monitoring due to frequent hypo- or hyperglycemia events should consider transitioning to an alternative testing method until the updated devices with updated software becomes available. Firm will notify users when the upgrade program becomes available. The firm asked customers to handle dissemination of the updated notice and completion/return of the acknowledgement form as stated above. Affected serial number search tool is located at www.trividiahealth.com/E-5productnotice |
| Quantity in Commerce | 12,067,458 |
| Distribution | Worldwide distribution - US Nationwide including Puerto Rico and the countries of Mexico, United Kingdom, Jamaica, Trinidad & Tobago, and Australia.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = NBW
|
|
|
|