| | Class 2 Device Recall Gelsoft Plus, Gelweave, Thoraflex Hybrid |  |
| Date Initiated by Firm | December 18, 2025 |
| Date Posted | March 04, 2026 |
| Recall Status1 |
Open3, Classified |
| Recall Number | Z-1492-2026 |
| Recall Event ID |
98338 |
| 510(K)Number | K241070 K241550 |
| Product Classification |
Prosthesis, vascular graft, of 6mm and greater diameter - Product Code DSY
|
| Product | Gelsoft Plus, REF: 635024PE, 631514PE, 631222PE, 633008PE, 633010PE, 633007PE, 636010PE, 632522PE, 633012PE, 635018PE, 631516PE, 636016PE, 631220PE, 631608PE, 631206PE, 631407P50E, 631508PE, 636008PE, 636006PE, 632518PE, 634008PE, 634006PE, 631218PE, 632211PE, 634007PE, 633006PE, 631506PE, 632524PE, 636007PE, 631512PE, 636012PE, 636014PE, 632211P50E, 633016PE, 631407PE, 631809PE, 632010PE, 631608P50E, 632520PE, 632412PE, 631510PE, 635022PE, 631224PE, 632412P50E, 631507PE, 633014PE, 632010P50E, 635020PE, 631809P50E.
Gelweave, REF: 736038E, 736036E, 736034E, 736032ST/10DE, 736032E, 736030ST/10DE, 736030E, 736028ST/10DE, 736028E, 736026ST/10DE, 736026E, 736024ST/10DE, 736024E, 736022ST/10DE, 736022E, 736020E, 736018E, 736016E, 736014E, 736012E, 736010E, 736008E, 736006E, 735034ST/8AE, 735034/8RME, 735034/10SEE, 735034/10RME, 735032STE, 735032ST/8AE, 735032/8SE, 735032/8RME, 735032/10SEE, 735032/10RME, 735030STE, 735030ST/8AE, 735030/8SE, 735030/8RME, 735030/10SEE, 735030/10RME, 735028STE, 735028ST/8AE, 735028E, 735028/8SE, 735028/8RME, 735028/10RME, 735026STE, 735026ST/8AE, 735026E, 735026/8SE, 735026/8RME, 735026/10RME, 735024STE, 735024ST/8AE, 735024E, 735024/8SE, 735024/8RME, 735024/10SEE, 735024/10RME, 735022STE, 735022E, 735022/8SE, 735022/8RME, 735020E, 735020/8RME, 735018E, 734038E, 734036E, 734034E, 734034CX4RMSE, 734034/8E, 734034/10E, 734032E, 734032CX4RMSE, 734032/8E, 734032/10E, 734030E, 734030CX4RMSE, 734030/8E, 734030/10E, 734028CX4RMSE, 734028/8E, 734028/10E, 734026CX4RMSE, 734026/8E, 734026/10E, 734024CX4RMSE, 734024/8E, 734024/10E, 734022CX4RMSE, 734022/8E, 734022/10E, 734020CX4RMSE, 734020/8E, 734020/10E, 734018/8E, 73341088E, 73341088/8RME, 73341088/8E, 73341088/10RME, 73341088/10E, 73321488/8ARME, 7332148/10RMFE, 7332148/10RMEE, 733212810/10STDE, 7332128/10RMFE, 7332128/10RMEE, 73321088E, 73321088/8SDE, 73321088/8RME, 73321088/8E, 73321088/10RME, 73321088/10E, 7330888/8E, 7330668/6E, 733038E, 733036E, 733034E, 733032E, 733030E, 733028E, 733026E, 733024E, 733024/6AE, 733022E, 733022/6AE, 733020E, 733020/6AE, 733018E, 733018/6AE, 733016E, 733014E, 73301488/8ARME, 7330148/10RMFE, 7330148/10RMEE, 73301412/10ETE, 733012E, 733012810/10STDE, 7330128/10RMFE, 7330128/10RMEE, 733010E, 73301088E, 73301088/8SDE, 73301088/8RME, 73301088/8E, 73301088/10RME, 73301088/10E, 733010108/8S4E, 733010108/8RME, 733010/10RME, 733008E, 733006E, 7328888/8E, 7328668/6E, 73281488/8ARME, 7328148/10RMEE, 73281412/10ETE, 73281288/8E, 732812810/10STDE, 7328128/10RMFE, 7328128/10RMEE, 73281088E, 73281088/8SDE, 73281088/8RME, 73281088/8E, 73281088/10RME, 73281088/10E, 732810108/8S4E, 7326888/8E, 7326668/6E, 73261488/8ARME, 7326148/10RMEE, 732612810/10STDE, 7326128/10RMFE, 7326128/10RMEE, 73261088E, 73261088/8SDE, 73261088/8RME, 73261088/8E, 73261088/10RME, 73261088/10E, 732610108/8S4E, 732610/10RME, 732528E, 732526E, 732524E, 732522E, 732520E, 732518E, 7324888/8E, 7324668/6E, 73241488/8ARME, 7324148/10RMEE, 732412E, 732412810/10STDE, 7324128/10RMEE, 732412/6E, 73241088E, 73241088/8SDE, 73241088/8RME, 73241088/8E, 73241088/10RME, 73241088/10E, 732410108/8S4E, 73221488/8ARME, 732214/10/10CRME, 732212810/10STDE, 732212/10/8CRME, 732212/10/8ARME, 732211E, 73221088/8RME, 73221088/8E, 73221088/10E, 732210108/8S4E, 7320888/8E, 7320668/6E, 732038E, 732034E, 732034CX4RME, 732034/8E, 732034/10E, 732032E, 732032CX4RME, 732032/8E, 732032/10E, 732030E, 732030CX4RME, 732030/8E, 732030/10E, 732028CX4RME, 732028/8E, 732028/10E, 732026CX4RME, 732026/8E, 732026/10E, 732024CX4RME, 732024/8E, 732024/10E, 732022/10E, 732014/8X2E, 732014/8X2AE, 732014/10X2E, 732014/10X2AE, 732012810/10STDE, 732012/8X2E, 732012/8X2AE, 732012/10X2E, 732010E, 73201088/8RME, 73201088/8E, 732010108/8S4E, 732010/6E, 731809E, 731809/6E, 731608XL55E, 731608E, 731608/8/10RME, 731608/6E, 731608/10/10RME, 731516E, 731514E, 731512E, 731510E, 731508E, 731506E, 7314108/10RME, 731407XL55E, |
| Code Information |
All Gelsoft Plus devices (any item number ending 'E-B' e.g. 631206PE-B); REF, Item Number, UDI-DI(GTIN):
635024PE, 635024PE-B, 5037881166636;
631514PE, 631514PE-B, 5037881166858;
631222PE, 631222PE-B, 5037881166896;
633008PE, 633008PE-B, 5037881166698;
633010PE, 633010PE-B, 5037881166711;
633007PE, 633007PE-B, 5037881166681;
636010PE, 636010PE-B, 5037881166520;
632522PE, 632522PE-B, 5037881166780;
633012PE, 633012PE-B, 5037881166728;
635018PE, 635018PE-B, 5037881166605;
631516PE, 631516PE-B, 5037881166865;
636016PE, 636016PE-B, 5037881166551;
631220PE, 631220PE-B, 5037881166889;
631608PE, 631608PE-B, 5037881166933;
631206PE, 631206PE-B, 5037881166919;
631407P50E, 631407P50E-B, 5037881167008;
631508PE, 631508PE-B, 5037881166827;
636008PE, 636008PE-B, 5037881166513;
636006PE, 636006PE-B, 5037881166063;
632518PE, 632518PE-B, 5037881166759;
634008PE, 634008PE-B, 5037881166667;
634006PE, 634006PE-B, 5037881166643;
631218PE, 631218PE-B, 5037881166872;
632211PE, 632211PE-B, 5037881166971;
634007PE, 634007PE-B, 5037881166650;
633006PE, 633006PE-B, 5037881166674;
631506PE, 631506PE-B, 5037881166803;
632524PE, 632524PE-B, 5037881166797;
636007PE, 636007PE-B, 5037881166506;
631512PE,631512PE-B, 5037881166841;
636012PE, 636012PE-B, 5037881166537;
636014PE, 636014PE-B, 5037881166544;
632211P50E, 632211P50E-B, 5037881167046;
633016PE, 633016PE-B, 5037881166742;
631407PE, 631407PE-B, 5037881166926;
631809PE, 631809PE-B, 5037881166940;
632010PE, 632010PE-B, 5037881166964;
631608P50E, 631608P50E-B, 5037881167015;
632520PE, 632520PE-B, 5037881166773;
632412PE, 632412PE-B, 5037881166988;
631510PE, 631510PE-B, 5037881166834;
635022PE, 635022PE-B, 5037881166629;
631224PE, 631224PE-B, 5037881166902;
632412P50E, 632412P50E-B, 5037881167053;
631507PE, 631507PE-B, 5037881166810;
633014PE, 633014PE-B, 5037881166735;
632010P50E, 632010P50E-B, 5037881167039;
635020PE, 635020PE-B, 5037881166612;
631809P50E, 631809P50E-B, 5037881167022.
All Gelweave devices (any item number ending E-B e.g. 7320128/10RMEE-B); REF, Item Number, UDI-DI(GTIN):
736038E, 736038E-B, 5037881169415;
736036E, 736036E-B, 5037881169408;
736034E, 736034E-B, 5037881169392;
736032ST/10DE, 736032ST/10DE-B, 5037881172620;
736032E, 736032E-B, 5037881169385;
736030ST/10DE, 736030ST/10DE-B, 5037881172637;
736030E , 736030E-B, 5037881169378;
736028ST/10DE, 736028ST/10DE-B, 5037881172644;
736028E, 736028E-B, 5037881169361;
736026ST/10DE, 736026ST/10DE-B, 5037881172651;
736026E, 736026E-B, 5037881169354;
736024ST/10DE, 736024ST/10DE-B, 5037881172668;
736024E, 736024E-B, 5037881169347;
736022ST/10DE, 736022ST/10DE-B, 5037881172675;
736022E, 736022E-B, 5037881169330;
736020E, 736020E-B, 5037881169323;
736018E, 736018E-B, 5037881169316;
736016E, 736016E-B, 5037881114057;
736014E, 736014E-B, 5037881114040;
736012E, 736012E-B, 5037881168548;
736010E, 736010E-B, 5037881168517;
736008E, 736008E-B, 5037881168494;
736006E, 736006E-B, 5037881166056;
735034ST/8AE, 735034ST/8AE-B, 5037881173061;
735034/8RME, 735034/8RME-B, 5037881170039;
735034/10SEE, 735034/10SEE-B, 5037881171883;
735034/10RME, 735034/10RME-B, 5037881169958;
735032STE, 735032STE-B, 5037881175102;
735032ST/8AE, 735032ST/8AE-B, 5037881173078;
735032/8SE, 735032/8SE-B, 5037881172095;
735032/8RME, 735032/8RME-B, 5037881170046;
735032/10SEE, 735032/10SEE-B, 5037881171890;
735032/10RME, 735032/10RME-B, 5037881169965;
735030STE, 735030STE-B, 5037881175126;
735030ST/8AE, 735030ST/8AE-B, 5037881173085;
735030/8SE, 735030/8SE-B, 5037881172101;
735030/8RME, 735030/8RME-B, 5037881170053;
735030/10SEE, 735030/10SEE-B, 5037881171906;
735030/10RME, 735030/10RME-B, 5037881169972;
735028STE, 735028STE-B, 5037881175133;
735028ST/8AE, 735028ST/8AE-B, 5037881173092;
735028E, 735028E-B 5037881168876;
735028/8SE, 735028/8SE-B, 5037881172118;
735028/8RME, 735028/8RME-B, 5037881170060;
735028/10RME, 735028/10RME-B, 5037881169989;
735026STE, 735026STE-B, 5037881175140;
735026ST/8AE, 735026ST/8AE-B, 5037881173108;
735026E, 735026E-B, 5037881168869;
735026/8SE, 735026/8SE-B, 5037881172125;
735026/8RME, 735026/8RME-B, 5037881170077;
735026/10RME, 735026/10RME-B, 5037881169996;
735024STE, 735024STE-B, 5037881175164;
735024ST/8AE, 735024ST/8AE-B, 5037881173115;
735024E, 735024E-B, 5037881168845;
735024/8SE, 735024/8SE-B, 5037881172132;
735024/8RME, 735024/8RME-B, 5037881170084;
735024/10SEE, 735024/10SEE-B, 5037881171937;
735024/10RME, 735024/10RME-B, 5037881170008;
735022STE, 735022STE-B, 5037881175188;
735022E, 735022E-B, 5037881168838;
735022/8SE, 735022/8SE-B, 5037881172149;
735022/8RME, 735022/8RME-B, 5037881170091;
735020E, 735020E-B, 5037881168821;
735020/8RME, 735020/8RME-B, 5037881170107;
735018E, 735018E-B, 5037881168814;
734038E, 734038E-B, 5037881169309;
734036E, 734036E-B, 5037881169293;
734034E, 734034E-B, 5037881169286;
734034CX4RMSE, 734034CX4RMSE-B, 5037881170794;
734034/8E, 734034/8E-B, 5037881172460;
734034/10E, 734034/10E-B, 5037881172224;
734032E, 734032E-B, 5037881169279;
734032CX4RMSE, 734032CX4RMSE-B, 5037881170800;
734032/8E, 734032/8E-B, 5037881172477;
734032/10E, 734032/10E-B, 5037881172231;
734030E, 734030E-B, 5037881169262;
734030CX4RMSE, 734030CX4RMSE-B, 5037881170817;
734030/8E, 734030/8E-B, 5037881172484;
734030/10E, 734030/10E-B, 5037881172262;
734028CX4RMSE, 734028CX4RMSE-B, 5037881170824;
734028/8E, 734028/8E-B, 5037881172491;
734028/10E, 734028/10E-B, 5037881172286;
734026CX4RMSE, 734026CX4RMSE-B, 5037881170831;
734026/8E, 734026/8E-B, 5037881172507;
734026/10E, 734026/10E-B, 5037881172309;
734024CX4RMSE, 734024CX4RMSE-B, 5037881170848;
734024/8E, 734024/8E-B, 5037881172514;
734024/10E, 734024/10E-B, 5037881172323;
734022CX4RMSE, 734022CX4RMSE-B, 5037881170855;
734022/8E, 734022/8E-B, 5037881172521;
734022/10E, 734022/10E-B, 5037881172347;
734020CX4RMSE, 734020CX4RMSE-B, 5037881170862;
734020/8E, 734020/8E-B, 5037881172538;
734020/10E, 734020/10E-B, 5037881172354;
734018/8E, 734018/8E-B, 5037881172545;
73341088E, 73341088E-B, 5037881171562;
73341088/8RME, 73341088/8RME-B, 5037881169873;
73341088/8E, 73341088/8E-B, 5037881171371;
73341088/10RME, 73341088/10RME-B, 5037881169774;
73341088/10E, 73341088/10E-B, 5037881171296;
73321488/8ARME, 73321488/8ARME-B, 5037881171043;
7332148/10RMFE, 7332148/10RMFE-B, 5037881174136;
7332148/10RMEE, 7332148/10RMEE-B, 5037881174334;
733212810/10STDE, 733212810/10STDE-B, 5037881172552;
7332128/10RMFE, 7332128/10RMFE-B, 5037881174266;
7332128/10RMEE, 7332128/10RMEE-B, 5037881174433;
73321088E, 73321088E-B, 5037881171579;
73321088/8SDE, 73321088/8SDE-B, 5037881172699;
73321088/8RME, 73321088/8RME-B, 5037881169880;
73321088/8E, 73321088/8E-B, 5037881171388;
73321088/10RME, 73321088/10RME-B, 5037881169781;
73321088/10E, 73321088/10E-B, 5037881171302;
7330888/8E, 7330888/8E-B, 5037881170657;
7330668/6E, 7330668/6E-B, 5037881170718;
733038E, 733038E-B, 5037881169163;
733036E , 733036E-B, 5037881169156;
733034E, 733034E-B, 5037881169149;
733032E, 733032E-B, 5037881169132;
733030E, 733030E-B, 5037881169118;
733028E, 733028E-B, 5037881169101;
733026E, 733026E-B, 5037881169095;
733024E, 733024E-B, 5037881169088;
733024/6AE, 733024/6AE-B, 5037881171845;
733022E, 733022E-B, 5037881169071;
733022/6AE, 733022/6AE-B, 5037881171852;
733020E, 733020E-B, 5037881169064;
733020/6AE, 733020/6AE-B, 5037881171869;
733018E, 733018E-B, 5037881169057;
733018/6AE, 733018/6AE-B, 5037881171876;
733016E, 733016E-B, 5037881168715;
733014E, 733014E-B, 5037881168708;
73301488/8ARME, 73301488/8ARME-B, 5037881171050;
7330148/10RMFE, 7330148/10RMFE-B, 5037881174150;
7330148/10RMEE, 7330148/10RMEE-B, 5037881174358;
73301412/10ETE, 73301412/10ETE-B, 5037881170206;
733012E, 733012E-B, 5037881168692;
733012810/10STDE, 733012810/10STDE-B, 5037881172569;
7330128/10RMFE, 7330128/10RMFE-B, 5037881174273;
7330128/10RMEE, 7330128/10RMEE-B, 5037881174440;
733010E, 733010E-B, 5037881168678;
73301088E, 73301088E-B, 5037881171586;
73301088/8SDE, 73301088/8SDE-B, 5037881172705;
73301088/8RME, 73301088/8RME-B, 5037881169897;
73301088/8E, 73301088/8E-B, 5037881171395;
73301088/10RME, 73301088/10RME-B, 5037881169811;
73301088/10E, 73301088/10E-B, 5037881171319;
733010108/8S4E, 733010108/8S4E-B, 5037881170589;
733010108/8RME, 733010108/8RME-B, 5037881169576;
733010/10RME, 733010/10RME-B, 5037881169729;
733008E, 733008E-B, 5037881168654;
733006E, 733006E-B, 5037881168616;
7328888/8E, 7328888/8E-B, 5037881170664;
7328668/6E, 7328668/6E-B, 5037881170725;
73281488/8ARME, 73281488/8ARME-B, 5037881171067;
7328148/10RMEE, 7328148/10RMEE-B, 5037881174365;
73281412/10ETE, 73281412/10ETE-B, 5037881170213;
73281288/8E, 73281288/8E-B, 5037881171289;
732812810/10STDE, 732812810/10STDE-B, 5037881172576;
7328128/10RMFE, 7328128/10RMFE-B, 5037881174280;
7328128/10RMEE, 7328128/10RMEE-B, 5037881174457;
73281088E, 73281088E-B, 5037881171593;
73281088/8SDE, 73281088/8SDE-B, 5037881172712;
73281088/8RME, 73281088/8RME-B, 5037881169903;
73281088/8E, 73281088/8E-B, 5037881171401;
73281088/10RME, 73281088/10RME-B, 5037881169828;
73281088/10E, 73281088/10E-B, 5037881171326;
732810108/8S4E, 732810108/8S4E-B, 5037881170596;
7326888/8E, 7326888/8E-B, 5037881170671;
7326668/6E, 7326668/6E-B, 5037881170732;
73261488/8ARME, 73261488/8ARME-B, 5037881171074;
7326148/10RMEE, 7326148/10RMEE-B, 5037881174372;
732612810/10STDE, 732612810/10STDE-B, 5037881172583;
7326128/10RMFE, 7326128/10RMFE-B, 5037881174297;
7326128/10RMEE, 7326128/10RMEE-B, 5037881174464;
73261088E, 73261088E-B, 5037881171609;
73261088/8SDE, 73261088/8SDE-B, 5037881172729;
73261088/8RME, 73261088/8RME-B, 5037881169910;
73261088/8E, 73261088/8E-B, 5037881171418;
73261088/10RME, 73261088/10RME-B, 5037881169835;
73261088/10E, 73261088/10E-B, 5037881171333;
732610108/8S4E, 732610108/8S4E-B, 5037881170602;
732610/10RME, 732610/10RME-B, 5037881169743;
732528E, 732528E-B, 5037881168982;
732526E, 732526E-B, 5037881168975;
732524E, 732524E-B, 5037881168968;
732522E, 732522E-B, 5037881168951;
732520E, 732520E-B, 5037881168944;
732518E, 732518E-B, 5037881168937;
7324888/8E, 7324888/8E-B, 5037881170688;
7324668/6E, 7324668/6E-B, 5037881170749;
73241488/8ARME, 73241488/8ARME-B, 5037881171081;
7324148/10RMEE, 7324148/10RMEE-B, 5037881174389;
732412E, 732412E-B, 5037881174907;
732412810/10STDE, 732412810/10STDE-B, 5037881172590;
7324128/10RMEE, 7324128/10RMEE-B, 5037881174495;
732412/6E, 732412/6E-B, 5037881174983;
73241088E, 73241088E-B, 5037881171616;
73241088/8SDE, 73241088/8SDE-B, 5037881172736;
73241088/8RME, 73241088/8RME-B, 5037881169927;
73241088/8E, 73241088/8E-B, 5037881171425;
73241088/10RME, 73241088/10RME-B, 5037881169842;
73241088/10E, 73241088/10E-B, 5037881171340;
732410108/8S4E, 732410108/8S4E-B, 5037881170619;
73221488/8ARME, 73221488/8ARME-B, 5037881171098;
732214/10/10CRME, 732214/10/10CRME-B, 5037881171661;
732212810/10STDE, 732212810/10STDE-B, 5037881172606;
732212/10/8CRME, 732212/10/8CRME-B, 5037881171678;
732212/10/8ARME, 732212/10/8ARME-B, 5037881171692;
732211E, 732211E-B, 5037881174839;
73221088/8RME, 73221088/8RME-B, 5037881169934;
73221088/8E, 73221088/8E-B, 5037881171432;
73221088/10E, 73221088/10E-B, 5037881171357;
732210108/8S4E, 732210108/8S4E-B, 5037881170626;
7320888/8E, 7320888/8E-B, 5037881170701;
7320668/6E, 7320668/6E-B, 5037881170763;
732038E, 732038E-B, 5037881169255;
732034E, 732034E-B, 5037881169248;
732034CX4RME, 732034CX4RME-B, 5037881170879;
732034/8E, 732034/8E-B, 5037881172392;
732034/10E, 732034/10E-B, 5037881172156;
732032E, 732032E-B, 5037881169231;
732032CX4RME, 732032CX4RME-B, 5037881170886;
732032/8E, 732032/8E-B, 5037881172408;
732032/10E, 732032/10E-B, 5037881172163;
732030E, 732030E-B, 5037881169224;
732030CX4RME, 732030CX4RME-B, 5037881170893;
732030/8E, 732030/8E-B, 5037881172415;
732030/10E, 732030/10E-B, 5037881172170;
732028CX4RME, 732028CX4RME-B, 5037881170909;
732028/8E, 732028/8E-B, 5037881172422;
732028/10E, 732028/10E-B, 5037881172187;
732026CX4RME, 732026CX4RME-B, 5037881170916;
732026/8E, 732026/8E-B, 5037881172439;
732026/10E, 732026/10E-B, 5037881172194;
732024CX4RME, 732024CX4RME-B, 5037881170923;
732024/8E, 732024/8E-B, 5037881172446;
732024/10E, 732024/10E-B, 5037881172200;
732022/10E, 732022/10E-B, 5037881172217;
732014/8X2E, 732014/8X2E-B, 5037881171739;
732014/8X2AE, 732014/8X2AE-B, 5037881176277;
732014/10X2E, 732014/10X2E-B, 5037881171722;
732014/10X2AE, 732014/10X2AE-B, 5037881175355;
732012810/10STDE, 732012810/10STDE-B, 5037881172613;
732012/8X2E, 732012/8X2E-B, 5037881171753;
732012/8X2AE, 732012/8X2AE-B, 5037881175324;
732012/10X2E, 732012/10X2E-B, 5037881171746;
732010E, 732010E-B, 5037881174822;
73201088/8RME, 73201088/8RME-B, 5037881169941;
73201088/8E, 73201088/8E-B, 5037881171449;
732010108/8S4E, 732010108/8S4E-B, 5037881170633;
732010/6E, 732010/6E-B, 5037881174976;
731809E, 731809E-B, 5037881174792;
731809/6E, 731809/6E-B, 5037881174945;
731608XL55E, 731608XL55E-B, 5037881175348;
731608E, 731608E-B, 5037881174747;
731608/8/10RME, 731608/8/10RME-B, 5037881175195;
731608/6E, 731608/6E-B, 5037881174921;
731608/10/10RME, 731608/10/10RME-B, 5037881175201;
731516E, 731516E-B, 5037881168807;
731514E, 731514E-B, 5037881168791;
731512E, 731512E-B, 5037881168784;
731510E, 731510E-B, 5037881168760;
731508E, 731508E-B, 5037881168746;
731506E, 731506E-B, 5037881168722;
7314108/10RME, 7314108/10RME-B, 5037881176468;
731407XL55E, 731407XL55E-B, 5037881175362;
731407E, 731407E-B, 5037881174686;
731407/7E, 731407/7E-B, 5037881175119;
731407/10/10RME, 731407/10/10RME-B, 5037881175218;
731228E, 731228E-B, 5037881169040;
731226E, 731226E-B, 5037881169033;
731224E, 731224E-B, 5037881169026;
731222E, 731222E-B, 5037881169019;
731220E, 731220E-B, 5037881169002;
731218E, 731218E-B, 5037881168999;
7312108/10RME, 7312108/10RME-B, 5037881176444;
731210/8X2E, 731210/8X2E-B, 5037881171456;
731206XL55E, 731206XL55E-B, 5037881175379;
731206E, 731206E-B, 5037881174631;
731206/10/10RME, 731206/10/10RME-B, 5037881175225;
731038E, 731038E-B, 5037881169217;
731036E, 731036E-B, 5037881169200;
731034E, 731034E-B, 5037881169194;
731032E, 731032E-B, 5037881169187;
731030E, 731030E-B, 5037881169170;
731010/8X2E, 731010/8X2E-B, 5037881171463;
730034ADPE, 730034ADPE-B, 5037881173986;
730032ADPSLE, 730032ADPSLE-B, 5037881173603;
730032ADPE, 730032ADPE-B, 5037881173993;
730030ADPSLE, 730030ADPSLE-B, 5037881173634;
730030ADPE, 730030ADPE-B, 5037881174013;
730028ADPSLE, 730028ADPSLE-B, 5037881173641;
730028ADPE, 730028ADPE-B, 5037881174037;
730026ADPE, 730026ADPE-B, 5037881174044;
730024ADPE, 730024ADPE-B, 5037881174051;
730022ADPE, 730022ADPE-B, 5037881174068;
730020ADPE, 730020ADPE-B, 5037881174082;
730018ADPE, 730018ADPE-B, 5037881174099;
730016ADPE, 730016ADPE-B, 5037881174112.
|
| FEI Number |
3002808437
|
Recalling Firm/ Manufacturer |
Vascutek, Ltd.
Newmains Avenue
Inchinnan
United Kingdom
|
| For Additional Information Contact | Adrienne Day 441418125555 |
Manufacturer Reason for Recall | Gelatin sealed woven polyester prostheses devices, designed for vascular repair, were incorrectly labeled with an additional month of shelf life. Device expiry should have been calculated from gel impregnation (date of manufacture) but it was incorrectly calculated from device packaging. |
FDA Determined Cause 2 | Error in labeling |
| Action | On 12/18/2025, Field Safety Notices were emailed to customers who were asked to do the following:
Awareness of this Field Safety Notice is advised and shall be distributed to operating surgeons and users. The shelf life ranges (expiry dates) have now been corrected for all newly manufactured products.
Please share this information with anyone in your organization who needs to be aware or is a user of the affected devices.
Complete and return the Field Safety Notice Return Confirmation form to: taukfsn2025-003shelflife@terumoaortic.com Questions can be sent to the this email address as well. |
| Quantity in Commerce | 37430 |
| Distribution | US Nationwide distribution including in the states of FL, PA, TX, NC, NY, CA, NJ, OH, MI, DC, CO, VA, GA, IL, OR, ND, VT, MS, MA, IN, WA, AZ, MN, MD, IA, OK, MO, AL, WI, LA, AR, TN, CT, ME, NE, SC, KS, NV, KY, RI, WV, UT, MT, ID, AK, NH, DE, NM, SD, HI, WY.
|
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
|
| 510(K) Database | 510(K)s with Product Code = DSY
|
|
|
|