• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall NeuroStar Advanced Therapy System

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall NeuroStar Advanced Therapy Systemsee related information
Date Initiated by FirmAugust 24, 2026
Date PostedOctober 09, 2026
Recall Status1 Open3, Classified
Recall NumberZ-0142-2027
Recall Event ID 99812
510(K)NumberK231926 K233621 
Product Classification Transcranial magnetic stimulator - Product Code OBP
ProductNeuroStar Advanced Therapy System. Contains the NeuroStar HP Coil Assembly Part Number 81-01000-200. Affects the below NeuroStar Advanced Therapy Systems: 1. NeuroStar 2.0 US; Part Number: 81-01315-000. 2. NeuroStar 3.0 US; Part Number: 81-02315-000. 3. NeuroStar 2.X to 3.2 Upgrade Kit; Part Number: 81-25000-100. 4. NeuroStar Upgrade Kit 3.0 US; Part Number: 81-35000-000. 5. NeuroStar Mobile Console 3.6, 120V, US; Part Number: 81-03315-000. 6. NeuroStar Mobile Console 3.7, 120V, US; Part Number: 81-04315-000. 7. NeuroStar Mobile Console 3.8, 120V, US; Part Number: 81-05315-000.
Code Information NeuroStar Software Versions 3.6, 3.7, 3.8 1. NeuroStar 2.0 US; Part Number: 81-01315-000; UDI-DI: 00869378000100. 2. NeuroStar 3.0 US; Part Number: 81-02315-000; UDI-DI: 00869378000117. 3. NeuroStar 2.X to 3.2 Upgrade Kit; Part Number: 81-25000-100; UDI-DI: 00869378000193. 4. NeuroStar Upgrade Kit 3.0 US; Part Number: 81-35000-000; UDI-DI: 00869378000155. 5. NeuroStar Mobile Console 3.6, 120V, US; Part Number: 81-03315-000; UDI-DI: 00850005944058. 6. NeuroStar Mobile Console 3.7, 120V, US; Part Number: 81-04315-000; UDI-DI: 00850005944522. 7. NeuroStar Mobile Console 3.8, 120V, US; Part Number: 81-05315-000; UDI-DI: 00850005944621. NeuroStar HP Coil Assembly Part Number: 81-01000-200. Coil Serial Numbers: 40203, 40208, 40202, 40204, 40206, 40205, 40207, 40209, 40210, 40211, 40212, 40213, 40214, 40215, 40216, 40217, 40218, 40219, 40220, 40221, 40222, 40223, 40224, 40225, 40227, 40228, 40226, 40229, 40230, 40231, 40232, 40233, 40234, 40236, 40235, 40238, 40239, 40240, 40241, 40242, 40243, 40244, 40245, 40246, 40247, 40248, 40249, 40250, 40251, 40252, 40253, 40254, 40255, 40256, 40257, 40258, 40259, 40260, 40261, 40262, 40263, 40265, 40266, 40267, 40264, 40283, 40284, 40285, 40286, 40287, 40288, 40289, 40290, 40291, 40293, 40292, 40294, 40295, 40296, 40297, 40298, 40299, 40300, 40301, 40302, 40303, 40304, 40305, 40306, 40307, 40308, 40309, 40310, 40311, 40312, 40313, 40314, 40315, 40316, 40317, 40318, 40320, 40319, 40321, 40322, 40323, 40324, 40325, 40326, 40327, 40328, 40329, 40331, 40332, 40333, 40334, 40335, 40336, 40337, 40338, 40339, 40340, 40341, 40342, 40343, 40344, 40347, 40346, 40345, 40348, 40349, 40350, 40353, 40352, 40351, 40354, 40355, 40356, 40357, 40358, 40359, 40360, 40371, 40372, 40373, 40374, 40375, 40376, 30643, 40274, 30638, 40281, 40199, 30671, 40201, 40272, 30641, 40278, 30646, 30653, 30640, 30647, 30652, 40275, 30648, 40197, 40279, 30650, 30639, 30623, 40198, 40280, 40270, 30645, 30654, 30636, 40200, 30635, 40282, 30606, 30607, 30608, 30609, 30610, 30611, 30624, 30625, 30630, 30631, 30632, 30633, 30634, 30655, 30656, 30657, 30658, 30659, 30660, 30622, 30621, 30626, 30614, 30620, 30616, 30619, 30615, 30617, 30661, 30662, 30663, 30664, 30665, 30627, 30628, 30629, 30644, 40268, 40269.
FEI Number 3004824012
Recalling Firm/
Manufacturer
Neuronetics, Inc.
3222 Phoenixville Pike
Malvern PA 19355-9615
For Additional Information ContactAlexandra Williamson
610-640-4202
Manufacturer Reason
for Recall
Cables are not in conformance with the product specifications required in the production of the HP Coil Assembly.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionNeuroStar began notification to consignees on about 08/24/2026 via email. Consignees were instructed to check their NeuroStar System for Console and Coil for affected serial numbers and complete and return the response form. A NeuroStar Field Service Engineer will contact you to coordinate product inspection at no expense to you. If the HP Coild Assembly is not affected, no action is required. If it is determined the product is nonconforming, the affected product will be replaced with a HP Coil Assembly with the proper specifications to address the issue at no expense to you. The Field Service Engineer will handle the delivery and installation of the HP Coil Assembly and return of the affected coil. The actions are expected to be completed in the next 12 months.
Quantity in Commerce179 units
DistributionWorldwide distribution - US Nationwide and the countries of Japan and South Korea.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 The manufacturer has initiated the recall and not all products have been corrected or removed. This record will be updated as the status changes.
510(K) Database510(K)s with Product Code = OBP
510(K)s with Product Code = OBP
-
-