Class 2 Device Recall

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recallsee related information
Date Initiated by FirmSeptember 01, 2003
Date PostedJuly 30, 2004
Recall Status1 Terminated 3 on July 24, 2007
Recall NumberZ-1233-04
Recall Event ID 29387
Product Classification Light, Ultraviolet, Dermatological - Product Code FTC
Product3 Series Full Body Phototherapy Device with Smart Touch Control System, Models #PASNBSM-2424.
Code Information All Touch Control System, Model #PASNBSM-2424 units manufactured between 6/5/03 and 4/30/04.
FEI Number 1526255
Recalling Firm/
Manufacturer
Daavlin Distributing Company
205 West Bement Street
PO Box 626
Bryan OH 43506-0626
For Additional Information ContactDavid Swanson
419-636-6304
Manufacturer Reason
for Recall
Software errors cause the unit to not deliver the prescribed therapy and do not allow the patient records to be saved.
FDA Determined
Cause 2
Other
ActionThe firm has software patches to correct the programming problems as they are reported by the users. These patches are being installed either by a service representative for the manufacturer or by the customer
Quantity in Commerce2 units distributed domestically.
DistributionNation wide plus Canada, the Philippines, Saudi Arabia, Turkey & Germany
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
-
-