| | Class 2 Device Recall Advisor |  |
| Date Initiated by Firm | January 05, 2005 |
| Date Posted | February 10, 2005 |
| Recall Status1 |
Terminated 3 on November 14, 2005 |
| Recall Number | Z-0531-05 |
| Recall Event ID |
30811 |
| 510(K)Number | K011177 K030407 K982279 |
| Product Classification |
Electrocardiograph - Product Code DPS
|
| Product | Advisor Vital Signs Monitor model 9200, catalog number 925454325. Item Description 9200 II 3LD/RP/02/IP/TP/BAT MDD. The basic monitoring package includes ECG (3 lead / 5 lead), impedance respiration, non-invasive blood pressure, pulse oximetry, two invasive blood pressures, and tow temperature channels. The device permits patient monitoring with adjustable alarm limits as well as visible and audible alarm signals. |
| Code Information |
serial numbers: 706330000, 706431000, 706431001, 706431002, 706431003, 706431004, 706431005, 706431006, 706431008, 706431009, 706431010 |
| FEI Number |
2183502
|
Recalling Firm/ Manufacturer |
Smiths Medical PM, Inc.
N7w22025 Johnson Dr
Waukesha
WI
53186-1856
|
Manufacturer Reason for Recall | An issue with the pace detect function was discovered on a 9200 II monitor. Could result in a falsely high or low heart rate reading. |
FDA Determined Cause 2 | Other |
| Action | On 12/22/04 distribution of new units was placed on hold and customers were notified to recommend that they do not use the pace detect function. On 12/27/03 an email was sent to customers of the solution. On 12/29/04 a Technical Worksheet describing how to replace a resistor at R8 was emailed to customers. Between 12/29/04 and 01/03/05 kits containing instructions and new resistors were sent to customers to begin reworking of the devices. |
| Quantity in Commerce | 11 |
| Distribution | Internationally only. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = DPS 510(K)s with Product Code = DPS
|
|
|
|