| | Class 1 Device Recall |  |
| Date Initiated by Firm | June 27, 2005 |
| Date Posted | September 28, 2005 |
| Recall Status1 |
Terminated 3 on August 21, 2006 |
| Recall Number | Z-1459-05 |
| Recall Event ID |
32513 |
| 510(K)Number | K961847 |
| Product Classification |
Prosthesis, Vascular Graft, Of 6mm And Greater Diameter - Product Code DSY
|
| Product | Hemashield Vantage Vascular Graft. Vascular graft prosthesis. |
| Code Information |
Catalog Number, Product Description, Lot Number: M002023750060, HVVG 6mm x 100cm, 5649003; M002023750060, HVVG 6mm x 100cm, 5948031; M002023750060, HVVG 6mm x 100cm, 6639821; M002023750060, HVVG 6mm x 100cm, 6647296; M002023750060, HVVG 6mm x 100cm, 7030066; M002023750070, HVVG 7mm x 100cm, 5669601; M002023750070, HVVG 7mm x 100cm, 5728283; M002023750070, HVVG 7mm x 100cm, 5821740; M002023750070, HVVG 7mm x 100cm, 6032950; M002023750070, HVVG 7mm x 100cm, 6962284; M002023750080, HVVG 8mm x 100cm, 6007978; M002023750080, HVVG 8mm x 100cm, 6279546; M002023750080, HVVG 8mm x 100cm,6722180; M002023750080, HVVG 8mm x 100cm, 6919131; M002023750080, HVVG 8mm x 100cm, 7161034; M002023750080, HVVG 8mm x 100cm, 7249926; M002023750080, HVVG 8mm x 100cm, 7402224; M002023750100, HVVG 10mm x 100cm, 6876321; M002023750100, HVVG 10mm x 100cm, 6962286; M002023754060, HVVG 6mm x 40cm, 5721446; M002023754060, HVVG 6mm x 40cm, 5728284; M002023754060, HVVG 6mm x 40cm, 5780318; M002023754060, HVVG 6mm x 40cm, 5860237; M002023754060, HVVG 6mm x 40cm, 5899944; M002023754060, HVVG 6mm x 40cm, 6011811; M002023754060, HVVG 6mm x 40cm, 6239445; M002023754060, HVVG 6mm x 40cm, 6268215; M002023754060, HVVG 6mm x 40cm, 6304504; M002023754060, HVVG 6mm x 40cm, 6320785; M002023754060, HVVG 6mm x 40cm, 6364795; M002023754060, HVVG 6mm x 40cm, 6440114; M002023754060, HVVG 6mm x 40cm, 6615660; M002023754060, HVVG 6mm x 40cm, 6628696; M002023754060, HVVG 6mm x 40cm, 6647104; M002023754060, HVVG 6mm x 40cm, 6696492; M002023754060, HVVG 6mm x 40cm, 6807737; M002023754060, HVVG 6mm x 40cm, 6917823; M002023754060, HVVG 6mm x 40cm, 6979854; M002023754060, HVVG 6mm x 40cm, 7000352; M002023754060, HVVG 6mm x 40cm, 7071546; M002023754060, HVVG 6mm x 40cm, 7200982; M002023754070, HVVG 7mm x 40cm, 5691792; M002023754070, HVVG 7mm x 40cm, 5726049; M002023754070, HVVG 7mm x 40cm, 5779847; M002023754070, HVVG 7mm x 40cm, 5797094; M002023754070, HVVG 7mm x 40cm, 5955887; M002023754070, HVVG 7mm x 40cm, 6011812; M002023754070, HVVG 7mm x 40cm, 6037086; M002023754070, HVVG 7mm x 40cm, 6136721; M002023754070, HVVG 7mm x 40cm, 6611338; M002023754070, HVVG 7mm x 40cm, 6657664; M002023754070, HVVG 7mm x 40cm, 6691603; M002023754070, HVVG 7mm x 40cm, 6851214; M002023754070, HVVG 7mm x 40cm, 7161578; M002023754070, HVVG 7mm x 40cm, 7175592; M002023754070, HVVG 7mm x 40cm, 7249015; M002023754070, HVVG 7mm x 40cm, 7402368; M002023754070, HVVG 7mm x 40cm, 7403197; M002023754070, HVVG 7mm x 40cm, 7535410; M002023754070, HVVG 7mm x 40cm, 7565108; M002023754070, HVVG 7mm x 40cm, 7631134; M002023754080, HVVG 8mm x 40cm, 5669605; M002023754080, HVVG 8mm x 40cm, 5728288; M002023754080, HVVG 8mm x 40cm, 5783271; M002023754080, HVVG 8mm x 40cm, 5797165; M002023754080, HVVG 8mm x 40cm, 5841181; M002023754080, HVVG 8mm x 40cm, 5912738; M002023754080, HVVG 8mm x 40cm, 5976346; M002023754080, HVVG 8mm x 40cm, 6008050; M002023754080, HVVG 8mm x 40cm, 6107641; M002023754080, HVVG 8mm x 40cm, 6138907; M002023754080, HVVG 8mm x 40cm, 6238549; M002023754080, HVVG 8mm x 40cm, 6320790; M002023754080, HVVG 8mm x 40cm, 6361655; M002023754080, HVVG 8mm x 40cm, 6427209; M002023754080, HVVG 8mm x 40cm, 6598660; M002023754080, HVVG 8mm x 40cm, 6617325; M002023754080, HVVG 8mm x 40cm, 6646329; M002023754080, HVVG 8mm x 40cm, 6647152; M002023754080, HVVG 8mm x 40cm, 6691608; M002023754080, HVVG 8mm x 40cm, 6876382; M002023754080, HVVG 8mm x 40cm, 7114368; M002023754080, HVVG 8mm x 40cm, 7138064; M002023754080, HVVG 8mm x 40cm, 7161580; M002023754080, HVVG 8mm x 40cm, 7175587; M002023754080, HVVG 8mm x 40cm, 7276583; M002023754080, HVVG 8mm x 40cm, 7402360; M002023754080, HVVG 8mm x 40cm, 7535409; M002023754100, HVVG 10mm x 40cm, 5691791; M002023754100, HVVG 10mm x 40cm, 5731519; M002023754100, HVVG 10mm x 40cm, 5891897; M002023754100, HVVG 10mm x 40cm, 6052007; M002023754100, HVVG 10mm x 40cm, 6418381; M002023754100, HVVG 10mm x 40cm, 6926326; M002023754100, HVVG 10mm x 40cm, 7000766; M002023754100, HVVG 10mm x 40cm, 7504489; M002023756060, HVVG 6mm x 60cm, 5707200; M002023756060, HVVG 6mm x 60cm, 5724793; M002023756060, HVVG 6mm x 60cm, 5745982; M002023756060, HVVG 6mm x 60cm, 5767491; M002023756060, HVVG 6mm x 60cm, 5767495; M002023756060, HVVG 6mm x 60cm, 5852837; M002023756060, HVVG 6mm x 60cm, 5853224; M002023756060, HVVG 6mm x 60cm, 5891798; M002023756060, HVVG 6mm x 60cm, 5947960; M002023756060, HVVG 6mm x 60cm, 5968420; M002023756060, HVVG 6mm x 60cm, 5988577; M002023756060, HVVG 6mm x 60cm, 6050593; M002023756060, HVVG 6mm x 60cm, 6096033; M002023756060, HVVG 6mm x 60cm, 6268220; M002023756060, HVVG 6mm x 60cm, 6347541; M002023756060, HVVG 6mm x 60cm, 6388858; M002023756060, HVVG 6mm x 60cm, 6418565; M002023756060, HVVG 6mm x 60cm, 6611350; M002023756060, HVVG 6mm x 60cm, 6628703; M002023756060, HVVG 6mm x 60cm, 6654862; M002023756060, HVVG 6mm x 60cm, 6655453; M002023756060, HVVG 6mm x 60cm, 6770384; M002023756060, HVVG 6mm x 60cm, 6807126; M002023756060, HVVG 6mm x 60cm, 6850094; M002023756060, HVVG 6mm x 60cm, 6919450; M002023756060, HVVG 6mm x 60cm, 7010071; M002023756060, HVVG 6mm x 60cm, 7057154; M002023756060, HVVG 6mm x 60cm, 7075488; M002023756060, HVVG 6mm x 60cm, 7161035; M002023756060, HVVG 6mm x 60cm, 7173034; M002023756060, HVVG 6mm x 60cm, 7230491; M002023756060, HVVG 6mm x 60cm, 7259726; M002023756060, HVVG 6mm x 60cm, 7528090; M002023756070, HVVG 7mm x 60cm, 5633991; M002023756070, HVVG 7mm x 60cm, 5691790; M002023756070, HVVG 7mm x 60cm, 5745981; M002023756070, HVVG 7mm x 60cm, 5773786; M002023756070, HVVG 7mm x 60cm, 5793283; M002023756070, HVVG 7mm x 60cm, 5853511; M002023756070, HVVG 7mm x 60cm, 5903684; M002023756070, HVVG 7mm x 60cm, 5988575; M002023756070, HVVG 7mm x 60cm, 6011813; M002023756070, HVVG 7mm x 60cm, 6037090; M002023756070, HVVG 7mm x 60cm, 6109301; M002023756070, HVVG 7mm x 60cm, 6282450; M002023756070, HVVG 7mm x 60cm, 6383841; M002023756070, HVVG 7mm x 60cm, 6440361; M002023756070, HVVG 7mm x 60cm, 6611337; M002023756070, HVVG 7mm x 60cm, 6667242; M002023756070, HVVG 7mm x 60cm, 6754019; M002023756070, HVVG 7mm x 60cm, 6930145; M002023756070, HVVG 7mm x 60cm, 7137399; M002023756070, HVVG 7mm x 60cm, 7276579; M002023756070, HVVG 7mm x 60cm, 7403200; M002023756070, HVVG 7mm x 60cm, 7545048; M002023756070, HVVG 7mm x 60cm, 7545052; M002023756070, HVVG 7mm x 60cm, 7579389; M002023756080, HVVG 8mm x 60cm, 5633990; M002023756080, HVVG 8mm x 60cm, 5726141; M002023756080, HVVG 8mm x 60cm, 5856535; M002023756080, HVVG 8mm x 60cm, 5891889; M002023756080, HVVG 8mm x 60cm, 5948034; M002023756080, HVVG 8mm x 60cm, 6073565; M002023756080, HVVG 8mm x 60cm, 6185146; M002023756080, HVVG 8mm x 60cm, 6339117; M002023756080, HVVG 8mm x 60cm, 6388288; M002023756080, HVVG 8mm x 60cm, 6415673; M002023756080, HVVG 8mm x 60cm, 6598671; M002023756080, HVVG 8mm x 60cm, 6610550; M002023756080, HVVG 8mm x 60cm, 6628701; M002023756080, HVVG 8mm x 60cm, 6722144; M002023756080, HVVG 8mm x 60cm, 6754031; M002023756080, HVVG 8mm x 60cm, 6820115; M002023756080, HVVG 8mm x 60cm, 6861408; M002023756080, HVVG 8mm x 60cm, 6926327; M002023756080, HVVG 8mm x 60cm, 6948142; M002023756080, HVVG 8mm x 60cm, 7008949; M002023756080, HVVG 8mm x 60cm, 7071547; M002023756080, HVVG 8mm x 60cm, 7115438; M002023756080, HVVG 8mm x 60cm, 7528226; M002023756100, HVVG 10mm x 60cm, 5856641; M002023756100, HVVG 10mm x 60cm, 6051871; M002023756100, HVVG 10mm x 60cm, 6193600; M002023756100, HVVG 10mm x 60cm, 6524295; M002023756100, HVVG 10mm x 60cm, 6620113; M002023756100, HVVG 10mm x 60cm, 6691596; M002023756100, HVVG 10mm x 60cm, 6960306; M002023756100, HVVG 10mm x 60cm, 6961911; M002023758060, HVVG 6mm x 80cm, 5669380; M002023758060, HVVG 6mm x 80cm, 5779844; M002023758060, HVVG 6mm x 80cm, 5922480; M002023758060, HVVG 6mm x 80cm, 5964495; M002023758060, HVVG 6mm x 80cm, 6032948; M002023758060, HVVG 6mm x 80cm, 6216685; M002023758060, HVVG 6mm x 80cm, 6304505; M002023758060, HVVG 6mm x 80cm, 6347707; M002023758060, HVVG 6mm x 80cm, 6364796; M002023758060, HVVG 6mm x 80cm, 6388362; M002023758060, HVVG 6mm x 80cm, 6416301; M002023758060, HVVG 6mm x 80cm, 6615663; M002023758060, HVVG 6mm x 80cm, 6628554; M002023758060, HVVG 6mm x 80cm, 6646521; M002023758060, HVVG 6mm x 80cm, 6654376; M002023758060, HVVG 6mm x 80cm, 6696487; M002023758060, HVVG 6mm x 80cm, 6722189; M002023758060, HVVG 6mm x 80cm, 6774183; M002023758060, HVVG 6mm x 80cm, 6816154; M002023758060, HVVG 6mm x 80cm, 6930142; M002023758060, HVVG 6mm x 80cm, 7000351; M002023758060, HVVG 6mm x 80cm, 7010061; M002023758060, HVVG 6mm x 80cm, 7161036; M002023758060, HVVG 6mm x 80cm, 7230490; M002023758060, HVVG 6mm x 80cm, 7630848; M002023758070, HVVG 7mm x 80cm, 5724827; M002023758070, HVVG 7mm x 80cm, 5728287; M002023758070, HVVG 7mm x 80cm, 5955888; M002023758070, HVVG 7mm x 80cm, 6050597; M002023758070, HVVG 7mm x 80cm, 6230257; M002023758070, HVVG 7mm x 80cm, 6288273; M002023758070, HVVG 7mm x 80cm, 6440359; M002023758070, HVVG 7mm x 80cm, 6691602; M002023758070, HVVG 7mm x 80cm, 6770192; M002023758070, HVVG 7mm x 80cm, 6851207; M002023758070, HVVG 7mm x 80cm, 7161577; M002023758070, HVVG 7mm x 80cm, 7175590; M002023758070, HVVG 7mm x 80cm, 7248691; M002023758070, HVVG 7mm x 80cm, 7402229; M002023758070, HVVG 7mm x 80cm, 7402371; M002023758070, HVVG 7mm x 80cm, 7631133; M002023758080, HVVG 8mm x 80cm, 5700634; M002023758080, HVVG 8mm x 80cm, 5765239; M002023758080, HVVG 8mm x 80cm, 5815016; M002023758080, HVVG 8mm x 80cm, 5911567; M002023758080, HVVG 8mm x 80cm, 5972377; M002023758080, HVVG 8mm x 80cm, 6068815; M002023758080, HVVG 8mm x 80cm, 6238548; M002023758080, HVVG 8mm x 80cm, 6256385; M002023758080, HVVG 8mm x 80cm, 6320652; M002023758080, HVVG 8mm x 80cm, 6347706; M002023758080, HVVG 8mm x 80cm, 6416627; M002023758080, HVVG 8mm x 80cm, 6598409; M002023758080, HVVG 8mm x 80cm, 6617271; M002023758080, HVVG 8mm x 80cm, 6647222; M002023758080, HVVG 8mm x 80cm, 6691605; M002023758080, HVVG 8mm x 80cm, 6876377; M002023758080, HVVG 8mm x 80cm, 6962282; M002023758080, HVVG 8mm x 80cm, 7114366; M002023758080, HVVG 8mm x 80cm, 7138066; M002023758080, HVVG 8mm x 80cm, 7161579; M002023758080, HVVG 8mm x 80cm, 7220186; M002023758080, HVVG 8mm x 80cm, 7276584; M002023758080, HVVG 8mm x 80cm, 7317245; M002023758080, HVVG 8mm x 80cm, 7402219; M002023758100, HVVG 10mm x 80cm, 5783268; M002023758100, HVVG 10mm x 80cm, 5891786; M002023758100, HVVG 10mm x 80cm, 6165924; M002023758100, HVVG 10mm x 80cm, 6524301; M002023758100, HVVG 10mm x 80cm, 6598574; M002023758100, HVVG 10mm x 80cm, 7031876; M002023758100, HVVG 10mm x 80cm, 7317854; M002023758100, HVVG 10mm x 80cm, 7504480. |
| FEI Number |
3001451463
|
Recalling Firm/ Manufacturer |
Boston Scientific Corporation
1 Boston Scientific Pl
Natick
MA
01760-1536
|
| For Additional Information Contact | Jennifer Bolton 763-494-1163 |
Manufacturer Reason for Recall | During surgery, several reports of the grafts fraying and tearing upon suturing by the surgeon. |
FDA Determined Cause 2 | Other |
| Action | Recall notifications were sent out to all customers on 6/27/2005. Enclosed within the recall notification letters are Dear Doctor Letters with directions to the customers to supplyh a copy of it to each physician who has implanted the Hemashield Vantage Vascular Graft. |
| Quantity in Commerce | 3487 devices |
| Distribution | The devices were distributed to 375 customers in the US, Argentina, Australia, Belgium, China, Colombia, Czech Republic, Germany, Denmark, Spain, France, Great Britain, Italy, Kuwait, Lebanon, Mexico, Malaysia, Netherlands, Norway, New Zealand, Poland, Portugal, Sweden, Singapore and Taiwan. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = DSY
|
|
|
|