| | Class 2 Device Recall Respironics PLV Continuum |  |
| Date Initiated by Firm | November 18, 2005 |
| Date Posted | January 06, 2006 |
| Recall Status1 |
Terminated 3 on April 22, 2008 |
| Recall Number | Z-0354-06 |
| Recall Event ID |
34129 |
| 510(K)Number | K022750 |
| Product Classification |
Ventilator, Continuous, Facility Use - Product Code CBK
|
| Product | The PL V Continuum ventilator is a microprocessor controlled, compressorbased, mechanical ventilator. It is intended to control or assist breathing by delivering room air to the patient. PL V Continuum utilizes an internal compressor to generate compressed air for delivery to the patient. Breath delivery is controlled by software algorithms. The user interface on PL V Continuum has a membrane keypad with indicator Light Emitting Diodes (LED) for the selection and acceptance of patient settings and for the display of alarm conditions. PL V Continuum is capable of providing the following types of ventilatory support:
. Positive Pressure Ventilation, delivered either invasively (via endotracheal or tracheostomy tube) or non-invasively (via mask or mouthpiece).
. Assist/Control, Spontaneous Intermittent Mandatory Ventilation (SIMV) or Continuous Positive Airway Pressure (CP AP) modes of ventilation.
. Volume-Controlled (VC). Available in AlC and SIMV.
. Pressure-Controlled (PC). Available in AlC and SIMV.
. Pressure Support (PS). Available in SIMV and SPONT. |
| Code Information |
Domestic Distribution Serial Numbers: VS4700581, VS4700661, VS4700631, VS4702578, VS4702507, VS4702551, VS4702583, VS4702589, VS4702590, VS470259], VS4702574, VS4702580, VS4702566, VS4700824, VS4702593, VS4700835, VS4700536, 1VS4700539, VS4700540, VS4700543, VS4700544, S4700546, VS4700548, VS4700549, VS4700550, VS4700552,VS4700554, VS4700555, VS4700558, VS4700565, VS4700566,VS4700567, VS4700569, VS4700573, VS4700574, VS4700577,VS4700578, VS4700580, VS4700581, VS4700585, VS4700588VS4700590, VS4700593, VS4700594, VS4700595, VS4700596, VS4700605, VS4700607, VS4700608, VS4700613, VS4700614, VS4700615, VS4700617, VS4700620, VS4700622, VS4700624, VS4700625, VS4700632, VS4700635, VS4700637, VS4700642, VS4700643, VS4700646, VS4700647, VS4700649, VS4700650, VS4700653, VS4700660, VS4700661, VS4700663, VS4700668, VS4700670, VS4700673, VS4700681, VS4700687, VS4700693, VS4700694, VS4700698, VS4700699, VS4700708, VS470071O, VS4700711, VS4700718, VS4700719, VS4700721, VS4700724, VS4700725, VS4700729, VS4700730, VS4700736, VS4700738, VS4700739, VS4700740, VS4700743, VS4700744, VS4700750, VS4700751, VS4700754, VS4700760, VS4700761, VS4700762, VS4700767, VS4700770, VS4700772, VS4700773, VS4700774, VS4700775, VS4700776, VS4700777, VS4700780, VS4700781, VS4700782, VS4700783, VS4700785, VS4700786, VS4700789, VS4700790, VS4700791, VS4700796, VS4700797, VS4700798, VS4700802, VS4700804, VS4700808, VS4700809, VS4700811, VS4700813, VS4700814, VS4700816, VS4700817, VS4700818, VS4700819, VS4700821, VS4700822, VS4700823, VS4700827, VS4700828, VS4700829, VS4700832, VS4700833, VS4700834, VS4700836, VS4700837, VS4700838, VS4700840, VS4700841, VS4700842, VS4700843, VS4700844, VS4700846, VS4700847, VS4700849, VS4700850, VS4700852, VS4700853, VS4700854, VS4700855, VS4700856, VS4702501, VS4702502, VS4702505, VS4702512, VS4702516, VS4702519, VS4702523, VS4702524, VS4702525, VS4702528, VS4702531, VS4702532, VS4702535, VS4702543, VS4702544, VS4702545, VS4702548, VS4702549, VS4702552, VS4702553, VS4702555, VS4702556, VS4702557, VS4702567, VS4702568, VS4702570, VS4702571, VS4702572, VS4702573, VS4702575, VS4702586, VS4702594, VS4702599, VS4702600, VS4702604, VS4702612, VS4702613, VS4702623, VS4702636, VS4702647, VS4702650, VS4702670, VS4702677, VS4702688, VS4702689, VS4702709, VS4702734, VS4702747, VS4702748, VS4702750, VS4702776, VS4702784, VS4702800, VS4702805, VS4702809, VS4702811, VS4702813, VS4702825, VS4702835, VS4702848, VS4702854, VS4702856, VS4702857, VS4702859, VS4702883, VS4702885, VS4702981, VS4703079, VS4703103 International Distribution Serial Numbers: VS4702605, VS4702608, VS4702609, VS4702631, VS4702632, VS4702634, VS4702642, VS4702643, VS4702659, VS4702719, VS4702754, VS4702759, VS4700792, VS4702722, VS4700563, VS4700696, VS4700820, VS4700601, VS4702645, VS4700839, VS4700644, VS4700589, VS4700703, VS4700723, VS4700746, VS4700553, VS4700628, VS4702962, VS4702983, VS4702995, VS4703006, VS4703023, VS4703041, VS4703056, VS4703064, VS4700717, VS4700645, VS4702915, VS4702500, VS4702828, VS4702858 |
| FEI Number |
2518422
|
Recalling Firm/ Manufacturer |
Respironics California Inc
2271 Cosmos Ct
Carlsbad
CA
92009-1517
|
| For Additional Information Contact | Anil Bhalani 760-918-7458 |
Manufacturer Reason for Recall | During product testing of the PLV Continuum Ventilator's Safety Valve/Pressure Relief Valve (SV/PRV) at our manufacturing facility, Respironics found that when the PLY Continuum Ventilator is operated in the 'Face Up' orientation, the SV /PRV may stick in the open position, which may result in loss of delivered gas volume to the patient.
|
FDA Determined Cause 2 | Other |
| Action | On November 18, 2005 all affected customers received a personal phone call from Respironics explaining the recall situation in detail. On November 21, 2005, an advisory notice was sent to customers via fax, email, or mail, instructing customers to only use your PLV Continuum Ventilator in the 'Face Foward' orientation.
The firm will contact customers with further information shortly to assure that this situation is positively resolved.
Customers were instructed to Only use the PL V Continuum Ventilator in the 'Face Forward' orientation.
In the event that a customer requires the application of the PLV Continuum Ventilator in the Face Up Orientation, Respironics will provide the customer, at no charge, with a rental PLV-l00, PLV-IO2b or PLV-I02 Ventilator for the customer to use until their PLV Continuum Ventilator is upgraded.
If additional information is needed the customer was instructed to contact Respironics'' Customer Service at 800 345-6443.
|
| Quantity in Commerce | 268 |
| Distribution | Nationwide, Taiwan, Saudi Arabia, Netherlands, Hong Kong, Canada, Brazil, Australia, and Argentina. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = CBK
|
|
|
|