| | Class 2 Device Recall Bayer Advia 70 Hematology Analyzer |  |
| Date Initiated by Firm | September 20, 2006 |
| Date Posted | January 05, 2007 |
| Recall Status1 |
Terminated 3 on March 10, 2008 |
| Recall Number | Z-0450-2007 |
| Recall Event ID |
36728 |
| 510(K)Number | K991539 |
| Product Classification |
Hematology Analyzer - Product Code GKZ
|
| Product | Bayer Advia 70 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX. |
| Code Information |
All lots. |
| FEI Number |
3003071743
|
Recalling Firm/ Manufacturer |
Drew Scientific Group PLC
4230 Shilling Way
Dallas
TX
75237-1023
|
Manufacturer Reason for Recall | System has given inaccurate readings when running samples that trigger an extra rinse due to very high WBC, RBC or Platlet counts. |
FDA Determined Cause 2 | Other |
| Action | Firm sent 'Technical Bulletin #39' to all customers on 09/20/06. Users advised of potential for inaccurate readings and provided with instructions on how to eliminate problem. |
| Quantity in Commerce | 545 units. |
| Distribution | Worldwide distribution ---- including USA states of FL, MA, MN, ND, NM, NY, Puerto Rico and TX and countries of Austria, Bangladesh, Chile, China, Czech Republic, Egypt, France, Hong Kong, Italy, India, Iran, Korea, Mexico, Morocco, Phillipines, Portugal, Russia, Serbia and Montenegro, Spain, Syria, Thailand, Turkey, United Arab Emerates (Dubai), United Kingdom and Vietnam, West Indies and Yemen. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
| 510(K) Database | 510(K)s with Product Code = GKZ
|
|
|
|