 Class 2 Device Recall Lumexsee related information
Date Initiated by FirmApril 02, 2007
Date PostedMay 05, 2007
Recall Status1 Terminated 3 on January 12, 2010
Recall NumberZ-0772-2007
Recall Event ID 37747
Product Classification Rollator - Product Code ITJ
ProductLumex Walkabout Lite Four-Wheel Rollator, Manufactured for: Graham-Field Health Products, Atlanta, Georgia 30360.
Code Information Model numbers: RJ4300B, RJ4300D, RF4300G, RJ4300K, RJ4300R, and Imperial Model RF4400K. (Note: Models distributed after January 2007, serial numbers 120791 through 145254, are not subject to this correction)
FEI Number 2427360
Recalling Firm/
Manufacturer
Graham Field Atlanta
2935 Northeast Pkwy
Atlanta GA 30360-2808
For Additional Information ContactKatie Johnson
678-291-3166
Manufacturer Reason
for Recall
The caster bolt stem/screw may loosen over time causing the wheel to come off.
FDA Determined
Cause 2
Other
ActionGraham-Field notified all of its Durable Medical Equipment (DME) dealers involved in the distribution of these products by certified US Mail on April 3, 2007. They were made aware of the need to adjust the product to lessen the risk of injury and were provided a re-work kit. The distribution of DME dealers includes the announcement of the field correction to the DME dealer, the letter to send to the end-users, rework supplies, rework instructions and proof of compliance cards. Graham-Field instructed the DME dealers to adjust any impacted products on site, as well as, any delivered by end-users, customers or other impacted third parties to the DME dealer for correction. DME dealers will be compensated for performing the corrective action. The installation of the rework kit involves the installation of lock washers and Locitite and is described in more detail in the attached rework instructions. Graham-Field does not have direct access to end user names, however it plans to reach end-users by financially supporting a certified mailing from DME dealers to their individual customers notifying them of this field correction. A follow-up 60 day certified mailing will be performed for non-responders.
Quantity in Commerce13,300 units
DistributionU.S. Nationwide, Costa Rica, Ecuador, El Salvador, Guatemala, and Santo Domingo
Total Product Life CycleTPLC Device Report