| | Class 2 Device Recall King Systems, DatexOhmeda, Universal Flex2 Breathing Circuit |  |
| Date Initiated by Firm | August 01, 2008 |
| Date Posted | February 03, 2009 |
| Recall Status1 |
Terminated 3 on November 12, 2009 |
| Recall Number | Z-0675-2009 |
| Recall Event ID |
48991 |
| Product Classification |
Breathing Circuit - Product Code CAI
|
| Product | King Systems, Datex-Ohmeda, Universal Flex2 Breathing Circuit, Latex Free, 50 per case, King Systems Corporation, Noblesville, IN; REF # CP-05204.
This breathing circuit is a device intended to administer medical gases and/or anesthetic gases to a patient during anesthesia for inhalation, or medical gases to a patient during respiratory care inhalation. |
| Code Information |
Lots IKAM2, IKED8, IKG46. IKK37, IKMB8, IKP25, IKRG9, IKT44, IKWR0, IL077, IL2P6, IL480 and IL6J4. |
| FEI Number |
1824226
|
Recalling Firm/ Manufacturer |
King Systems Corp.
15011 Herriman Blvd
Noblesville
IN
46060
|
| For Additional Information Contact | 317-776-6823 |
Manufacturer Reason for Recall | The inner line may be severed or broken into two pieces, thus hindering the delivery of oxygen and anesthetics to the patient and possibly elevated patient carbon dioxide levels. |
FDA Determined Cause 2 | Pending |
| Action | Consignees were notified by an Urgent Product Recall letter on or about 7/29/2008 to cease sales, notify their customers to the hospital level, and to return product on hand. |
| Quantity in Commerce | Approx. 500,000 units of all products. |
| Distribution | Worldwide Distribution - including USA and countries of Australia, Colombia, Germany, Netherlands, and Peru. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|
|
|
|