| | Class 2 Device Recall LIFEPAK 12 DEFIBRILLATOR/MONITOR |  |
| Date Initiated by Firm | March 31, 2009 |
| Date Posted | February 22, 2010 |
| Recall Status1 |
Terminated 3 on August 08, 2013 |
| Recall Number | Z-0811-2010 |
| Recall Event ID |
54005 |
| 510(K)Number | K063510 |
| Product Classification |
automated external defibrillators (non-wearable) - Product Code MKJ
|
| Product | Biphasic LIFEPAK 12 defibrillator / monitor. |
| Code Information |
Part Number: LP12, VLP12-02-(various dash numbers). Serial Numbers (non-contiguous): 8529728, 8529728, 9210418, 9210419, 9599334, 10128750, 10132851, 10248200, 10389064, 11120818, 11244332, 11252979, 11252980, 11252981, 11289300, 11311777, 11311778, 11311782, 11311787, 11316922, 11372139, 11424301, 11424819, 11471546, 11568711, 11610477, 11766215, 11766216, 11766222, 11767531, 11767565, 11877154, 12007118, 12007124, 12007129, 12007131, 12007135, 12009524, 12009528, 12030152, 12209828, 12209830, 12209832, 12209834, 12551932, 12551934, 12551936, 12551937, 12551939, 12776733, 12776734, 12779439, 12779439, 12789243, 13165363, 13269232, 13423039, 13434816, 13434837, 13522303, 13536842, 13704769, 13704770, 13870608, 13993048, 14008074, 14008075, 14201185, 14201186, 14201187, 30646561, 32117838, 32117838, 32117839, 32117839, 32117840, 32117840, 37125611, 37125612, 37125617, 37125743, 37125751, 37125752, 37125753, 37126327, 37126328, 37126329, 37126330, 37126331, 37126332, 37126333, 37126334, 37126335, 37126336, 37126342, 37126369, 37128467, 37128482, 37128485, 37128642, 37128643, 37128644, 37128646, 37128651, 37128652, 37128653, 37128660, 37128661, 37128662, 37128663, 37128666, 37128667, 37128668, 37128669, 37128670, 37128674, 37128678, 37135341, 37135394, 37135396, 37135397, 37135398, 37135418, 37135423, 37135424, 37135425, 37136603, 37136622, 37136623, 37136670, 37138749, 37138751, 37138756, 37138759, 37138769, 37138777, 37139181, 37139182, 37139183, 37139184, 37139186, 37139187, 37139188, 37139189, 37139190, 37139191, 37139192, 37139193, 37139194, 37139202, 37139203, 37139204, 37139205, 37139206, 37139207, 37139252, 37139254, 37139315, 37139317, 37139318, 37139324, 37139932, 37139933, 37140146, 37140147, 37143216, 37143418, 37143420, 37143457, 37143531, 37143636, 37146769, 37148809, 37148812, 37148863, 37148863, 37148864, 37148864, 37148865, 37148865, 37148867, 37148868, 37148868, 37148872, 37148878, 37148881, 37148881, 37148882, 37148882, 37154928, 37154936, 37158763, 37158766, 37158767, 37158768, 37158772, 37158773, 37158774, 37159196, 37159205, 37159231, 37159249, 37159250, 37159251, 37159264, 37159272, 37159273, 37159329, 37159330, 37159334, 37159337, 37159338, 37159339, 37159340, 37159341, 37170142, 37170143, 37170147, 37170149, 37170152, 37170750, 37170993, 37178695, 37178738, 37178742, 37190161, 37190162, 37190163, 37190165, 37190166, 37190167, 37190168, 37190169, 37190172, 37190173, 37190174, 37190175, 37190176, 37190177, 37190178, 37190179, 37190180, 37190181, 37190182, 37190183, 37190184, 37190185, 37190186, 37190187, 37190979, 37190981, 37193009, 37193409, 37193415, 37193417, 37193419, 37193420, 37193424, 37193425, 37193434, 37193435, 37193438, 37193439, 37193509, 37193512, 37193513, 37193514, 37193515, 37193516, 37193517, 37193518, 37193518, 37193519, 37193522, 37193527, 37193528, 37193529, 37193530, 37193531, 37193643, 37193650, 37193660, 37193716, 37193717, 37193719, 37193750, 37193753, 37193754, 37193755, 37193757, 37193758, 37193759, 37193760, 37193761, 37193776, 37193781, 37193783, 37193786, 37193788, 37193790, 37193791, 37200151, 37205122, 37205127, 37205127, 37205175, 37205175, 37205178, 37205185, 37205189, 37205190, 37205194, 37205198, 37205297, 37205298, 37205299, 37205301, 37205302, 37205303, 37205304, 37205305, 37205306, 37205307, 37205308, 37205309, 37205310, 37205311, 37205312, 37205313, 37205350, 37205591, 37205594, 37205594, 37205597, 37205597, 37205598, 37205598, 37207554, 37207577, 37207578, 37207579, 37207580, 37207581, 37207582, 37207583, 37207584, 37207585, 37207586, 37207587, 37207588, 37207589, 37207590, 37207591, 37207592, 37207690, 37208504, 37208516, 37208519, 37208520, 37208521, 37208524, 37208527, 37208563, 37208564, 37208565, 37208566, 37208567, 37208568, 37208569, 37208570, 37208654, 37208655, 37208656, 37208712, 37208713, 37208714, 37208715, 37208716, 37208717, 37208722, 37208723, 37208724, 37208767, 37208768, 37211021, 37211041, 37211044, 37211045, 37211098, 37211099, 37211100, 37211386, 37211389, 37211392, 37211406, 37211414, 37211642, 37211643, 37211644, 37211645, 37211646, 37211647, 37211648, 37212070, 37212391, 37212409, 37212412, 37212414, 37212424, 37212429, 37212827, 37213069, 37213095, 37213129, 37213174, 37213188, 37213220, 37213249, 37213502, 37213503, 37213504, 37213505, 37213541, 37213542, 37213574, 37213575, 37213576, 37216010, 37216669, 37216670, 37216774, 37216775, 37216883, 37216884, 37216886, 37216905, 37216906, 37216907, 37216908, 37216911, 37216916, 37216934, 37217051, 37217190, 37217191, 37217192, 37217194, 37217198, 37217199, 37217203, 37219856, 37219882, 37219883, 37219884, 37219922, 37219924, 37219927, 37219936, 37219981, 37219991, 37219998, 37219999, 37220007, 37220008, 37220009, 37220010, 37220011, 37220024, 37220091, 37220092, 37220210, 37220210, 37220211, 37220211, 37223009, 37223012, 37223036, 37223092, 37223096, 37223098, 37223099, 37223100, 37223101, 37223102, 37223103, 37223104, 37223105, 37223106, 37223107, 37223108, 37223109, 37223110, 37223114, 37223115, 37223116, 37223118, 37223122, 37223123, 37223124, 37223192, 37223198, 37223199, 37223202, 37223203, 37223544, 37223545, 37224052, 37224063, 37224064, 37224134, 37224135, 37224142, 37224147, 37224148, 37224160, 37226315, 37226316, 37226330, 37226331, 37226332, 37226333, 37226334, 37226335, 37226336, 37226337, 37226338, 37226339, 37226348, 37226349, 37226350, 37226351, 37226352, 37226353, 37226410, 37226411, 37226412, 37226413, 37226416, 37226420, 37226511, 37231960, 37231961, 37232006, 37234694, 37234695, 37234696, 37234697, 37234698, 37234699, 37234700, 37234701, 37234702, 37234703, 37234704, 37234705, 37234706, 37234708, 37234709, 37234710, 37234711, 37234712, 37234746, 37235337, 37235338, 37235365, 37235368, 37237465, 37237552, 37237553, 37237554, 37237563, 37237571, 37237818, 37237821, 37237824, 37237825, 37240785, 37240792, 37240793, 37240794, 37240795, 37240796, 37240798, 37240799, 37240800, 37240801, 37240802, 37240803, 37240804, 37240805, 37240806, 37240808, 37240810, 37240811, 37240812, 37240813, 37241075, 37241087, 37241117, 37245319, 37245771, 37245772, 37245780, 37245781, 37246017, 37246018, 37246354, 37246355, 37246356, 37246493, 37246499, 37249345, 37250362, 37250365, 37250372, 37250375, 37250379, 37250382, 37259407, 37259408, 37259409, 37259410, 37259411, 37259412, 37259413, 37259414, 37259418, 37259420, 37259421, 37259422, 37259423, 37259424, 37259426, 37259427, 37259428, 37259429, 37259430, 37259434, 37259442, 37259450, 37259451, 37259455, 37259457, 37259460, 37259463, 37259644, 37259668, 37259933, 37260513, 37260525, 37260526, 37260527, 37260589, 37260590, 37260591, 37260594, 37260596, 37260600, 37260606, 37260617, 37260623, 37260630, 37263770, 37263771, 37263772, 37263773, 37263774, 37263874, 37263875, 37263876, 37263877, 37263878, 37263920, 37263922, 37263931, 37263941, 37263949, 37263950, 37263951, 37263952, 37263953, 37263954, 37263955, 37263956, 37264340, 37264341, 37264342, 37264345, 37264346, 37264347, 37264350, 37264351, 37264358, 37264365, 37264365, 37264374, 37264375, 37264376, 37264377, 37264378, 37264379, 37264698, 37264714, 37268732, 37268860, 37268861, 37268862, 37268863, 37268864, 37268865, 37268866, 37268878, 37268879, 37268880, 37268881, 37268882, 37268896, 37268898, 37268915, 37273926, 37274137, 37274138, 37274139, 37274140, 37274141, 37274142, 37274143, 37274144, 37274145, 37274146, 37274147, 37274148, 37274149, 37274150, 37274151, 37274152, 37274153, 37274154, 37274155, 37274156, 37274157, 37274158, 37274159, 37274160, 37274161, 37274162, 37274510, 37274511, 37274512, 37274513, 37274514, 37274515, 37274516, 37274517, 37275262, 37275272, 37275272, 37275273, 37275273, 37275274, 37275274, 37275275, 37275275, 37275282, 37275312, 37288407, 37288408, 37288409, 37288410, 37288411, 37288433, 37288453, 37288454, 37288455, 37288456, 37288474, 37288475, 37288478, 37288598, 37288603, 37288610, 37288614, 37288647, 37288649, 37288650, 37288653, 37288654, 37288655, 37288656, 37288657, 37288660, 37288661, 37288663, 37288664, 37288665, 37288666, 37288667, 37288668, 37288669, 37288670, 37288671, 37288672, 37288819, 37288821, 37288822, 37288823, 37288825, 37288826, 37288894, 37298226, 37298227, 37298229, 37298232, 37298236, 37298243, 37298246, 37305892, 37305897, 37305897, 37305901, 37305901, 37305902, 37305902, 37305903, 37305903, 37305904, 37305904, 37306051, 37306051, 37306052, 37306052, 37306053, 37306053, 37306054, 37306054, 37306058, 37306058, 37306059, 37306059, 37306455, 37306456, 37306457, 37306458, 37306459, 37306460, 37306461, 37306462, 37315264, 37357985, 37357986, 37357987, 37357988, 37357989, 37357990, 37357991, 37357992, 37357993, 37357994, 37358010, 37358011, 37358012, 37358020, 37358021, 37358026, 37358027, 37358028, 37358033, 37358035, 37358053, 37358066, 37358069, 37358070, 37358071, 37358074, 37358075, 37358082, 37358083, 37358116, 37358147, 37358150, 37358155, 37358158, 37358167, 37358204, 37358207, 37358235, 37358239, 37358241, 37358285, 37358286, 37358287, 37358449, 37358450, 37358451, 37358460, 37358463, 37358464, 37358465, 37358480, 37358481, 37358547, 37359127, 37359128, 37359129, 37359396, 37359410, 37359428, 37359487, 37359491, 37359495, 37359496, 37359497, 37359498, 37359499, 37359500, 37359501, 37359502, 37359503, 37359505, 37359506, 37359507, 37359508, 37359509, 37359510, 37359511, 37361533, 37361849, 37361883, 37361934, 37361935, 37361939, 37361940, 37361941, 37361947, 37362070, 37362071, 37362072, 37362265, 37362266, 37362270, 37362276, 37362281, 37362282, 37362286, 37362300, 37362301, 37362302, 37362302, 37362303, 37362304, 37362305, 37362941, 37362941, 37362942, 37362942, 37362968, 37362968, 37362976, 37362976, 37363070, 37363070, 37363071, 37363146, 37363213, 37363224, 37363224, 37363229, 37363231, 37363231, 37363234, 37363237, 37363249, 37363252, 37363456, 37363456, 37363465, 37363465, 37363466, 37363467, 37363467, 37363472, 37363486, 37363487, 37363494, 37363500, 37363629, 37363633, 37363635, 37363636, 37363637, 37363638, 37363650, 37365624, 37365631, 37365672, 37365676, 37365678, 37365679, 37365681, 37365682, 37365684, 37365698, 37365699, 37365704, 37365705, 37365706, 37365707, 37365708, 37365709, 37365712, 37365723, 37366277, 37366278, 37366282, 37366342, 37366816, 37366817, 37366984, 37366998, 37367004, 37367019, 37367025, 37367037, 37367063, 37367069, 37367328, 37367567, 37367582, 37367613, 37367615, 37367615, 37367621, 37367622, 37367623, 37367624, 37367625, 37370902, 37370903, 37370914, 37370915, 37370919, 37370922, 37370925, 37370928, 37370929, 37370930, 37370931, 37370942, 37371050, 37371092, 37371096, 37371099, 37371101, 37371104, 37371107, 37371116, 37371118, 37371122, 37371123, 37371130, 37371131, 37371133, 37371134, 37371134, 37371135, 37371135, 37371136, 37371137, 37371139, 37371154, 37371155, 37371159, 37371164, 37371165, 37371174, 37371178, 37371180, 37371500, 37371500, 37371547, 37371548, 37372195, 37372585, 37372586, 37372589, 37372592, 37372736, 37372941, 37372942, 37372943, 37372946, 37373092, 37373294, 37377735, 37377816, 37380563, 37380589, 37380590, 37380611, 37380612, 37380613, 37380614, 37380615, 37380616, 37380617, 37380664, 37380680, 37380681, 37381174, 37381175, 37381682, 37381690, 37381692, 37381693, 37381696, 37381697, 37381701, 37381702, 37381703, 37381704, 37381707, 37381708, 37381711, 37381714, 37381719, 37381722, 37381724, 37381730, 37381734, 37385933, 37386729, 37386730, 37386731, 37386732, 37398412, 37404258, 37404270, 37410107, 37410669, 37410737, 37426060, 37441282, 37475899, 37475900, 37488841, 37488908, 37488910, 37494355, 37494355, 37494356, 37494356, 37494447, 37494454, 37494458, 37494461, 37494462, 37494680, 37494682, 37494685, 37494689, 37494733, 37494736, 37494738, 37494741, 37494744, 37494747, 37494763, 37494809, 37494890, 37494964, 37494966, 37495059, 37495123, 37495131, 37495441, 37495483, 37495484, 37495486, 37495487, 37495488, 37499017, 37499017, 37499200, 37499201, 37499202, 37499203, 37499204, 37499205, 37499224, 37499345, 37499351, 37499497, 37499557, 37499657, 37499659, 37499661, 37502162, 37502164, 37502167, 37502168, 37502169, 37502170, 37502171, 37502894, 37502897, 37502898, 37502901, 37503118, 37503121, 37503122, 37503124, 37503126, 37503127, 37503128, 37503129, 37503130, 37503131, 37503133, 37503196, 37503244, 37503245, 37503246, 37503247, 37503248, 37503423, 37503423, 37505870, 37506943, 37506944, 37506945, 37506946, 37506947, 37506948, 37506953, 37506975, 37513568, 37513569, 37513596, 37513666, 37513669, 37513671, 37513673, 37513675, 37513677, 37513680, 37513685, 37513686, 37513687, 37513689, 37513698, 37513701, 37518148, 37518174, 37518175, 37518180, 37518181, 37518186, 37518189, 37518189, 37518190, 37518190, 37518198, 37518228, 37518229, 37518316, 37521298, 37521299, 37521304, 37521305, 37521311, 37521328, 37521330, 37521331, 37521332, 37521333, 37521334, 37521335, 37521336, 37521337, 37521338, 37521339, 37521340, 37521341, 37521343, 37521344, 37521345, 37521346, 37524484, 37524489, 37524491, 37524531, 37524538, 37524557, 37524559, 37524560, 37524563, 37524569, 37524575, 37524581, 37524582, 37524583, 37524636, 37524639, 37525432, 37525433, 37525434, 37525435, 37525437, 37525438, 37525442, 37525870, 37534823, 37534968, 37534969, 37534970, 37535205, 37535206, 37536810, 37536816, 37536831, 37536832, 37536971, 37537170, 37537175, 37539183, 37539184, 37539185, 37539486, 37539487, 37539488, 37540470, 37540471, 37540558, 37540841, 37540842, 37542689, 37542990, 37542994, 37542995, 37542996, 37543000, 37543001, 37543005, 37543038, 37543062, 37543065, 37543067, 37543077, 37543080, 37543081, 37543082, 37543083, 37543084, 37543087, 37543088, 37543090, 37543091, 37543093, 37543100, 37543106, 37543108, 37543231, 37543563, 37543564, 37543565, 37546843, 37546844, 37546845, 37546846, 37546847, 37546850, 37546851, 37546853, 37546854, 37546855, 37546857, 37546858, 37546859, 37546862, 37553469, 37553474, 37553475, 37553476, 37553482, 37554331, 37563996, 37563997, 37563998, 37563999, 37564505, 37564506, 37564507, 37564969, 37564969, 37565641, 37569249, 37569256, 37569257, 37569258, 37569259, 37569260, 37569261, 37569265, 37569266, 37569327, 37569341, 37569391, 37569689, 37573492, 37573493, 37574329, 37574330, 37574331, 37574333, 37574334, 37574336, 37574340, 37574341, 37584475, 37585125, 37585128, 37586088, 37586089, 37589197, 37589397, 37592029, 37592042, 37592042, 37592043, 37592043, 37592092, 37592093, 37592094, 37593313, 37593313, 37593314, 37593314, 37593316, 37593316, 37593318, 37593318, 37593319, 37593319, 37595719, 37595720, 37595721, 37595722, 37596072, 37596073, 37596074, 37596076, 37596077, 37596093, 37596094, 37596099, 37596100, 37596104, 37596106, 37596111, 37596115, 37596119, 37596120, 37596962, 37604552, 37604553, 37604554, 37604569, 37604582, 37604583, 37604584, 37604585, 37604586, 37604640, 37604661, 37604698, 37604739, 37604758, 37604778, 37604779, 37604789, 37604816, 37604823, 37604827, 37604834, 37604851, 37604852, 37604857, 37604873, 37605119, 37605120, 37605133, 37605134, 37605140, 37605144, 37605145, 37605145, 37605153, 37605219, 37607894, 37607897, 37607917, 37607918, 37607919, 37607921, 37607922, 37607923, 37607924, 37607925, 37607926, 37607927, 37607928, 37607930, 37607931, 37607932, 37607933, 37607934, 37607935, 37607936, 37607938, 37607939, 37607940, 37607941, 37607942, 37607943, 37607945, 37607946, 37607947, 37607949, 37607950, 37607951, 37607952, 37607953, 37607954, 37607955, 37607956, 37607957, 37607958, 37607959, 37607960, 37607961, 37607962, 37607963, 37607964, 37607965, 37607966, 37607967, 37607968, 37607969, 37607970, 37608253, 37608276, 37608277, 37608278, 37609291, 37609294, 37609303, 37611851, 37611857, 37611858, 37611859, 37611863, 37611867, 37611870, 37611871, 37611872, 37611877, 37611881, 37611907, 37611908, 37611909, 37611910, 37611911, 37611912, 37611913, 37611914, 37611915, 37611916, 37611917, 37611918, 37611933, 37612135, 37612136, 37613013, 37613014, 37613014, 37613015, 37613015, 37613016, 37613016, 37613017, 37613017, 37613018, 37613018, 37613019, 37613019, 37613031, 37613032, 37613033, 37613034, 37613035, 37613036, 37613037, 37613038, 37613039, 37613040, 37613043, 37613048, 37613049, 37613050, 37613051, 37613052, 37613055, 37613072, 37613109, 37613515, 37613515, 37613516, 37613516, 37613517, 37613517, 37613518, 37613518, 37613519, 37613519, 37613520, 37613520, 37616722, 37616723, 37616724, 37616725, 37616726, 37616727, 37616728, 37616729, 37616730, 37616731, 37620258, 37620259, 37620260, 37620261, 37620262, 37620263, 37620264, 37620265, 37620266, 37620267, 37620268, 37620269, 37620270, 37620271, 37620272, 37620273, 37620274, 37620275, 37620276, 37620277, 37620278, 37620279, 37620280, 37620281, 37620282, 37620283, 37620284, 37620285, 37620286, 37620287, 37620288, 37624306, 37624310, 37624311, 37624312, 37624315, 37624318, 37624613, 37624979, 37624979, 37625061, 37625062, 37625124, 37625125, 37625126, 37625127, 37625145, 37625146, 37625147, 37625161, 37625163, 37625167, 37628407, 37628550, 37628551, 37628552, 37628553, 37628554, 37628555, 37628556, 37628557, 37628558, 37628559, 37628560, 37628561, 37628562, 37628563, 37628564, 37628565, 37628566, 37628567, 37628568, 37628569, 37628570, 37628571, 37628572, 37628573, 37628574, 37628575, 37628576, 37628577, 37628578, 37628579, 37628580, 37628581, 37628582, 37628583, 37628584, 37628585, 37628586, 37628587, 37628588, 37628589, 37628590, 37628591, 37628592, 37628593, 37628594, 37628595, 37628596, 37628597, 37628598, 37628599, 37628600, 37628601, 37628602, 37628603, 37628604, 37628605, 37628606, 37628607, 37628608, 37628609, 37628610, 37628611, 37628612, 37628613, 37632499, 37632500, 37632501, 37632502, 37632503, 37632504, 37632918, 37632922, 37632941, 37632945, 37632946, 37632947, 37632948, 37632951, 37632952, 37632953, 37632955, 37632956, 37632958, 37632959, 37632960, 37632961, 37632963, 37632964, 37632967, 37633634, 37633636, 37636570, 37636571, 37636572, 37636573, 37636574, 37636575, 37636576, 37636577, 37636578, 37636579, 37636659, 37652880, 37652883, 37652883, 37660898, 37660902, and 37661630. |
| FEI Number |
3015876
|
Recalling Firm/ Manufacturer |
Physio Control, Inc.
11811 Willows Rd Ne
Redmond
WA
98052-2003
|
| For Additional Information Contact | 425-867-4000 |
Manufacturer Reason for Recall | Affected devices contain a Biphasic PCBA that may have a solder defect on the H Bridge component that could prevent the device from providing a defibrillation shock. There could be a delay in therapy or prevention of defibrillation therapy if the defibrillator were needed in a cardiac arrest situation. Delay or prevention of defibrillation therapy may contribute to the death of a viable patient. |
FDA Determined Cause 2 | Nonconforming Material/Component |
| Action | Consignees were sent a notification letter dated March 2009 that explained firm representatives would follow up with each customer within 30 days to arrange an update to affected devices.
Affected Biphasic PCBAs will be replaced. Customers advised to keep the devices in use and to continue to perform daily User Test. |
| Quantity in Commerce | 1718 |
| Distribution | Distributed domestically and internationally to hospitals, clinics, etc.
Foreign distribution to the following countries:
Australia, Brazil, Canada, Colombia, Costa Rica, Germany, Hong Kong, Mexico, Netherlands, and Puerto Rico. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|