Class 2 Device Recall 3M Micropore Surgical Tape

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall 3M Micropore Surgical Tapesee related information
Date Initiated by FirmJanuary 29, 2010
Date PostedApril 20, 2010
Recall Status1 Terminated 3 on February 04, 2012
Recall NumberZ-1413-2010
Recall Event ID 54640
Product Classification Surgical Tape - Product Code KGX
Product3M Micropore Single Use Rolls, Surgical Tape, REF 1530S-1, 1 in x 1-1/2 yd.
Code Information Lot numbers containing "2014" and lot numbers beginning with "2015-01" followed by two-character letters: AA, AB, AC, AD, AE, AF, AH, AI, AL, AN, BW.
FEI Number 2110898
Recalling Firm/
Manufacturer
3M Company / Medical Division
3M Center, Bldg 275-05-W-06
Saint Paul MN 55144
For Additional Information ContactCustomer Helpline
800-228-3957
Manufacturer Reason
for Recall
Reduced adhesion: Micropore Surgical Tape for use as the primary securement device for 1) dialysis needles used to access A-V fistulas or grafts; or 2) pressure dressings used after dialysis treatments. If the tape does not perform, the dialysis needle may loosen and could then dislodge. In the interim, please offer 3M Micropore Surgical Tape, 1 inch x 10 yard rolls (CAT #1530-1) as a substitutio
FDA Determined
Cause 2
Nonconforming Material/Component
ActionAn Urgent Medical Device Recall letter was sent to 3M customers beginning 01/29/2010. The letter described the issue, asked distributors to check existing inventory, quarantine affected lots. Distributors were asked to notify their customers and end-users (3M has provided example letters and return forms to facilitate). End users can return product either to the distributor or contact 3M directly. A Product Recall Form is asked to be faxed back to 3M. A second 3M "Urgent Medical Device Recall" letter dated February 22, 2010 was sent to customers. This letter informed the cusotmers that 3M was expanding the recall to include bulk-packaged rolls of Micropore Surgical Tape. The letter addressed to "Valued Kit manufacturer" described the problem and the product, provided instrucitions for "Action Requried by the Kit Manufacturer", "Action Requried by the Kit Manufacturer to the End-User", and "Action Required for Future Tape Orders from 3M". A Product Recall Form is aked to be faxed back to 3M.
Quantity in Commerce165,826 boxes, 100 rolls per box (16,582,600 rolls)
DistributionNationwide distribution, and Puerto Rico.
Total Product Life CycleTPLC Device Report