 Class 2 Device Recall Varian brand Cseries Clinac, Trilogy and Novalis Tx, Software Versions 6.X and 7.X,see related information
Date Initiated by FirmJune 10, 2010
Date PostedJanuary 08, 2011
Recall Status1 Terminated 3 on December 23, 2011
Recall NumberZ-0886-2011
Recall Event ID 56347
510(K)NumberK070094 K904364 K913119 
Product Classification Accelerator, linear, medical - Product Code GKZ
ProductVarian brand C-series Clinac, Trilogy and Novalis Tx, Software Versions 6.X and 7.X, Model Numbers: H14, H27, H29, HCX, Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Trilogy Radiotherapy Delivery System is a radiation therapy accelerator intended to deliver megavoltage x-ray treatments for conventional radiotherapy (three dimensional conformal radiotherapy and intensity modulated radiotherapy) and stereotactic radiosurgery and radiotherapy. Stereotactic treatments are intended for therapy of lesions, e.g., arteriovenous malformations, primary tumors and metastases. Stereotactic treatments may be intracranial or extra cranial and consist of single-session or fractionated delivery.
Code Information CODE(S):   H294145

H293620 ,H271973 ,H272328 ,H293393 ,H294467 ,H293944 ,H293764 ,H293725 H140117

H140734 ,H272492 ,H140070 ,H293653 ,H293998 ,H294049 ,H294199 ,H293726 H271877

H293666 ,H273065 ,H272283 ,H293823 ,H293999 ,H293697 ,H294367 ,H293728 H272434

H140754 ,H273130 ,H272370 ,H293529 ,H290201 ,H294103 ,H294440 ,H293698 H272828

H293724 ,H273170 ,H272327 ,H293787 ,H290555 ,H293760 ,H294474 ,H293774 H272889

H293817 ,H293381 ,H272329 ,H140772 ,H290553 ,H293682 ,H294496 ,H293776 H272922

H140768 ,H140709 ,H272299 ,H293883 ,H290549 ,H293009 ,H294597 ,H293773 H140203

H140775 ,H293411 ,H272425 ,H294113 ,H290552 ,H293942 ,H294641 ,H293829 H140204

H293843 ,H293412 ,H272421 ,H294129 ,H290551 ,H293765 ,H294650 ,H293830 H291054

H293847 ,H293463 ,H272407 ,H293908 ,H290559 ,H293758 ,H294699 ,H293854 H273072

H140786 ,H140729 ,H272857 ,H293909 ,H293396 ,H290557 ,H291075 ,H293839 H140208

H293850 ,H293499 ,H272859 ,H294134 ,H290547 ,H293770 ,H293368 ,H294073 H291098

H293968 ,H293500 ,H291022 ,H294186 ,H290541 ,H294136 ,H140756 ,H294118 H291114

H294022 ,H293668 ,H291011 ,H294092 ,H293365 ,H294143 ,H294034 ,H294104 H273027

H293970 ,H293841 ,H291015 ,H294273 ,H293435 ,H294239 ,H294527 ,H294200 H273107

H294017 ,H294164 ,H291082 ,H294204 ,H293387 ,H294185 ,H294107 ,H294203 H273102

H140796 ,H140882 ,H291021 ,H294210 ,H293482 ,H294248 ,H294535 ,H294256 H291160

H294065 ,H293567 ,H291016 ,H294317 ,H290554 ,H294290 ,H273048 ,H140856 H291191

H294062 ,H290278 ,H291085 ,H140836 ,H293434 ,H294390 ,H273140 ,H294471 H291164

H294175 ,H273187 ,H291018 ,H294364 ,H293555 ,H293766 ,H293371 ,H273150 H291165

H294232 ,H293595 ,H291154 ,H294403 ,H293506 ,H294153 ,H273208 ,H293898 H291163

H140819 ,H140763 ,H291166 ,H294383 ,H293569 ,H293961 ,H293305 ,H271763 H273103

H294409 ,H293929 ,H291190 ,H294414 ,H273031 ,H294250 ,H293681 ,H273117 H140229

H140843 ,H293958 ,H291172 ,H293906 ,H293853 ,H294422 ,H293842 ,H273182 H140234

H294448 ,H293450 ,H291231 ,H294347 ,H293455 ,HCX4138 ,H293924 ,H293374 H140231

H140844 ,H271159 ,H291179 ,H294585 ,H293498 ,H294011 ,H294462 ,H293601 H291212

H294584 ,H271157 ,H291178 ,H294586 ,H293633 ,H294445 ,H294525 ,H293602 H140236

H140865 ,H271569 ,H293307 ,H294553 ,H293477 ,H294480 ,H294532 ,H293645 H140235

H294667 ,H271624 ,H291195 ,H294551 ,H293489 ,H294400 ,H294608 ,H293661 H291261

H272188 ,H271662 ,H291194 ,H294674 ,H293626 ,HCX4400 ,H140096 ,H294098 H140232

H272946 ,H271663 ,H291180 ,H294683 ,H293705 ,H294490 ,H140097 ,H294123 H291262

H293494 ,H271941 ,H293509 ,H294684 ,H293676 ,H294488 ,H291115 ,H294296 H293422

H294593 ,H271926 ,H293598 ,H294675 ,H293710 ,H294449 ,H293424 ,H294297 H293442

H294594 ,H272157 ,H293549 ,H294352 ,H293907 ,H294568 ,H293444 ,H294304 H140713

H293796 ,H272216 ,H140728 ,H294750 ,H293586 ,H294634 ,H140751 ,H294543 H291259

H293940 ,H272159 ,H293521 ,H294751 ,H293745 ,H294611 ,H293662 ,H294461 H291260

H294526 ,H272162 ,H293656 ,H294758 ,H293838 ,H294015 ,H293663 ,H294592 H293512

H294502 ,H272311 ,H293654 ,H294766 ,H293978 ,H293495 ,H293650 ,H270794 H140732

H293324 ,H272309 ,H293655 ,H293336 ,H293738 ,H293759 ,H293727 ,H271582 H293561

H273186 ,H272310 ,H293784 ,H293337 ,H293753 ,H293761 ,H293612 ,H272277 H272372

H293407 ,H294201 ,H273164 ,H294146 ,H294140 ,H294602 ,H294428 ,H293363 H272473

H293474 ,H294109 ,H273162 ,H294114 ,H294133 ,H294636 ,H294460 ,H293362 H272458

H140711 ,H294228 ,H273191 ,H294368 ,H140124 ,H273185 ,H140857 ,H293380 H272494

H140704 ,H140822 ,H273200 ,H294356 ,H290533 ,H273036 ,H294432 ,H293410 H273011

H140717 ,H294241 ,H273203 ,H294346 ,H294329 ,H273168 ,H294587 ,H293440 H140119

H293578 ,H294230 ,H293321 ,H294264 ,H294433 ,H291199 ,H294647 ,H293408 H140114

H140154 ,H294262 ,H273202 ,H294233 ,H292832 ,H291268 ,H294183 ,H293452 H273077

H293533 ,H294242 ,H293379 ,H294357 ,H272834 ,H140715 ,H294371 ,H293430 H273080

H140707 ,H294280 ,H293388 ,H140842 ,H291174 ,H293502 ,H290282 ,H140719 H273106

H140742 ,H140850 ,H293378 ,H294084 ,H273172 ,H294074 ,H273064 ,H293409 H273110

H140731 ,H294057 ,H293376 ,H294231 ,H290556 ,H293353 ,H140129 ,H293454 H273121

H293576 ,H294529 ,H293392 ,H294274 ,H273066 ,H293947 ,H273089 ,H293312 H273126

H293625 ,H294419 ,H273210 ,H294415 ,H293333 ,H294042 ,H273039 ,H293449 H273131

H140740 ,H294519 ,H293464 ,H294160 ,H293304 ,H294043 ,H273042 ,H293451 H273104

H293644 ,H140863 ,H293389 ,H294477 ,H273178 ,H271802 ,H273092 ,H293527 H273128

H293675 ,H294581 ,H293597 ,H294310 ,H293524 ,H272195 ,H273134 ,H293523 H273147

H140759 ,H294627 ,H293640 ,H294536 ,H293665 ,H272985 ,H273127 ,H293566 H273160

H140748 ,H294605 ,H293686 ,H294601 ,H293844 ,H140125 ,H273122 ,H293557 H140140

H293624 ,H140872 ,H293619 ,H294591 ,H293897 ,H272932 ,H290548 ,H293558 H140142

H293651 ,H294646 ,H140747 ,H294545 ,H293755 ,H273113 ,H273111 ,H293623 H273177

H272905 ,H140875 ,H140758 ,H294511 ,H140774 ,H273083 ,H273116 ,H293641 H273167

H293779 ,H294678 ,H293530 ,H294369 ,H293946 ,H273115 ,H273119 ,H293634 H273175

H293646 ,H140871 ,H293735 ,H294547 ,H293793 ,H273145 ,H273144 ,H293719 H273171

H140764 ,H140887 ,H293763 ,H294548 ,H294121 ,H272964 ,H273132 ,H293720 H273165

H293861 ,H294068 ,H293736 ,H294473 ,H293876 ,H293370 ,H273133 ,H293721 H273204

H140766 ,H294067 ,H140760 ,H294589 ,H294124 ,H272973 ,H140138 ,H293730 H140144

H140779 ,H293534 ,H293845 ,H140866 ,H293677 ,H273183 ,H273124 ,H293729 H140148

H293860 ,H271370 ,H293732 ,H294645 ,H293310 ,H140724 ,H273156 ,H293762 H140146

H140788 ,H271378 ,H293465 ,H294671 ,H294351 ,H140712 ,H140137 ,H293788 H140145

H293918 ,H271531 ,H293955 ,H294685 ,H294387 ,H293419 ,H140141 ,H293797 H140151

H293922 ,H271593 ,H293982 ,H294724 ,H294212 ,H293443 ,H273123 ,H293792 H140150

H293949 ,H272137 ,H293575 ,H293643 ,H294374 ,H293491 ,H273125 ,H293799 H293351

H293954 ,H272111 ,H294036 ,H293857 ,H294465 ,H140762 ,H273163 ,H140767 H293373

H293963 ,H272497 ,H293917 ,H294198 ,H294618 ,H293781 ,H273194 ,H293813 H140139

H140794 ,H272875 ,H293791 ,H273166 ,H294423 ,H140769 ,H291257 ,H293800 H140153

H293992 ,H272909 ,H294085 ,H140862 ,H294060 ,H294100 ,H273197 ,H293809 H140708

H294023 ,H273101 ,H294167 ,H294292 ,H294542 ,H294161 ,H273138 ,H293832 H293391

H140798 ,H140136 ,H294119 ,H290561 ,H294515 ,H294223 ,H273193 ,H293798 H293334

H294173 ,H273151 ,H294005 ,H140152 ,H294664 ,H140789 ,H273189 ,H293807 H293369

H140815 ,H273174 ,H294117 ,H293585 ,H294238 ,H140841 ,H140238 ,H293816 H293815

H294342 ,H293808 ,H140839 ,H291083 ,H273082 ,H273180 ,H294215 ,H294349 H293814

H294334 ,H294656 ,H294335 ,H291127 ,H293306 ,H293320 ,H294229 ,H294588 H140773

H140828 ,H294659 ,H140739 ,H291128 ,H273181 ,H294144 ,H294315 ,H294705 H293827

H140834 ,H140873 ,H291064 ,H291141 ,H293375 ,H294205 ,H294370 ,H293618 H293828

H294318 ,H294610 ,H291227 ,H291168 ,H293532 ,H294401 ,H294569 ,H271265 H293881

H294309 ,H140889 ,H291044 ,H291204 ,H294080 ,H294402 ,H272033 ,H272219 H293895

H294293 ,H294600 ,H291132 ,H291205 ,H294435 ,H294495 ,H273084 ,H290386 H293896

H294307 ,H294693 ,H293540 ,H291223 ,H294434 ,H140886 ,H273108 ,H272405 H293920

H294338 ,H294723 ,H140755 ,H293447 ,H294746 ,H291130 ,H140147 ,H272362 H293911

H294363 ,H140892 ,H293959 ,H271860 ,H271881 ,H291137 ,H273179 ,H272333 H293943

H294283 ,H294635 ,H293678 ,H271976 ,H272023 ,H291055 ,H273201 ,H291037 H293887

H294396 ,H294643 ,H293997 ,H271681 ,H293349 ,H140135 ,H273198 ,H291238 H293812

H294308 ,H294666 ,H293616 ,H140847 ,H293545 ,H291198 ,H291254 ,H291237 H140785

H294450 ,H294668 ,H294097 ,H291250 ,H293801 ,H273159 ,H273184 ,H293395 H293884

H294447 ,H294676 ,H294247 ,H293469 ,H140840 ,H273153 ,H273195 ,H293448 H294002

H140849 ,H294700 ,H294313 ,H293468 ,H140874 ,H293421 ,H293479 ,H293394 H294027

H294319 ,H294701 ,H290208 ,H294344 ,H140700 ,H293456 ,H293572 ,H293423 H294046

H294451 ,H294706 ,H290274 ,HCX4344 ,H294044 ,H293364 ,H293559 ,H293604 H294152

H294312 ,H294713 ,H290400 ,H140090 ,H140878 ,H293704 ,H293570 ,H293657 H140804

H294407 ,H294715 ,H290399 ,H140101 ,H140803 ,H293553 ,H293390 ,H293767 H294071

H294260 ,H294726 ,H272359 ,H290517 ,H273112 ,H293631 ,H293599 ,H290372 H140814

H294066 ,H294728 ,H290501 ,H140121 ,H273176 ,H293628 ,H293446 ,H294037 H294154

H294463 ,H294731 ,H291211 ,H140062 ,H294108 ,H293679 ,H293382 ,H294082 H294195

H294503 ,H294732 ,H293596 ,H140143 ,H294190 ,H293794 ,H293383 ,H140790 H294220

H294087 ,H294736 ,H293741 ,H293384 ,H294191 ,H293386 ,H140730 ,H294219 H294176

H294213 ,H294737 ,H293866 ,H140705 ,H294520 ,H293471 ,H293712 ,H294681 H294165

H294580 ,H293603 ,H293964 ,H293328 ,H294606 ,H293629 ,H293497 ,H294682 H294258

H294583 ,H294093 ,H294331 ,H140134 ,H290256 ,H293785 ,H293849 ,H271400 H294271

H294492 ,H294491 ,H140877 ,H293637 ,H271890 ,H293786 ,H293852 ,H272397 H294270

H294582 ,H294279 ,H272418 ,H140745 ,H273173 ,H293870 ,H293951 ,H272427 H294276

H294508 ,H294533 ,H291009 ,H140750 ,H273155 ,H294038 ,H293941 ,H290511 H294277

H294590 ,H293550 ,H291103 ,H140741 ,H293372 ,H293904 ,H293972 ,H272911 H294278

H294632 ,H293938 ,H293531 ,H140749 ,H293977 ,H293905 ,H294193 ,H291033 H294289

H294339 ,H294489 ,H140752 ,H294070 ,H294055 ,H293510 ,H294192 ,H290540 H294216

H294094 ,H290262 ,H293574 ,H294072 ,H140743 ,H293869 ,H294208 ,H291213 H294217

H294637 ,H290263 ,H293996 ,H140805 ,H140771 ,H294047 ,H294211 ,H291232 H294281

H294537 ,H290352 ,H294081 ,H294069 ,H294214 ,H294059 ,H294389 ,H293577 H294272

H294649 ,H291046 ,H294523 ,H140859 ,H294779 ,H140818 ,H294014 ,H293593 H294306

H294619 ,H293630 ,H290423 ,H140861 ,H294565 ,H293878 ,H293990 ,H293492 H294288

H294663 ,H140823 ,H291007 ,H291059 ,H273142 ,H294147 ,H293991 ,H293649 H293667

H293511 ,H272908 ,H294102 ,H290249 ,H140001 ,H272014 ,H140046 ,H272401 H293742

H293339 ,H273004 ,H294076 ,H271393 ,H271760 ,H290362 ,H272238 ,H272394 H293652

H290560 ,H273049 ,H294077 ,H271407 ,H271762 ,H272009 ,H272237 ,H272801 H293818

H294639 ,H272469 ,H294078 ,H271413 ,H271785 ,H271975 ,H272236 ,H272456 H293834

H294640 ,H273063 ,H294079 ,H271385 ,H271780 ,H271964 ,H272259 ,H272470 H293778

H291233 ,H273105 ,H294218 ,H271439 ,H271797 ,H272052 ,H290392 ,H272500 H293914

H293879 ,H291152 ,H293867 ,H271452 ,H271835 ,H272000 ,H272271 ,H272803 H293928

H273149 ,H291151 ,H294257 ,H271438 ,H271838 ,H272043 ,H140058 ,H272454 H293965

H273152 ,H140224 ,H294354 ,H271425 ,H271831 ,H272035 ,H272301 ,H272466 H293981

H293323 ,H140225 ,H294263 ,H271502 ,H271783 ,H140038 ,H272306 ,H272479 H293993

H273114 ,H140226 ,H294323 ,H271515 ,H271858 ,H272045 ,H290397 ,H272491 H140797

H293315 ,H140227 ,H294321 ,H271519 ,H271856 ,H271951 ,H272307 ,H140088 H294305

H293584 ,H291271 ,H294320 ,H271526 ,H271862 ,H272049 ,H272291 ,H272480 H140864

H293739 ,H273206 ,H294322 ,H271507 ,H271833 ,H290376 ,H272321 ,H140092 H273143

H293803 ,H293325 ,H294268 ,H271523 ,H140011 ,H271942 ,H272325 ,H140201 H271994

H293687 ,H293326 ,H294539 ,H290287 ,H271879 ,H272062 ,H272293 ,H290505 H291008

H293795 ,H293207 ,H294541 ,H271497 ,H271825 ,H272069 ,H272326 ,H272816 H291182

H293865 ,H293356 ,H294171 ,H271559 ,H271870 ,H272001 ,H272302 ,H290503 H291108

H294110 ,H293327 ,H294540 ,H271587 ,H271893 ,H272086 ,H272346 ,H272476 H293311

H294111 ,H293416 ,H294538 ,H271588 ,H271896 ,H272078 ,H290405 ,H272811 H293331

H294040 ,H293417 ,H294621 ,H271420 ,H271900 ,H272089 ,H272345 ,H290509 H293556

H293301 ,H293377 ,H294620 ,H271594 ,H290341 ,H140043 ,H140073 ,H272837 H293507

H293314 ,H273055 ,H294716 ,H271607 ,H271903 ,H290371 ,H290402 ,H272822 H293747

H294345 ,H293441 ,H140893 ,H271626 ,H140018 ,H272081 ,H140071 ,H272824 H140795

H294412 ,H293583 ,H270788 ,H271669 ,H271921 ,H272129 ,H272336 ,H272488 H293740

H294411 ,H293674 ,H271041 ,H271664 ,H290326 ,H272134 ,H140067 ,H272854 H293980

H294379 ,H293671 ,H270862 ,H290304 ,H271906 ,H140045 ,H272354 ,H272860 H293872

H294438 ,H293672 ,H271066 ,H271691 ,H271918 ,H272143 ,H272365 ,H140098 H293875

H294439 ,H293670 ,H271105 ,H290313 ,H271895 ,H272172 ,H272341 ,H272841 H293960

H294388 ,H293716 ,H271169 ,H290308 ,H271935 ,H272187 ,H272355 ,H272862 H140829

H294644 ,H293715 ,H290241 ,H271678 ,H271966 ,H272178 ,H272377 ,H290510 H291272

H273056 ,H293717 ,H271167 ,H271618 ,H140026 ,H290379 ,H140076 ,H272873 H293496

H293752 ,H293718 ,H271242 ,H140020 ,H140027 ,H272202 ,H272388 ,H290508 H293600

H294255 ,H293669 ,H290222 ,H271711 ,H140031 ,H272174 ,H272410 ,H272866 H293638

H293751 ,H293804 ,H271362 ,H271693 ,H271993 ,H272209 ,H272422 ,H272868 H293848

H273148 ,H293820 ,H271194 ,H271712 ,H271961 ,H272220 ,H272428 ,H272872 H294141

H294657 ,H293782 ,H271308 ,H271720 ,H271974 ,H272204 ,H140078 ,H272847 H294265

H271954 ,H293826 ,H290224 ,H271729 ,H140025 ,H140047 ,H140077 ,H272885 H294284

H271872 ,H293713 ,H271334 ,H271714 ,H272029 ,H272231 ,H272426 ,H272914 H290550

H272826 ,H293714 ,H290243 ,H271724 ,H272003 ,H272223 ,H272444 ,H291002 H290600

H273029 ,H291076 ,H291048 ,H273135 ,H291219 ,H293303 ,H293318 ,H293432 H272893

H291028 ,H291077 ,H291088 ,H291140 ,H140233 ,H291236 ,H293357 ,H293425 H290520

H273030 ,H291052 ,H140211 ,H291045 ,H291202 ,H293332 ,H291229 ,H293300 H291006

H291032 ,H291079 ,H291099 ,H291109 ,H291169 ,H291266 ,H291249 ,H293438 H140109

H272971 ,H272978 ,H140215 ,H291117 ,H291200 ,H291247 ,H140149 ,H140706 H272899

H291027 ,H291073 ,H140133 ,H291111 ,H291218 ,H291256 ,H293313 ,H293366 H291001

H291025 ,H291067 ,H291120 ,H291116 ,H291170 ,H291270 ,H291252 ,H293404 H291004

H291026 ,H291078 ,H291136 ,H291157 ,H291196 ,H291253 ,H140703 ,H293348 H272926

H291038 ,H291063 ,H140219 ,H291161 ,H291221 ,H293316 ,H293345 ,H293486 H272938

H273016 ,H140216 ,H273120 ,H291113 ,H291222 ,H140240 ,H293429 ,H293397 H290519

H272982 ,H140213 ,H273109 ,H140220 ,H291207 ,H293308 ,H293431 ,H293490 H291003

H291023 ,H291090 ,H291149 ,H291092 ,H291209 ,H293322 ,H293385 ,H293503 H140111

H272990 ,H291068 ,H291134 ,H273137 ,H140228 ,H291224 ,H293426 ,H291176 H140202

H291029 ,H291074 ,H291096 ,H273136 ,H291133 ,H291243 ,H293342 ,H293439 H272960

H291020 ,H140217 ,H291139 ,H291162 ,H291181 ,H291225 ,H293360 ,H293508 H272981

H291042 ,H273026 ,H273118 ,H291183 ,H291171 ,H291267 ,H293418 ,H293485 H272842

H272825 ,H291072 ,H291142 ,H291150 ,H291185 ,H291245 ,H293428 ,H293518 H291013

H291035 ,H291080 ,H291094 ,H291131 ,H291216 ,H291258 ,H293414 ,H293406 H291005

H291039 ,H291097 ,H140218 ,H291053 ,H291203 ,H291255 ,H291226 ,H293354 H140205

H291043 ,H291093 ,H140221 ,H140223 ,H291210 ,H290558 ,H291248 ,H293459 H291010

H272987 ,H291100 ,H291101 ,H273158 ,H273199 ,H293343 ,H291263 ,H293476 H272846

H140207 ,H291105 ,H291146 ,H291159 ,H291208 ,H293358 ,H293436 ,H140720 H272848

H291041 ,H291066 ,H140222 ,H291135 ,H291234 ,H293329 ,H293403 ,H293466 H290504

H291049 ,H291069 ,H140209 ,H291186 ,H291235 ,H293340 ,H293460 ,H293483 H290512

H273052 ,H291081 ,H291119 ,H291192 ,H291215 ,H293335 ,H293413 ,H293405 H272956

H273040 ,H291061 ,H291123 ,H291184 ,H291220 ,H293350 ,H140239 ,H293488 H272963

H291050 ,H273075 ,H291129 ,H291187 ,H291206 ,H293309 ,H293402 ,H293420 H272993

H291056 ,H140214 ,H291121 ,H291144 ,H291241 ,H293346 ,H293472 ,H293433 H272999

H291065 ,H140206 ,H291122 ,H291143 ,H140237 ,H291230 ,H293398 ,H293484 H291024

H140212 ,H291106 ,H291112 ,H291167 ,H291175 ,H293352 ,H140714 ,H140722 H291012

H291057 ,H291102 ,H291145 ,H291158 ,H291217 ,H293302 ,H293453 ,H293401 H272986

H291058 ,H291084 ,H273141 ,H291177 ,H291264 ,H291240 ,H293445 ,H293541 H291019

H291070 ,H291086 ,H291153 ,H273157 ,H273190 ,H293361 ,H293437 ,H293548 H140115

H291071 ,H291087 ,H291118 ,H291148 ,H273209 ,H291269 ,H293427 ,H293344 H291017

H273053 ,H291089 ,H291155 ,H291188 ,H291242 ,H291251 ,H293457 ,H293480 H291014

H291047 ,H291107 ,H291125 ,H291197 ,H291173 ,H293317 ,H140716 ,H293481 H291030

H291051 ,H272996 ,H291126 ,H140230 ,H291246 ,H293330 ,H293461 ,H293478 H291034

H291062 ,H291104 ,H291138 ,H291189 ,H291214 ,H291239 ,H293462 ,H293536 H290530

H291031 ,H291040 ,H291147 ,H291193 ,H291273 ,H291265 ,H293355 ,H293581 H291036

H291060 ,H140210 ,H291156 ,H291124 ,H291228 ,H293399 ,H293415 ,H293319 H293501

H293591 ,H293673 ,H293695 ,H293873 ,H293919 ,H294032 ,H140806 ,H294184 H293517

H293562 ,H293690 ,H140733 ,H293891 ,H140791 ,H294045 ,H294105 ,H294166 H293519

H293504 ,H293587 ,H293771 ,H293890 ,H293926 ,H294030 ,H294112 ,H294178 H293473

H293590 ,H293689 ,H293756 ,H293880 ,H293932 ,H294006 ,H294053 ,H294209 H293458

H293592 ,H293706 ,H140736 ,H293855 ,H293925 ,H294033 ,H293986 ,H140813 H293367

H293538 ,H293647 ,H293611 ,H293868 ,H293948 ,H294041 ,H294125 ,H294194 H293568

H293608 ,H293700 ,H293783 ,H140780 ,H293956 ,H294051 ,H140807 ,H294234 H293564

H293539 ,H293696 ,H293790 ,H293877 ,H293930 ,H294026 ,H140808 ,H294227 H293579

H140701 ,H293694 ,H293750 ,H293885 ,H140793 ,H294048 ,H140810 ,H294207 H293475

H293610 ,H293554 ,H293806 ,H293864 ,H293962 ,H294039 ,H294130 ,H294116 H291095

H140746 ,H293702 ,H140765 ,H293882 ,H293966 ,H293974 ,H294095 ,H294101 H293359

H140718 ,H293692 ,H293708 ,H140784 ,H293899 ,H293994 ,H294106 ,H294197 H293544

H140735 ,H293693 ,H293658 ,H293701 ,H293967 ,H294004 ,H294122 ,H294181 H293552

H293522 ,H293699 ,H293769 ,H293862 ,H293936 ,H294031 ,H294142 ,H294196 H293565

H293543 ,H293703 ,H293821 ,H293913 ,H293953 ,H294008 ,H294132 ,H294188 H293589

H293594 ,H293746 ,H293819 ,H140770 ,H293971 ,H294063 ,H294157 ,H294237 H293347

H293505 ,H291201 ,H293824 ,H293888 ,H293957 ,H294003 ,H294151 ,H294131 H293487

H293642 ,H293733 ,H140776 ,H293902 ,H293934 ,H293984 ,H294099 ,H294202 H293514

H293648 ,H140757 ,H293789 ,H293900 ,H293950 ,H294024 ,H294054 ,H294225 H293516

H293588 ,H293709 ,H293825 ,H140783 ,H293969 ,H294028 ,H294135 ,H294029 H293467

H293606 ,H293734 ,H293754 ,H293886 ,H293995 ,H294025 ,H140809 ,H294148 H293513

H293537 ,H293685 ,H293802 ,H293910 ,H293975 ,H294061 ,H294137 ,H294162 H293515

H293520 ,H293535 ,H293573 ,H293912 ,H293987 ,H294052 ,H140811 ,H294163 H293546

H293551 ,H293659 ,H293805 ,H293889 ,H140782 ,H294050 ,H294156 ,H294240 H293614

H293621 ,H293707 ,H293836 ,H293874 ,H293952 ,H294056 ,H294159 ,H294243 H140725

H293622 ,H293743 ,H293822 ,H293893 ,H293988 ,H293983 ,H294168 ,H140821 H293525

H293660 ,H293723 ,H293846 ,H293903 ,H293985 ,H294086 ,H294058 ,H294244 H293617

H293639 ,H291244 ,H140778 ,H293915 ,H293921 ,H140799 ,H294179 ,H294252 H140721

H140738 ,H293744 ,H140777 ,H293935 ,H294009 ,H294088 ,H140801 ,H294149 H140726

H293636 ,H293748 ,H293856 ,H140787 ,H293979 ,H294075 ,H294169 ,H294126 H140727

H293607 ,H140753 ,H293831 ,H293772 ,H294010 ,H294064 ,H294172 ,H294127 H273196

H293571 ,H140761 ,H293851 ,H293931 ,H294021 ,H294000 ,H294155 ,H294128 H293615

H293683 ,H293737 ,H293840 ,H293894 ,H293989 ,H294089 ,H294177 ,H294245 H140723

H293563 ,H293341 ,H293859 ,H293923 ,H293976 ,H294091 ,H294187 ,H294254 H293609

H140710 ,H140737 ,H293835 ,H293901 ,H294016 ,H294120 ,H140800 ,H294269 H293560

H293684 ,H293749 ,H293863 ,H293892 ,H294012 ,H140802 ,H140817 ,H140820 H293470

H293493 ,H293688 ,H293664 ,H140781 ,H294018 ,H294007 ,H294182 ,H294267 H293635

H293613 ,H293780 ,H293858 ,H293916 ,H294013 ,H294083 ,H294170 ,H294226 H293627

H293691 ,H293775 ,H293837 ,H293933 ,H294001 ,H294035 ,H294189 ,H294266 H140744

H293605 ,H293777 ,H293871 ,H140792 ,H294020 ,H294090 ,H294180 ,H294285 H294261

H294360 ,H294397 ,H294521 ,H294572 ,H140883 ,H294282 ,H294394 ,H293833 H140825

H294158 ,H294421 ,H294443 ,H294577 ,H294698 ,H294294 ,H294365 ,H294486 H294275

H294326 ,H294427 ,H294509 ,H294458 ,H294660 ,H294328 ,H294393 ,H294475 H294150

H294366 ,H294425 ,H294406 ,H294507 ,H294679 ,H294324 ,H294381 ,H294430 H294235

H294358 ,H294429 ,H294530 ,H294571 ,H294696 ,H294325 ,H294380 ,H294479 H140824

H140838 ,H140848 ,H294534 ,H294574 ,H294711 ,H140831 ,H294395 ,H294494 H140827

H294246 ,H294436 ,H294517 ,H294236 ,H294720 ,H294336 ,H294399 ,H294418 H294287

H294333 ,H294437 ,H294470 ,H294598 ,H294719 ,H140832 ,H294362 ,H294478 H294286

H294355 ,H294424 ,H294510 ,H140855 ,H294712 ,H294332 ,H294372 ,H294482 H294295

H294361 ,H294426 ,H294513 ,H294603 ,H294677 ,H294348 ,H294398 ,H294504 H294224

H294373 ,H294469 ,H294516 ,H294604 ,H294686 ,H294350 ,H294420 ,H294485 H294299

H294340 ,H140851 ,H294524 ,H294616 ,H140897 ,H294353 ,H294408 ,H294481 H294298

H294392 ,H294444 ,H294441 ,H140816 ,H294738 ,H294341 ,H294385 ,H294499 H294301

H294376 ,H294468 ,H294544 ,H294599 ,H294735 ,H140833 ,H140846 ,H294483 H294316

H294174 ,H140852 ,H294528 ,H294595 ,H294729 ,H294337 ,H294413 ,H294514 H294302

H294377 ,H294359 ,H294518 ,H294612 ,H294742 ,H294343 ,H294442 ,H140854 H140812

H140830 ,H294472 ,H294206 ,H294629 ,H294727 ,H294330 ,H294404 ,H294484 H294303

H294378 ,H140826 ,H294546 ,H294625 ,H140891 ,H140835 ,H294410 ,H294505 H294314

H294384 ,H294416 ,H140845 ,H140867 ,H294743 ,H140837 ,H294417 ,H294501 H294476

H294615 ,H140881 ,H294493 ,H140868 ,H294747 ,H294221 ,H294386 ,H140853 H294554

H294576 ,H140895 ,H294531 ,H294555 ,H294753 ,H294327 ,H294382 ,H294497 H294556

H294630 ,H140896 ,H294487 ,H140869 ,H294689 ,H294498 ,H294617 ,H294690 H294552

H294613 ,H294730 ,H294300 ,H294624 ,H294761 ,H294500 ,H294642 ,H294763 H294506

H294653 ,H294771 ,H294431 ,H294652 ,H294783 ,H294391 ,H294567 ,H140702 H294549

H294658 ,H294773 ,H294311 ,H294654 ,H294794 ,H294573 ,H294694 ,H294253 H294575

H294680 ,H294782 ,H294446 ,H294665 ,H291091 ,H140858 ,H294626 ,HCX4019 H294550

H294695 ,H294291 ,H140876 ,H293542   
FEI Number 2916710
Recalling Firm/
Manufacturer
Varian Medical Systems Oncology Systems
911 Hansen Way
Palo Alto CA 94304-1028
For Additional Information Contact
650-424-5731
Manufacturer Reason
for Recall
The Auto goto or Auto setup functions in C-Series version 7 software ignore the couch angle and exceed motion zone (unless a tolerance is defined by the user); it has the potential for collision with the patient on the couch.
FDA Determined
Cause 2
Other
ActionVarian sent a Customer Technical Bulletin letter to all affected users on May 27, 2010, and was also posted to Varian's customer support site: http://www.MyVarian.com. The letter identified the product, the problem, and provided instructions on the recall. Customers were instructed to return the proof of notification (via completed "Proof of Notification Form or appropriate Postal system certified receipt of delivery to Varian. For questions regarding this recall call 650-483-3153.
DistributionWorldwide Distribution - USA including CA, TX, IL, GA, NJ, and NV and the countries of Australia, Belgium, Brazil, Canada, China, Denmark, France, Germany, Hong Kong, India, Italy, Japan, Middle East, Netherlands, Russia, Spain, Switzerland, and the UK
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.