| Class 1 Device Recall Powerheart AED, CardioVive AED, GE Responder AED, Nihon Kohden CardioLife AED |  |
Date Initiated by Firm | January 13, 2012 |
Date Posted | February 27, 2012 |
Recall Status1 |
Terminated 3 on September 27, 2013 |
Recall Number | Z-1057-2012 |
Recall Event ID |
60992 |
510(K)Number | K031987 K102496 |
Product Classification |
Automated external defibrillators (non-wearable) - Product Code MKJ
|
Product | Powerheart G3 Automated External Defibrillators (Non-Wearable) and include the following models: Powerheart 9300A, 9300E, 9300P, 9390A, 9390E; CardioVive 92532, 92533; Nihon-Kohden 9200G, 9231; and GE Responder 2019198, 2023440 AEDs.
The Powerheart AED G3 is indicated for emergency treatment of victims exhibiting symptoms of sudden cardiac arrest who are unresponsive and not breathing. Post-resuscitation, if the victim is breathing, the AED should be left attached to allow for acquisition and detection of the ECG rhythm. If a shockable ventricular tachyarrhythmia recurs, the device will charge automatically and be able to deliver therapy. The device is intended to be used by personnel who have been trained it is operation. The user should be qualified by training in basic life support or other physician-authorized medical response. |
Code Information |
Serial Numbers: 4372713, 4373884, 4373898, 4373902, 4373904, 4373906, 4373908, 4373913, 4373915, 4373917, 4373920, 4373922, 4373923, 4373925, 4373927, 4373931, 4373938, 4373944, 4373945, 4373946, 4373947, 4373949, 4373951, 4373952, 4373955, 4373956, 4373958, 4373959, 4373960, 4373965, 4373991, 4373993, 4373998, 4374002, 4374009, 4374010, 4374011, 4374012, 4374013, 4374014, 4374015, 4374018, 4374021, 4374023, 4374032, 4374035, 4374038, 4374050, 4374052, 4374053, 4374055, 4374056, 4374061, 4374065, 4374077, 4374082, 4374086, 4374096, 4374098, 4374102, 4374106, 4374110, 4374111, 4374115, 4374119, 4374137, 4374143, 4374151, 4374154, 4374159, 4374161, 4374165, 4374174, 4374175, 4374180, 4374181, 4374182, 4374183, 4374186, 4374188, 4374191, 4374193, 4374196, 4374203, 4374204, 4374209, 4374210, 4374212, 4374215, 4374220, 4374230, 4374235, 4374240, 4374244, 4374257, 4374275, 4374283, 4374285, 4374286, 4374288, 4374294, 4374297, 4374299, 4374368, 4374369, 4374370, 4374372, 4374375, 4374377, 4374378, 4374380, 4374389, 4374390, 4374394, 4374396, 4374399, 4374427, 4374428, 4374429, 4374431, 4374435, 4374440, 4374442, 4374446, 4374454, 4374459, 4374463, 4374465, 4374487, 4374488, 4374491, 4374493, 4374495, 4374497, 4374504, 4374506, 4374508, 4374509, 4374511, 4374515, 4374516, 4374519, 4374520, 4374522, 4374529, 4374530, 4374539, 4374542, 4374544, 4374546, 4374547, 4374548, 4374549, 4374550, 4374551, 4374552, 4374554, 4374555, 4374559, 4374560, 4374562, 4374564, 4374571, 4374575, 4374578, 4374579, 4374590, 4374597, 4374599, 4374600, 4374601, 4374602, 4374604, 4374607, 4374612, 4374614, 4374618, 4374619, 4374620, 4374622, 4374623, 4374626, 4374627, 4374628, 4374630, 4374636, 4374660, 4374663, 4374671, 4374677, 4374678, 4374683, 4374698, 4374701, 4374703, 4374705, 4374714, 4374717, 4374719, 4374721, 4374722, 4374723, 4374724, 4374725, 4374728, 4374731, 4374733, 4374735, 4374740, 4374742, 4374745, 4374758, 4374761, 4374762, 4374774, 4374775, 4374785, 4374787, 4374796, 4374801, 4374802, 4374805, 4374806, 4374813, 4374816, 4374819, 4374820, 4374821, 4374822, 4374836, 4375055, 4375059, 4375061, 4375066, 4375107, 4375112, 4375114, 4375118, 4375122, 4375125, 4375128, 4375129, 4375131, 4375134, 4375136, 4375180, 4375184, 4375187, 4375659, 4375679, 4376016, 4376018, 4376019, 4376020, 4376021, 4376022, 4376023, 4376024, 4376025, 4376026, 4376027, 4376028, 4376029, 4376030, 4376031, 4376032, 4376033, 4376034, 4376035, 4376036, 4376037, 4376038, 4376039, 4376040, 4376041, 4376042, 4376043, 4376044, 4376045, 4376046, 4376047, 4376048, 4376049, 4376050, 4376051, 4376052, 4376053, 4376054, 4376055, 4376056, 4376057, 4376058, 4376059, 4376060, 4376061, 4376062, 4376064, 4376065, 4376066, 4376067, 4376068, 4376070, 4376071, 4376072, 4376073, 4376075, 4376077, 4376078, 4376079, 4376080, 4376081, 4376082, 4376083, 4376084, 4376085, 4376086, 4376087, 4376088, 4376089, 4376090, 4376091, 4376092, 4376093, 4376094, 4376095, 4376096, 4376097, 4376100, 4376101, 4376102, 4376103, 4376104, 4376105, 4376107, 4376110, 4376111, 4376113, 4376114, 4376115, 4376117, 4376118, 4376119, 4376120, 4376121, 4376122, 4376123, 4376124, 4376125, 4376126, 4376127, 4376128, 4376129, 4376130, 4376132, 4376136, 4376138, 4376139, 4376140, 4376142, 4376143, 4376144, 4376145, 4376146, 4376147, 4376149, 4376150, 4376151, 4376152, 4376153, 4376154, 4376155, 4376156, 4376157, 4376158, 4376159, 4376160, 4376162, 4376163, 4376164, 4376166, 4376168, 4376169, 4376171, 4376173, 4376175, 4376176, 4376179, 4376180, 4376181, 4376183, 4376184, 4376185, 4376187, 4376188, 4376191, 4376193, 4376197, 4376198, 4376200, 4376201, 4376202, 4376205, 4376206, 4376208, 4376209, 4376210, 4376211, 4376212, 4376214, 4376216, 4376217, 4376218, 4376219, 4376220, 4376224, 4376227, 4376244, 4376247, 4376248, 4376251, 4376252, 4376253, 4376254, 4376255, 4376256, 4376257, 4376258, 4376259, 4376260, 4376261, 4376262, 4376263, 4376264, 4376265, 4376266, 4376268, 4376269, 4376271, 4376273, 4376276, 4376278, 4376279, 4376280, 4376281, 4376282, 4376285, 4376291, 4376294, 4376295, 4376296, 4376297, 4376298, 4376299, 4376300, 4376301, 4376302, 4376303, 4376304, 4376305, 4376306, 4376307, 4376308, 4376311, 4376312, 4376313, 4376315, 4376316, 4376317, 4376318, 4376319, 4376320, 4376321, 4376322, 4376323, 4376325, 4376326, 4376328, 4376329, 4376331, 4376332, 4376333, 4376334, 4376335, 4376336, 4376337, 4376338, 4376339, 4376340, 4376341, 4376342, 4376343, 4376344, 4376345, 4376346, 4376347, 4376348, 4376349, 4376351, 4376352, 4376353, 4376354, 4376355, 4376356, 4376357, 4376358, 4376359, 4376360, 4376361, 4376362, 4376363, 4376364, 4376365, 4376366, 4376367, 4376368, 4376369, 4376370, 4376371, 4376372, 4376373, 4376374, 4376375, 4376376, 4376377, 4376378, 4376379, 4376380, 4376381, 4376382, 4376383, 4376384, 4376385, 4376386, 4376387, 4376388, 4376389, 4376390, 4376391, 4376392, 4376393, 4376394, 4376395, 4376396, 4376397, 4376398, 4376399, 4376404, 4376405, 4376406, 4376407, 4376408, 4376409, 4376410, 4376411, 4376412, 4376413, 4376414, 4376415, 4376416, 4376417, 4376418, 4376419, 4376420, 4376421, 4376425, 4376426, 4376427, 4376428, 4376429, 4376430, 4376431, 4376432, 4376433, 4376436, 4376438, 4376440, 4376442, 4376444, 4376445, 4376447, 4376448, 4376450, 4376451, 4376452, 4376453, 4376455, 4376456, 4376457, 4376460, 4376463, 4376464, 4376465, 4376466, 4376467, 4376468, 4376469, 4376470, 4376471, 4376472, 4376474, 4376475, 4376476, 4376477, 4376478, 4376479, 4376480, 4376482, 4376483, 4376484, 4376485, 4376486, 4376487, 4376488, 4376489, 4376490, 4376491, 4376492, 4376495, 4376496, 4376498, 4376505, 4376507, 4376509, 4376511, 4376512, 4376513, 4376514, 4376515, 4376516, 4376517, 4376518, 4376520, 4376521, 4376522, 4376523, 4376524, 4376525, 4376527, 4376528, 4376529, 4376530, 4376531, 4376533, 4376534, 4376535, 4376537, 4376538, 4376539, 4376540, 4376542, 4376544, 4376545, 4376546, 4376547, 4376548, 4376549, 4376550, 4376551, 4376552, 4376553, 4376554, 4376555, 4376556, 4376557, 4376558, 4376559, 4376561, 4376563, 4376565, 4376566, 4376567, 4376568, 4376569, 4376845, 4376846, 4376847, 4376848, 4376849, 4376850, 4376851, 4376852, 4376853, 4376854, 4376855, 4376856, 4376857, 4376858, 4376861, 4377088, 4377095, 4377096, 4377097, 4377098, 4377099, 4377100, 4377101, 4377102, 4377103, 4377104, 4377105, 4377106, 4377107, 4377108, 4377109, 4377110, 4377111, 4377112, 4377113, 4377114, 4377115, 4377116, 4377117, 4377118, 4377119, 4377120, 4377122, 4377123, 4377124, 4377125, 4377126, 4377129, 4377130, 4377131, 4377132, 4377133, 4377134, 4377135, 4377137, 4377138, 4377139, 4377140, 4377141, 4377142, 4377143, 4377144, 4377145, 4377146, 4377147, 4377148, 4377149, 4377150, 4377151, 4377152, 4377153, 4377154, 4377155, 4377156, 4377157, 4377158, 4377159, 4377160, 4377161, 4377162, 4377163, 4377164, 4377165, 4377166, 4377167, 4377168, 4377169, 4377170, 4377171, 4377172, 4377173, 4377174, 4377175, 4377176, 4377177, 4377178, 4377179, 4377180, 4377181, 4377182, 4377183, 4377184, 4377185, 4377186, 4377187, 4377188, 4377189, 4377190, 4377191, 4377192, 4377193, 4377194, 4377195, 4377196, 4377197, 4377198, 4377199, 4377200, 4377201, 4377202, 4377204, 4377222, 4377224, 4377225, 4377226, 4377227, 4377228, 4377229, 4377230, 4377231, 4377232, 4377233, 4377234, 4377235, 4377236, 4377237, 4377238, 4377239, 4377240, 4377241, 4377242, 4377243, 4377244, 4377245, 4377246, 4377247, 4377248, 4377249, 4377250, 4377251, 4377252, 4377253, 4377254, 4377255, 4377256, 4377257, 4377258, 4377259, 4377260, 4377261, 4377262, 4377263, 4377264, 4377265, 4377266, 4377267, 4377268, 4377269, 4377270, 4377271, 4377272, 4377273, 4377274, 4377275, 4377276, 4377277, 4377305, 4377306, 4377307, 4377308, 4377309, 4377311, 4377312, 4377313, 4377314, 4377315, 4377316, 4377317, 4377318, 4377319, 4377320, 4377321, 4377322, 4377323, 4377324, 4377325, 4377326, 4377327, 4377328, 4377329, 4377330, 4377331, 4377332, 4377333, 4377334, 4377335, 4377336, 4377337, 4377338, 4377339, 4377340, 4377341, 4377342, 4377343.4377344, 4377345, 4377346, 4377347, 4377348, 4377349, 4377350, 4377351, 4377352, 4377353, 4377354, 4377355, 4377356, 4377358, 4377359, 4377360, 4377361, 4377362, 4377363, 4377364, 4377365, 4377366, 4377367, 4377368, 4377370, 4377372, 4377373, 4377374, 4377375, 4377376, 4377377, 4377378, 4377379, 4377380, 4377381, 4377382, 4377383, 4377384, 4377385, 4377386, 4377387, 4377388, 4377389, 4377390, 4377391, 4377392, 4377393, 4377394, 4377395, 4377396, 4377397, 4377398, 4377399, 4377400, 4377401, 4377402, 4377403, 4377404, 4377405, 4377406, 4377407, 4377408, 4377409, 4377410, 4377411, 4377412, 4377413, 4377414, 4377415, 4377416, 4377417, 4377418, 4377419, 4377420, 4377421, 4377422, 4377423, 4377424, 4377425, 4377426, 4377427, 4377428, 4377429, 4377430, 4377431, 4377432, 4377433, 4377434, 4377435, 4377436, 4377437, 4377438, 4377439, 4377440, 4377441, 4377442, 4377443, 4377444, 4377445, 4377446, 4377447, 4377448, 4377449, 4377450, 4377451, 4377452, 4377453, 4377454, 4377455, 4377456, 4377457, 4377458, 4377459, 4377460, 4377461, 4377462, 4377463, 4377464, 4377465, 4377466, 4377467, 4377468, 4377469, 4377470, 4377471, 4377472, 4377473, 4377474, 4377475, 4377476, 4377477, 4377478, 4377479, 4377480, 4377481, 4377482, 4377483, 4377484, 4377485, 4377486, 4377487, 4377488, 4377489, 4377490, 4377491, 4377492, 4377493, 4377494, 4377495, 4377496, 4377497, 4377498, 4377499, 4377500, 4377501, 4377502, 4377503, 4377504, 4377505, 4377506, 4377507, 4377508, 4377509, 4377510, 4377511, 4377512, 4377513, 4377514, 4377515, 4377516, 4377517, 4377518, 4377519, 4377520, 4377521, 4377522, 4377523, 4377524, 4377525, 4377526, 4377527, 4377542, 4377543, 4377544, 4377545, 4377547, 4377549, 4377550, 4377551, 4377552, 4377553, 4377578, 4377579, 4377580, 4377581, 4377583, 4377584, 4377596, 4377598, 4377599, 4377600, 4377601, 4377602, 4377603, 4377604, 4377605, 4377606, 4377607, 4377608, 4377609, 4377610, 4377611, 4377612, 4377613, 4377614, 4377615, 4377616, 4377617, 4377618, 4377619, 4377620, 4377621, 4377622, 4377623, 4377624, 4377625, 4377626, 4377627, 4377628, 4377629, 4377631, 4377632, 4377633, 4377634, 4377635, 4377636, 4377637, 4377638, 4377639, 4377640, 4377641, 4377642, 4377643, 4377644, 4377645, 4377646, 4377647, 4377648, 4377649, 4377650, 4377651, 4377652, 4377653, 4377654, 4377655, 4377656, 4377657, 4377658, 4377659, 4377660, 4377661, 4377662, 4377663, 4377664, 4377665, 4377666, 4377667, 4377668, 4377670, 4377671, 4377672, 4377673, 4377674, 4377675, 4377676, 4377677, 4377678, 4377679, 4377680, 4377681, 4377682, 4377683, 4377684, 4377685, 4377686, 4377687, 4377688, 4377690, 4377691, 4377692, 4377693, 4377694, 4377695, 4377696, 4377697, 4377698, 4377699, 4377700, 4377701, 4377702, 4377703, 4377704, 4377705, 4377706, 4377707, 4377708, 4377709, 4377710, 4377711, 4377712, 4377713, 4377714, 4377715, 4377716, 4377717, 4377718, 4377719, 4377720, 4377721, 4377722, 4377723, 4377724, 4377725, 4377726, 4377727, 4377728, 4377729, 4377730, 4377731, 4377732, 4377733, 4377734, 4377735, 4377736, 4377738, 4377739, 4377740, 4377741, 4377742, 4377743, 4377744, 4377745, 4377746, 4377747, 4377748, 4377749, 4377750, 4377751, 4377752, 4377753, 4377754, 4377755, 4377756, 4377757, 4377758, 4377759, 4377760, 4377761, 4377762, 4377763, 4377765, 4377766, 4377767, 4377768, 4377769, 4377770, 4377771, 4377772, 4377773, 4377774, 4377775, 4377776, 4377777, 4377778, 4377779, 4377780, 4377781, 4377782, 4377783, 4377784, 4377785, 4377786, 4377787, 4377788, 4377789, 4377790, 4377791, 4377792, 4377793, 4377794, 4377795, 4377796, 4377797, 4377798, 4377799, 4377800, 4377801, 4377802, 4377803, 4377804, 4377805, 4377806, 4377807, 4377808, 4377809, 4377810, 4377811, 4377812, 4377813, 4377814, 4377815, 4377816, 4377817, 4377818, 4377819, 4377820, 4377821, 4377822, 4377824, 4377826, 4377827, 4377828, 4377829, 4377830, 4377831, 4377832, 4377833, 4377834, 4377835, 4377836, 4377837, 4377838, 4377839, 4377840, 4377841, 4377842, 4377843, 4377844, 4377845, 4377846, 4377847, 4377848, 4377849, 4377851, 4377852, 4377853, 4377854, 4377855, 4377856, 4377857, 4377858, 4377859, 4377860, 4377861, 4377862, 4377863, 4377864, 4377865, 4377866, 4377867, 4377868, 4377869, 4377870, 4377871, 4377872, 4377873, 4377874, 4377875, 4377876, 4377877, 4377878, 4377879, 4377880, 4377881, 4377882, 4377883, 4377884, 4377885, 4377886, 4377887, 4377888, 4377889, 4377890, 4377891, 4377892, 4377893, 4377894, 4377895, 4377897, 4377898, 4377899, 4377900, 4377901, 4377902, 4377903, 4377904, 4377905, 4377906, 4377907, 4377908, 4377909, 4377910, 4377911, 4377912, 4377913, 4377914, 4377915, 4377916, 4377917, 4377918, 4377919, 4377920, 4377921, 4377922, 4377923, 4377924, 4377925, 4377926, 4377927, 4377928, 4377929, 4377931, 4377932, 4377933, 4377934, 4377935, 4377936, 4377937, 4377938, 4377939, 4377940, 4377941, 4377942, 4377943, 4377944, 4377945, 4377946, 4377947, 4377948, 4377949, 4377950, 4377951, 4377952, 4377953, 4377954, 4377955, 4377956, 4377957, 4377958, 4377959, 4377960, 4377961, 4377962, 4377963, 4377964, 4377965, 4377966, 4377967, 4377968, 4377969, 4377970, 4377971, 4377972, 4377973, 4377974, 4377975, 4377976, 4377977, 4377978, 4377979, 4377980, 4377981, 4377982, 4377983, 4377984, 4377985, 4377986, 4377987, 4377988, 4377989, 4377990, 4377991, 4377992, 4377993, 4377994, 4377995, 4377996, 4377997, 4377998, 4377999, 4378000, 4378001, 4378002, 4378003, 4378004, 4378005, 4378006, 4378007, 4378008, 4378009, 4378010, 4378011, 4378013, 4378014, 4378015, 4378016, 4378017, 4378018, 4378019, 4378020, 4378021, 4378022, 4378023, 4378024, 4378025, 4378026, 4378028, 4378029, 4378030, 4378031, 4378032, 4378033, 4378034, 4378035, 4378036, 4378037, 4378038, 4378039, 4378040, 4378041, 4378042, 4378043, 4378044, 4378045, 4378046, 4378047, 4378048, 4378049, 4378050, 4378051, 4378052, 4378053, 4378054, 4378055, 4378056, 4378057, 4378058, 4378059, 4378060, 4378061, 4378062, 4378063, 4378064, 4378065, 4378066, 4378067, 4378068, 4378069, 4378070, 4378071, 4378072, 4378073, 4378074, 4378075, 4378076, 4378077, 4378091, 4378092, 4378093, 4378094, 4378095, 4378096, 4378097, 4378098, 4378099, 4378100, 4378101, 4378102, 4378103, 4378105, 4378106, 4378107, 4378108, 4378109, 4378110, 4378111, 4378112, 4378113, 4378114, 4378115, 4378116, 4378117, 4378118, 4378119, 4378120, 4378121, 4378122, 4378123, 4378124, 4378125, 4378126, 4378127, 4378128, 4378129, 4378130, 4378131, 4378133, 4378134, 4378135, 4378136, 4378137, 4378138, 4378139, 4378140, 4378141, 4378142, 4378143, 4378144, 4378145, 4378146, 4378147, 4378148, 4378149, 4378150, 4378151, 4378152, 4378153, 4378154, 4378155, 4378156, 4378157, 4378158, 4378160, 4378161, 4378162, 4378163, 4378164, 4378165, 4378166, 4378167, 4378168, 4378169, 4378170, 4378171, 4378172, 4378173, 4378174, 4378175, 4378176, 4378177, 4378178, 4378179, 4378180, 4378181, 4378182, 4378183, 4378184, 4378185, 4378186, 4378187, 4378188, 4378189, 4378190, 4378191, 4378192, 4378193, 4378194, 4378195, 4378196, 4378197, 4378198, 4378199, 4378200, 4378201, 4378204, 4378205, 4378206, 4378207, 4378208, 4378209, 4378210, 4378211, 4378212, 4378213, 4378214, 4378215, 4378216, 4378217, 4378218, 4378219, 4378220, 4378221, 4378222, 4378223, 4378224, 4378225, 4378226, 4378227, 4378228, 4378229, 4378230, 4378231, 4378232, 4378233, 4378234, 4378235, 4378236, 4378237, 4378238, 4378239, 4378240, 4378241, 4378242, 4378243, 4378244, 4378245, 4378246, 4378247, 4378248, 4378249, 4378250, 4378251, 4378252, 4378253, 4378254, 4378255, 4378256, 4378257, 4378258, 4378259, 4378260, 4378261, 4378262, 4378263, 4378264, 4378265, 4378266, 4378267, 4378268, 4378269, 4378270, 4378271, 4378272, 4378273, 4378274, 4378275, 4378276, 4378277, 4378278, 4378279, 4378280, 4378281, 4378282, 4378283, 4378284, 4378285, 4378286, 4378287, 4378288, 4378289, 4378290, 4378291, 4378292, 4378293, 4378294, 4378295, 4378296, 4378297, 4378298, 4378299, 4378300, 4378301, 4378302, 4378303, 4378304, 4378306, 4378307, 4378308, 4378309, 4378310, 4378311, 4378312, 4378313, 4378314, 4378315, 4378316, 4378317, 4378318, 4378319, 4378321, 4378322, 4378323, 4378325, 4378326, 4378327, 4378328, 4378329, 4378330, 4378332, 4378333, 4378334, 4378335, 4378336, 4378337, 4378338, 4378339, 4378340, 4378341, 4378343, 4378344, 4378345, 4378347, 4378348, 4378350, 4378352, 4378353, 4378354, 4378355, 4378356, 4378357, 4378358, 4378359, 4378360, 4378361, 4378362, 4378363, 4378364, 4378365, 4378366, 4378367, 4378369, 4378370, 4378371, 4378372, 4378373, 4378374, 4378375, 4378376, 4378377, 4378378, 4378379, 4378380, 4378381, 4378382, 4378383, 4378384, 4378385, 4378386, 4378387, 4378388, 4378389, 4378390, 4378391, 4378392, 4378393, 4378394, 4378395, 4378396, 4378397, 4378398, 4378399, 4378400, 4378401, 4378402, 4378403, 4378404, 4378405, 4378406, 4378407, 4378408, 4378409, 4378410, 4378411, 4378412, 4378413, 4378414, 4378415, 4378416, 4378417, 4378418, 4378419, 4378420, 4378421, 4378422, 4378423, 4378424, 4378425, 4378426, 4378427, 4378428, 4378429, 4378430, 4378431, 4378432, 4378433, 4378434, 4378435, 4378436, 4378437, 4378438, 4378439, 4378440, 4378441, 4378442, 4378443, 4378444, 4378445, 4378446, 4378447, 4378448, 4378449, 4378450, 4378451, 4378452, 4378453, 4378454, 4378455, 4378456, 4378457, 4378458, 4378459, 4378460, 4378461, 4378462, 4378463, 4378464, 4378465, 4378466, 4378467, 4378468, 4378469, 4378470, 4378471, 4378472, 4378473, 4378474, 4378475, 4378476, 4378477, 4378478, 4378479, 4378480, 4378481, 4378482, 4378483, 4378484, 4378485, 4378486, 4378487, 4378488, 4378489, 4378490, 4378491, 4378492, 4378493, 4378494, 4378495, 4378496, 4378497, 4378498, 4378499, 4378500, 4378501, 4378502, 4378503, 4378504, 4378505, 4378506, 4378507, 4378508, 4378509, 4378510, 4378511, 4378512, 4378513, 4378514, 4378515, 4378516, 4378517, 4378518, 4378519, 4378520, 4378521, 4378522, 4378523, 4378524, 4378525, 4378526, 4378527, 4378528, 4378529, 4378530, 4378531, 4378532, 4378533, 4378534, 4378535, 4378536, 4378537, 4378538, 4378539, 4378540, 4378541, 4378542, 4378543, 4378544, 4378545, 4378546, 4378547, 4378548, 4378549, 4378550, 4378551, 4378552, 4378553, 4378554, 4378555, 4378556, 4378557, 4378558, 4378559, 4378560, 4378561, 4378562, 4378563, 4378564, 4378565, 4378566, 4378567, 4378568, 4378569, 4378570, 4378571, 4378572, 4378573, 4378574, 4378575, 4378576, 4378577, 4378578, 4378579, 4378580, 4378581, 4378582, 4378583, 4378584, 4378585, 4378586, 4378587, 4378588, 4378589, 4378590, 4378591, 4378592, 4378593, 4378594, 4378595, 4378596, 4378597, 4378598, 4378599, 4378600, 4378601, 4378602, 4378603, 4378604, 4378605, 4378606, 4378607, 4378608, 4378610, 4378611, 4378613, 4378614, 4378615, 4378616, 4378617, 4378618, 4378620, 4378621, 4378622, 4378623, 4378624, 4378625, 4378626, 4378627, 4378628, 4378629, 4378630, 4378631, 4378632, 4378633, 4378634, 4378635, 4378636, 4378637, 4378638, 4378639, 4378640, 4378641, 4378642, 4378643, 4378644, 4378645, 4378646, 4378647, 4378648, 4378649, 4378650, 4378651, 4378652, 4378653, 4378654, 4378655, 4378656, 4378657, 4378658, 4378659, 4378660, 4378661, 4378662, 4378663, 4378664, 4378665, 4378666, 4378667, 4378668, 4378669, 4378670, 4378671, 4378672, 4378674, 4378675, 4378676, 4378677, 4378678, 4378679, 4378681, 4378682, 4378683, 4378684, 4378685, 4378686, 4378689, 4378690, 4378691, 4378692, 4378693, 4378694, 4378695, 4378696, 4378697, 4378698, 4378699, 4378700, 4378701, 4378702, 4378703, 4378704, 4378705, 4378706, 4378707, 4378708, 4378709, 4378710, 4378711, 4378712, 4378713, 4378714, 4378715, 4378716, 4378717, 4378718, 4378719, 4378720, 4378721, 4378722, 4378723, 4378724, 4378725, 4378726, 4378727, 4378728, 4378729, 4378730, 4378731, 4378732, 4378733, 4378734, 4378735, 4378736, 4378737, 4378738, 4378739, 4378740, 4378741, 4378742, 4378743, 4378744, 4378745, 4378746, 4378747, 4378748, 4378749, 4378750, 4378752, 4378753, 4378754, 4378756, 4378757, 4378758, 4378759, 4378760, 4378761, 4378762, 4378763, 4378764, 4378765, 4378766, 4378767, 4378768, 4378769, 4378770, 4378771, 4378772, 4378773, 4378774, 4378775, 4378776, 4378777, 4378778, 4378779, 4378780, 4378781, 4378782, 4378783, 4378784, 4378785, 4378786, 4378787, 4378788, 4378789, 4378790, 4378791, 4378792, 4378793, 4378794, 4378795, 4378796, 4378797, 4378798, 4378799, 4378800, 4378801, 4378802, 4378803, 4378804, 4378805, 4378806, 4378807, 4378808, 4378809, 4378810, 4378811, 4378812, 4378813, 4378814, 4378815, 4378816, 4378817, 4378818, 4378819, 4378820, 4378821, 4378822, 4378823, 4378824, 4378825, 4378826, 4378827, 4378828, 4378829, 4378830, 4378832, 4378833, 4378834, 4378835, 4378836, 4378837, 4378838, 4378839, 4378840, 4378841, 4378842, 4378843, 4378844, 4378845, 4378846, 4378847, 4378848, 4378849, 4378850, 4378851, 4378852, 4378853, 4378854, 4378855, 4378856, 4378857, 4378858, 4378859, 4378860, 4378861, 4378862, 4378863, 4378864, 4378865, 4378866, 4378867, 4378868, 4378869, 4378870, 4378871, 4378872, 4378873, 4378874, 4378875, 4378876, 4378877, 4378878, 4378879, 4378880, 4378881, 4378882, 4378883, 4378884, 4378885, 4378886, 4378887, 4378888, 4378889, 4378890, 4378891, 4378892, 4378893, 4378894, 4378895, 4378896, 4378897, 4378898, 4378899, 4378900, 4378901, 4378902, 4378903, 4378904, 4378905, 4378906, 4378907, 4378908, 4378909, 4378910, 4378911, 4378912, 4378913, 4378914, 4378915, 4378916, 4378917, 4378918, 4378919, 4378920, 4378921, 4378922, 4378923, 4378924, 4378925, 4378926, 4378927, 4378928, 4378929, 4378930, 4378931, 4378932, 4378933, 4378934, 4378935, 4378936, 4378937, 4378938, 4378939, 4378940, 4378941, 4378942, 4378943, 4378944, 4378945, 4378946, 4378947, 4378948, 4378949, 4378950, 4378951, 4378952, 4378953, 4378954, 4378955, 4378956, 4378957, 4378958, 4378959, 4378960, 4378961, 4378962, 4378963, 4378964, 4378965, 4378966, 4378967, 4378968, 4378969, 4378970, 4378971, 4378972, 4378973, 4378974, 4378975, 4378976, 4378977, 4378978, 4378979, 4378980, 4378981, 4378982, 4378983, 4378984, 4378985, 4378986, 4378987, 4378988, 4378989, 4378990, 4378991, 4378992, 4378993, 4378994, 4378995, 4378996, 4378997, 4378998, 4378999, 4379000, 4379001, 4379002, 4379003, 4379004, 4379005, 4379006, 4379007, 4379008, 4379009, 4379010, 4379015, 4379031, 4379032, 4379033, 4379034, 4379035, 4379036, 4379037, 4379038, 4379039, 4379040, 4379041, 4379042, 4379043, 4379044, 4379045, 4379046, 4379047, 4379048, 4379049, 4379050, 4379055, 4379066, 4379069, 4379070, 4379071, 4379072, 4379073, 4379074, 4379075, 4379076, 4379077, 4379078, 4379079, 4379080, 4379081, 4379082, 4379083, 4379084, 4379085, 4379086, 4379087, 4379088, 4379089, 4379090, 4379091, 4379092, 4379093, 4379094, 4379095, 4379096, 4379097, 4379098, 4379117, 4379118, 4379119, 4379120, 4379121, 4379122, 4379123, 4379124, 4379125, 4379126, 4379128, 4379129, 4379131, 4379132, 4379133, 4379134, 4379135, 4379136, 4379139, 4379140, 4379141, 4379142, 4379143, 4379144, 4379145, 4379146, 4379147, 4379148, 4379149, 4379150, 4379151, 4379152, 4379153, 4379154, 4379155, 4379156, 4379157, 4379158, 4379159, 4379160, 4379161, 4379163, 4379165, 4379166, 4379167, 4379168, 4379169, 4379170, 4379171, 4379172, 4379173, 4379174, 4379175, 4379176, 4379177, 4379178, 4379179, 4379180, 4379181, 4379182, 4379183, 4379184, 4379185, 4379186, 4379187, 4379189, 4379191, 4379192, 4379193, 4379194, 4379195, 4379196, 4379197, 4379198, 4379199, 4379200, 4379201, 4379202, 4379203, 4379204, 4379205, 4379206, 4379207, 4379208, 4379209, 4379210, 4379211, 4379212, 4379214, 4379215, 4379216, 4379217, 4379218, 4379219, 4379220, 4379221, 4379222, 4379223, 4379224, 4379225, 4379226, 4379227, 4379228, 4379229, 4379230, 4379231, 4379232, 4379233, 4379234, 4379235, 4379236, 4379237, 4379238, 4379239, 4379240, 4379241, 4379242, 4379243, 4379244, 4379245, 4379246, 4379247, 4379248, 4379249, 4379250, 4379251, 4379252, 4379254, 4379255, 4379256, 4379257, 4379258, 4379259, 4379260, 4379261, 4379262, 4379263, 4379264, 4379265, 4379266, 4379267, 4379268, 4379269, 4379270, 4379271, 4379272, 4379273, 4379274, 4379275, 4379276, 4379277, 4379278, 4379279, 4379280, 4379281, 4379282, 4379283, 4379284, 4379285, 4379286, 4379287, 4379288, 4379289, 4379290, 4379291, 4379292, 4379293, 4379294, 4379295, 4379296, 4379297, 4379298, 4379299, 4379300, 4379301, 4379302, 4379303, 4379304, 4379305, 4379306, 4379307, 4379308, 4379309, 4379310, 4379311, 4379312, 4379313, 4379314, 4379315, 4379316, 4379317, 4379359, 4379360, 4379368, 4379369, 4379370, 4379371, 4379372, 4379373, 4379374, 4379375, 4379376, 4379377, 4379378, 4379379, 4379381, 4379382, 4379383, 4379384, 4379385, 4379386, 4379387, 4379389, 4379390, 4379403, 4379415, 4379432, 4379439, 4379445, 4379446, 4379447, 4379448, 4379449, 4379451, 4379452, 4379453, 4379454, 4379456, 4379461, 4379462, 4379463, 4379465, 4379466, 4379467, 4379472, 4379474, 4379480, 4379481, 4379482, 4379483, 4379484, 4379485, 4379486, 4379487, 4379488, 4379489, 4379490, 4379491, 4379492, 4379493, 4379494, 4379495, 4379496, 4379497, 4379498, 4379499, 4379500, 4379501, 4379502, 4379503, 4379504, 4379505, 4379506, 4379507, 4379508, 4379509, 4379510, 4379511, 4379512, 4379513, 4379514, 4379515, 4379516, 4379517, 4379518, 4379519, 4379520, 4379521, 4379522, 4379523, 4379524, 4379525, 4379526, 4379527, 4379528, 4379529, 4379530, 4379531, 4379532, 4379533, 4379535, 4379537, 4379538, 4379539, 4379540, 4379541, 4379542, 4379543, 4379544, 4379545, 4379546, 4379547, 4379548, 4379549, 4379550, 4379551, 4379552, 4379553, 4379554, 4379555, 4379556, 4379557, 4379558, 4379559, 4379560, 4379561, 4379562, 4379564, 4379565, 4379566, 4379567, 4379568, 4379569, 4379570, 4379571, 4379572, 4379573, 4379574, 4379575, 4379576, 4379577, 4379578, 4379579, 4379580, 4379581, 4379582, 4379583, 4379584, 4379585, 4379586, 4379587, 4379588, 4379589, 4379590, 4379591, 4379592, 4379593, 4379594, 4379595, 4379598, 4379599, 4379601, 4379602, 4379603, 4379604, 4379605, 4379606, 4379608, 4379609, 4379610, 4379611, 4379612, 4379613, 4379614, 4379615, 4379616, 4379617, 4379618, 4379619, 4379620, 4379621, 4379622, 4379623, 4379624, 4379625, 4379626, 4379627, 4379628, 4379629, 4379630, 4379631, 4379632, 4379633, 4379634, 4379636, 4379637, 4379638, 4379639, 4379641, 4379642, 4379643, 4379644, 4379646, 4379647, 4379648, 4379649, 4379650, 4379651, 4379652, 4379653, 4379654, 4379655, 4379656, 4379658, 4379659, 4379660, 4379662, 4379663, 4379664, 4379665, 4379666, 4379667, 4379668, 4379669, 4379670, 4379671, 4379672, 4379673, 4379674, 4379675, 4379676, 4379677, 4379678, 4379679, 4379680, 4379681, 4379682, 4379683, 4379684, 4379685, 4379686, 4379687, 4379688, 4379689, 4379690, 4379691, 4379692, 4379693, 4379694, 4379696, 4379698, 4379699, 4379700, 4379701, 4379702, 4379703, 4379704, 4379705, 4379706, 4379707, 4379708, 4379709, 4379710, 4379711, 4379712, 4379713, 4379714, 4379715, 4379716, 4379717, 4379718, 4379719, 4379720, 4379721, 4379722, 4379724, 4379725, 4379726, 4379727, 4379728, 4379729, 4379730, 4379731, 4379732, 4379733, 4379734, 4379736, 4379737, 4379738, 4379739, 4379740, 4379741, 4379743, 4379744, 4379745, 4379747, 4379748, 4379749, 4379750, 4379751, 4379753, 4379754, 4379755, 4379756, 4379757, 4379758, 4379759, 4379760, 4379761, 4379762, 4379763, 4379764, 4379765, 4379766, 4379767, 4379768, 4379769, 4379770, 4379771, 4379772, 4379773, 4379774, 4379775, 4379776, 4379777, 4379778, 4379779, 4379780, 4379781, 4379782, 4379783, 4379784, 4379786, 4379787, 4379788, 4379789, 4379790, 4379791, 4379792, 4379793, 4379795, 4379796, 4379798, 4379799, 4379800, 4379801, 4379802, 4379803, 4379804, 4379805, 4379806, 4379807, 4379808, 4379809, 4379810, 4379811, 4379812, 4379813, 4379814, 4379815, 4379816, 4379817, 4379819, 4379820, 4379821, 4379822, 4379823, 4379824, 4379825, 4379826, 4379828, 4379829, 4379830, 4379831, 4379832, 4379833, 4379834, 4379836, 4379837, 4379838, 4379839, 4379840, 4379841, 4379842, 4379843, 4379844, 4379845, 4379846, 4379847, 4379848, 4379849, 4379850, 4379851, 4379852, 4379853, 4379854, 4379855, 4379856, 4379857, 4379858, 4379859, 4379860, 4379861, 4379862, 4379864, 4379865, 4379866, 4379867, 4379868, 4379869, 4379870, 4379871, 4379872, 4379873, 4379874, 4379875, 4379876, 4379877, 4379878, 4379879, 4379880, 4379881, 4379882, 4379883, 4379884, 4379885, 4379886, 4379887, 4379888, 4379889, 4379890, 4379891, 4379892, 4379893, 4379894, 4379895, 4379896, 4379897, 4379898, 4379899, 4379900, 4379901, 4379902, 4379903, 4379904, 4379905, 4379906, 4379907, 4379908, 4379909, 4379910, 4379911, 4379912, 4379913, 4379914, 4379915, 4379916, 4379917, 4379918, 4379919, 4379920, 4379921, 4379922, 4379923, 4379924, 4379925, 4379926, 4379927, 4379928, 4379929, 4379930, 4379931, 4379932, 4379933, 4379934, 4379935, 4379936, 4379937, 4379938, 4379941, 4379942, 4379943, 4379944, 4379945, 4379946, 4379947, 4379948, 4379949, 4379950, 4379951, 4379952, 4379954, 4379955, 4379956, 4379957, 4379958, 4379959, 4379960, 4379961, 4379962, 4379963, 4379964, 4379965, 4379966, 4379967, 4379968, 4379969, 4379970, 4379971, 4379972, 4379973, 4379974, 4379975, 4379976, 4379977, 4379978, 4379979, 4379980, 4379982, 4379983, 4379984, 4379985, 4379986, 4379987, 4379988, 4379989, 4379990, 4379991, 4379992, 4379993, 4379994, 4379995, 4379997, 4379998, 4379999, 4380000, 4380001, 4380002, 4380003, 4380004, 4380005, 4380006, 4380007, 4380009, 4380010, 4380011, 4380012, 4380013, 4380014, 4380015, 4380016, 4380017, 4380018, 4380019, 4380020, 4380021, 4380022, 4380023, 4380024, 4380025, 4380026, 4380027, 4380028, 4380029, 4380030, 4380031, 4380032, 4380033, 4380034, 4380035, 4380036, 4380037, 4380038, 4380040, 4380041, 4380042, 4380043, 4380044, 4380045, 4380046, 4380047, 4380048, 4380049, 4380050, 4380051, 4380052, 4380053, 4380054, 4380055, 4380056, 4380057, 4380058, 4380060, 4380063, 4380064, 4380066, 4380067, 4380069, 4380070, 4380072, 4380073, 4380074, 4380076, 4380078, 4380079, 4380080, 4380081, 4380083, 4380084, 4380086, 4380089, 4380090, 4380092, 4380093, 4380094, 4380095, 4380096, 4380097, 4380098, 4380099, 4380102, 4380103, 4380104, 4380105, 4380106, 4380107, 4380109, 4380110, 4380111, 4380112, 4380113, 4380114, 4380116, 4380117, 4380118, 4380119, 4380120, 4380121, 4380122, 4380123, 4380124, 4380125, 4380126, 4380127, 4380128, 4380129, 4380130, 4380131, 4380132, 4380133, 4380134, 4380135, 4380136, 4380138, 4380139, 4380140, 4380141, 4380142, 4380143, 4380144, 4380145, 4380146, 4380147, 4380148, 4380149, 4380150, 4380151, 4380152, 4380153, 4380154, 4380155, 4380156, 4380157, 4380158, 4380159, 4380160, 4380161, 4380162, 4380163, 4380164, 4380165, 4380166, 4380167, 4380168, 4380169, 4380170, 4380171, 4380172, 4380173, 4380174, 4380175, 4380176, 4380177, 4380178, 4380179, 4380180, 4380181, 4380182, 4380183, 4380184, 4380185, 4380186, 4380187, 4380188, 4380189, 4380190, 4380191, 4380192, 4380193, 4380194, 4380195, 4380196, 4380197, 4380198, 4380199, 4380200, 4380201, 4380202, 4380203, 4380204, 4380205, 4380206, 4380207, 4380208, 4380209, 4380210, 4380211, 4380212, 4380213, 4380214, 4380215, 4380216, 4380217, 4380218, 4380219, 4380220, 4380221, 4380222, 4380223, 4380224, 4380225, 4380226, 4380227, 4380228, 4380229, 4380230, 4380231, 4380232, 4380233, 4380234, 4380235, 4380236, 4380237, 4380238, 4380239, 4380240, 4380241, 4380242, 4380243, 4380244, 4380245, 4380246, 4380247, 4380248, 4380249, 4380250, 4380251, 4380252, 4380253, 4380254, 4380255, 4380256, 4380257, 4380258, 4380259, 4380260, 4380261, 4380262, 4380263, 4380264, 4380265, 4380266, 4380267, 4380268, 4380269, 4380270, 4380271, 4380272, 4380273, 4380274, 4380275, 4380276, 4380277, 4380278, 4380279, 4380280, 4380281, 4380282, 4380283, 4380284, 4380285, 4380286, 4380287, 4380288, 4380289, 4380292, 4380293, 4380294, 4380295, 4380296, 4380297, 4380298, 4380299, 4380300, 4380301, 4380302, 4380303, 4380304, 4380305, 4380306, 4380307, 4380308, 4380309, 4380310, 4380311, 4380312, 4380313, 4380314, 4380315, 4380316, 4380317, 4380318, 4380319, 4380320, 4380321, 4380322, 4380323, 4380324, 4380325, 4380326, 4380327, 4380328, 4380329, 4380330, 4380331, 4380332, 4380333, 4380334, 4380335, 4380337, 4380338, 4380339, 4380340, 4380341, 4380342, 4380343, 4380344, 4380345, 4380346, 4380347, 4380348, 4380349, 4380350, 4380351, 4380352, 4380353, 4380354, 4380355, 4380356, 4380357, 4380358, 4380359, 4380360, 4380361, 4380362, 4380363, 4380364, 4380365, 4380366, 4380367, 4380368, 4380369, 4380370, 4380371, 4380372, 4380373, 4380374, 4380375, 4380376, 4380377, 4380378, 4380379, 4380380, 4380381, 4380382, 4380383, 4380384, 4380385, 4380386, 4380387, 4380388, 4380389, 4380390, 4380391, 4380392, 4380394, 4380395, 4380396, 4380397, 4380398, 4380399, 4380400, 4380401, 4380402, 4380403, 4380404, 4380405, 4380406, 4380407, 4380408, 4380409, 4380410, 4380411, 4380412, 4380413, 4380414, 4380415, 4380416, 4380417, 4380418, 4380419, 4380420, 4380421, 4380422, 4380423, 4380424, 4380425, 4380426, 4380427, 4380428, 4380429, 4380430, 4380431, 4380432, 4380433, 4380434, 4380435, 4380436, 4380437, 4380438, 4380439, 4380440, 4380441, 4380442, 4380443, 4380444, 4380445, 4380446, 4380447, 4380448, 4380449, 4380450, 4380451, 4380452, 4380453, 4380454, 4380455, 4380456, 4380457, 4380458, 4380459, 4380460, 4380461, 4380462, 4380463, 4380464, 4380465, 4380466, 4380467, 4380468, 4380469, 4380470, 4380471, 4380472, 4380473, 4380474, 4380475, 4380476, 4380477, 4380478, 4380479, 4380480, 4380481, 4380482, 4380483, 4380484, 4380485, 4380486, 4380487, 4380488, 4380489, 4380490, 4380492, 4380493, 4380494, 4380495, 4380496, 4380497, 4380498, 4380499, 4380500, 4380501, 4380502, 4380503, 4380504, 4380505, 4380506, 4380507, 4380508, 4380509, 4380510, 4380511, 4380512, 4380513, 4380514, 4380515, 4380516, 4380517, 4380518, 4380519, 4380520, 4380524, 4380526, 4380527, 4380529, 4380534, 4380535, 4380539, 4380540, 4380541, 4380542, 4380543, 4380544, 4380545, 4380546, 4380547, 4380548, 4380549, 4380550, 4380551, 4380552, 4380555, 4380557, 4380558, 4380559, 4380560, 4380561, 4380562, 4380564, 4380565, 4380566, 4380567, 4380568, 4380569, 4380571, 4380572, 4380573, 4380574, 4380575, 4380576, 4380577, 4380578, 4380579, 4380580, 4380581, 4380582, 4380583, 4380584, 4380585, 4380586, 4380587, 4380588, 4380589, 4380590, 4380591, 4380592, 4380593, 4380594, 4380595, 4380596, 4380597, 4380598, 4380599, 4380601, 4380602, 4380603, 4380604, 4380605, 4380606, 4380607, 4380608, 4380609, 4380610, 4380611, 4380612, 4380613, 4380614, 4380615, 4380616, 4380617, 4380618, 4380619, 4380620, 4380621, 4380622, 4380623, 4380624, 4380625, 4380626, 4380627, 4380628, 4380629, 4380630, 4380631, 4380632, 4380633, 4380635, 4380636, 4380637, 4380638, 4380639, 4380640, 4380641, 4380642, 4380643, 4380644, 4380645, 4380646, 4380647, 4380648, 4380649, 4380650, 4380651, 4380652, 4380653, 4380654, 4380655, 4380659, 4380663, 4380666, 4380676, 4380677, 4380678, 4380679, 4380680, 4380681, 4380682, 4380683, 4380684, 4380685, 4380686, 4380687, 4380688, 4380689, 4380690, 4380691, 4380692, 4380694, 4380695, 4380696, 4380697, 4380698, 4380699, 4380700, 4380701, 4380702, 4380703, 4380704, 4380705, 4380706, 4380708, 4380709, 4380710, 4380711, 4380712, 4380713, 4380719, 4380720, 4380721, 4380722, 4380723, 4380726, 4380727, 4380728, 4380729, 4380731, 4380732, 4380733, 4380734, 4380735, 4380736, 4380737, 4380738, 4380739, 4380740, 4380741, 4380742, 4380743, 4380744, 4380745, 4380746, 4380747, 4380748, 4380749, 4380750, 4380751, 4380752, 4380753, 4380754, 4380755, 4380756, 4380757, 4380758, 4380759, 4380760, 4380761, 4380762, 4380763, 4380764, 4380765, 4380766, 4380767, 4380768, 4380769, 4380770, 4380771, 4380772, 4380773, 4380774, 4380775, 4380776, 4380777, 4380778, 4380780, 4380781, 4380782, 4380783, 4380784, 4380785, 4380786, 4380787, 4380788, 4380789, 4380790, 4380791, 4380792, 4380793, 4380794, 4380795, 4380796, 4380797, 4380798, 4380799, 4380800, 4380801, 4380802, 4380803, 4380804, 4380805, 4380806, 4380807, 4380808, 4380809, 4380810, 4380811, 4380812, 4380813, 4380814, 4380815, 4380816, 4380817, 4380818, 4380819, 4380820, 4380821, 4380822, 4380823, 4380824, 4380825, 4380826, 4380827, 4380828, 4380829, 4380830, 4380831, 4380832, 4380833, 4380834, 4380835, 4380836, 4380837, 4380838, 4380840, 4380841, 4380842, 4380843, 4380844, 4380845, 4380846, 4380847, 4380848, 4380849, 4380850, 4380851, 4380852, 4380853, 4380854, 4380855, 4380856, 4380857, 4380858, 4380859, 4380860, 4380861, 4380862, 4380863, 4380864, 4380865, 4380866, 4380867, 4380868, 4380869, 4380870, 4380871, 4380872, 4380873, 4380874, 4380875, 4380876, 4380877, 4380878, 4380879, 4380880, 4380881, 4380882, 4380883, 4380884, 4380888, 4380889, 4380891, 4380893, 4380894, 4380895, 4380896, 4380897, 4380902, 4380904, 4380907, 4380909, 4380910, 4380911, 4380912, 4380913, 4380914, 4380916, 4380917, 4380918, 4380919, 4380920, 4380921, 4380922, 4380923, 4380924, 4380925, 4380926, 4380927, 4380929, 4380932, 4380933, 4380935, 4380937, 4380939, 4380940, 4380941, 4380942, 4380943, 4380944, 4380945, 4380947, 4380948, 4380949, 4380950, 4380951, 4380952, 4380955, 4380956, 4380957, 4380958, 4380959, 4380960, 4380961, 4380962, 4380963, 4380964, 4380965, 4380966, 4380967, 4380968, 4380969, 4380970, 4380971, 4380972, 4380973, 4380974, 4380975, 4380976, 4380977, 4380978, 4380982, 4380983, 4380984, 4380990, 4380991, 4380992, 4380993, 4380994, 4380996, 4380999, 4381000, 4381001, 4381002, 4381003, 4381004, 4381005, 4381006, 4381007, 4381008, 4381009, 4381010, 4381011, 4381012, 4381013, 4381014, 4381015, 4381016, 4381017, 4381018, 4381019, 4381020, 4381021, 4381022, 4381023, 4381024, 4381025, 4381026, 4381027, 4381028, 4381029, 4381030, 4381031, 4381032, 4381034, 4381035, 4381036, 4381037, 4381038, 4381039, 4381040, 4381041, 4381042, 4381043, 4381044, 4381045, 4381046, 4381047, 4381048, 4381049, 4381050, 4381051, 4381052, 4381053, 4381054, 4381055, 4381056, 4381057, 4381058, 4381059, 4381060, 4381061, 4381062, 4381063, 4381064, 4381065, 4381066, 4381067, 4381068, 4381069, 4381070, 4381071, 4381072, 4381073, 4381074, 4381075, 4381076, 4381077, 4381078, 4381079, 4381080, 4381081, 4381082, 4381083, 4381084, 4381085, 4381086, 4381087, 4381088, 4381089, 4381090, 4381091, 4381092, 4381093, 4381094, 4381095, 4381096, 4381097, 4381098, 4381099, 4381100, 4381101, 4381102, 4381103, 4381104, 4381105, 4381107, 4381108, 4381109, 4381110, 4381111, 4381112, 4381113, 4381114, 4381115, 4381116, 4381117, 4381118, 4381119, 4381120, 4381121, 4381122, 4381123, 4381124, 4381125, 4381126, 4381127, 4381128, 4381129, 4381130, 4381131, 4381132, 4381133, 4381134, 4381135, 4381136, 4381137, 4381138, 4381139, 4381140, 4381141, 4381142, 4381143, 4381144, 4381145, 4381146, 4381147, 4381148, 4381149, 4381150, 4381151, 4381152, 4381153, 4381154, 4381155, 4381156, 4381157, 4381158, 4381159, 4381160, 4381161, 4381162, 4381163, 4381164, 4381165, 4381166, 4381167, 4381168, 4381169, 4381170, 4381171, 4381172, 4381173, 4381174, 4381175, 4381176, 4381177, 4381178, 4381179, 4381180, 4381181, 4381182, 4381183, 4381184, 4381185, 4381186, 4381187, 4381188, 4381189, 4381190, 4381191, 4381192, 4381193, 4381194, 4381195, 4381196, 4381197, 4381198, 4381199, 4381200, 4381201, 4381202, 4381203, 4381204, 4381205, 4381206, 4381207, 4381208, 4381209, 4381210, 4381211, 4381212, 4381213, 4381214, 4381215, 4381216, 4381217, 4381218, 4381219, 4381220, 4381221, 4381222, 4381223, 4381224, 4381225, 4381226, 4381227, 4381228, 4381229, 4381230, 4381231, 4381232, 4381233, 4381234, 4381235, 4381236, 4381237, 4381238, 4381239, 4381240, 4381241, 4381242, 4381243, 4381244, 4381245, 4381246, 4381247, 4381248, 4381249, 4381250, 4381251, 4381252, 4381253, 4381254, 4381255, 4381256, 4381257, 4381258, 4381259, 4381260, 4381261, 4381262, 4381263, 4381264, 4381265, 4381266, 4381267, 4381268, 4381269, 4381270, 4381271, 4381272, 4381273, 4381274, 4381275, 4381276, 4381277, 4381278, 4381279, 4381280, 4381281, 4381282, 4381283, 4381284, 4381285, 4381286, 4381287, 4381288, 4381289, 4381290, 4381291, 4381292, 4381293, 4381294, 4381295, 4381296, 4381297, 4381298, 4381299, 4381300, 4381301, 4381302, 4381303, 4381304, 4381305, 4381306, 4381307, 4381308, 4381309, 4381310, 4381311, 4381312, 4381313, 4381314, 4381315, 4381316, 4381317, 4381318, 4381319, 4381320, 4381321, 4381322, 4381323, 4381324, 4381325, 4381326, 4381327, 4381328, 4381329, 4381330, 4381331, 4381332, 4381333, 4381334, 4381335, 4381336, 4381337, 4381338, 4381339, 4381340, 4381341, 4381342, 4381343, 4381344, 4381345, 4381346, 4381347, 4381348, 4381349, 4381351, 4381359, 4381360, 4381361, 4381364, 4381365, 4381366, 4381367, 4381368, 4381373, 4381376, 4381379, 4381381, 4381382, 4381383, 4381384, 4381386, 4381387, 4381388, 4381390, 4381391, 4381392, 4381393, 4381394, 4381395, 4381396, 4381399, 4381400, 4381401, 4381402, 4381403, 4381404, 4381408, 4381431, 4381452, 4381453, 4381454, 4381455, 4381456, 4381457, 4381458, 4381459, 4381460, 4381461, 4381462, 4381463, 4381464, 4381465, 4381466, 4381467, 4381468, 4381469, 4381470, 4381471, 4381472, 4381473, 4381474, 4381475, 4381476, 4381477, 4381478, 4381479, 4381480, 4381481, 4381482, 4381483, 4381484, 4381485, 4381486, 4381487, 4381488, 4381489, 4381490, 4381491, 4381492, 4381493, 4381494, 4381495, 4381496, 4381497, 4381498, 4381499, 4381500, 4381501, 4381502, 4381503, 4381504, 4381505, 4381506, 4381507, 4381508, 4381509, 4381510, 4381511, 4381512, 4381513, 4381514, 4381515, 4381516, 4381517, 4381518, 4381519, 4381520, 4381521, 4381522, 4381523, 4381524, 4381525, 4381526, 4381527, 4381528, 4381529, 4381530, 4381531, 4381532, 4381533, 4381534, 4381535, 4381536, 4381537, 4381538, 4381539, 4381540, 4381541, 4381542, 4381543, 4381544, 4381545, 4381546, 4381547, 4381548, 4381549, 4381553, 4381561, 4381563, 4381569, 4381570, 4381571, 4381572, 4381573, 4381574, 4381577, 4381578, 4381579, 4381580, 4381581, 4381582, 4381583, 4381584, 4381585, 4381586, 4381587, 4381588, 4381589, 4381590, 4381591, 4381592, 4381593, 4381594, 4381595, 4381596, 4381597, 4381598, 4381599, 4381600, 4381601, 4381602, 4381603, 4381604, 4381605, 4381606, 4381607, 4381608, 4381609, 4381610, 4381611, 4381612, 4381613, 4381614, 4381615, 4381616, 4381617, 4381618, 4381619, 4381620, 4381621, 4381622, 4381623, 4381624, 4381625, 4381626, 4381627, 4381628, 4381629, 4381630, 4381631, 4381632, 4381633, 4381634, 4381635, 4381636, 4381637, 4381638, 4381639, 4381640, 4381641, 4381642, 4381643, 4381644, 4381645, 4381646, 4381647, 4381648, 4381650, 4381651, 4381652, 4381653, 4381654, 4381655, 4381656, 4381659, 4381660, 4381661, 4381662, 4381663, 4381664, 4381665, 4381666, 4381667, 4381668, 4381669, 4381670, 4381671, 4381672, 4381673, 4381674, 4381675, 4381676, 4381677, 4381678, 4381679, 4381680, 4381681, 4381682, 4381683, 4381684, 4381685, 4381686, 4381687, 4381688, 4381689, 4381690, 4381691, 4381693, 4381694, 4381695, 4381696, 4381697, 4381698, 4381699, 4381700, 4381701, 4381702, 4381703, 4381704, 4381705, 4381706, 4381707, 4381708, 4381709, 4381710, 4381711, 4381712, 4381713, 4381714, 4381715, 4381716, 4381717, 4381718, 4381719, 4381720, 4381721, 4381723, 4381724, 4381726, 4381727, 4381728, 4381729, 4381730, 4381731, 4381732, 4381733, 4381734, 4381735, 4381736, 4381737, 4381738, 4381739, 4381740, 4381741, 4381742, 4381743, 4381744, 4381745, 4381746, 4381747, 4381748, 4381749, 4381750, 4381751, 4381752, 4381753, 4381754, 4381755, 4381756, 4381757, 4381758, 4381759, 4381760, 4381761, 4381762, 4381763, 4381764, 4381765, 4381766, 4381767, 4381768, 4381769, 4381770, 4381771, 4381772, 4381773, 4381774, 4381775, 4381776, 4381777, 4381778, 4381779, 4381780, 4381781, 4381782, 4381783, 4381784, 4381785, 4381786, 4381787, 4381788, 4381789, 4381790, 4381791, 4381792, 4381793, 4381794, 4381795, 4381796, 4381797, 4381798, 4381799, 4381800, 4381801, 4381802, 4381803, 4381804, 4381805, 4381806, 4381807, 4381808, 4381809, 4381810, 4381811, 4381812, 4381813, 4381814, 4381815, 4381816, 4381817, 4381818, 4381819, 4381820, 4381821, 4381822, 4381823, 4381824, 4381825, 4381826, 4381827, 4381828, 4381829, 4381830, 4381831, 4381832, 4381833, 4381834, 4381835, 4381836, 4381837, 4381838, 4381839, 4381840, 4381841, 4381842, 4381843, 4381845, 4381846, 4381847, 4381848, 4381849, 4381850, 4381851, 4381852, 4381853, 4381855, 4381856, 4381857, 4381858, 4381859, 4381860, 4381862, 4381863.4381864, 4381865, 4381866, 4381867, 4381868, 4381869, 4381870, 4381871, 4381872, 4381873, 4381874, 4381875, 4381876, 4381877, 4381878, 4381879, 4381880, 4381881, 4381882, 4381883, 4381884, 4381886, 4381887, 4381888, 4381889, 4381890, 4381891, 4381892, 4381893, 4381894, 4381895, 4381896, 4381897, 4381898, 4381899, 4381900, 4381901, 4381902, 4381903, 4381904, 4381905, 4381906, 4381907, 4381908, 4381909, 4381910, 4381911, 4381912, 4381913, 4381914, 4381915, 4381916, 4381917, 4381918, 4381919, 4381920, 4381921, 4381922, 4381924, 4381925, 4381926, 4381927, 4381928, 4381929, 4381930, 4381931, 4381932, 4381933, 4381934, 4381935, 4381936, 4381937, 4381938, 4381939, 4381940, 4381941, 4381942, 4381943, 4381944, 4381945, 4381947, 4381948, 4381949, 4381950, 4381951, 4381952, 4381953, 4381954, 4381955, 4381956, 4381957, 4381958, 4381959, 4381960, 4381961, 4381962, 4381963, 4381964, 4381965, 4381966, 4381968, 4381969, 4381971, 4381972, 4381973, 4381974, 4381975, 4381976, 4381977, 4381978, 4381979, 4381980, 4381981, 4381982, 4381984, 4381985, 4381986, 4381987, 4381988, 4381989, 4381990, 4381991, 4381992, 4381993, 4381994, 4381995, 4381996, 4381997, 4381998, 4381999, 4382000, 4382002, 4382003, 4382004, 4382005, 4382006, 4382007, 4382008, 4382009, 4382010, 4382011, 4382012, 4382013, 4382014, 4382015, 4382016, 4382017, 4382018, 4382019, 4382020, 4382021, 4382022, 4382023, 4382024, 4382025, 4382026, 4382027, 4382028, 4382029, 4382030, 4382031, 4382032, 4382033, 4382034, 4382035, 4382036, 4382037, 4382038, 4382039, 4382040, 4382041, 4382042, 4382043, 4382044, 4382045, 4382046, 4382047, 4382048, 4382049, 4382050, 4382052, 4382053, 4382054, 4382055, 4382056, 4382057, 4382058, 4382059, 4382060, 4382061, 4382062, 4382063, 4382064, 4382066, 4382067, 4382068, 4382069, 4382070, 4382071, 4382072, 4382073, 4382074, 4382075, 4382076, 4382077, 4382078, 4382079, 4382080, 4382081, 4382082, 4382083, 4382084, 4382085, 4382086, 4382087, 4382090, 4382092, 4382093, 4382094, 4382095, 4382096, 4382098, 4382099, 4382100, 4382101, 4382104, 4382105, 4382107, 4382108, 4382110, 4382111, 4382113, 4382114, 4382115, 4382118, 4382120, 4382121, 4382122, 4382123, 4382125, 4382126, 4382127, 4382130, 4382132, 4382133, 4382134, 4382135, 4382136, 4382137, 4382138, 4382139, 4382140, 4382141, 4382142, 4382143, 4382144, 4382145, 4382146, 4382147, 4382148, 4382149, 4382150, 4382151, 4382152, 4382153, 4382154, 4382155, 4382156, 4382158, 4382159, 4382160, 4382161, 4382162, 4382163, 4382164, 4382165, 4382166, 4382167, 4382168, 4382169, 4382170, 4382171, 4382172, 4382173, 4382175, 4382177, 4382178, 4382181, 4382183, 4382185, 4382188, 4382192, 4382193, 4382195, 4382196, 4382199, 4382201, 4382204, 4382205, 4382207, 4382209, 4382210, 4382211, 4382214, 4382219, 4382225, 4382228, 4382229, 4382230, 4382232, 4382233, 4382234, 4382235, 4382239, 4382240, 4382241, 4382243, 4382245, 4382246, 4382247, 4382249, 4382251, 4382252, 4382253, 4382254, 4382255, 4382256, 4382257, 4382260, 4382261, 4382262, 4382263, 4382264, 4382265, 4382266, 4382267, 4382268, 4382269, 4382271, 4382273, 4382274, 4382276, 4382278, 4382280, 4382281, 4382282, 4382283, 4382285, 4382286, 4382287, 4382288, 4382291, 4382295, 4382296, 4382301, 4382304, 4382305, 4382306, 4382309, 4382311, 4382313, 4382314, 4382316, 4382318, 4382319, 4382321, 4382324, 4382325, 4382326, 4382327, 4382328, 4382330, 4382331, 4382336, 4382338, 4382340, 4382344, 4382345, 4382346, 4382347, 4382348, 4382349, 4382351, 4382352, 4382353, 4382354, 4382355, 4382356, 4382357, 4382359, 4382360, 4382361, 4382362, 4382363, 4382364, 4382365, 4382366, 4382367, 4382368, 4382369, 4382370, 4382371, 4382372, 4382373, 4382374, 4382375, 4382376, 4382377, 4382378, 4382379, 4382380, 4382381, 4382382, 4382383, 4382384, 4382385, 4382386, 4382387, 4382388, 4382389, 4382390, 4382391, 4382392, 4382393, 4382394, 4382395, 4382396, 4382397, 4382398, 4382399, 4382400, 4382402, 4382404, 4382406, 4382407, 4382408, 4382410, 4382412, 4382413, 4382418, 4382421, 4382422, 4382423, 4382424, 4382425, 4382426, 4382427, 4382428, 4382429, 4382430, 4382431, 4382432, 4382433, 4382434, 4382435, 4382436, 4382437, 4382438, 4382439, 4382440, 4382441, 4382442, 4382443, 4382444, 4382445, 4382446, 4382447, 4382448, 4382449, 4382450, 4382451, 4382452, 4382453, 4382454, 4382455, 4382456, 4382457, 4382458, 4382459, 4382460, 4382461, 4382462, 4382463, 4382464, 4382465, 4382466, 4382467, 4382468, 4382469, 4382470, 4382471, 4382472, 4382473, 4382474, 4382475, 4382476, 4382477, 4382478, 4382479, 4382480, 4382481, 4382482, 4382483, 4382484, 4382485, 4382486, 4382487, 4382488, 4382489, 4382490, 4382491, 4382492, 4382493, 4382494, 4382495, 4382496, 4382497, 4382498, 4382499, 4382500, 4382501, 4382502, 4382503, 4382504, 4382505, 4382506, 4382507, 4382508, 4382509, 4382510, 4382511, 4382512, 4382513, 4382514, 4382515, 4382516, 4382517, 4382518, 4382519, 4382520, 4382521, 4382522, 4382523, 4382524, 4382525, 4382526, 4382527, 4382528, 4382529, 4382530, 4382531, 4382532, 4382533, 4382534, 4382535, 4382536, 4382537, 4382538, 4382539, 4382540, 4382541, 4382542, 4382543, 4382544, 4382545, 4382546, 4382547, 4382548, 4382549, 4382550, 4382551, 4382552, 4382553, 4382554, 4382555, 4382556, 4382557, 4382558, 4382559, 4382560, 4382561, 4382562, 4382563, 4382565, 4382567, 4382568, 4382569, 4382570, 4382571, 4382572, 4382573, 4382574, 4382577, 4382578, 4382579, 4382580, 4382581, 4382582, 4382583, 4382584, 4382585, 4382586, 4382587, 4382588, 4382589, 4382590, 4382591, 4382593, 4382594, 4382597, 4382598, 4382599, 4382600, 4382602, 4382603, 4382606, 4382608, 4382611, 4382615, 4382616, 4382617, 4382619, 4382620, 4382621, 4382622, 4382624, 4382627, 4382629, 4382630, 4382632, 4382633, 4382635, 4382636, 4382637, 4382638, 4382639, 4382640, 4382641, 4382642, 4382643, 4382644, 4382645, 4382646, 4382647, 4382648, 4382649, 4382650, 4382651, 4382652, 4382653, 4382654, 4382656, 4382667, 4382670, 4382671, 4382673, 4382675, 4382676, 4382677, 4382678, 4382679, 4382680, 4382681, 4382682, 4382683, 4382684, 4382685, 4382686, 4382687, 4382688, 4382689, 4382690, 4382691, 4382692, 4382693, 4382694, 4382695, 4382696, 4382697, 4382698, 4382699, 4382700, 4382701, 4382702, 4382703, 4382704, 4382705, 4382706, 4382707, 4382708, 4382709, 4382710, 4382711, 4382712, 4382713, 4382714, 4382715, 4382716, 4382717, 4382718, 4382719, 4382720, 4382721, 4382722, 4382723, 4382724, 4382725, 4382726, 4382727, 4382728, 4382729, 4382730, 4382731, 4382732, 4382733, 4382734, 4382735, 4382736, 4382737, 4382738, 4382739, 4382740, 4382741, 4382742, 4382743, 4382744, 4382745, 4382746, 4382747, 4382748, 4382749, 4382750, 4382751, 4382752, 4382753, 4382754, 4382755, 4382756, 4382757, 4382758, 4382759, 4382760, 4382761, 4382762, 4382763, 4382764, 4382765, 4382766, 4382767, 4382768, 4382769, 4382770, 4382771, 4382772, 4382773, 4382774, 4382775, 4382776, 4382777, 4382778, 4382779, 4382780, 4382781, 4382782, 4382783, 4382784, 4382785, 4382786, 4382787, 4382788, 4382789, 4382790, 4382791, 4382792, 4382793, 4382794, 4382795, 4382796, 4382797, 4382798, 4382799, 4382800, 4382801, 4382802, 4382803, 4382804, 4382805, 4382806, 4382807, 4382808, 4382809, 4382810, 4382811, 4382812, 4382813, 4382814, 4382815, 4382816, 4382817, 4382818, 4382819, 4382820, 4382821, 4382822, 4382823, 4382824, 4382825, 4382826, 4382827, 4382828, 4382829, 4382830, 4382831, 4382832, 4382833, 4382834, 4382835, 4382836, 4382837, 4382838, 4382839, 4382840, 4382841, 4382842, 4382843, 4382844, 4382845, 4382846, 4382847, 4382848, 4382849, 4382851, 4382852, 4382853, 4382854, 4382855, 4382856, 4382857, 4382858, 4382859, 4382860, 4382861, 4382862, 4382863, 4382864, 4382865, 4382866, 4382867, 4382868, 4382869, 4382870, 4382871, 4382872, 4382873, 4382874, 4382875, 4382876, 4382877, 4382878, 4382879, 4382880, 4382881, 4382882, 4382883, 4382884, 4382885, 4382886, 4382888, 4382889, 4382890, 4382891, 4382892, 4382893, 4382894, 4382895, 4382896, 4382897, 4382898, 4382899, 4382901, 4382902, 4382903, 4382904, 4382905, 4382906, 4382907, 4382908, 4382909, 4382910, 4382911, 4382912, 4382913, 4382914, 4382915, 4382916, 4382917, 4382918, 4382919, 4382920, 4382921, 4382922, 4382923, 4382924, 4382925, 4382926, 4382927, 4382928, 4382929, 4382930, 4382931, 4382932, 4382933, 4382934, 4382935, 4382936, 4382937, 4382938, 4382939, 4382940, 4382941, 4382942, 4382943, 4382944, 4382945, 4382946, 4382947, 4382948, 4382949, 4382950, 4382951, 4382952, 4382953, 4382954, 4382955, 4382956, 4382957, 4382958, 4382959, 4382960, 4382961, 4382962, 4382963, 4382964, 4382965, 4382966, 4382967, 4382968, 4382969, 4382970, 4382971, 4382972, 4382973, 4382974, 4382975, 4382976, 4382977, 4382978, 4382979, 4382980, 4382981, 4382982, 4382983, 4382984, 4382985, 4382986, 4382987, 4382988, 4382989, 4382990, 4382991, 4382992, 4382994, 4382995, 4382996, 4382997, 4382998, 4382999, 4383000, 4383001, 4383002, 4383003, 4383004, 4383005, 4383006, 4383007, 4383008, 4383009, 4383010, 4383011, 4383012, 4383013, 4383014, 4383015, 4383016, 4383017, 4383018, 4383019, 4383020, 4383021, 4383022, 4383023, 4383024, 4383026, 4383027, 4383028, 4383029, 4383030, 4383031, 4383032, 4383033, 4383034, 4383035, 4383036, 4383037, 4383038, 4383039, 4383040, 4383041, 4383042, 4383043, 4383044, 4383045, 4383046, 4383047, 4383048, 4383049, 4383050, 4383051, 4383052, 4383053, 4383055, 4383056, 4383057, 4383058, 4383059, 4383060, 4383061, 4383062, 4383063, 4383064, 4383065, 4383066, 4383067, 4383068, 4383069, 4383070, 4383071, 4383072, 4383073, 4383074, 4383075, 4383076, 4383077, 4383078, 4383079, 4383080, 4383081, 4383082, 4383083, 4383084, 4383085, 4383086, 4383087, 4383088, 4383089, 4383090, 4383091, 4383092, 4383093, 4383094, 4383095, 4383145, 4383147, 4383148, 4383150, 4383152, 4383155, 4383156, 4383157, 4383158, 4383159, 4383160, 4383161, 4383162, 4383163, 4383164, 4383165, 4383166, 4383167, 4383168, 4383169, 4383170, 4383171, 4383172, 4383173, 4383174, 4383175, 4383176, 4383177, 4383178, 4383179, 4383180, 4383181, 4383182, 4383183, 4383184, 4383185, 4383186, 4383187, 4383188, 4383189, 4383190, 4383191, 4383192, 4383193, 4383194, 4383195, 4383196, 4383198, 4383199, 4383200, 4383201, 4383202, 4383203, 4383204, 4383205, 4383206, 4383207, 4383208, 4383209, 4383210, 4383226, 4383227, 4383228, 4383229, 4383230, 4383231, 4383232, 4383233, 4383234, 4383235, 4383236, 4383237, 4383238, 4383239, 4383240, 4383241, 4383242, 4383243, 4383245, 4383246, 4383248, 4383249, 4383250, 4383252, 4383255, 4383256, 4383257, 4383259, 4383262, 4383263, 4383264, 4383265, 4383266, 4383267, 4383268, 4383269, 4383270, 4383271, 4383272, 4383273, 4383274, 4383275, 4383276, 4383277, 4383278, 4383279, 4383280, 4383281, 4383282, 4383283, 4383285, 4383286, 4383287, 4383288, 4383293, 4383294, 4383295, 4383298, 4383301, 4383302, 4383303, 4383305, 4383306, 4383307, 4383308, 4383309, 4383310, 4383311, 4383312, 4383313, 4383314, 4383315, 4383316, 4383317, 4383318, 4383319, 4383320, 4383321, 4383322, 4383323, 4383325, 4383327, 4383328, 4383329, 4383331, 4383335, 4383336, 4383338, 4383340, 4383341, 4383342, 4383343, 4383344, 4383345, 4383346, 4383352, 4383353, 4383354, 4383356, 4383361, 4383362, 4383364, 4383365, 4383366, 4383367, 4383368, 4383369, 4383370, 4383371, 4383372, 4383373, 4383374, 4383375, 4383376, 4383377, 4383378, 4383379, 4383380, 4383381, 4383382, 4383383, 4383384, 4383385, 4383386, 4383387, 4383388, 4383389, 4383390, 4383391, 4383392, 4383393, 4383394, 4383395, 4383396, 4383397, 4383398, 4383399, 4383400, 4383401, 4383402, 4383403, 4383404, 4383405, 4383406, 4383407, 4383408, 4383409, 4383410, 4383411, 4383412, 4383413, 4383414, 4383415, 4383416, 4383417, 4383418, 4383419, 4383420, 4383421, 4383422, 4383423, 4383424, 4383425, 4383426, 4383427, 4383428, 4383429, 4383430, 4383431, 4383432, 4383433, 4383434, 4383435, 4383436, 4383437, 4383438, 4383439, 4383440, 4383441, 4383442, 4383443, 4383444, 4383445, 4383446, 4383447, 4383448, 4383449, 4383450, 4383451, 4383452, 4383453, 4383454, 4383455, 4383456, 4383457, 4383458, 4383459, 4383460, 4383461, 4383462, 4383463, 4383464, 4383658, 4383659, 4383661, 4383662, 4383664, 4383666, 4383667, 4383668, 4383669, 4383670, 4383671, 4383672, 4383673, 4383674, 4383675, 4383677, 4383678, 4383681, 4383683, 4383685, 4383686, 4383688, 4383689, 4383690, 4383691, 4383692, 4383693, 4383694, 4383695, 4383696, 4383698, 4383699, 4383700, 4383701, 4383703, 4383704, 4383705, 4383707, 4383708, 4383710, 4383712, 4383715, 4383717, 4383718, 4383720, 4383721, 4383722, 4383723, 4383724, 4383725, 4383726, 4383727, 4383728, 4383729, 4383730, 4383731, 4383732, 4383733, 4383734, 4383735, 4383736, 4383737, 4383738, 4383740, 4383741, 4383742, 4383743, 4383744, 4383745, 4383746, 4383747, 4383748, 4383749, 4383750, 4383751, 4383753, 4383755, 4383756, 4383817, 4383821, 4383822, 4383827, 4383828, 4383830, 4383833, 4383837, 4383839, 4383840, 4383841, 4383842, 4383844, 4383847, 4383848, 4383849, 4383850, 4383851, 4383852, 4383854, 4383855, 4383856, 4383857, 4383858, 4383860, 4383861, 4383862, 4383863, 4383864, 4383865, 4383866, 4383867, 4383868, 4383869, 4383870, 4383871, 4383872, 4383873, 4383874, 4383875, 4383876, 4383877, 4383878, 4383879, 4383880, 4383881, 4383882, 4383883, 4383884, 4383885, 4383886, 4383887, 4383888, 4383889, 4383890, 4383891, 4383892, 4383893, 4383894, 4383895, 4383896, 4383897, 4383899, 4383900, 4383901, 4383902, 4383903, 4383904, 4383905, 4383906, 4383907, 4383908, 4383909, 4383910, 4383911, 4383912, 4383913, 4383914, 4383916, 4383917, 4383918, 4383919, 4383921, 4383922, 4383923, 4383924, 4383925, 4383926, 4383928, 4383930, 4383931, 4383934, 4383938, 4383940, 4383941, 4383942, 4383945, 4383946, 4383947, 4383948, 4383949, 4383950, 4383951, 4383952, 4383953, 4383954, 4383955, 4383956, 4383957, 4383958, 4383959, 4383960, 4383961, 4383962, 4383963, 4383965, 4383966, 4383967, 4383968, 4383969, 4383970, 4383971, 4383972, 4383973, 4383975, 4383976, 4383978, 4383980, 4383981, 4383983, 4383991, 4383994, 4383995, 4384067, 4384068, 4384069, 4384070, 4384071, 4384074, 4384077, 4384078, 4384079, 4384080, 4384081, 4384082, 4384087, 4384088, 4384095, 4384111, 4384122, 4384123, 4384125, 4384126, 4384127, 4384128, 4384129, 4384130, 4384132, 4384133, 4384134, 4384135, 4384136, 4384138, 4384139, 4384141, 4384144, 4384146, 4384148, 4384150, 4384151, 4384152, 4384153, 4384154, 4384155, 4384156, 4384157, 4384158, 4384159, 4384160, 4384161, 4384162, 4384163, 4384164, 4384165, 4384166, 4384167, 4384168, 4384169, 4384170, 4384171, 4384172, 4384173, 4384175, 4384176, 4384177, 4384178, 4384179, 4384180, 4384181, 4384182, 4384183, 4384184, 4384185, 4384186, 4384187, 4384188, 4384189, 4384190, 4384191, 4384192, 4384193, 4384194, 4384195, 4384196, 4384197, 4384198, 4384199, 4384200, 4384201, 4384202, 4384203, 4384204, 4384205, 4384206, 4384207, 4384208, 4384269, 4384270, 4384271, 4384278, 4384279, 4384280, 4384282, 4384283, 4384284, 4384285, 4384286, 4384287, 4384288, 4384289, 4384290, 4384295, 4384296, 4384297, 4384298, 4384300, 4384301, 4384302, 4384304, 4384305, 4384306, 4384307, 4384308, 4384310, 4384315, 4384316, 4384317, 4384318, 4384319, 4384320, 4384321, 4384322, 4384323, 4384324, 4384325, 4384326, 4384328, 4384329, 4384330, 4384333, 4384335, 4384336, 4384337, 4384338, 4384339, 4384340, 4384341, 4384342, 4384343, 4384344, 4384345, 4384346, 4384347, 4384348, 4384350, 4384351, 4384353, 4384354, 4384355, 4384356, 4384357, 4384358, 4384359, 4384360, 4384361, 4384362, 4384363, 4384364, 4384365, 4384366, 4384368, 4384369, 4384370, 4384371, 4384374, 4384376, 4384380, 4384382, 4384383, 4384384, 4384385, 4384386, 4384390, 4384392, 4384394, 4384398, 4384400, 4384401, 4384403, 4384404, 4384405, 4384406, 4384407, 4384408, 4384409, 4384410, 4384411, 4384412, 4384413, 4384414, 4384415, 4384416, 4384417, 4384418, 4384419, 4384420, 4384421, 4384423, 4384424, 4384425, 4384426, 4384428, 4384429, 4384430, 4384431, 4384432, 4384433, 4384435, 4384436, 4384441, 4384444, 4384446, 4384447, 4384450, 4384451, 4384455, 4384462, 4384465, 4384466, 4384467, 4384468, 4384469, 4384473, 4384476, 4384481, 4384482, 4384486, 4384497, 4384499, 4384500, 4384501, 4384502, 4384504, 4384506, 4384510, 4384515, 4384516, 4384518, 4384520, 4384521, 4384522, 4384525, 4384527, 4384528, 4384529, 4384533, 4384534, 4384535, 4384537, 4384538, 4384539, 4384540, 4384541, 4384542, 4384543, 4384544, 4384548, 4384549, 4384551, 4384552, 4384553, 4384559, 4384564, 4384565, 4384567, 4384569, 4384570, 4384571, 4384572, 4384573, 4384574, 4384575, 4384576, 4384577, 4384579, 4384580, 4384581, 4384582, 4384583, 4384584, 4384585, 4384586, 4384587, 4384604, 4384609, 4384611, 4384613, 4384614, 4384615, 4384616, 4384617, 4384618, 4384619, 4384620, 4384621, 4384622, 4384623, 4384624, 4384625, 4384626, 4384629, 4384630, 4384632, 4384634, 4384638, 4384639, 4384640, 4384643, 4384645, 4384658, 4384662, 4384665, 4384669, 4384671, 4384673, 4384674, 4384676, 4384679, 4384680, 4384681, 4384682, 4384683, 4384685, 4384686, 4384687, 4384688, 4384690, 4384691, 4384692, 4384693, 4384694, 4384695, 4384696, 4384697, 4384699, 4384700, 4384701, 4384702, 4384704, 4384708, 4384709, 4384711, 4384712, 4384713, 4384714, 4384715, 4384716, 4384718, 4384719, 4384720, 4384721, 4384722, 4384723, 4384724, 4384725, 4384726, 4384727, 4384728, 4384732, 4384733, 4384734, 4384735, 4384737, 4384738, 4384739, 4384741, 4384743, 4384745, 4384746, 4384747, 4384748, 4384749, 4384750, 4384751, 4384752, 4384753, 4384754, 4384755, 4384756, 4384758, 4384759, 4384760, 4384762, 4384763, 4384765, 4384766, 4384767, 4384768, 4384770, 4384771, 4384772, 4384773, 4384774, 4384775, 4384776, 4384777, 4384779, 4384780, 4384781, 4384782, 4384783, 4384784, 4384785, 4384786, 4384787, 4384788, 4384789, 4384790, 4384791, 4384792, 4384793, 4384794, 4384795, 4384796, 4384797, 4384798, 4384799, 4384800, 4384801, 4384802, 4384803, 4384804, 4384806, 4384807, 4384808, 4384810, 4384811, 4384812, 4384813, 4384814, 4384816, 4384817, 4384818, 4384819, 4384821, 4384822, 4384823, 4384826, 4384827, 4384828, 4384829, 4384830, 4384831, 4384832, 4384833, 4384834, 4384835, 4384836, 4384837, 4384839, 4384842, 4384844, 4384845, 4384848, 4384853, 4384854, 4384860, 4384861, 4384864, 4384865, 4384866, 4384867, 4384869, 4384870, 4384871, 4384872, 4384873, 4384874, 4384875, 4384877, 4384878, 4384879, 4384880, 4384881, 4384886, 4384888, 4384890, 4384891, 4384893, 4384900, 4384901, 4384903, 4384904, 4384905, 4384907, 4384908, 4384909, 4384912, 4384913, 4384914, 4384916, 4384917, 4384921, 4384922, 4384923, 4384926, 4384927, 4384928, 4384929, 4384930, 4384931, 4384932, 4384934, 4384936, 4384937, 4384942, 4384943, 4384946, 4384948, 4384949, 4384953, 4384954, 4384958, 4384959, 4384961, 4384962, 4384963, 4384966, 4384967, 4384968, 4384971, 4384973, 4384977, 4384978, 4384979, 4384980, 4384981, 4384987, 4384988, 4384989, 4384992, 4384993, 4384994, 4384995, 4384997, 4384998, 4385000, 4385001, 4385003, 4385004, 4385010, 4385011, 4385013, 4385015, 4385016, 4385018, 4385019, 4385020, 4385022, 4385023, 4385024, 4385026, 4385027, 4385028, 4385029, 4385030, 4385036, 4385037, 4385042, 4385043, 4385044, 4385046, 4385047, 4385055, 4385056, 4385057, 4385058, 4385059, 4385060, 4385061, 4385062, 4385063, 4385064, 4385066, 4385069, 4385077, 4385091, 4385093, 4385094, 4385095, 4385096, 4385099, 4385101, 4385110, 4385111, 4385112, 4385113, 4385114, 4385115, 4385117, 4385118, 4385119, 4385120, 4385121, 4385122, 4385123, 4385124, 4385125, 4385126, 4385127, 4385128, 4385130, 4385132, 4385133, 4385134, 4385136, 4385138, 4385141, 4385146, 4385147, 4385148, 4385150, 4385151, 4385152, 4385153, 4385154, 4385155, 4385156, 4385237, 4385238, 4385239, 4385240, 4385241, 4385242, 4385243, 4385244, 4385245, 4385246, 4385247, 4385248, 4385249, 4385250, 4385251, 4385252, 4385253, 4385254, 4385255, 4385256, 4385390, 4385426, 4385427, 4385447, 4385477, 4385478, 4385479, 4385480, 4385481, 4385482, 4385483, 4385484, 4385485, 4385486, 4385487, 4385488, 4385489, 4385490, 4385491, 4385492, 4385493, 4385494, 4385495, 4385497, 4385499, 4385500, 4385501, 4385502, 4385504, 4385505, 4385506, 4385507, 4385509, 4385510, 4385511, 4385513, 4385514, 4385515, 4385516, 4385517, 4385518, 4385519, 4385520, 4385521, 4385523, 4385525, 4385526, 4385527, 4385528, 4385529, 4385531, 4385532, 4385533, 4385534, 4385535, 4385538, 4385539, 4385540, 4385541, 4385542, 4385543, 4385544, 4385545, 4385546, 4385547, 4385548, 4385549, 4385550, 4385551, 4385552, 4385553, 4385554, 4385555, 4385557, 4385558, 4385559, 4385560, 4385565, 4385569, 4385570, 4385592, 4385593, 4385596, 4385597, 4385598, 4385600, 4385601, 4385602, 4385603, 4385604, 4385606, 4385607, 4385608, 4385609, 4385610, 4385613, 4385615, 4385616, 4385617, 4385618, 4385619, 4385621, 4385622, 4385623, 4385624, 4385628, 4385629, 4385630, 4385632, 4385633, 4385634, 4385635, 4385636, 4385637, 4385638, 4385640, 4385641, 4385642, 4385644, 4385646, 4385650, 4385653, 4385658, 4385661, 4385663, 4385664, 4385665, 4385672, 4385673, 4385674, 4385676, 4385678, 4385680, 4385684, 4385688, 4385689, 4385690, 4385693, 4385696, 4385697, 4385698, 4385701, 4385702, 4385703, 4385704, 4385705, 4385706, 4385707, 4385709, 4385710, 4385711, 4385712, 4385713, 4385714, 4385715, 4385716, 4385717, 4385718, 4385719, 4385720, 4385721, 4385722, 4385723, 4385724, 4385725, 4385726, 4385727, 4385728, 4385731, 4385733, 4385734, 4385741, 4385742, 4385743, 4385744, 4385745, 4385746, 4385747, 4385748, 4385749, 4385751, 4385753, 4385756, 4385758, 4385759, 4385761, 4385762, 4385763, 4385764, 4385769, 4385771, 4385775, 4385779, 4385780, 4385781, 4385785, 4385786, 4385788, 4385789, 4385790, 4385791, 4385794, 4385796, 4385797, 4385798, 4385799, 4385800, 4385801, 4385803, 4385804, 4385805, 4385814, 4385815, 4385817, 4385820, 4385822, 4385824, 4385825, 4385826, 4385827, 4385828, 4385829, 4385830, 4385831, 4385832, 4385833, 4385834, 4385835, 4385836, 4385837, 4385838, 4385839, 4385840, 4385841, 4385842, 4385843, 4385846, 4385847, 4385848, 4385849, 4385850, 4385851, 4385853, 4385854, 4385855, 4385856, 4385857, 4385859, 4385860, 4385861, 4385862, 4385863, 4385865, 4385866, 4385867, 4385869, 4385871, 4385872, 4385873, 4385875, 4385876, 4385877, 4385882, 4385885, 4385887, 4385888, 4385889, 4385894, 4385895, 4385898, 4385899, 4385900, 4385901, 4385902, 4385903, 4385904, 4385905, 4385906, 4385907, 4385908, 4385909, 4385910, 4385911, 4385912, 4385913, 4385914, 4385915, 4385916, 4385917, 4385918, 4385919, 4385922, 4385923, 4385924, 4385925, 4385927, 4385928, 4385929, 4385930, 4385931, 4385932, 4385934, 4385935, 4385938, 4385939, 4385940, 4385942, 4385943, 4385944, 4385945, 4385946, 4385949, 4385950, 4385951, 4385952, 4385953, 4385954, 4385955, 4385956, 4385957, 4385959, 4385960, 4385962, 4385963, 4385964, 4385965, 4385966, 4385967, 4385968, 4385969, 4385970, 4385971, 4385972, 4385973, 4385975, 4385976, 4385977, 4385978, 4385980, 4385982, 4385983, 4385984, 4385988, 4385990, 4385992, 4385993, 4385997, 4385998, 4385999, 4386000, 4386001, 4386002, 4386007, 4386008, 4386010, 4386011, 4386012, 4386013, 4386014, 4386015, 4386017, 4386019, 4386020, 4386024, 4386025, 4386026, 4386027, 4386028, 4386029, 4386030, 4386031, 4386032, 4386033, 4386034, 4386035, 4386036, 4386037, 4386038, 4386039, 4386040, 4386041, 4386042, 4386043, 4386044, 4386045, 4386046, 4386047, 4386048, 4386049, 4386050, 4386051, 4386052, 4386053, 4386054, 4386055, 4386056, 4386057, 4386058, 4386060, 4386061, 4386062, 4386063, 4386064, 4386065, 4386066, 4386068, 4386069, 4386070, 4386071, 4386072, 4386073, 4386074, 4386075, 4386076, 4386077, 4386078, 4386079, 4386080, 4386081, 4386082, 4386083, 4386084, 4386085, 4386086, 4386087, 4386088, 4386089, 4386090, 4386091, 4386092, 4386096, 4386098, 4386099, 4386101, 4386103, 4386104, 4386105, 4386106, 4386108, 4386109, 4386111, 4386114, 4386116, 4386117, 4386118, 4386119, 4386120, 4386121, 4386122, 4386125, 4386127, 4386130, 4386131, 4386132, 4386135, 4386136, 4386137, 4386139, 4386140, 4386142, 4386144, 4386145, 4386148, 4386150, 4386151, 4386155, 4386157, 4386159, 4386160, 4386163, 4386164, 4386165, 4386166, 4386167, 4386168, 4386170, 4386171, 4386172, 4386174, 4386176, 4386179, 4386180, 4386181, 4386182, 4386184, 4386185, 4386186, 4386187, 4386188, 4386190, 4386191, 4386192, 4386194, 4386199, 4386204, 4386205, 4386212, 4386213, 4386214, 4386218, 4386219, 4386220, 4386221, 4386222, 4386227, 4386229, 4386230, 4386231, 4386232, 4386233, 4386235, 4386237, 4386238, 4386239, 4386241, 4386242, 4386243, 4386246, 4386248, 4386249, 4386251, 4386255, 4386256, 4386257, 4386258, 4386259, 4386260, 4386261, 4386262, 4386263, 4386264, 4386265, 4386266, 4386267, 4386268, 4386269, 4386270, 4386271, 4386273, 4386274, 4386275, 4386276, 4386277, 4386278, 4386279, 4386280, 4386281, 4386282, 4386284, 4386285, 4386286, 4386287, 4386288, 4386289, 4386290, 4386291, 4386292, 4386293, 4386294, 4386295, 4386296, 4386297, 4386298, 4386299, 4386300, 4386303, 4386304, 4386305, 4386306, 4386307, 4386308, 4386309, 4386310, 4386311, 4386312, 4386313, 4386314, 4386315, 4386316, 4386317, 4386318, 4386319, 4386320, 4386321, 4386322, 4386323, 4386324, 4386325, 4386326, 4386327, 4386328, 4386329, 4386330, 4386331, 4386332, 4386333, 4386334, 4386335, 4386336, 4386337, 4386338, 4386339, 4386340, 4386341, 4386342, 4386343, 4386344, 4386345, 4386346, 4386347, 4386348, 4386349, 4386350, 4386351, 4386352, 4386353, 4386354, 4386355, 4386356, 4386357, 4386358, 4386359, 4386360, 4386361, 4386362, 4386363, 4386364, 4386365, 4386366, 4386367, 4386368, 4386369, 4386370, 4386371, 4386372, 4386373, 4386374, 4386375, 4386376, 4386377, 4386378, 4386379, 4386380, 4386381, 4386382, 4386383, 4386384, 4386385, 4386386, 4386387, 4386388, 4386389, 4386390, 4386391, 4386392, 4386398, 4386402, 4386407, 4386408, 4386409, 4386410, 4386411, 4386412, 4386413, 4386414, 4386415, 4386416, 4386417, 4386418, 4386435, 4386443, 4386444, 4386446, 4386447, 4386448, 4386450, 4386453, 4386455, 4386456, 4386457, 4386458, 4386459, 4386460, 4386461, 4386462, 4386463, 4386464, 4386465, 4386466, 4386467, 4386468, 4386469, 4386470, 4386471, 4386473, 4386474, 4386475, 4386476, 4386477, 4386478, 4386483, 4386485, 4386486, 4386491, 4386494, 4386496, 4386497, 4386498, 4386499, 4386500, 4386501, 4386502, 4386503, 4386504, 4386506, 4386507, 4386508, 4386509, 4386511, 4386512, 4386513, 4386514, 4386515, 4386516, 4386518, 4386519, 4386520, 4386522, 4386524, 4386525, 4386527, 4386528, 4386529, 4386530, 4386531, 4386532, 4386533, 4386534, 4386535, 4386540, 4386541, 4386543, 4386544, 4386545, 4386546, 4386547, 4386548, 4386549, 4386550, 4386551, 4386552, 4386553, 4386554, 4386555, 4386556, 4386557, 4386558, 4386559, 4386560, 4386561, 4386562, 4386563, 4386564, 4386565, 4386566, 4386567, 4386568, 4386569, 4386570, 4386571, 4386572, 4386573, 4386574, 4386575, 4386576, 4386577, 4386578, 4386579, 4386580, 4386581, 4386582, 4386583, 4386584, 4386585, 4386586, 4386587, 4386588, 4386589, 4386590, 4386591, 4386592, 4386593, 4386594, 4386595, 4386596, 4386597, 4386598, 4386599, 4386600, 4386601, 4386602, 4386603, 4386604, 4386605, 4386606, 4386607, 4386608, 4386609, 4386610, 4386611, 4386612, 4386613, 4386614, 4386615, 4386616, 4386617, 4386618, 4386619, 4386620, 4386621, 4386622, 4386623, 4386624, 4386625, 4386626, 4386627, 4386628, 4386629, 4386630, 4386631, 4386632, 4386633, 4386634, 4386635, 4386636, 4386637, 4386638, 4386639, 4386640, 4386641, 4386642, 4386643, 4386644, 4386645, 4386646, 4386647, 4386648, 4386649, 4386650, 4386651, 4386652, 4386654, 4386655, 4386657, 4386658, 4386660, 4386661, 4386663, 4386664, 4386665, 4386666, 4386667, 4386668, 4386669, 4386670, 4386671, 4386672, 4386673, 4386674, 4386675, 4386676, 4386677, 4386678, 4386679, 4386680, 4386681, 4386682, 4386683, 4386684, 4386685, 4386686, 4386687, 4386688, 4386689, 4386690, 4386691, 4386692, 4386693, 4386694, 4386695, 4386696, 4386697, 4386698, 4386699, 4386700, 4386701, 4386702, 4386703, 4386704, 4386705, 4386706, 4386707, 4386708, 4386709, 4386710, 4386711, 4386712, 4386713, 4386714, 4386715, 4386716, 4386717, 4386718, 4386719, 4386720, 4386721, 4386722, 4386723, 4386724, 4386725, 4386726, 4386727, 4386728, 4386729, 4386730, 4386731, 4386732, 4386733, 4386734, 4386735, 4386736, 4386737, 4386738, 4386739, 4386740, 4386741, 4386742, 4386743, 4386744, 4386745, 4386746, 4386747, 4386748, 4386749, 4386750, 4386751, 4386752, 4386753, 4386754, 4386755, 4386756, 4386757, 4386758, 4386759, 4386760, 4386761, 4386762, 4386763, 4386764, 4386765, 4386766, 4386767, 4386768, 4386769, 4386770, 4386771, 4386772, 4386773, 4386774, 4386775, 4386776, 4386777, 4386778, 4386779, 4386780, 4386781, 4386782, 4386783, 4386784, 4386785, 4386786, 4386787, 4386788, 4386789, 4386790, 4386791, 4386792, 4386793, 4386794, 4386795, 4386796, 4386797, 4386798, 4386799, 4386800, 4386801, 4386802, 4386803, 4386804, 4386805, 4386806, 4386807, 4386808, 4386809, 4386810, 4386811, 4386812, 4386813, 4386814, 4386815, 4386816, 4386817, 4386818, 4386819, 4386820, 4386821, 4386822, 4386823, 4386824, 4386825, 4386826, 4386827, 4386828, 4386829, 4386830, 4386831, 4386832, 4386833, 4386834, 4386835, 4386836, 4386837, 4386838, 4386839, 4386840, 4386841, 4386842, 4386843, 4386844, 4386845, 4386846, 4386847, 4386848, 4386849, 4386850, 4386851, 4386852, 4386853, 4386854, 4386855, 4386856, 4386857, 4386858, 4386859, 4386860, 4386861, 4386862, 4386863, 4386864, 4386865, 4386866, 4386867, 4386868, 4386869, 4386870, 4386871, 4386872, 4386873, 4386874, 4386875, 4386876, 4386877, 4386878, 4386879, 4386880, 4386881, 4386882, 4386883, 4386884, 4386885, 4386886, 4386887, 4386888, 4386889, 4386890, 4386891, 4386892, 4386893, 4386894, 4386895, 4386896, 4386897, 4386898, 4386899, 4386900, 4386901, 4386902, 4386903, 4386904, 4386905, 4386906, 4386907, 4386908, 4386909, 4386910, 4386911, 4386912, 4386913, 4386914, 4386915, 4386916, 4386917, 4386918, 4386919, 4386920, 4386921, 4386922, 4386923, 4386924, 4386925, 4386926, 4386927, 4386928, 4386929, 4386930, 4386931, 4386932, 4386933, 4386934, 4386935, 4386936, 4386937, 4386938, 4386939, 4386940, 4386941, 4386942, 4386943, 4386945, 4386946, 4386947, 4386948, 4386949, 4386950, 4386951, 4386952, 4386953, 4386954, 4386955, 4386956, 4386957, 4386958, 4386959, 4386960, 4386961, 4386962, 4386963, 4386964, 4386965, 4386966, 4386967, 4386968, 4386969, 4386970, 4386971, 4386972, 4386973, 4386974, 4386975, 4386976, 4386977, 4386978, 4386979, 4386980, 4386981, 4386982, 4386983, 4386984, 4386985, 4386986, 4386987, 4386988, 4386989, 4386990, 4386991, 4386992, 4386993, 4386994, 4386995, 4386996, 4386997, 4386998, 4386999, 4387000, 4387001, 4387002, 4387003, 4387004, 4387005, 4387006, 4387007, 4387008, 4387009, 4387010, 4387011, 4387012, 4387013, 4387014, 4387015, 4387016, 4387017, 4387018, 4387019, 4387020, 4387021, 4387022, 4387023, 4387024, 4387025, 4387026, 4387027, 4387028, 4387029, 4387030, 4387031, 4387032, 4387034, 4387035, 4387036, 4387037, 4387038, 4387039, 4387040, 4387041, 4387042, 4387043, 4387044, 4387045, 4387046, 4387047, 4387048, 4387049, 4387050, 4387051, 4387052, 4387053, 4387054, 4387055, 4387056, 4387058, 4387059, 4387060, 4387061, 4387064, 4387067, 4387068, 4387069, 4387070, 4387071, 4387072, 4387073, 4387074, 4387075, 4387076, 4387077, 4387078, 4387079, 4387080, 4387081, 4387082, 4387083, 4387084, 4387085, 4387086, 4387087, 4387088, 4387089, 4387090, 4387091, 4387092, 4387093, 4387094, 4387095, 4387096, 4387097, 4387098, 4387099, 4387100, 4387101, 4387102, 4387103, 4387104, 4387105, 4387106, 4387107, 4387108, 4387109, 4387110, 4387111, 4387112, 4387113, 4387114, 4387115, 4387116, 4387117, 4387118, 4387119, 4387120, 4387121, 4387122, 4387123, 4387124, 4387125, 4387126, 4387127, 4387128, 4387129, 4387130, 4387131, 4387132, 4387133, 4387134, 4387135, 4387136, 4387137, 4387138, 4387139, 4387140, 4387141, 4387142, 4387143, 4387144, 4387145, 4387146, 4387147, 4387148, 4387149, 4387150, 4387151, 4387152, 4387153, 4387154, 4387155, 4387156, 4387157, 4387158, 4387159, 4387160, 4387161, 4387162, 4387163, 4387164, 4387165, 4387166, 4387167, 4387168, 4387169, 4387170, 4387171, 4387172, 4387173, 4387174, 4387175, 4387176, 4387177, 4387178, 4387179, 4387180, 4387181, 4387182, 4387183, 4387184, 4387185, 4387186, 4387187, 4387188, 4387189, 4387190, 4387191, 4387192, 4387193, 4387194, 4387195, 4387196, 4387197, 4387198, 4387199, 4387200, 4387202, 4387203, 4387204, 4387205, 4387206, 4387207, 4387208, 4387209, 4387210, 4387211, 4387214, 4387215, 4387216, 4387217, 4387218, 4387219, 4387220, 4387221, 4387222, 4387223, 4387224, 4387225, 4387226, 4387227, 4387229, 4387230, 4387231, 4387232, 4387234, 4387235, 4387236, 4387237, 4387238, 4387239, 4387240, 4387241, 4387242, 4387243, 4387244, 4387245, 4387246, 4387247, 4387248, 4387249, 4387250, 4387251, 4387252, 4387253, 4387254, 4387255, 4387256, 4387257, 4387258, 4387259, 4387260, 4387261, 4387262, 4387263, 4387264, 4387265, 4387266, 4387267, 4387268, 4387269, 4387271, 4387272, 4387273, 4387274, 4387275, 4387276, 4387277, 4387278, 4387280, 4387281, 4387282, 4387283, 4387284, 4387286, 4387287, 4387288, 4387289, 4387290, 4387291, 4387292, 4387293, 4387294, 4387295, 4387296, 4387297, 4387298, 4387299, 4387300, 4387301, 4387302, 4387304, 4387305, 4387306, 4387307, 4387308, 4387309, 4387311, 4387312, 4387313, 4387314, 4387315, 4387316, 4387317, 4387318, 4387319, 4387320, 4387321, 4387322, 4387323, 4387324, 4387325, 4387326, 4387327, 4387328, 4387329, 4387330, 4387331, 4387332, 4387333, 4387334, 4387335, 4387336, 4387337, 4387338, 4387339, 4387340, 4387341, 4387342, 4387343, 4387344, 4387346, 4387347, 4387348, 4387349, 4387350, 4387351, 4387352, 4387353, 4387354, 4387355, 4387356, 4387357, 4387358, 4387359, 4387360, 4387361, 4387362, 4387363, 4387364, 4387365, 4387366, 4387367, 4387368, 4387369, 4387370, 4387371, 4387372, 4387373, 4387374, 4387375, 4387376, 4387377, 4387378, 4387379, 4387380, 4387381, 4387382, 4387383, 4387384, 4387385, 4387386, 4387387, 4387388, 4387389, 4387390, 4387391, 4387392, 4387393, 4387394, 4387395, 4387396, 4387397, 4387398, 4387399, 4387400, 4387401, 4387402, 4387403, 4387404, 4387405, 4387406, 4387407, 4387408, 4387409, 4387410, 4387411, 4387412, 4387413, 4387414, 4387415, 4387416, 4387417, 4387418, 4387420, 4387421, 4387422, 4387423, 4387424, 4387426, 4387427, 4387428, 4387429, 4387430, 4387431, 4387432, 4387433, 4387434, 4387435, 4387436, 4387437, 4387438, 4387439, 4387440, 4387441, 4387442, 4387443, 4387444, 4387445, 4387446, 4387447, 4387448, 4387449, 4387450, 4387451, 4387452, 4387453, 4387454, 4387455, 4387456, 4387457, 4387458, 4387459, 4387460, 4387461, 4387462, 4387463, 4387464, 4387465, 4387466, 4387467, 4387468, 4387469, 4387470, 4387471, 4387472, 4387473, 4387474, 4387475, 4387476, 4387477, 4387478, 4387479, 4387480, 4387481, 4387482, 4387483, 4387484, 4387485, 4387486, 4387487, 4387488, 4387489, 4387490, 4387491, 4387492, 4387493, 4387494, 4387495, 4387496, 4387497, 4387498, 4387499, 4387500, 4387501, 4387502, 4387503, 4387504, 4387505, 4387506, 4387507, 4387508, 4387509, 4387510, 4387511, 4387512, 4387513, 4387514, 4387515, 4387516, 4387517, 4387518, 4387519, 4387520, 4387521, 4387522, 4387523, 4387524, 4387525, 4387526, 4387527, 4387528, 4387529, 4387530, 4387531, 4387532, 4387533, 4387534, 4387535, 4387536, 4387537, 4387538, 4387539, 4387540, 4387541, 4387542, 4387543, 4387544, 4387545, 4387546, 4387547, 4387548, 4387549, 4387550, 4387551, 4387552, 4387553, 4387554, 4387555, 4387556, 4387557, 4387558, 4387559, 4387560, 4387561, 4387562, 4387563, 4387564, 4387565, 4387566, 4387567, 4387568, 4387569, 4387570, 4387571, 4387572, 4387573, 4387574, 4387575, 4387576, 4387577, 4387578, 4387579, 4387580, 4387581, 4387582, 4387583, 4387584, 4387585, 4387586, 4387587, 4387588, 4387589, 4387590, 4387591, 4387592, 4387594, 4387595, 4387596, 4387597, 4387598, 4387599, 4387600, 4387601, 4387602, 4387604, 4387605, 4387606, 4387607, 4387608, 4387610, 4387611, 4387612, 4387613, 4387615, 4387616, 4387617, 4387618, 4387620, 4387621, 4387622, 4387623, 4387625, 4387626, 4387627, 4387628, 4387629, 4387630, 4387631, 4387632, 4387634, 4387635, 4387636, 4387637, 4387638, 4387639, 4387640, 4387641, 4387642, 4387643.4387644, 4387645, 4387646, 4387647, 4387648, 4387649, 4387650, 4387651, 4387652, 4387653, 4387654, 4387655, 4387656, 4387657, 4387658, 4387659, 4387660, 4387661, 4387662, 4387663, 4387664, 4387665, 4387666, 4387667, 4387668, 4387669, 4387670, 4387671, 4387672, 4387673, 4387674, 4387675, 4387676, 4387677, 4387678, 4387679, 4387680, 4387681, 4387682, 4387683, 4387684, 4387685, 4387686, 4387687, 4387688, 4387689, 4387690, 4387691, 4387692, 4387693, 4387694, 4387695, 4387696, 4387697, 4387698, 4387699, 4387700, 4387701, 4387702, 4387703, 4387704, 4387705, 4387706, 4387707, 4387708, 4387709, 4387710, 4387711, 4387712, 4387713, 4387714, 4387715, 4387716, 4387717, 4387718, 4387719, 4387720, 4387721, 4387722, 4387724, 4387730, 4387732, 4387733, 4387734, 4387735, 4387736, 4387737, 4387738, 4387739, 4387740, 4387741, 4387742, 4387743, 4387744, 4387745, 4387746, 4387747, 4387748, 4387749, 4387750, 4387751, 4387752, 4387753, 4387754, 4387755, 4387756, 4387757, 4387758, 4387759, 4387760, 4387761, 4387762, 4387764, 4387765, 4387766, 4387767, 4387768, 4387769, 4387770, 4387771, 4387772, 4387773, 4387774, 4387775, 4387776, 4387777, 4387778, 4387779, 4387780, 4387781, 4387782, 4387783, 4387784, 4387785, 4387786, 4387787, 4387788, 4387789, 4387790, 4387791, 4387792, 4387793, 4387794, 4387795, 4387796, 4387797, 4387798, 4387800, 4387832, 4387834, 4387838, 4387843, 4387844, 4387847, 4387848, 4387849, 4387850, 4387851, 4387852, 4387853, 4387854, 4387855, 4387856, 4387857, 4387858, 4387859, 4387860, 4387861, 4387862, 4387863, 4387864, 4387865, 4387866, 4387867, 4387868, 4387869, 4387870, 4387871, 4387874, 4387875, 4387876, 4387877, 4387878, 4387879, 4387880, 4387882, 4387883, 4387885, 4387887, 4387888, 4387889, 4387890, 4387891, 4387894, 4387896, 4387897, 4387898, 4387899, 4387901, 4387902, 4387903, 4387904, 4387905, 4387906, 4387907, 4387908, 4387909, 4387910, 4387911, 4387912, 4387913, 4387914, 4387916, 4387917, 4387918, 4387919, 4387923, 4387924, 4387925, 4387926, 4387927, 4387928, 4387929, 4387930, 4387931, 4387932, 4387933, 4387934, 4387935, 4387936, 4387937, 4387938, 4387939, 4387940, 4387941, 4387942, 4387943, 4387944, 4387945, 4387946, 4387947, 4387949, 4387950, 4387951, 4387952, 4387954, 4387955, 4387956, 4387957, 4387959, 4387960, 4387961, 4387962, 4387963, 4387964, 4387965, 4387966, 4387967, 4387968, 4387969, 4387970, 4387971, 4387974, 4387975, 4387976, 4387977, 4387978, 4387979, 4387980, 4387981, 4387982, 4387983, 4387984, 4387985, 4387986, 4387987, 4387988, 4387989, 4387992, 4387994, 4387995, 4387996, 4387997, 4387998, 4387999, 4388000, 4388001, 4388002, 4388006, 4388007, 4388008, 4388009, 4388010, 4388011, 4388012, 4388014, 4388015, 4388016, 4388017, 4388018, 4388019, 4388021, 4388022, 4388023, 4388024, 4388025, 4388026, 4388027, 4388028, 4388029, 4388030, 4388031, 4388032, 4388033, 4388034, 4388035, 4388036, 4388037, 4388038, 4388039, 4388040, 4388041, 4388042, 4388043, 4388044, 4388045, 4388046, 4388047, 4388048, 4388049, 4388050, 4388052, 4388053, 4388055, 4388057, 4388058, 4388061, 4388062, 4388064, 4388066, 4388071, 4388072, 4388073, 4388074, 4388075, 4388076, 4388077, 4388078, 4388079, 4388080, 4388081, 4388082, 4388083, 4388084, 4388085, 4388086, 4388087, 4388088, 4388089, 4388090, 4388091, 4388092, 4388093, 4388094, 4388095, 4388096, 4388097, 4388098, 4388099, 4388100, 4388101, 4388102, 4388103, 4388104, 4388105, 4388106, 4388107, 4388108, 4388109, 4388110, 4388111, 4388112, 4388113, 4388114, 4388115, 4388116, 4388117, 4388118, 4388119, 4388120, 4388121, 4388122, 4388123, 4388124, 4388125, 4388126, 4388127, 4388128, 4388129, 4388130, 4388131, 4388132, 4388133, 4388134, 4388135, 4388136, 4388137, 4388138, 4388139, 4388140, 4388141, 4388142, 4388143, 4388144, 4388145, 4388146, 4388147, 4388148, 4388149, 4388150, 4388151, 4388152, 4388153, 4388154, 4388155, 4388156, 4388157, 4388158, 4388159, 4388160, 4388161, 4388162, 4388163, 4388164, 4388165, 4388166, 4388167, 4388168, 4388169, 4388170, 4388171, 4388172, 4388173, 4388174, 4388175, 4388176, 4388177, 4388178, 4388179, 4388180, 4388181, 4388182, 4388183, 4388184, 4388185, 4388186, 4388187, 4388188, 4388189, 4388190, 4388191, 4388192, 4388193, 4388194, 4388195, 4388196, 4388197, 4388198, 4388199, 4388200, 4388201, 4388202, 4388203, 4388204, 4388205, 4388206, 4388207, 4388208, 4388209, 4388210, 4388211, 4388212, 4388213, 4388214, 4388215, 4388216, 4388217, 4388218, 4388219, 4388220, 4388221, 4388222, 4388223, 4388224, 4388225, 4388226, 4388227, 4388228, 4388229, 4388230, 4388231, 4388232, 4388233, 4388234, 4388235, 4388236, 4388237, 4388238, 4388239, 4388240, 4388241, 4388247, 4388249, 4388252, 4388253, 4388257, 4388264, 4388271, 4388273, 4388279, 4388282, 4388284, 4388285, 4388286, 4388287, 4388288, 4388289, 4388290, 4388291, 4388296, 4388297, 4388304, 4388306, 4388308, 4388311, 4388313, 4388314, 4388315, 4388316, 4388320, 4388323, 4388325, 4388332, 4388335, 4388336, 4388338, 4388339, 4388341, 4388537, 4388543, 4388545, 4388547, 4388548, 4388550, 4388551, 4388552, 4388553, 4388554, 4388555, 4388556, 4388557, 4388558, 4388559, 4388560, 4388561, 4388562, 4388563, 4388564, 4388565, 4388566, 4388567, 4388568, 4388569, 4388570, 4388571, 4388572, 4388573, 4388574, 4388575, 4388576, 4388577, 4388578, 4388579, 4388580, 4388581, 4388582, 4388583, 4388584, 4388585, 4388586, 4388587, 4388588, 4388589, 4388590, 4388591, 4388592, 4388593, 4388594, 4388595, 4388596, 4388597, 4388598, 4388599, 4388600, 4388601, 4388602, 4388603, 4388604, 4388605, 4388606, 4388607, 4388608, 4388609, 4388610, 4388611, 4388612, 4388613, 4388614, 4388615, 4388616, 4388617, 4388619, 4388620, 4388621, 4388622, 4388623, 4388624, 4388625, 4388626, 4388627, 4388628, 4388629, 4388630, 4388631, 4388632, 4388633, 4388634, 4388635, 4388636, 4388637, 4388638, 4388639, 4388640, 4388641, 4388642, 4388643, 4388644, 4388645, 4388646, 4388647, 4388648, 4388649, 4388650, 4388651, 4388652, 4388654, 4388655, 4388656, 4388657, 4388658, 4388659, 4388660, 4388661, 4388662, 4388664, 4388666, 4388669, 4388670, 4388671, 4388672, 4388674, 4388675, 4388676, 4388677, 4388678, 4388679, 4388680, 4388681, 4388682, 4388683, 4388684, 4388685, 4388686, 4388687, 4388688, 4388689, 4388690, 4388691, 4388692, 4388693, 4388694, 4388695, 4388696, 4388697, 4388698, 4388699, 4388700, 4388701, 4388702, 4388703, 4388704, 4388705, 4388706, 4388707, 4388708, 4388709, 4388710, 4388711, 4388712, 4388713, 4388714, 4388715, 4388716, 4388717, 4388718, 4388719, 4388720, 4388721, 4388722, 4388723, 4388724, 4388725, 4388726, 4388727, 4388728, 4388729, 4388730, 4388731, 4388733, 4388734, 4388735, 4388738, 4388744, 4388750, 4388754, 4388755, 4388756, 4388763, 4388764, 4388765, 4388766, 4388767, 4388768, 4388769, 4388770, 4388771, 4388772, 4388773, 4388774, 4388775, 4388776, 4388777, 4388778, 4388780, 4388781, 4388782, 4388783, 4388784, 4388785, 4388786, 4388787, 4388788, 4388789, 4388790, 4388791, 4388792, 4388793, 4388794, 4388795, 4388796, 4388797, 4388798, 4388799, 4388800, 4388801, 4388802, 4388803, 4388804, 4388805, 4388806, 4388807, 4388808, 4388809, 4388810, 4388811, 4388812, 4388813, 4388814, 4388815, 4388816, 4388817, 4388818, 4388819, 4388820, 4388821, 4388822, 4388823, 4388824, 4388825, 4388826, 4388827, 4388828, 4388829, 4388830, 4388831, 4388832, 4388833, 4388834, 4388837, 4388838, 4388839, 4388840, 4388841, 4388842, 4388843, 4388844, 4388845, 4388846, 4388847, 4388848, 4388849, 4388850, 4388851, 4388852, 4388853, 4388855, 4388857, 4388858, 4388860, 4388861, 4388862, 4388863, 4388864, 4388866, 4388867, 4388868, 4388869, 4388870, 4388871, 4388872, 4388873, 4388874, 4388875, 4388876, 4388877, 4388878, 4388879, 4388883, 4388884, 4388885, 4388888, 4388889, 4388891, 4388892, 4388893, 4388894, 4388895, 4388901, 4388902, 4388904, 4388906, 4388910, 4388912, 4388920, 4388921, 4388922, 4388923, 4388924, 4388925, 4388927, 4388930, 4388931, 4388932, 4388933, 4388934, 4388935, 4388936, 4388937, 4388938, 4388939, 4388940, 4388941, 4388942, 4388943, 4388944, 4388945, 4388948, 4388950, 4388951, 4388952, 4388953, 4388954, 4388955, 4388958, 4388959, 4388961, 4388966, 4388967, 4388968, 4388969, 4388970, 4388971, 4388972, 4388973, 4388974, 4388975, 4388976, 4388977, 4388978, 4388979, 4388980, 4388981, 4388983, 4388984, 4388986, 4388987, 4388989, 4388991, 4388992, 4388993, 4388995, 4388996, 4388998, 4388999, 4389001, 4389002, 4389004, 4389005, 4389006, 4389007, 4389008, 4389009, 4389010, 4389011, 4389012, 4389013, 4389015, 4389016, 4389017, 4389018, 4389022, 4389028, 4389029, 4389030, 4389033, 4389034, 4389035, 4389036, 4389038, 4389040, 4389041, 4389043, 4389044, 4389045, 4389046, 4389047, 4389048, 4389051, 4389052, 4389058, 4389061, 4389065, 4389066, 4389070, 4389072, 4389073, 4389075, 4389077, 4389080, 4389081, 4389082, 4389083, 4389085, 4389086, 4389088, 4389089, 4389091, 4389092, 4389093, 4389097, 4389098, 4389101, 4389103, 4389104, 4389106, 4389111, 4389115, 4389117, 4389118, 4389119, 4389120, 4389121, 4389122, 4389123, 4389124, 4389125, 4389126, 4389127, 4389128, 4389129, 4389130, 4389131, 4389132, 4389133, 4389134, 4389135, 4389136, 4389138, 4389139, 4389142, 4389143, 4389144, 4389147, 4389148, 4389149, 4389150, 4389152, 4389153, 4389154, 4389155, 4389156, 4389157, 4389158, 4389159, 4389160, 4389161, 4389162, 4389163, 4389164, 4389165, 4389166, 4389167, 4389168, 4389170, 4389171, 4389172, 4389173, 4389174, 4389175, 4389176, 4389177, 4389178, 4389179, 4389180, 4389181, 4389182, 4389183, 4389184, 4389185, 4389186, 4389187, 4389188, 4389189, 4389190, 4389191, 4389192, 4389193, 4389194, 4389195, 4389196, 4389197, 4389198, 4389199, 4389200, 4389201, 4389202, 4389203, 4389204, 4389205, 4389206, 4389207, 4389208, 4389209, 4389210, 4389211, 4389212, 4389213, 4389214, 4389215, 4389216, 4389217, 4389218, 4389219, 4389220, 4389221, 4389222, 4389223, 4389224, 4389225, 4389226, 4389227, 4389228, 4389229, 4389230, 4389231, 4389232, 4389233, 4389234, 4389235, 4389236, 4389237, 4389238, 4389239, 4389240, 4389241, 4389242, 4389243, 4389244, 4389245, 4389246, 4389247, 4389248, 4389249, 4389250, 4389251, 4389253, 4389254, 4389255, 4389256, 4389257, 4389258, 4389259, 4389260, 4389261, 4389262, 4389263, 4389264, 4389265, 4389266, 4389267, 4389268, 4389269, 4389270, 4389271, 4389272, 4389273, 4389274, 4389275, 4389276, 4389277, 4389278, 4389279, 4389280, 4389284, 4389285, 4389286, 4389287, 4389288, 4389289, 4389290, 4389293, 4389294, 4389295, 4389297, 4389298, 4389299, 4389303, 4389304, 4389305, 4389306, 4389307, 4389309, 4389310, 4389311, 4389312, 4389313, 4389314, 4389315, 4389316, 4389317, 4389318, 4389319, 4389321, 4389325, 4389327, 4389328, 4389329, 4389330, 4389331, 4389333, 4389335, 4389336, 4389337, 4389338, 4389339, 4389340, 4389341, 4389342, 4389343, 4389344, 4389345, 4389346, 4389347, 4389348, 4389349, 4389350, 4389351, 4389352, 4389353, 4389354, 4389355, 4389356, 4389357, 4389358, 4389359, 4389360, 4389361, 4389362, 4389363, 4389364, 4389365, 4389366, 4389367, 4389369, 4389370, 4389372, 4389373, 4389374, 4389375, 4389376, 4389377, 4389378, 4389379, 4389381, 4389382, 4389383, 4389384, 4389385, 4389386, 4389387, 4389388, 4389389, 4389390, 4389391, 4389392, 4389393, 4389394, 4389395, 4389396, 4389397, 4389398, 4389399, 4389400, 4389401, 4389402, 4389403, 4389404, 4389405, 4389406, 4389407, 4389408, 4389409, 4389410, 4389411, 4389412, 4389413, 4389414, 4389415, 4389416, 4389417, 4389418, 4389419, 4389420, 4389421, 4389422, 4389423, 4389424, 4389425, 4389426, 4389427, 4389428, 4389430, 4389431, 4389432, 4389433, 4389434, 4389435, 4389436, 4389437, 4389438, 4389458, 4389460, 4389461, 4389462, 4389463, 4389464, 4389465, 4389466, 4389467, 4389468, 4389469, 4389470, 4389471, 4389472, 4389473, 4389474, 4389475, 4389476, 4389477, 4389478, 4389480, 4389482, 4389483, 4389484, 4389485, 4389486, 4389487, 4389488, 4389489, 4389491, 4389492, 4389493, 4389494, 4389495, 4389496, 4389497, 4389499, 4389500, 4389501, 4389502, 4389503, 4389504, 4389505, 4389506, 4389507, 4389508, 4389509, 4389510, 4389511, 4389512, 4389513, 4389514, 4389515, 4389516, 4389517, 4389518, 4389519, 4389520, 4389521, 4389522, 4389523, 4389524, 4389525, 4389526, 4389527, 4389528, 4389529, 4389530, 4389531, 4389532, 4389533, 4389534, 4389535, 4389536, 4389537, 4389538, 4389539, 4389540, 4389541, 4389542, 4389543, 4389544, 4389545, 4389546, 4389547, 4389548, 4389549, 4389600, 4389601, 4389602, 4389603, 4389604, 4389605, 4389606, 4389607, 4389608, 4389609, 4389610, 4389611, 4389613, 4389614, 4389615, 4389616, 4389625, 4389626, 4389627, 4389628, 4389629, 4389630, 4389631, 4389632, 4389633, 4389634, 4389639, 4389640, 4389641, 4389642, 4389643, 4389644, 4389645, 4389646, 4389647, 4389648, 4389649, 4389650, 4389651, 4389652, 4389653, 4389654, 4389655, 4389656, 4389657, 4389658, 4389659, 4389660, 4389661, 4389662, 4389663, 4389664, 4389665, 4389666, 4389667, 4389668, 4389669, 4389670, 4389671, 4389672, 4389673, 4389674, 4389675, 4389676, 4389677, 4389678, 4389679, 4389680, 4389681, 4389682, 4389683, 4389684, 4389685, 4389686, 4389687, 4389688, 4389689, 4389690, 4389691, 4389692, 4389693, 4389694, 4389695, 4389696, 4389697, 4389698, 4389699, 4389700, 4389701, 4389702, 4389703, 4389704, 4389705, 4389706, 4389707, 4389708, 4389709, 4389710, 4389711, 4389712, 4389713, 4389714, 4389715, 4389716, 4389717, 4389718, 4389719, 4389720, 4389721, 4389722, 4389723, 4389724, 4389725, 4389726, 4389727, 4389728, 4389729, 4389730, 4389731, 4389732, 4389733, 4389734, 4389735, 4389736, 4389737, 4389738, 4389739, 4389740, 4389741, 4389742, 4389743, 4389744, 4389745, 4389746, 4389747, 4389748, 4389749, 4389750, 4389751, 4389752, 4389753, 4389754, 4389755, 4389756, 4389757, 4389758, 4389759, 4389760, 4389761, 4389762, 4389763, 4389764, 4389765, 4389766, 4389767, 4389768, 4389769, 4389770, 4389771, 4389772, 4389773, 4389774, 4389775, 4389776, 4389777, 4389778, 4389779, 4389780, 4389781, 4389782, 4389783, 4389784, 4389785, 4389786, 4389787, 4389788, 4389789, 4389790, 4389791, 4389792, 4389793, 4389794, 4389795, 4389796, 4389797, 4389798, 4389799, 4389800, 4389801, 4389802, 4389803, 4389804, 4389805, 4389806, 4389807, 4389808, 4389809, 4389810, 4389811, 4389812, 4389813, 4389814, 4389815, 4389816, 4389817, 4389818, 4389819, 4389820, 4389821, 4389822, 4389823, 4389824, 4389825, 4389826, 4389827, 4389828, 4389829, 4389830, 4389831, 4389832, 4389834, 4389839, 4389842, 4389846, 4389847, 4389848, 4389849, 4389851, 4389852, 4389853, 4389854, 4389855, 4389856, 4389857, 4389858, 4389859, 4389860, 4389861, 4389862, 4389863, 4389864, 4389865, 4389866, 4389867, 4389868, 4389869, 4389870, 4389871, 4389879, 4389881, 4389886, 4389887, 4389888, 4389889, 4389890, 4389891, 4389892, 4389893, 4389894, 4389895, 4389896, 4389897, 4389898, 4389899, 4389900, 4389901, 4389902, 4389903, 4389904, 4389905, 4389906, 4389907, 4389908, 4389909, 4389910, 4389911, 4389912, 4389914, 4389915, 4389916, 4389917, 4389918, 4389919, 4389920, 4389923, 4389927, 4389930, 4389932, 4389933, 4389934, 4389935, 4389937, 4389938, 4389944, 4389945, 4389947, 4389949, 4389950, 4389951, 4389952, 4389953, 4389954, 4389955, 4389956, 4389957, 4389958, 4389959, 4389960, 4389962, 4389963, 4389964, 4389965, 4389968, 4389970, 4389971, 4389972, 4389973, 4389974, 4389975, 4389976, 4389977, 4389978, 4389979, 4389980, 4389981, 4389982, 4389983, 4389984, 4389985, 4389986, 4389987, 4389988, 4389990, 4389992, 4389993, 4389995, 4389996, 4389997, 4389999, 4390000, 4390001, 4390002, 4390003, 4390005, 4390006, 4390007, 4390009, 4390010, 4390011, 4390012, 4390013, 4390014, 4390015, 4390016, 4390018, 4390019, 4390020, 4390021, 4390022, 4390023, 4390024, 4390025, 4390026, 4390027, 4390028, 4390029, 4390030, 4390031, 4390032, 4390033, 4390034, 4390035, 4390036, 4390037, 4390038, 4390039, 4390040, 4390041, 4390042, 4390043, 4390044, 4390045, 4390046, 4390047, 4390048, 4390049, 4390050, 4390051, 4390052, 4390053, 4390054, 4390055, 4390056, 4390057, 4390058, 4390059, 4390060, 4390061, 4390100, 4390101, 4390102, 4390103, 4390104, 4390105, 4390106, 4390107, 4390108, 4390109, 4390110, 4390111, 4390112, 4390113, 4390114, 4390115, 4390116, 4390117, 4390118, 4390119, 4390120, 4390121, 4390122, 4390123, 4390124, 4390125, 4390126, 4390127, 4390128, 4390129, 4390130, 4390131, 4390132, 4390133, 4390134, 4390135, 4390136, 4390137, 4390138, 4390139, 4390140, 4390141, 4390143, 4390144, 4390145, 4390146, 4390147, 4390148, 4390149, 4390150, 4390151, 4390152, 4390153, 4390154, 4390155, 4390156, 4390157, 4390158, 4390159, 4390160, 4390161, 4390162, 4390163, 4390164, 4390165, 4390166, 4390167, 4390168, 4390169, 4390170, 4390171, 4390172, 4390173, 4390174, 4390175, 4390176, 4390177, 4390178, 4390179, 4390180, 4390181, 4390182, 4390183, 4390184, 4390185, 4390186, 4390187, 4390188, 4390189, 4390192, 4390201, 4390204, 4390223, 4390225, 4390254, 4390274, 4390276, 4390278, 4390279, 4390283, 4390284, 4390285, 4390289, 4390339, 4390340, 4390341, 4390342, 4390345, 4390346, 4390347, 4390348, 4390353, 4390355, 4390356, 4390357, 4390360, 4390373, 4390375, 4390376, 4390377, 4390381, 4390382, 4390383, 4390385, 4390386, 4390388, 4390390, 4390392, 4390394, 4390395, 4390396, 4390397, 4390398, 4390399, 4390400, 4390436, 4390438, 4390442, 4390452, 4390495, 4390499, 4390500, 4390504, 4390505, 4390508, 4390509, 4390510, 4390511, 4390512, 4390513, 4390514, 4390515, 4390570, 4390571, 4390573, 4390574, 4390576, 4390577, 4390578, 4390579, 4390580, 4390581, 4390582, 4390583, 4390595, 4390605, 4390617, 4390618, 4390619, 4390620, 4390621, 4390623, 4390625, 4390633, 4390636, 4390637, 4390638, 4390640, 4390641, 4390645, 4390647, 4390648, 4390649, 4390650, 4390651, 4390652, 4390655, 4390657, 4390665, 4390666, 4390667, 4390674, 4390675, 4390686, 4390687, 4390691, 4390693, 4390694, 4390696, 4390697, 4390699, 4390700, 4390702, 4390705, 4390706, 4390707, 4390709, 4390712, 4390713, 4390715, 4390754, 4390762, 4390790, 4390792, 4390793, 4390794, 4390795, 4390798, 4390799, 4390800, 4390801, 4390803, 4390804, 4390805, 4390806, 4390807, 4390808, 4390809, 4390810, 4390811, 4390812, 4390813, 4390814, 4390815, 4390816, 4390817, 4390818, 4390819, 4390820, 4390821, 4390822, 4390823, 4390824, 4390825, 4390826, 4390827, 4390828, 4390829, 4390830, 4390831, 4390832, 4390833, 4390834, 4390835, 4390836, 4390837, 4390838, 4390839, 4390840, 4390841, 4390842, 4390843, 4390844, 4390845, 4390846, 4390847, 4390848, 4390849, 4390850, 4390852, 4390853, 4390854, 4390856, 4390857, 4390858, 4390859, 4390860, 4390862, 4390863, 4390866, 4390867, 4390868, 4390871, 4390873, 4390877, 4390878, 4390880, 4390882, 4390883, 4390884, 4390885, 4390886, 4390889, 4390890, 4390892, 4390895, 4390896, 4390898, 4390900, 4390901, 4390902, 4390903, 4390904, 4390906, 4390907, 4390909, 4390911, 4390912, 4390914, 4390915, 4390916, 4390917, 4390918, 4390919, 4390921, 4390922, 4390923, 4390924, 4390925, 4390926, 4390927, 4390928, 4390932, 4390933, 4390935, 4390938, 4390939, 4390940, 4390941, 4390942, 4390943, 4390944, 4390945, 4390947, 4390950, 4390954, 4390956, 4390959, 4390960, 4390961, 4390962, 4390963, 4390968, 4390969, 4390970, 4390971, 4390972, 4390973, 4390974, 4390975, 4390976, 4390977, 4390978, 4390979, 4390981, 4390982, 4390984, 4390985, 4390986, 4390987, 4390989, 4390990, 4390991, 4390992, 4390993, 4391194, and 4391195. |
Recalling Firm/ Manufacturer |
Cardiac Science Corporation 3303 Monte Villa Pkwy Bothell WA 98021-8969
|
For Additional Information Contact | Bob Odell 888-402-2484 |
Manufacturer Reason for Recall | The affected Automated External Defibrillators (AEDs) contains a circuit board manufactured with a component that may fail unexpectedly due to a supplier manufacturing defect. If the component were to fail during a rescue attempt, the AED may not deliver defibrillation therapy. |
FDA Determined Cause 2 | Process control |
Action | Cardiac Science sent an "URGENT-Voluntary Medical Device Recall" letter dated January 13, 2012 to all affected customers.
The letter described the product, problem, and actions to be taken by the customers. Customers are advised to immediately return the device to the manufacturing location for repair. The affected devices will receive a hardware correction, and the same serial number device returned to the consignee in most cases.
Customers in the US can call 1-888-402-2484 and customers outside the US can call 1-425-402-2482 or email at aed210@cardiacscience.com to arrange delivery of the shipping materials.
Customers can also visit www.cardiacscience.com/aed210 for more information about AEDs affected by this recall. |
Quantity in Commerce | 11,019 units (6142 units in the US and 4877 units international) |
Distribution | Worldwide distribution: USA (nationwide) and Puerto Rico; and to the following countries: Australia, Belgium, Brazil, Canada, China, Colombia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Jamaica, Japan, Korea, Kuwait, Malaysia, Maldives, Malta, Netherlands, New Zealand, Norway, Panama, Peru, Poland, Portugal, Qatar, Romania, Russia, Saint Lucia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, Ukraine, and United Kingdom. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = MKJ 510(K)s with Product Code = MKJ
|
|
|
|