• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall Syngo RT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall Syngo RTsee related information
Date Initiated by FirmJanuary 22, 2012
Date PostedFebruary 23, 2012
Recall Status1 Terminated 3 on November 07, 2012
Recall NumberZ-1092-2012
Recall Event ID 61121
510(K)NumberK060226 K072485 K993425 
Product Classification Accelerator, linear, medical - Product Code IYE
ProductARTISTE, ONCOR and PRIMUS Linear Accelerators with RT Therapist Workspace, version 4.1. Manufactured by Siemens AG, Kemnath, Germany. The indication is to deliver X-ray, photon and electron radiation for the therapeutic treatment of cancer.
Code Information Syngo RT Therapist v4.1, part number 8162815; Syngo RT Therapist Assist (off line system), part number 8162807; Syngo RT Express Basic (stand alone system), part number 8151289; and Syngo RT Express Assist (offline system), part number 8151297.
FEI Number 2937457
Recalling Firm/
Manufacturer
Siemens Medical Solutions USA, Inc
4040 Nelson Ave
Concord CA 94520-1200
For Additional Information ContactChristine Dunbar
925-602-8157
Manufacturer Reason
for Recall
Software upgrade to installed base to the latest release of the RT Therapist version 4.1.144 which corrects multiple safety issues that could lead to patient mistreatment.
FDA Determined
Cause 2
Software design
ActionSiemens customers with active devices will receive the software update through the distribution of the "Update Instructions" with a release date of January 22, 2012. The software update will be delivered and installed by a qualified Field Service Engineer utilizing a SW kit and instruction distributed by the Update instruction. Consignee notification of the Update Instructions will be delivered by Customer Service or by certified mail.
Quantity in Commerce182 active devices
DistributionWorldwide Distribution - USA, including the states of WI, UT, PA, TN, OH, LA, DE, MI, NJ, FL, NE, WV, MO and the countries of Germany, Spain, Poland, Australia, Malaysia, New Zealand, India, Republic of Korea, Canada, Italy, Egypt, Philippines, France, Belgium, Saudi Arabia, South Africa, Norway, Turkey , Hungary, Russian Federation, P.R. China, Lebanon, Iran, and the United Kingdom.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = IYE
-
-