 Class 2 Device Recall Servoi ventilator systemsee related information
Date Initiated by FirmDecember 22, 2011
Date PostedApril 03, 2012
Recall Status1 Terminated 3 on October 22, 2013
Recall NumberZ-1363-2012
Recall Event ID 61327
510(K)NumberK041223 
Product Classification Ventilator, continuous, facility use - Product Code CBK
ProductServo-s ventilator system Product Usage: Intended for treatment and monitoring of patients in the range of neonates, infants and adults with respiratory failure or respiratory insufficiency.
Code Information 510k k041223 B052264 Article 66 40 440 Serial numbers 19841-21664
FEI Number 3008355164
Recalling Firm/
Manufacturer
Maquet Inc.
45 Barbour Pond Drive
Wayne NJ 07470
For Additional Information ContactMs. Whitney Torning
973-709-7994
Manufacturer Reason
for Recall
The Servo-i and Servo-s ventilator systems equipped with specific revision os of the AC/DC convertor can result in electric shock during servicing.
FDA Determined
Cause 2
Device Design
ActionMaquet sent a Product Recall Urgent-Medical Device Corrective Action letter dated December 22, 2011 to all affected customers. The letter identified the affected products, patient risk, product issue and actions to be taken. Customers were instructed to complete and return the attached Customer Fax Response form as soon as possible. Maquet Field Service Technicians will update all affected products at all the concerned customer sites. Customers are advised untill Maquet has updated all the affected products (all of which are still under warranty), inform all the concerned biomedical staff about the content of this Device Correction Letter. For technical support call 1-888-627-8383.
Quantity in Commerce50 units US
DistributionUSA Nationwide Distribution
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = CBK
-
-